FEMTEX JUNIOR/SLENDER REGULAR OPEN-END TAMPONS

K031067 · First Quality Hygienic, Inc. · HEB · Jul 14, 2003 · Obstetrics/Gynecology

Device Facts

Record IDK031067
Device NameFEMTEX JUNIOR/SLENDER REGULAR OPEN-END TAMPONS
ApplicantFirst Quality Hygienic, Inc.
Product CodeHEB · Obstetrics/Gynecology
Decision DateJul 14, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5470
Device ClassClass 2

Indications for Use

Femtex Open-End Junior and Slender Regular Tampons are menstrual tampons that are inserted into the vagina and used to absorb menstrual fluid.

Device Story

Menstrual tampons; used to absorb menstrual fluid or vaginal discharge. Device consists of absorbent plug made from rayon and cotton, with cotton cord and sewing thread. Inserted into vagina by user. No electronic, mechanical, or software components. Device functions via passive absorption of fluid. Used in home setting by consumer.

Clinical Evidence

Bench testing only. Biocompatibility testing (irritation, oral toxicity, cytotoxicity) performed on components. Standard Syngyna testing conducted to confirm absorbency. No clinical trials performed.

Technological Characteristics

Materials: rayon absorbent, cotton cord, cotton sewing thread. Form factor: plug for vaginal insertion. Principle: passive fluid absorption. No energy source, connectivity, or software.

Indications for Use

Indicated for use as a menstrual tampon, inserted into the vagina to absorb menstrual or other vaginal discharge. Intended for use by individuals requiring menstrual hygiene products.

Regulatory Classification

Identification

An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## JUL 1 4 2003 ## K031067 510K SUMMARY Device Name: Private Label and Femtex Open-End Tampons Legally marketed device: These Femtex Junior and Slender Regular Tampons are substantially equivalent to legally marketed Femtex tampons with open-end applicators. Device description: Femtex Junior and Slender Regular Tampons are menstrual tampons used to absorb menstrual fluid. These Femtex Open-End Tampons will be provided with 2 absorbencies: Slender Regular and Junior. Femtex Open-End Junior and Slender Regular Tampons are made from rayon absorbent, cotton cord and cotton sewing thread. The material used in Femtex Open-End Junior and Slender Regular Tampons are similar to those used in other legally marketed tampons. Intended Use: Femtex Open-End Junior and Slender Regular Tampons are menstrual tampons that are inserted into the vagina and used to absorb menstrual fluid. ## Assessment of Performance Standards: Not Applicable Non-Clinical Testing: Biocompatability testing of the Femtex Junior and Slender Regular Tampons' components was reviewed. The results of these tests demonstrate that the Femtex Tampons are equivalent in terms of safety and effectiveness to legally marketed tampons. Standard Syngyna testing confirmed the absorbency of the Femtex Tampons. In addition to the review of existing toxicological data in the public literature, the following tests have been conducted on components relevant to the safety of Femtex tampons. - ◇ irritation testing - 0 oral toxicity testing - cytotoxicity testing � *** Image /page/0/Picture/14 description: The image shows a series of zeros, specifically six zeros in a row. The zeros are written in a slightly irregular font, giving them a somewhat handwritten appearance. The image is in black and white, with the zeros appearing as dark shapes against a light background. {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## JUL 1 4 2003 First Quality Hygienic, Inc. % Robert J. Stabb. Ph.D. Official Correspondent Regulatory and Technical Associates 73 Franklin TPK. ALLENDALE NJ 07401 Re: K031067 Trade/Device Name: First Quality Hygienic and Other Private Label Slender Regular and Junior Tampons Regulation Number: 21 CFR 884.5470 Regulation Name: Unscented menstrual tampons Regulatory Class: II Product Code: 85 HEB Dated: April 2, 2003 Received: May 30, 2003 Dear Dr. Stabb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ | Ver/ 3 - 4/24/96 | K031067 | |---------------------------|--------------------------------------------------------------------------------------| | Applicant: | First Quality Hygienic | | 510(k) Number (if known): | applied for | | Device Name: | First Quality Hygienic and other Private Label<br>Slender Regular and Junior tampons | | Indications For Use: | | As a Class II device, the menstrual tampon is defined as follows (21CFR884.5460 and 21CFR884.5470) First Quality Hygienic slender regular and Junior tampons are a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other v vaginal discharge (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Per 21 CFR 801.109) (Optional Format 1-2-96) David A. Symm (Division Sig ar 510k Over-The-Counter Use
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...