← Product Code [HGX](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX) · K241004

# Wearable Breast Pump (ABP-1508PRO, ABP-1506, ABP-1306, ABP-1306 PRO, M1, M3, M8, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) (K241004)

_Huizhou Lvb Maternal and Infant Supplies Co., Ltd. · HGX · Sep 11, 2024 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX/K241004

## Device Facts

- **Applicant:** Huizhou Lvb Maternal and Infant Supplies Co., Ltd.
- **Product Code:** [HGX](/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX.md)
- **Decision Date:** Sep 11, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

Wearable Breast Pump (ABP-1508PRO, ABP-1506, ABP-1306, ABP-1306 PRO, M1, M3, M8, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The device is intended for a single user.

## Device Story

Wearable/tabletop powered breast pump for home use by lactating women; single-user. Device uses DC motor to drive membrane vacuum pump; generates cyclic negative pressure (suction) to mimic infant suckling. User selects stimulation, expression, or auto modes via interface; 9 suction levels available. Milk-contacting components use physical silicone diaphragm barrier for backflow protection. Powered by internal rechargeable Li-ion battery or AC power. Output is expressed milk collected in device container. Healthcare provider does not operate; patient self-operates. Output informs milk collection volume; benefits patient by facilitating lactation management.

## Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1, IEC 60601-1-11), battery safety (IEC 6213-2), electromagnetic compatibility (IEC 60601-1-2), vacuum level verification, backflow protection, use-life, and battery performance.

## Technological Characteristics

Powered breast pump; DC motor-driven membrane vacuum pump. Materials: milk-contacting components compliant with 21 CFR 174-179. Suction: -60 to -240 mmHg; 26-108 cycles/min. Power: internal rechargeable Li-ion battery or AC mains. Connectivity: none. Sterilization: non-sterile. Software: microcontroller-based, basic documentation level.

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Wearable Breast Pump (ABP-1508) ([K230102](/device/K230102.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

September 11, 2024

HuiZhou LVB Maternal and Infant Supplies Co., Ltd. % Youshan Gong RA Specialist Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90, Oianhai Road Shenzhen, Guangdong 518052 CHINA

Re: K241004

> Trade/Device Name: Wearable Breast Pump (ABP-1508PRO, ABP-1506, ABP-1306, ABP-1306 PRO, M1, M3, M8, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: August 12, 2024 Received: April 12, 2024

Dear Youshan Gong:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/uniquedevice-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice

{2}------------------------------------------------

(https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K241004

Device Name

Wearable Breast Pump (ABP-1508PRO, ABP-1306, ABP-1306, ABP-1306, M1, M3, M8, M5, M6, K1, K2, K3, ABP-1306PLUS, ABP-1308, M2)

#### Indications for Use (Describe)

Wearable Breast Pump (ABP-1508PRO, ABP-1306, ABP-1306, ABP-1306 PRO, M1, M3, M8, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The device is intended for a single user.

| Type of Use (Select one or both, as applicable)                                      |                                                                                                |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary - K241004

#### 1. Submitter Information

| Applicant: | HuiZhou LVB maternal and infant supplies<br>Co.,Ltd.                                                              |
|------------|-------------------------------------------------------------------------------------------------------------------|
| Phone:     | 0752-2156859                                                                                                      |
| Address:   | NO. 114,Xin Xing North Road, Sandong<br>Town, Huicheng District, Huizhou City,<br>Guangdong, Province, P.R. China |

#### 2. Correspondent Information

| Contact: | Dongan Zhang<br>Head of Firm, |
|----------|-------------------------------|
| Email:   | zhda7311@163.com              |
| Phone:   | 86 (13075) 210-606            |

#### 3. Date prepared: September 5, 2024

#### 4. Device Information

| Device Name:       | Wearable Breast Pump (ABP-1508PRO, ABP-1506, ABP-1306, ABP-1306 PRO, M1, M3, M8, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:       | Powered breast pump                                                                                                                  |
| Regulation Number: | 21 CFR 884.5160                                                                                                                      |
| Regulation Name:   | Powered Breast Pump                                                                                                                  |
| Product Code:      | HGX (Pump, Breast, Powered)                                                                                                          |
| Regulatory Class:  | Class II                                                                                                                             |

#### 5. Predicate Device Information

| Device Name:   | Wearable Breast Pump (ABP-1508)                  |
|----------------|--------------------------------------------------|
| 510(k) Number: | K230102                                          |
| Manufacturer:  | Huizhou LVB Maternal And Infant Supplies Co,Ltd. |

The predicate device has not been subject to a design-related recall.

