Fit Pro Breast Pump

K212955 · Hygeia II Medical Group, Inc. · HGX · Mar 15, 2022 · Obstetrics/Gynecology

Device Facts

Record IDK212955
Device NameFit Pro Breast Pump
ApplicantHygeia II Medical Group, Inc.
Product CodeHGX · Obstetrics/Gynecology
Decision DateMar 15, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5160
Device ClassClass 2

Indications for Use

The FIT Pro Breast Pump is a single-user, powered breast pump intended to express and collect milk from the breasts of lactating women.

Device Story

Electric breast pump for lactating women; home use. Device utilizes DC-powered motor driving diaphragm-type vacuum pump and electromechanical solenoid. User interface includes LCD display and soft-touch buttons for mode selection (stimulation/expression), vacuum strength (6 levels stimulation, 12 levels expression), and timer. Device operates via AC-DC power or internal rechargeable lithium-ion battery. Hygeia PAS pumping kit connects to breast; diaphragm in backflow protection assembly isolates pump/tubing from milk to prevent contamination. User selects vacuum/cycle frequency; device provides suction to express milk into collection bottles. Benefits include efficient milk expression and protection of breast milk from contamination.

Clinical Evidence

Bench testing only. Performance testing verified vacuum levels, cycle speeds, battery life, battery indicator accuracy, and backflow/overflow protection under both AC and battery power. Compliance with ISO 14971, IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11 confirmed.

Technological Characteristics

Powered diaphragm vacuum pump; polypropylene and silicone patient-contacting materials. Powered by 5VDC AC/DC adapter or 3.7V 1500 mAh lithium-ion battery. Features LCD display, soft-touch controls, and backflow protection assembly. Vacuum range 40-270 mmHg; cycle range 25-120 cycles/minute. Software-controlled electromechanical solenoid and motor. Non-sterile.

Indications for Use

Indicated for lactating women requiring expression and collection of breast milk. Intended for single-user, home environment use.

