K161877 · First Quality Hygienic, Inc. · HEB · Aug 5, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K161877
Device Name
Opal by Femtex Pull & Snap and Private Label Unscented Compact Applicator Tampon, RegularOpal by Femtex Pull & Snap and Private Label Unscented Compact Applicator Tampon, Super, Opal by Femtex Pull & Snap and Private Label Unscented Compact Applicator Tampon, Super Pl
Applicant
First Quality Hygienic, Inc.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Aug 5, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.5470
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina.
Device Story
Opal by Femtex Pull & Snap is an unscented, compact menstrual tampon designed for over-the-counter use. The device consists of a compressed viscose rayon fiber pledget with an inserted 100% cotton withdrawal cord, covered by a nonwoven fiber overwrap. It includes a compact plastic applicator (polyethylene and polypropylene) with a barrel and two-piece plunger to facilitate insertion into the vagina. The device is used by the patient for personal hygiene to absorb menstrual or vaginal discharge. It functions as a passive absorbent barrier. The device is provided non-sterile.
Clinical Evidence
Bench testing only. Testing included Syngyna absorbency per 21 CFR 801.430 (all absorbencies met requirements), withdrawal cord attachment strength per AS/NZS 2869:1998 (all met ≥ 22.4N individual and ≥ 28N average requirements), fiber shedding (comparable to predicate), and extractables testing per USP 1663 (no safety concerns).
Technological Characteristics
Materials: Viscose rayon (pledget), 100% cotton (cord), PE/PET nonwoven overwrap, polyethylene (PE) and polypropylene (PP) applicator. Standards: 21 CFR 177-178 (resins/additives), 21 CFR 178.3297 (colorants), USP 1663 (extractables), AS/NZS 2869:1998 (cord strength). Form factor: Compact applicator with two-piece plunger. Non-sterile.
Indications for Use
Indicated for women for personal hygiene via intravaginal absorption of menstrual or other vaginal discharge. Available in Regular, Super, and Super Plus absorbencies.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Light, Regular, Super, and Super Plus (K151470)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2016
First Quality Hygienic, Inc. Melanie Leibowitz Director of Regulatory Affairs 121 North Road McElhattan, PA 17748
Re: K161877
> Trade/Device Name: Opal by Femtex Pull & Snap and Private Label Unscented Compact Applicator Tampon, Regular, Super, and Super Plus Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB Dated: July 6, 2016 Received: July 8, 2016
Dear Melainie Leibowitz,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Douglas Silverstein -S 2016.08.05 10:43:37 -04'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161877
### Device Name
Opal by Femtex Pull & Snap and Private Label Unscented Compact Applicator Tampon, Regular, Super, and Super Plus
Indications for Use (Describe)
Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for First Quality Hygienic. The logo features a stylized blue square with a smaller square inside, connected to the text "First Quality" in a bold, blue font. Below "First Quality" is the word "HYGIENIC" in a smaller, gray font.
### 510(k) Summary K161877
#### 1 Submitter's Information:
### Name and Address
First Quality Hygienic, Inc. 121 North Road McElhattan, PA 17748
### Establishment Registration Number
2529605
### Telephone Number and Fax Number:
+1 (516) 498-3665 Tel: Fax: +1 (516) 829-4949
### Contact Person:
Melanie Leibowitz Director of Regulatory Affairs
- 2 Date 510(k) Summary Prepared: 04/08/2016
#### Regulatory Description: 3
Trade Name: Opal by Femtex Pull & Snap and Private Label Unscented Compact Applicator Tampon, Regular, Super, and Super Plus Common Name: Unscented Menstrual Tampon Classification Name: Tampon, Menstrual, Unscented Regulation Description: Unscented Menstrual Tampon 21 CFR 884.5470 Regulation Number: Class: II Product Code: HEB (tampon, menstrual, unscented)
#### 4 Legally Marketed Device to Which Substantial Equivalence is Claimed:
Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Light, Regular, Super, and Super Plus (K151470)
#### 5 Description of the Device:
The Opal by Femtex Pull & Snap Unscented Compact Applicator Tampon is an unscented, compact menstrual tampon. It is intended for over-the-counter use for women's personal hygiene with respect to intra-vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina. The tampon itself is made of a strip of viscose rayon fibers in which a cord is inserted. A thin strip of nonwoven fiber covers the tampon, and the tampon is pressed firmly together. The applicator tubes are made of polyethylene (PE) and Polypropylene (PP) and include a barrel and a two-piece plunger. The assembled
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for First Quality Hygienic. The logo features a blue square with a white "Q" inside of it. To the right of the square is the text "First Quality" in a bold, blue font. Below that is the word "HYGIENIC" in a smaller, gray font.
> tampon with applicator is wrapped in a printed PE wrapper. The tampons are marketed in Regular, Super, and Super Plus absorbencies.
#### 6 Indications for Use:
### Indications for Use
Unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina.
These are the same Indications for Use as the Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Light, Regular, Super, and Super Plus (K151470) predicate device.