## 6. Device Description

The Wearable Breast Pump (ABP-1508PRO, ABP-1506, ABP-1306 PRO, M1, M3, M8, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) is an electrically powered breast pump to be used in a home environment by a single user. The device is provided non-sterile and can be used on one breast (single pumping) or on both breasts at the same time (double pumping).

The subject devices are capable of expression, stimulation, and auto modes with nine associated suction levels for each. The subject devices feature various combinations of on/off buttons, level up/down buttons, mode selection buttons, suction level/time indicators, battery visual indicator displays. The pumps are powered by internal, non-replaceable, rechargeable lithium-ion batteries which are charged using the included AC power supply and cable. The subject device can be operated while plugged into AC power.

The breast pumps use cyclic negative pressure (suction) to mimic the suckling patterns of a feeding infant. A

{5}------------------------------------------------

DC motor drives a membrane vacuum pump to generate the suction required to stimulate and express breast milk. The timing of this pattern is dependent upon the suction/speed settings selected by the user and is preprogrammed in the devices. The devices are capable of producing peak suction levels between -60 and -240 mmHg at speeds between 26 and 108 cycles per minute. There available pumping modes with 9 distinct levels of vacuum and cycle speed. The subject device ensures backflow protection between the breast shield and the electronic components via a physical barrier (silicone diaphragm) mechanism

All other components (i.e., motor unit) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 174-179.

## 7. Indications for Use

Wearable Breast Pump (ABP-1508PRO, ABP-1306, ABP-1306, ABP-1306 PRO, M1, M3, M5, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The device is intended for a single user.

#### 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

#### Table 1: Comparator Table for Subject and Predicate Devices

|                                              | Wearable Breast Pump<br>(ABP-1508PRO, ABP-<br>1506, ABP-1306, ABP-<br>1306 PRO, M1, M3,<br>M8, M5, M6, K1, K2,<br>K3, K5, ABP-<br>1306PLUS, ABP-1308,<br>M2)<br>K241004<br>Subject Device                                                                                                                          | Electric Breast Pump<br>K230102<br>Predicate Device                                                                                                                                  | Comparison |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code                                 | HGX                                                                                                                                                                                                                                                                                                                | HGX                                                                                                                                                                                  | Same       |
| Regulation Number                            | 21 CFR 884.5160                                                                                                                                                                                                                                                                                                    | 21 CFR 884.5160                                                                                                                                                                      | Same       |
| Regulatory Class                             | Class II                                                                                                                                                                                                                                                                                                           | Class II                                                                                                                                                                             | Same       |
| Patient Population                           | Lactating Women                                                                                                                                                                                                                                                                                                    | Lactating Women                                                                                                                                                                      | Same       |
| Indications for Use                          | Wearable Breast Pump (ABP-<br>1508PRO, ABP-1506, ABP-<br>1306, ABP-1306 PRO, M1, M3,<br>M8, M5, M6, K1, K2, K3, K5,<br>ABP-1306PLUS, ABP-1308,<br>M2) is a powered breast pump to<br>be used by lactating women to<br>express and collect milk from<br>their breasts. The device is<br>intended for a single user. | Wearable Breast Pump is a<br>powered breast pump to be used<br>by lactating women to express<br>and collect milk from their<br>breasts. The device is intended<br>for a single user. | Same       |
| Single/double pump                           | Single or double                                                                                                                                                                                                                                                                                                   | Single                                                                                                                                                                               | Different  |
| Media separation<br>(backflow<br>protection) | Yes                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                  | Same       |

