The Annabella Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Annabella Breast Pump is intended for a single user.
Device Story
Annabella Breast Pump is a single-user, electric, rechargeable Li-ion battery-operated device for home use by lactating women to express and collect breast milk. The system includes a vacuum unit, tubing, and a massage unit designed to mimic a baby's sucking motions. Users operate buttons on the vacuum unit to adjust vacuum intensity, suction rate, and the speed/location of the massage unit on the breast. The device features a 7-segment display and LED indicators. It includes a reusable, adjustable breast shield and bottle lids. The device cannot operate while connected to AC mains. By providing controlled suction and massage, the device assists in milk expression, potentially benefiting the user by facilitating lactation management. The healthcare provider does not interact with the device; the patient operates it independently.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1, IEC 60601-1-11), battery safety (IEC 62133-2), EMC (IEC 60601-1-2), software verification (IEC 62304), and biocompatibility (ISO 10993-1, -5, -10). Non-clinical performance testing verified vacuum levels, cycle speeds, backflow protection, battery performance, and device use-life.
Technological Characteristics
Powered breast pump; Li-ion battery or AC adapter; microcontroller-based control. Features 9 suction levels and 9 massage speed/location levels. Includes massage unit with 4 rotating cams. Materials evaluated per ISO 10993. Software verification per IEC 62304. Home-use environment.
Indications for Use
Indicated for lactating women to express and collect milk from their breasts. Intended for single-user, home-use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 4, 2023
Annabella Ltd. % Dalia Dickman, Ph.D. Regulatory Consultant Dalia Dickman Consulting Manof Misgav, 20184 Israel
Re: K230672 Trade/Device Name: Annabella Breast Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: July 6, 2023 Received: July 7, 2023
Dear Dalia Dickman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230672
Device Name Annabella Breast Pump
Indications for Use (Describe)
The Annabella Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Annabella Breast Pump is intended for a single user.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
#### Annabella Tech Ltd's Annabella Breast Pump
#### Submitter
Annabella Tech Ltd. 23/5 Hataas KFAR SABA, 4442525 Israel. +972-9-8845513
Contact Person: Dalia Dickman, PhD. Date Prepared: August 2, 2023
Device Name: Annabella Breast Pump
Common or Usual Name: Powered breast pump
Regulation Number: 21 CFR 884.5160
Regulation Name: Powered breast pump
Regulatory Class: II
Product Code: HGX (Powered, Breast, Pump)
#### Predicate Device
Medela Freestyle (K150499)
The predicate device has not been subject to a design-related recall.
#### Device Description
The Annabella Breast Pump is a single, electric breast pump system intended to express and collect milk from the breasts of lactating women. It is comprised of a vacuum unit including tubing and a massage unit that mimics the baby's sucking motions. The user employs buttons on the vacuum unit to control the vacuum intensity levels, suction rate as well as the massage unit speed and location on the breast. The device is supplied with a reusable, washable and adjustable breast shield, a bottle lids and a charger. The Annabella is a rechargeable Li-ion battery operated device that contains software. The device cannot be operated while connected to the mains AC Power core.
The battery compartment is at the bottom of the pump motor unit and is covered by a battery door. The runtime of the removable lithium-ion battery is influenced by the number and duration of pumping sessions and lasts usually for one day.
The device is to be used in the home.
{4}------------------------------------------------
#### Intended Use / Indications for Use
The Annabella Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Annabella Breast Pump is intended for a single user.