#### 7 Summary of Technological Characteristics Compared to the Predicate Device:
The Opal by Femtex Pull & Snap Compact Applicator Tampon is a modification of the Opal by Femtex and Private Label Unscented Compact Applicator Tampon K151470 device. The modifications do not change the technological characteristics of the predicate device in that they do not change the tampon pledget design, the tampon pledget materials or chemical composition, or the way the tampon pledget is physically constructed (i.e., formation of fiber, method of nonwoven application, and method of compression). The modifications include material and physical design changes to the applicator and minor modifications to the pledget weights and dimensions. The table below provides a detailed comparison.
| Characteristic | Opal by Femtex Pull & Snap and<br>Private Brand Unscented<br>Compact Applicator Tampon | Opal by Femtex and Private<br>Brand Unscented Compact<br>Applicator Tampon 510(k)<br>K151470 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Regulation | 884.5470 | 884.5470 |
| Product Code | HEB | HEB |
| Indications for Use | For Over-The-Counter Use:<br>Unscented tampon for women's<br>personal hygiene with respect to<br>intra vaginal absorption of<br>menstrual or other vaginal<br>discharge. The plastic applicator is<br>for easing the placement of the<br>tampon correctly into the vagina. | For Over-The-Counter Use:<br>Unscented tampon for women's<br>personal hygiene with respect to<br>intra vaginal absorption of<br>menstrual or other vaginal<br>discharge. The plastic applicator is<br>for easing the placement of the<br>tampon correctly into the vagina. |
| Sterile? | No | No |
### Subject and Predicate Device Technological Characteristic Comparisons
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for First Quality Hygienic. The logo features a blue square with a smaller square inside of it. To the right of the square is the text "First Quality" in a stylized font, with the word "Hygienic" underneath in a smaller font. The colors used in the logo are blue and white.
| | Tampon Design | Digital tampon with round tip and<br>straight grooves | Digital tampon with mushroom<br>shape tip and straight grooves |
|----------------------------|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Applicator Design | Compact style applicator with<br>cylindrical barrel with finger grip<br>and two-piece plunger | Compact style applicator with<br>cylindrical barrel and one-piece<br>plunger |
| Absorbency | Light | N/A | ≤ 6 g |
| | Regular | 6.0 - 9.0 g | 6.0 - 9.0 g |
| | Super | 9.0 - 12.0 g | 9.0 - 12.0 g |
| | Super Plus | 12.0 - 15.0 g | 12.0 - 15.0 g |
| | Absorbent Material | Viscose Rayon | Viscose Rayon |
| Nonwoven Overwrap Material | | PE/PET Nonwoven—Conforms<br>with 21 CFR 177-178 | |
| | Withdrawal Cord | 100% Cotton | 100% Cotton |
| | Applicator Tube | Toxic Substance Control Act Inventory (TSCA) Listed Resins and additives are 21 CFR 177-178 compliant Colorants are 21 CFR 178.3297 compliant USP 1663 Extraction testing confirmed that no harmful compounds are bioavailable | Toxic Substance Control Act Inventory (TSCA) Listed Resins and additives are 21 CFR 177-178 compliant Colorants are 21 CFR 178.3297 and/or EU Resolution (89)1 compliant |
| | Primary Packaging (i.e.<br>wrapper) | Printed PE Toxic Substance Control Act Inventory (TSCA) Listed Resins and additives are 21 CFR 177-178 compliant Colorants are 21 CFR 178.3297 compliant | Printed PP Toxic Substance Control Act Inventory (TSCA) Listed Resins and additives are 21 CFR 177-178 compliant Colorants are 21 CFR 178.3297 compliant |
#### 8 Non-Clinical Performance Data:
The following testing was performed:
### Syngyna Absorbency
Absorbency testing for the new device was performed per the 21 CFR 801.430 for Syngina absorbency testing, and it was confirmed that the Regular, Super, and Super Plus absorbencies met the absorbency range requirements defined by 21 CFR 801.430.
### Extractables Testing and Biocompatibility Review
Extractables testing was performed on all applicator components per <USP 1663> Assessment of Extractables Associated with Pharmaceutical Packaging/Delivery Systems. Results of Extraction testing did not raise safety concerns.
### Withdrawal cord attachment per AS 2869-2008, Tampons-Menstrual
Withdrawal cord attachment strength testing for the new device was performed per
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for First Quality Hygienic. The logo features a blue square with a white "Q" inside of it. To the right of the square is the text "First Quality" in a bold, blue font. Below the text "First Quality" is the word "HYGIENIC" in a smaller, gray font.
> Australian Standard AS/NZS 2869:1998, Tampons-Menstrual. Each tampon tested for every absorbency level met the Standard's minimum attachment strength requirement of 22.4 Newtons (N). Each absorbency data set met an average value of ≥ 28N, which is the Standard's requirement for each production lot.
### Fiber shedding
Fiber shed testing for the subject device demonstrate that it does not exhibit the potential to shed more fiber than the predicate device.
#### 9 Conclusion:
The performance testing demonstrate that the Opal by Femtex Pull & Snap Unscented Compact Applicator Tampon is substantially equivalent to the Opal by Femtex and Private Brand Unscented Compact Applicator Tampon.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.