K040238 · Modern Medical Equipment Mfg., Ltd. · LKF · Jun 22, 2004 · Obstetrics/Gynecology
Device Facts
Record ID
K040238
Device Name
SOFTSEAL HSG CATHETER
Applicant
Modern Medical Equipment Mfg., Ltd.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Jun 22, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
Softseal Catheter (a uterine manipulator/injector) is used to access the uterine cavity for the injection and retention of contrast fluid (or gas) into the uterus to enable the visualization of the uterine cavity, not for intrafallopian procedure. The following clinical indications are some of the applications: 1. Infertility 2. Menorrhagia 3. Dysmenorrhca Due to Uterine Synechia 4. Suspected Endometrial Polyp 5. Submucous Fibroid 6. Uterine Synechia (Asherman Syndrome) 7. Congenital Uterine Abnormality
Device Story
Softseal HSG Catheter is a uterine manipulator/injector used to access the uterine cavity. Device facilitates injection and retention of contrast media or gas to enable visualization of the uterine cavity during diagnostic procedures. Operated by clinicians in a clinical setting. Device provides a pathway for contrast delivery; aids in diagnosing uterine pathologies such as fibroids, polyps, and congenital abnormalities. Benefits include improved diagnostic visualization of the uterus for gynecological assessment.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Uterine manipulator/injector catheter. Mechanical device for fluid/gas delivery. No software or electronic components.
Indications for Use
Indicated for patients requiring uterine cavity visualization via contrast fluid or gas injection for conditions including infertility, menorrhagia, dysmenorrhea due to uterine synechia, suspected endometrial polyps, submucous fibroids, Asherman syndrome, and congenital uterine abnormalities. Not indicated for intrafallopian procedures.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
K022503 — UTERINE INJECTOR · Thomas Medical, Inc. · Oct 25, 2002
K041094 — HSG CATHETER SET · Bioteque America, Inc. · Jun 17, 2004
K063611 — SALINE/ CONTRAST MEDIA INFUSION CATHETER · Utah Medical Products, Inc. · May 23, 2007
K023051 — INTRAUTERINE DELIVERY SYSTEM - INJECTOR · Prosurg, Inc. · Dec 12, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUN 2 2 2004
Al Muzzammel, MD Official Correspondent Modern Medical Equipment Mfg., Ltd. E & M Engineering 1705 Dabney Road RICHMOND VA 23230
## Re: K040238
Trade/Device Name: Softseal HSG Catheter Regulation Number: None Regulatory Class: Unclassified Product Code: 85 LKF Dated: June 2, 2004 Received: June 4, 2004
Dear Dr. Muzzammel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are carsion to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that a Printeral statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product qadiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and my you to ough finding of substantial equivalence of your device to a legally premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spoolite at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Addisonaliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding Other of Comphanes at (2017-2211) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small miormation on your responsional Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 1. Indications for Use
510(k) Number (if known): K040238
Device Name: Softseal HSG Catheter
## Indications For Use:
Indications I VI OSC.
Softseal Catheter (a uterine manipulator/injector) is used to access the uterine cavity for Softseal Catheter (a atention of contrast fluid (or gas) into the uterus to enable the the injection and retemator of contrast not for intrafallopian procedure. The perforing clinical indications are some of the applications
- 1. Infertility
- 2. Menorrhagia
- 3. Dysmenorrhca Due to Uterine Synechia
- 4. Suspected Endometrial Polyp
- 5. Submucous Fibroid
- 5. Submutorus I lorean Uterine Synechia (Asherman Syndrome)
- 7. Congenital Uterine Abnormality
Prescription Use
> AND/OR Over-The Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1 .
Daniel G. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.