#### Predicate Device Comparison
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Device Name | Annabella<br>Breast Pump | Predicate Device:<br>FreeStyle;<br>K150499 | Discussion |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Annabella<br>Breast Pump is a<br>powered breast<br>pump to be used<br>by lactating women<br>to express and<br>collect milk from<br>their breasts. The<br>Annabella Breast<br>Pump is intended<br>for a single user. | The Freestyle® is a<br>powered breast<br>pump to be used by<br>lactating women to<br>express and collect<br>milk from their<br>breasts. The<br>Freestyle® is<br>intended for a<br>single user. | Identical |
| Intended Use | Express and<br>collect milk | Express and collect<br>milk | Identical |
| Single User Device | Yes | Yes | Identical |
| Environment of<br>Use | Home | Home | Identical |
| Over the Counter | Yes | Yes | Identical |
| User Interface<br>Hardware Interfaces | | | |
| User Control | • On-off switch<br>• Vacuum<br>adjustment<br>Control<br>• Cycle-<br>adjustment<br>Control<br>• Tongue<br>mechanism | On-off switch<br>Vacuum/Cycle-<br>adjustment<br>control | Different -<br>Annabella has<br>two independent<br>controls<br>for vacuum and<br>cycles.<br>Freestyle® uses<br>a single<br>control to adjust<br>vacuum<br>and cycles<br>together. |
| | control (height<br>and pace) | | |
| Visual Indicator | 7 Segment<br>Display and LED<br>Indications | LCD Display | Different - Both<br>provide indications<br>display |
| Pumping<br>Options | Single only | Single or Double | Different -<br>Pumping<br>mechanism is the<br>same |
| Accessories | A variety of<br>accessories for:<br>• Collection of<br>milk<br>• Storage of milk<br>• Breast care<br>• Providing<br>Power | A variety of<br>accessories for:<br>• Collection of milk<br>• Storage of milk<br>• Breast care<br>• Providing Power<br>• Carrying<br>• Cleaning<br>• Feeding | Different - both<br>systems come<br>with or make<br>available a<br>variety of<br>accessories that<br>can be used with<br>the pump for<br>collection and<br>storage of<br>breast milk,<br>providing<br>power and breast<br>pump. Freestyle®<br>has<br>additional<br>accessories for<br>carrying and<br>cleaning its<br>components<br>and feeding<br>stored milk. |
| Media Separation | Yes | Yes | Identical |
| Specifications | | | |
| Power Supply | • Li-Ion battery or<br>• AC adaptor<br>provided | • Li-Ion battery or<br>• AC adaptor<br>provided | Identical |
| Suction Levels<br>(stimulation) | 27-105 mmHg | 40 - 140 mmHg | Different |
| Cycles per second<br>(stimulation) | 1.66-2.0 | 1.7-1.93 | Different |
| Suction Levels | 26-217 mmHg | 45 – 245 mmHg | Different |
| (expression) | | | |
| Cycles per second<br>(expression) | 0.8-1.44 | 0.83-1.36 | Different |
| Maximum Vacuum | 250 mmHg | 270 mmHg | Different |
| Suction Settings | 9 | 9 | Identical |
| Adjustable Suction<br>Levels | Yes | Yes | Identical |
| Let-Down Button | Yes | Yes | Identical |
| Massage (Tongue<br>Mechanism) | - 4 cams rotating at a<br>typical angular<br>speed of 0-40rpm<br>- 9 levels of speed<br>- 9 levels for location on<br>the breast (distance<br>from the breast center) | NA | Annabella has an<br>additional massage unit –<br>both the speed and location on<br>the breast is controlled by the<br>user for maximum<br>comfort; the user can stop its<br>operation at any time. The<br>addition of the massage unit<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Cycling Control<br>Mechanism | Microcontroller | Microcontroller | Identical |
| Back Flow<br>Protection | Yes | Yes | Identical |
| 2-phase<br>expression | Yes | Yes | Equivalent. Both<br>devices offer an initial<br>simulation phase that moves<br>to expression phase<br>after two |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
| | | | minutes. |
|--|--|--|----------|
| | | | |
The indications for use of the subject and predicate devices are identical and they have the same intended use (i.e., the collection of breast milk from the breasts of lactating women).
The subject and predicate devices have different technological features, including differences in the pumping options, cycle speeds, vacuum strengths, visual indicators, and the inclusion of a massage unit in the subject device. The different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.
#### Performance Data
Non-clinical tests were conducted to verify that the subject device met all design specifications to be considered substantially equivalent to the predicate device.
#### Electrical Safety and Electromagnetic Compatibility:
- -Electrical Safety Testing was performed in accordance with IEC 60601-1:2005 (3rd Edition)
- | Electrical safety testing for use in home was performed in accordance with IEC 60601-1-11:2015, General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- -Battery safety testing was performed in accordance with the applicable standards for Lithium-ion batteries: Cell and pack: 62133-2:2017-02: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary lithium cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- Electromagnetic compatibility testing was performed in accordance with IEC | 60601-1-2:2021. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
#### Software Verification:
Software Verification was performed in accordance with IEC 62304:2015 Ed.1.1: Medical Device Software – Software Life Cycle Processes and according to the FDA Guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.
#### Biocompatibility:
{8}------------------------------------------------
Biocompatibility evaluation was performed according to the following requirements of ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process for patient-contacting components of the subject device as follow:
- -Cytotoxicity testing per ISO 10993-5:2009
- -Guinea Pig Maximization Sensitization testing per ISO 10993-10:2010
- -Irritation testing per ISO 10993-10:2010
## Non-clinical Performance Testing:
- Vacuum level/suction strength of subject devices was tested for each mode/ cycle.
- -Backflow protection testing was conducted to ensure that even if the bottle is over-filled, no liquid will backflow into the tubing, and therefore no liquid can backflow into the pump motor.
- -Cycle speed of subject devices was tested.
- Battery performance testing was conducted to demonstrate that the battery | remains functional during its stated use-life.
- -Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
- | Device use life to demonstrate that the device maintains its specifications throughout its proposed use life.
## Conclusions
The results of the performance testing described above demonstrate that the Annabella Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.