The Tosama 100% Organic Cotton Menstrual Tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.
Device Story
Device is an unscented menstrual tampon consisting of an absorbent 100% organic cotton pledget, a withdrawal cord, and a thermoplastic olefin (TPO) applicator. Designed for vaginal insertion to absorb menstrual or vaginal discharge. Available in Regular, Super, and Super Plus absorbencies. Intended for over-the-counter use by patients. Operates via passive absorption of fluid. Withdrawal cord treated with Sevophob (paraffin-based resin) to prevent wicking. Safety and efficacy established through non-clinical bench testing, including biocompatibility (ISO 10993), microbiological testing, and Syngina absorbency testing per 21 CFR 801.430(f)(2).
Clinical Evidence
Bench testing only. No clinical data. Testing included chemical analysis (pesticides, dioxins, phthalates, PCBs), microbiological testing (Ph. Eur. 2.6.12/2.6.13, TSST-1, S. aureus growth), biocompatibility (ISO 10993-5, -10, -11), and efficacy testing (Syngina absorbency, fiber loss, pull-out resistance, tampon integrity).
Indicated for females requiring absorption of menstrual or other vaginal discharge.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure in profile, with three faces overlapping to symbolize health, human services, and the relationship between them. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 2, 2015
TOSAMA, d.o.o. % Dave Yungvirt CEO Third Party Review Group, LLC 45 Rockefeller Plaza. Suite 2000 New York, NY 10111
Re: K151170
> Trade/Device Name: Tosama 100% Organic Cotton Menstrual Tampon Regulation Number: 21 CFR 884.5470 Regulation Name: Unscented menstrual tampon Regulatory Class: II Product Code: HEB Dated: June 16, 2015 Received: June 18, 2015
Dear Dave Yungvirt,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151170
Device Name
Tosama 100% Organic Cotton Menstrual Tampon
Indications for Use (Describe)
The Tosama 100% Organic Cotton Menstrual Tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
DATE OF PREPARATION: July 1, 2015
| COMPANY/OWNER: | TOSAMA, d.o.o.<br>Saranoviceva cesta 35, Vir<br>SI-1230 Domzale<br>SLOVENIA, EU |
|----------------|---------------------------------------------------------------------------------|
| CONTACT: | Marjetka Kralj Kuncic, PhD<br>Head of Quality Control |
| TELEPHONE: | 00386 1 7290370 |
| FAX: | 00386 1 7291244 |
| EMAIL: | marjetka.kralj-kuncic@tosama.si |
Tosama 100% Organic Cotton Menstrual Tampon DEVICE TRADE NAME: Unscented Menstrual Tampon COMMON NAME: Tampon, Menstrual, Unscented CLASSIFICATION NAME: REGULATION NUMBER: 21 CFR §884.5470 PRODUCT CODE: HEB DEVICE CLASS: =
PREDICATE DEVICE: Naturalena 100% Cotton Tampon (K130609)
DEVICE DESCRIPTION: The device will be offered as a traditional unscented menstrual 100% organic cotton tampon consisting of an absorbent pledget, a withdrawal cord, and an applicator. The pledget is of the traditional cylindrical, bullet-like shape and the applicator has a standard rounded tip to ease insertion. Each tampon is individually wrapped and packaged in multi-unit containers for retail sale. It will be offered in three absorbencies: Regular, Super, and Super Plus.
INDICATION FOR USE: The Tosama 100% Organic Cotton Menstrual Tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.
## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The Tosama 100% Orqanic Cotton Menstrual Tampon demonstrates substantial equivalence to the Naturalena 100% Cotton Tampon (K130609). Table 5-1 summarizes the key technological characteristics and features of both the predicates and the new device.
