Browse hierarchy: [Obstetrics/Gynecology (OB)](/submissions/OB) → [Subpart E — Obstetrical and Gynecological Surgical Devices](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices) → [21 CFR 884.4350](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/884.4350) → PWB — Fetal Head Elevator

# PWB · Fetal Head Elevator

_Obstetrics/Gynecology · 21 CFR 884.4350 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/PWB

## Overview

- **Product Code:** PWB
- **Device Name:** Fetal Head Elevator
- **Regulation:** [21 CFR 884.4350](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/884.4350)
- **Device Class:** 2
- **Review Panel:** [Obstetrics/Gynecology](/submissions/OB)

## Identification

A fetal head elevator is a prescription device consisting of a mechanism that elevates the fetal head to facilitate delivery during a Caesarean section.

## Classification Rationale

Class II (special controls). The special controls for this device are:

## Special Controls

In combination with the general controls of the FD&C Act, the fetal head elevator is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of patient-contacting components of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(4) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Reliability testing of device deployment and retrieval under relevant use conditions must be conducted.
(ii) Testing of the maximum force applied to the fetal head in an anatomic model must be conducted.
(iii) Testing of uniform application of the elevator mechanism on the fetal head must be conducted.
(5) Labeling must include the following:
(i) Contraindication for use in the presence of active genital infection;
(ii) Specific instructions regarding the proper placement and use of the device; and
(iii) A shelf life.

## Recent Cleared Devices (2 of 2)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K243799](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/PWB/K243799.md) | Fetal Pillow | CooperSurgical, Inc. | Aug 29, 2025 | SESE |
| [DEN150053](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/PWB/DEN150053.md) | FETAL PILLOW | Safe Obstetric Systems, Ltd. | Jul 27, 2017 | DENG |

## Top Applicants

- CooperSurgical, Inc. — 1 clearance
- Safe Obstetric Systems, Ltd. — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/PWB](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/PWB)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
