The MINUET LCD and MINUET LCD eco are multiple-user, powered breast pumps intended to express and collect milk from the breasts of lactating women.
Device Story
Portable, multiple-user, powered breast pump for home or healthcare settings; operates via microprocessor-controlled diaphragm pump and solenoid. Inputs: user-selected settings via control unit. Outputs: vacuum suction (50-280 mmHg) to express milk into collection bottles. Features two modes: massage (7 suction levels, 23-120 cycles/min) and expression (9 suction levels, 13-60 cycles/min). MINUET LCD includes rechargeable battery; MINUET LCD eco does not. Users operate device to express milk; output is collected milk. Benefits include efficient milk expression for lactating women.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including vacuum level/speed verification, backflow/cross-contamination protection, battery operation, and use-life testing. Biocompatibility (ISO 10993-5, -10), electrical safety (IEC 60601-1, -11), EMC (IEC 60601-1-2), and battery safety (IEC 62133) testing performed.
Technological Characteristics
Powered breast pump; microprocessor-controlled diaphragm pump and solenoid. Vacuum range 50-280 mmHg. Power: AC/DC adapter with micro USB or rechargeable Li-ion battery. Connectivity: None. Sterilization: Not specified. Software: Microprocessor-controlled. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 62133, AAMI/ANSI ES60601-1.
Indications for Use
Indicated for lactating women requiring expression and collection of breast milk.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 6, 2018
UNIMOM CO. Sang-Hyun Hong President 110-19 Gajangsaneopseobuk-ro Osan-si, Gyeonggi-do 18102 Korea
Re: K180075
Trade/Device Name: MINUET LCD, MINUET LCD eco Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: June 27, 2018 Received: July 5, 2018
Dear Sang-Hyun Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180075
Device Name MINUET LCD, MINUET LCD eco
Indications for Use (Describe)
The MINUET LCD and MINUET LCD eco are multiple-user, powered breast pumps intended to express and collect milk from the breasts of lactating women.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (K180075)
# 1. Submitter Information
| Submitter: | UNIMOM CO. |
|---------------|----------------------------------------------------------------------|
| Address: | 110-19, Gajangsaneopseobuk-ro, Osan-si,<br>Gyeonggi-do, 18102, KOREA |
| Contact Name: | Hong, Sang-Hyun |
| Telephone: | +82-70-8798-5577 |
| Fax: | +82-31-372-6782 |
| Email | infounimom@naver.com |
#### 2. Date Prepared: August 3, 2018
#### 3. Device Information
| Trade/Device Name | MINUET LCD, MINUET LCD eco |
|-------------------|-----------------------------|
| Common Name | Powered breast pump |
| Regulation Number | 21 CFR 884.5160 |
| Regulation Name | Powered Breast Pump |
| Product Code | HGX (Pump, Breast, Powered) |
| Regulatory Class | II |
#### 4. Predicate Device
Spectra S1 Plus and Spectra S2 Plus (K150476) manufactured by Uzinmedicare Co. The predicate device has not been subject to any design related recalls.
# 5. Device Description
The MINUET LCD and MINUET LCD eco are portable, multiple-user, powered breast pumps to be used in home or health care settings. These devices are the same, except the MINUET LCD has a rechargeable battery whereas the MINUET LCD eco does not. The devices have two major components: control unit (pump) and breast shield kit (breast shield, diaphragm, tubing, silicone massager, and bottle components). Users have the option of single or double pumping. The devices have two operating modes: massage and expression. The massage mode has seven suction levels whereas the expression mode offers nine suction levels. The vacuum strength range is 50-280 mmHq.
# 6. Indications for Use
The MINUET LCD and MINUET LCD eco are multiple-user, powered breast pumps intended to express and collect milk from the breasts of lactating women.
{4}------------------------------------------------
# 7. Comparison of Intended Use and Technological Characteristics of the Subject Device and Predicate Device
| Devices | K180075 (subject devices) | K150476 (predicate device) |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The MINUET LCD and MINUET LCD<br>eco are multiple-user, powered breast<br>pumps intended to express and<br>collect milk from the breasts of<br>lactating women. | The Spectra S1 Plus and Spectra S2 Plus<br>are single-user, powered breast pumps<br>intended to express and collect milk from<br>the breasts of lactating women. |
| Pump type | Same as the predicate | Diaphragm |
| Vacuum strength | Same as the predicate | 50-280 mmHg |
| Suction level | Massage Mode: 7 levels | Massage Mode: 5 levels |
| Suction level | Expression Mode: 9 Levels | Expression Mode: 12 Levels |
| Cycle speed | Massage Mode: 23-120 cycles/min<br>Expression Mode: 13-60 cycles/min | 38-70 cycles/min |
| Back flow protection | Same as the predicate | Backflow protector (diaphragm) on the top<br>of milk collection assembly |
| Visual indicator | Same as the predicate | LCD |
| Software control | Same as the predicate | Yes |
| Power supply - Mains | AC/DC adapter with micro USB cable | AC/DC converter |
| Power supply - Battery | Same as the predicate | Rechargeable lithium ion battery |
| Pump option | Same as the predicate | Single or double |
The subject and predicate devices have intended use – expressing mill from the breasts of lactating women. They also use the same fundamental technoloqy - a microprocessor-controlled diaphragm pump and a solenoid. They have the same vacuum strength range, backflow prevention mechanism, and pump options. They also have comparable power supplies.
The subject and predicate devices have different technological characteristics, as they have different suction levels under each operational mode. Also, the subject devices have high cycle speed up to 120 cycles/min under massage mode. These differences between the subject and predicate devices do not raise different questions of safety and effectiveness.
# 8. Summary of Non-Clinical Performance Testing
The following studies have been conducted on the subject devices to support substantial equivalence to the predicate device:
- Biocompatibility studies on the breast shield, as follows: .
- Cytotoxicity testing per 10993-5:2009 O
- Guinea Pig Maximization Sensitization testing per ISO 10993-10:2010 O
- Irritation testing per ISO 10993-10:2010 O
- Electrical safety testing in accordance with: .
- IEC 60601-1:2005+ CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) O
- IEC 60601-1-11:2015 O
{5}------------------------------------------------
- Electromagnetic compatibility testing per IEC 60601-1-2:2014 .
- Battery safety testing in accordance with: .
- AAMI/ANSI ES60601-1:2005 o
- IEC 62133:2012 o
- . Software verification and validation testing in accordance with FDA quidance: "The content of premarket submissions for software contained in medical devices" dated May 11, 2005
- . Bench tests using internal test protocols to demonstrate device performance, including vacuum level settings, speed settings, backflow/cross-contamination protection, battery operation, and use-life
#### 9. Conclusion
The subject and predicate devices have the same intended use. Although there are differences in technological characteristics between the subject and predicate devices, these differences do not raise different questions of safety or effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.