{5}------------------------------------------------
| Feature | Tosama 100%<br>Organic Cotton<br>Menstrual<br>Tampon | Predicate<br>Naturalena<br>100% Cotton<br>Tampon |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | K130609 |
| Intended Use | The device is<br>intended for insertion<br>into the vagina for the<br>absorption of<br>menstrual or other<br>vaginal discharge. | The device is<br>intended for insertion<br>into the vagina for the<br>absorption of<br>menstrual or other<br>vaginal discharge. |
| Design | Tampon with<br>cylindrical shape and<br>bullet-like tip.<br>Applicator with<br>smooth, rounded tip. | Tampon with<br>cylindrical shape and<br>bullet-like tip.<br>Applicator with<br>smooth, rounded tip. |
| Dimensions &<br>Weights | Digital Tampons | Digital Tampons |
| Description: | Regular Digital | Regular Digital |
| Length: | 43.0mm – 46.0mm | 43.0mm – 46.0mm |
| Diameter: | 11.0mm – 12.0mm | 11.0mm – 12.0mm |
| Mass: | 2.30g – 2.70g | 2.30g – 2.70g |
| Withdrawal Cord: | 130mm – 160mm | 130mm – 160mm |
| Syngina Absorbency: | 6.0g – 9.0g | 6.0g – 9.0g |
| Description: | Super Digital | Super Digital |
| Length: | 48.0mm – 51.0mm | 48.0mm – 51.0mm |
| Diameter: | 12.0mm – 13.0mm | 12.0mm – 13.0mm |
| Mass: | 2.70g – 3.10g | 2.70g – 3.10g |
| Withdrawal Cord: | 130mm – 160mm | 130mm – 160mm |
| Syngina Absorbency: | 9.0g – 12.0g | 9.0g – 12.0g |
| Description: | Super Plus Digital | Super Plus Digital |
| Length: | 48.0mm – 51.0mm | 48.0mm – 51.0mm |
| Diameter: | 14.0mm – 15.0mm | 14.0mm – 15.0mm |
| Mass: | 3.30g – 3.80g | 3.30g – 3.80g |
| Withdrawal Cord: | 130mm – 160mm | 130mm – 160mm |
| Syngina Absorbency: | 12.0g – 15.0g | 12.0g – 15.0g |
| Dimensions &<br>Weights | Applicator<br>Tampons | Applicator<br>Tampons |
| Description: | Regular Applicator | Regular Applicator |
| Tampon Length: | 43.0mm – 46.0mm | 43.0mm – 46.0mm |
| Applicator Length: | 120mm ± 5% | 120mm ± 5% |
| Tampon Diameter: | 11.0mm – 12.0mm | 11.0mm – 13.0mm |
| Applicator Diameter: | 13mm | 16mm |
| Tampon Mass: | 2.00g – 2.30g | 2.00g – 2.30g |
| Applicator Mass: | 3.40g – 3.60g | 3.40g – 3.60g |
| Total Mass: | 5.40g – 5.90g | 5.40g – 5.90g |
| Withdrawal Cord: | 130mm – 160mm | 130mm – 160mm |
| Feature | Tosama 100%<br>Organic Cotton<br>Menstrual<br>Tampon | Predicate<br>Naturalena<br>100% Cotton<br>Tampon |
| Syngina Absorbency: | 6.0g - 9.0g | 6.0g - 9.0g |
| Description: | Super Applicator | Super Applicator |
| Tampon Length: | 43.0mm - 46.0mm | 48.0mm - 51.0mm |
| Applicator Length: | 120mm ± 5% | 120mm ± 5% |
| Tampon Diameter: | 12.0mm - 13.0mm | 13.0mm - 15.0mm |
| Applicator Diameter: | 16mm | 19mm |
| Tampon Mass: | 2.60g - 3.00g | 2.60g - 3.00g |
| Applicator Mass: | 4.10g - 4.30g | 4.20g - 4.40g |
| Total Mass: | 6.70g - 7.30g | 6.80g - 7.40g |
| Withdrawal Cord: | 130mm - 160mm | 130mm - 160mm |
| Syngina Absorbency: | 9.0g - 12.0g | 9.0g - 12.0g |
| Description: | Super Plus Applicator | Super Plus Applicator<br>Not included in<br>submission. |
| Tampon Length: | 43.0mm - 46.0mm | |
| Applicator Length: | 120mm ± 5% | |
| Tampon Diameter: | 14.0mm - 15.0mm | |
| Applicator Diameter: | 18mm | |
| Tampon Mass: | 3.00g - 3.50g | |
| Applicator Mass: | 4.20g - 4.40g | |
| Total Mass: | 7.20g - 7.90g | |
| Withdrawal Cord: | 130mm - 160mm | |
| Syngina Absorbency: | 12.0g - 15.0g | |