{6}------------------------------------------------

| Cycling control mechanism       | Microcontroller                                                                                                                         | Microcontroller                               | Same      |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|-----------|
| Expression pattern              | 2-Phase                                                                                                                                 | 2-Phase                                       | Same      |
| Power supply                    | Li-Ion Battery and mains powered                                                                                                        | Li-Ion Battery and mains powered              | Same      |
| Suction levels (stimulation)    | Single Pumping:<br>$80-200 \pm 5$ mmHg<br>Double Pumping:<br>$60-180 \pm 5$ mmHg                                                        | $80-200 \pm 5$ mmHg                           | Different |
| Suction levels (expression)     | Single Pumping:<br>$90-240 \pm 5$ mmHg<br>Double Pumping:<br>$70-220 \pm 5$ mmHg                                                        | $90-240 \pm 5$ mmHg                           | Different |
| Cycles per minute (stimulation) | $64-108 \pm 2$ cycle/min                                                                                                                | $64-108 \pm 2$ cycle/min                      | Same      |
| Cycles per minute (expression)  | $26-48 \pm 2$ cycle/min                                                                                                                 | $26-48 \pm 2$ cycle/min                       | Same      |
| Available modes                 | 3 modes (stimulation/expression/auto)                                                                                                   | 2 modes (stimulation/expression)              | Different |
| Suction levels                  | All models: 9 Levels each mode                                                                                                          | Both models: 9 Levels each mode               | Similar   |
| User Interface                  | On-Off buttons/switches, vacuum adjustment, mode change                                                                                 | On-Off switch, vacuum adjustment, mode change | Similar   |
| Adjustable Suction Levels       | Yes                                                                                                                                     | Yes                                           | Same      |
| Mobile Application              | No                                                                                                                                      | No                                            | Same      |
| Design                          | Wearable (ABP-1508 PRO, ABP-1506, ABP-1306, ABP-1306 PRO, M1, M2, M3, M8, M5, M6, K1, K2, K3, K5)<br>Tabletop (ABP-1308, ABP-1306 Plus) | Wearable (ABP-1508)                           | Different |

The indications for use of the subject and predicate device are identical and they have the same intended use - the expression and collection of breast milk.

The subject and predicate devices have similar technological features, including available number of available vacuum levels, single/double pumping operation, and cycle specifications. However, as shown in the table above, there are technological differences between the subject and predicate device, including different overall vacuum pressure specifications, available pumping modes, tabletop/wearable design, etc. However, the different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.

#### 9. Summary of Non-Clinical Performance Testing

#### Biocompatibility

Biocompatibility information was provided in accordance with Attachment G of the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process."

{7}------------------------------------------------

## Electrical Safety

Testing was conducted in accordance with IEC 60601- 1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance). IEC 6213-2:2017. Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, and IEC 60601-1-11 Edition 3.1 2020-07 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

## Electromagnetic Compatibility

Testing was conducted in accordance with IEC 60601-1-2:2014+A1: 2020 Medical Electrical Equipment -Part 1-2: "General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests."

#### Software

Software was evaluated at the Basic Documentation level as recommended in the 2023 FDA guidance document "Content of Premarket Submissions for Device Software Functions."

#### Performance Testing

Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:

- . Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
- Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
- Use life testing was conducted to demonstrate that the device maintains its specifications ● throughout its proposed use life.
- Battery performance testing was conducted to demonstrate that the battery remains functional ● during its stated battery use-life.
- Battery status indicator testing was conducted to demonstrate that the battery status indicator ● remains functional during its stated battery life.

#### 10. Conclusion

The results of the performance testing described above demonstrate that the Wearable Breast Pump (ABP-1508PRO, ABP-1506, ABP-1306, ABP-1306 PRO, M1, M3, M6, K1, K2, K3, K5, ABP-1306PLUS, ABP-1308, M2) is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX/K241004](https://fda.innolitics.com/submissions/OB/subpart-f%E2%80%94obstetrical-and-gynecological-therapeutic-devices/HGX/K241004)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
