K090819 · Ontex International NV · HEB · Dec 23, 2009 · Obstetrics/Gynecology
Device Facts
Record ID
K090819
Device Name
ONTEX TAMPON (UNSCENTED)
Applicant
Ontex International NV
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Dec 23, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
Ontex Tampon is a tampon that is inserted into the vagina and used to absorb menstrual fluid. The intended use of the organic cotton tampon is the same as all other products that are legally marketed.
Device Story
Ontex Tampon is a conventional unscented menstrual tampon, available in digital and applicator versions. The device consists of an absorbent pledget made of 100% organic cotton, with a 100% organic cotton withdrawal cord and non-woven cover. Applicator versions utilize a cardboard applicator. The device is intended for over-the-counter use by females to absorb menstrual fluid. It functions by insertion into the vagina, where the absorbent material collects fluid. Safety and effectiveness are established through bench testing, including Syngina absorbency rate testing, expulsion force testing for applicators, fiber loss testing, and stability checks for digital tampons.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including Syngina absorbency rate, expulsion force for applicator tampons, fiber loss (ATS testing method), and stability checks for digital tampons.
Technological Characteristics
Materials: 100% organic cotton wadding, non-woven cover, and withdrawal cord. Applicator: Cardboard. Dimensions: Applicator tampons (45-50mm length, 14.2-15.7mm diameter); Digital tampons (42-26mm length, 11.8-12.2mm diameter). Absorbency: 9.0-12.0g (applicator) and 6.0-9.0g (digital) via Syngina test.
Indications for Use
Indicated for use by females for the absorption of menstrual fluid.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
{0}------------------------------------------------
## DEC 2 3 2009
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMVDA 1990 and 21 CFR §807.92.
510(k) number:
1. Submitter's Identification:
Ontex International NV Spinnerijstraat 12 9240 Zele Belgium Submitter Phone: 0032 / 52 454 651 Submitter Fax: 0032 /52 454 610 Submitter Contact: Steven Smet Title:
2. Name of the Device: Digital and Applicator Menstrual Tampons Ontex Tampons (Unscented)
Device Class: II ·gulation: 884.5470 r roduct Code: HEB
#### 3. Predicate Device Information:
Maxim Hygeine Organically Grown Cotton Tampon 510(k) Number: K090098 Maxim Hygeine Products. 39 Maple Street Roslyn Heights, NY 11577-1941
### 4. Device Description:
The tampons to be marketed are conventional unscented menstrual tampons consisting of an absorbent pledget, with and without applicator.
#### 5.0 Device Comparison to Predicate Device(s):
The following tests were performed to support substantial equivalence (see Appendix 2 - Testing):
- . Determination of Absorbency Rate of Tampons = Syngina Test
- Expulsion Force Applicator Tampons
- . Fiber Loss ATS Testing Method
- . Stability Check on Digital Tampons
{1}------------------------------------------------
# Applicator Tampon
| Parameters | Ontex International | Maxim Hygiene<br>Products |
|--------------------------------|---------------------|--------------------------------------|
| 510(k) Number | | K090098 |
| Dimensions | | |
| Total Weight | 5,0-5,9g | 5,0-5,9g |
| Weight without<br>applicator | 2,7-3,2g | 2,7-3,2g |
| Withdrawal Cord | 115-175 mm | 115-175 mm |
| Length with<br>Applicator | 120-125 mm | 120-125 mm |
| Length without Applicator | 45-50 mm | 45-50 mm |
| Diameter with Applicator | 15,9-16,1 mm | 15,9-16,1 mm |
| Diameter without Applicator | 14,2-15,7 mm | 14,2-15,7 mm |
| Syngina Absorption | 9,0-12,0g | 9,0-12,0g |
| Wadding | 100% organic cotton | 100% organic cotton |
| Non Woven | 100% organic cotton | 100% organic cotton |
| Withdrawal Cord | 100% organic cotton | 100% organic cotton |
| Applicator | Cardboard | Cardboard |
| Parameters | Ontex International | Maxim Hygiene<br>Products<br>K090098 |
| Digital Tampon | | |
| Dimensions | | |
| Weight with Single Packaging | 2,1-2,5g | 2,1-2,5g |
| Weight Tampon | 2,0-2,4g | 2,0-2,4g |
| Length with Single Packaging | 42-26 mm | 42-26 mm |
| Diameter with Single Packaging | 11,8-12,2 mm | 11,8-12,2 mm |
| Withdrawal Cord | 130-160 mm | 130-160 mm |
| Syngina Absorption | 6,0-9,0g | 6,0-9,0g |
| Wadding | 100% organic cotton | 100% organic cotton |
| Withdrawal Cord | 100% organic cotton | 100% organic cotton |
{2}------------------------------------------------
## Digital Tampon
#### Conclusions:
The subject device has the same intended use and technical characteristics as the predicate device. Similarly, there are no differences in materials used to fabricate the subject device and the predicate device. Therefore no new questions of safety or effectiveness are raised by this submission. Thus, the Ontex Tampon (Unscented) is substantially equivalent to the predicate device, the Maxim Hygeine Products Tampon (K090098).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ontex International NV % Mr. Neal Kolber Project Manager Emergo Group 1705 S. Capital of Texas Highway, Suite 500 AUSTIN TX 78746
DEC 2 3 2009
Re: K090819
Trade/Device Name: Ontex Tampon (Unscented) Regulation Number: 21 CFR §884.5470 Regulation Name: Menstrual tampon Regulatory Class: II Product Code: HEB Dated: November 30, 2009 Received: December 1, 2009
Dear Mr. Kolber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janina M. Morris
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{5}------------------------------------------------
## INDICATIONS FOR USE
090819
## 510(k) Number (if known):
Device Name: Ontex Tampon (Unscented)
## Indications for Use:
Ontex Tampon is a tampon that is inserted into the vagina and used to absorb menstrual fluid.
The intended use of the organic cotton tampon is the same as all other products that are legally marketed.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
hopinothy
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Dev 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.