| Materials | All Products | All Products |
| Pledget: | 100% Organic Cotton | 100% Cotton |
| Withdrawal Cord: | 100% Organic Cotton | 100% Cotton |
| Applicator: | TPO | Polyethylene |
| Individual Wrapping: | Polypropylene | Polypropylene |
| Additives &<br>Finishing<br>Agents | Hydrophobic<br>withdrawal cord. | Hydrophobic coating<br>on withdrawal cord. |
Table 5-1: Comparison of Proposed Device with Predicate Devices
{6}------------------------------------------------
The proposed Tosama 100% Organic Cotton Menstrual Tampon and the Naturalena predicate device have the following two differences:
- Naturalena uses Repellan T anti-wicking agent in its withdrawal cord, while the 1. proposed device uses Sevophob. Both are a paraffin-based resin used to ensure the cord does not wick menstrual fluid out of the vaqina.
- 2. Naturalena uses a polyethylene plastic for the applicator material, while the proposed device uses a thermoplastic olefin plastic for the applicator material.
The proposed device is essentially identical to the Naturalena predicate device other than the differences listed above. It has been demonstrated through biocompatibility and performance testing that neither of the two differences have an impact on the safety and effectiveness of the device. Although the addition of the wicking agent to the withdrawal cord could potentially impact the safety (biocompatibility) of the
{7}------------------------------------------------
device, it has been confirmed as safe through appropriate biocompatibility testing. SUMMARY OF NON-CLINICAL DATA:
All materials used in the construction of Tosama 100% Organic Cotton Menstrual Tampons as well as complete devices have been subjected to the following tests in accordance with recommended requirements and proper consideration of its intended use:
- · Chemical analysis for pesticide, dioxin, phthalate, and PCB residues;
- Microbiological testing for enumeration per Ph. Eur. 2.6.12/2.6.13. Toxic Shock Syndrome Toxin-1 production, vaginal microflora growth alteration, and S. aureus growth enhancement:
- · Biocompatibility testing of the pledget as follows:
- o ISO 10993-5:2009: Tests for in vitro cytotoxicity;
- ISO 10993-10:2010: Tests for irritation and skin sensitization;
- o ISO 10993-11:2006: Tests for systemic toxicity;
- · Biocompatibility testing of the applicator as follows:
- o ISO 10993-5:2009: Tests for in vitro cytotoxicity;
- o ISO 10993-10:2010: Tests for irritation and skin sensitization:
- Efficacy testing according to the FDA Guidance on Menstrual Tampons and Pads:
- Syngina absorbency according to 21 CFR 801.430(f)(2); o
- fiber loss: o
- withdrawal cord pull-out resistance; o
- o tampon integrity.
The results of these tests support the conclusion that the proposed device is safe for its intended use.
## CONCLUSIONS:
The intended use and indications for use, as well as the safety and effectiveness, of Tosama 100% Organic Cotton Menstrual Tampons and that of the Naturalena predicate device as unscented menstrual tampons are identical. Chemical, biocompatibility, microbiological, and efficacy testing show this to be verifiable. Thus, the Tosama 100% Organic Cotton Menstrual Tampon is substantially equivalent to the legally marketed predicate device Naturalena 100% Cotton Tampon (K130609).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.