Regulatory Classification

Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 15, 2022 Hygeia II Medical Group, Inc. John Conklin VP Product Development 6241 Yarrow Drive Carlsbad, CA 92011 Re: K212955 Trade/Device Name: FIT Pro Breast Pump Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: February 14, 2022 Received: February 15, 2022 #### Dear John Conklin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K212955 Device Name FIT Pro Breast Pump Indications for Use (Describe) The FIT Pro Breast Pump is a single-user, powered breast pump intended to express and collect milk from the breasts of lactating women. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(K) SUMMARY K212955 In accordance with 21 CFR 807.92(a) the following summary is provided: ### DATE PREPARED: March 14, 2022 #### SUBMITTER: Hygeia II Medical Group, Inc. 6241 Yarrow Drive Carlsbad, CA 92011 Phone: 714-615-7571 ### PRIMARY CONTACT PERSON: Brett Nakfoor CEO Hygeia II Medical Group, Inc. 847-964-2620 #### Device information Trade Name: FIT Pro Breast Pump Common Name: Powered breast pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered breast pump Product Code: HGX (Pump, Breast, Powered) Classification Panel: Obstetrics/Gynecology Regulatory Class: II #### Predicate Device Information Ameda Mya Joy PLUS: K203570 The predicate device has not been subject to a design-related recall. Reference Device Information Evolve Breast Pump: K190465 #### Device Description The FIT Pro Breast Pump is an electric breast pump powered by an external AC-DC power supply or by an internal rechargeable lithium-ion battery. The device is provided non-sterile. {4}------------------------------------------------ The device is intended to be used by lactating women to express and collect milk from their breasts. Pumping can be performed on either one breast (single pumping) or both breasts at the same time (double pumping). The FIT Pro Breast Pump utilizes a DC-powered motor driving a diaphragmtype vacuum pump and an electromechanical solenoid which are controlled electronically to provide a range of user-selectable vacuum (suction) levels at various pre-determined cycle frequencies. The FIT Pro Breast Pump has a backlit LCD display, which shows pumping mode, suction level, timer, and battery level. The device also has four softtouch buttons allowing the user to power the device on/off, switch between stimulation and expression pumping modes, and control vacuum strength within each mode (6 levels of vacuum strength in stimulation mode and 12 levels in expression mode). The FIT Pro Breast Pump is intended for a single user in the home environment. When properly connected, the Hygeia PAS pumping kit transfers the vacuum generated by the powered pump to the breast, enabling expression and collection of milk. A diaphragm in the backflow protection assembly physically isolates pump and tubing from the space where milk is expressed and collected, protecting the breast milk from contamination. The base model of the FIT Pro Breast Pump contains a pump, pumping kit, AC adapter with detachable USB-C cable, and a lanyard. # Indications for Use The FIT Pro Breast Pump is a single-user, powered breast pump intended to express and collect milk from the breasts of lactating women. # Comparison of Intended Use and Technological Characteristics of the Subiect and Predicate Device The intended use and key technological characteristics of the FIT Pro Breast Pump and thepredicate device are compared side-by-side in the table below. {5}------------------------------------------------ | | Proposed Device | Predicate Device | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device name | FIT Pro Breast Pump | Mya Joy PLUS | | 510(k) Number | K212955 | K203570 | | Manufacturer | Hygeia II Medical Group,<br>Inc. | Ameda, Inc. | | Product Code | HGX | HGX | | Device Class | 2 | 2 | | Indications for<br>Use | The FIT Pro Breast Pump is a<br>single-user, powered breast<br>pump intended to express<br>and collect milk from the<br>breasts of lactating women. | The Ameda Mya Joy PLUS<br>breast pump is a single-<br>user, powered breast pump<br>intended to express and<br>collect milk from the<br>breasts of lactating women. | | Intended Use<br>Environment | Home Environment | Home Environment | | POWER | | | | Power Source<br>(external) | AC/DC adapter; 5VDC | AC/DC adapter; 5VDC | | Power Source<br>(internal) | rechargeable lithium-ion<br>battery<br>3.7 V, 1500 mAh | rechargeable lithium-ion<br>battery<br>3.7 V, 1500 mAh | | Battery<br>performance | At least 2 hours on full<br>charge | At least 2 hours on full<br>charge | | Auto Power-off | After 45 minutes in<br>expression mode | After 45 minutes in<br>expression mode | | VACUUM<br>PERFORMANCE | | | | Pump Type | Diaphragm | Diaphragm | | Pump Options | Single or Double | Single or Double | | Modes | Stimulation and Expression | Stimulation and Expression | | | Proposed Device | Predicate Device | | Vacuum levels | 6 (stimulation)<br>12 (expression) | 6 (stimulation)<br>12 (expression) | | Vacuum range | 40 – 270 mm Hg | 40 – 250 mm Hg | | Cycle range | 80 – 120 cycles/minute<br>(stimulation)<br>25 – 63 cycles/minute<br>(expression) | 80 – 120 cycles/minute<br>(stimulation)<br>25 – 63 cycles/minute<br>(expression) | | Cycle speed | Pre-programmed | Pre-programmed | | USER<br>INTERFACE | | | | LCD Display | Mode, Time, Vacuum Level,<br>Battery | Mode, Time, Vacuum Level,<br>Battery | | Button controls | Power, Mode, Increase,<br>Decrease | Power, Mode, Increase,<br>Decrease | | Pumping Kit | | | | Backflow<br>Protection | Circular silicone diaphragm | Cylindrical silicone<br>diaphragm | | Flange Design | Multi-piece | Single Piece | | Flange and<br>Bottle Material | Polypropylene | Polypropylene | | Valve Design | Duckbill | Duckbill | | Valve Material | Silicone | Silicone | | Single-pumping<br>plug shape | Truncated Cylindrical | Cylindrical | {6}------------------------------------------------ The subject and predicate device have similar indications for use statement and have the same intended use – to express and collect milk from lactating women. The subject and predicate devices are based on the same vacuum pump platform. The designs, both hardware and software, are identical. Both devices utilize identical lithium-ion batteries of equal capacity. The time before automated shut-off is identical in either device. While both subject and predicate devices share a vacuum range from 40 to 250 mm Hq, the subject device is capable of 270 mm Hg maximum vacuum in expression mode while double pumping. The pumping speeds for both the subject and predicate devices are identical at equivalent settings in either mode. {7}------------------------------------------------ The subject device has identical patient contacting materials when compared to the reference device, K190465. The principal differences in technological characteristics between the subject device and the predicate are limited to differences in the pumping kit. Similarities include shape of the breast shield, size and shape of the milk bottle, and valve material and design. Key differences are backflow protector shape, multi-piece (versus integrated) connector and the design of the pluq that allows single pumping. Functionally similar kit parts are made of the same type of material, either polypropylene or silicone. In the subject device, tubing points away from the user, while tubing in the predicate device rises vertically from the flange assembly. This difference in design does not affect pump performance and does not raise different questions of safety and effectiveness as compared to the predicate device. The circular diaphragm of the subject device flexes under vacuum in a manner different from the cylindrical diaphragm of the predicate. The multipiece connector assembly has more connecting surfaces than the predicate, which presents more areas of potential air leaks. The difference in plug shape for single pumping provides a controlled air leak, resulting in a slightly lower vacuum at any setting. These differences do not raise different questions of safety and effectiveness as compared to the predicate device. # Summary Of Non-Clinical Testing The FIT Pro Breast Pump complies with the following recognized voluntary standards for risk management, electrical safety, electromagnetic compatibility, and use in the home healthcare environment. The data below were provided in support of the substantial equivalence determination. Risk Analysis in accordance with ISO 14971:2007 Electrical safety testing per IEC 60601-1:2005 (3rd Edition) with US deviations per AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 standard Electromagnetic compatibility testing per IEC 60601-1-2: 2014 (4th edition) standard, Medical Electrical Equipment Part 1 - 2 # Use in the home healthcare environment per IEC 60601-1-11:2010 Software verification and validation was conducted in accordance with the 2005 FDA quidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" to support a moderate level of concern. {8}------------------------------------------------ # Non-clinical Performance Testing The FIT Pro Breast Pump was tested to demonstrate it meets stated performance specifications. Testing involved measurement of vacuum and cycle speed at userselectable settings in both stimulation and expression modes for pumping at a sinqle breast (single pumping) and both breasts simultaneously(double pumping). Testing was conducted separately under two states of power: (1) externally supplied by an AC/DC adapter and (2) internally supplied from a rechargeable lithium-ion battery. Specifications were met under all conditions. Testing confirmed device life and battery operating time, battery indicator accuracy, and verified protection against backflow and overflow. The test results for the non-clinical testing outlined above demonstrated that the FIT Pro Breast Pump met the predetermined acceptance criteria. # Conclusion The results of the performance testing described above demonstrate that the Fit Pro Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...