MITYVAC MERLIN, MODEL 10027

K020447 · Prism Enterprises, Inc. · HDB · Apr 12, 2002 · Obstetrics/Gynecology

Device Facts

Record IDK020447
Device NameMITYVAC MERLIN, MODEL 10027
ApplicantPrism Enterprises, Inc.
Product CodeHDB · Obstetrics/Gynecology
Decision DateApr 12, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.4340
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Merlin™ System and the MityOne™ System are intended to be used to facilitate the delivery of the fetus during childbirth. These devices are indicated for use during vaginal delivery and caesarean sections.

Device Story

Obstetrical vacuum assist delivery system; combines vacuum cup and pressure source into single unit for single-clinician operation. Input: manual activation of CO2 canister to generate negative pressure. Operation: clinician inserts cup through vaginal canal or transverse incision; attaches to fetal scalp flexion point; applies negative pressure and traction to assist delivery. Output: mechanical traction on fetal head. Used in clinical settings (labor/delivery) by physicians/clinicians. Benefits: facilitates delivery; allows single-operator use.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and design similarity to predicate.

Technological Characteristics

Fetal vacuum extractor; CO2 canister-powered vacuum source; integrated cup and pump unit; single-use design; sterile packaging.

Indications for Use

Indicated for use by clinicians to assist in infant delivery during vaginal or Cesarean section childbirth.

Regulatory Classification

Identification

A fetal vacuum extractor is a device used to facilitate delivery. The device enables traction to be applied to the fetal head (in the birth canal) by means of a suction cup attached to the scalp and is powered by an external vacuum source. This generic type of device may include the cup, hosing, vacuum source, and vacuum control.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K020447 # APR 1 2 2002 ### Attachment 9 ## 510(k) SUMMARY FOR PRISM ENTERPRISES LP MITYVAC® MERLIN™ Vacuum Assist Delivery System ## Submitter's Name, Address, Telephone Number, and Contact Person Prism Enterprises LP 6952 Fairgrounds Parkway San Antonio, TX 78238-4528 | Contact: | Frances D. Menard | |------------|----------------------| | | Prism Enterprises LP | | Phone: | (210) 256-3113 | | Facsimile: | (210) 509-7463 | Merle M. Smith or Prism Enterprises LP (210) 520-8051 (210) 520-8039 #### Date Prepared February 8, 2002 #### Name of the Device Mityvac® Merlin™ #### Common or Usual Name Obstetrical Vacuum Assist Delivery System #### Classification Name Fetal Vacuum Extractor (21 C.F.R. § 884.4340) #### Product Code HDB ### Predicate Devices The Mityvac® Merlin™ System is a modification of the Mityvac® MityOne™ Obstetrical Vacuum Assist Delivery System with M-Style® or MitySoft® ("MityOne™"). {1}------------------------------------------------ K020447 #### Intended Use The Merlin™ System and the MityOne™ System are intended to be used to facilitate the delivery of the fetus during childbirth. These devices are indicated for use during vaginal delivery and caesarean sections. Thus, Merlin™ and MityOne™ have the same intended use and the same indications. 2 ### Principles of Operation The Mityvac® Merlin™ and MityOne™ combine the cup and pump into one unit that allows one clinician to operate the system without an assistant. First, the user opens the sterile package and removes the Merlin™. Next, the user checks the integrity of the system by pressing the cup to the palm of a gloved hand and applying vacuum. Then, the user examines the fetus' presentation and position and inserts the extractor cup through the vaginal canal during vaginal delivery and through the transverse incision during caesarean delivery and attaches it to the flexion point on the fetus' scalp. After confirming that neither maternal tissue nor the placenta is under the cup, the user supplies the negative pressure and applies traction. The user releases vacuum pressure when either the fetus' head is delivered or the fetus' head has not been delivered but certain events have occurred. The user then removes the cup and discards the system per hospital protocol. Thus, the principles of operation of Mityvac® Merlin™ System and Mityvac MityOne™ System are very similar. The only difference in the principle of operation is that the user of the Merlin™ must puncture the device's Carbon Dioxide ("CO2") canister to activate its pressure source. ### Technical Characteristics The Mityvac® Merlin™ System is a modification of the cleared Mityvac® MityOne™ System with M-Style® or MitySoft® Bell Cups (K011532). The primary difference between the Mityvac® Merlin™ System and the MityOne™ System is that the Merlin™ System utilizes a CO2 canister to create the energy for vacuum, while the MityOne™ uses a builtin hand driven pump as the energy source. This difference does not raise new questions of safety or effectiveness. #### Summary Basis for the Finding of Substantial Equivalence The FDA has granted 510(k) clearance to the MityOne™. Merlin™ has the same intended use and indications and very similar principles of operation and technological characteristics as the predicate device. Therefore, Merlin™ is substantially equivalent. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # APR 1 2 2002 Prism Enterprises, Inc. % Mr. Howard M. Holstein Partner Hogan & Hartson, L.L.P. Columbia Square 555 Thirteenth Street, NW WASHINGTON DC 20004-1109 Re: K020447 Trade/Device Name: MityVac Merlin, Model 10027 Obstetrical Vacuum Cup Regulation Number: 21 CFR 884.4340 Regulation Name: Fetal Vacuum Extractor Regulatory Class: II Product Code: 85 HDB Dated: March 15, 2002 Received: March 15, 2002 Dear Mr. Holstein: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known):_KΩΩΟ 447 Device Name: Prism Enterprises, Inc.'s Mityvac® Merlin™ Indications for Use: Prism Enterprises, Inc.'s Mityvac® Merlin™ is intended to be utilized to assist a clinician in the delivery of an infant during childbirth. This device is indicated for use during vaginal delivery and Cesarean sections. ### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 C.F.R. 801.109) OR Over-The-Counter Use_ Nancy C. Gordon (Optional Format 1-2-96) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices K020447 510(k) Number \\\DC - 62868/1 - #1489299 v1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...