SWIFT DELIVERY PRODUCT #002-VAC EX CUP W/VAC REL VALVE
Device Facts
| Record ID | K970170 |
|---|---|
| Device Name | SWIFT DELIVERY PRODUCT #002-VAC EX CUP W/VAC REL VALVE |
| Applicant | Richard G. Lindsay |
| Product Code | HDB · Obstetrics/Gynecology |
| Decision Date | Mar 28, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.4340 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Indications for Use
Vacuum Extractor Cup used in child birth
Device Story
Vacuum extractor cup used during childbirth; assists physician in vaginal delivery. Device features modified vacuum tubing attachment point and vacuum release valve location compared to predicate. Design changes intended to reduce cup detachment (pop-offs), improve physician control, and enable single-handed vacuum regulation. Used by obstetricians in clinical delivery settings. Benefits include increased safety for neonate and reduced delivery time for mother.
Clinical Evidence
No clinical data. Non-clinical performance evidence consists of a physician survey across nine states confirming the need for design improvements.
Technological Characteristics
Vacuum extractor cup and stem assembly. Mechanical device; no energy source or software. Design modifications focus on ergonomic placement of vacuum tubing and release valve.
Indications for Use
Indicated for use in childbirth to assist in vaginal delivery.
Regulatory Classification
Identification
A fetal vacuum extractor is a device used to facilitate delivery. The device enables traction to be applied to the fetal head (in the birth canal) by means of a suction cup attached to the scalp and is powered by an external vacuum source. This generic type of device may include the cup, hosing, vacuum source, and vacuum control.
Predicate Devices
- Mityvac Vacuum Extractor with release valve (K890307)
Related Devices
- K973942 — SWIFT DELIVERY PRODUCT #300 · Swift Delivery Products · Dec 12, 1997
- K963287 — VACUUM CUP (004SD) · Swift Delivery Products · Oct 28, 1996
- K980976 — AMEDA DISPOSABLE VACUUM ASSIST CUP WITH FLUID RETENTION TRAP · Hollister, Inc. · Aug 21, 1998
- K980212 — MITYVAC · Prism Technologies, Inc. · Jul 9, 1998
- K021193 — FORTUNE SILICONE VACUUM SUCTION, MODEL #1300 · Fortune Medical Instrument Corp. · Feb 12, 2003