K120831 · Neomedic International S.L. · PBQ · Oct 12, 2012 · Obstetrics/Gynecology
Device Facts
Record ID
K120831
Device Name
ANCHORSURE
Applicant
Neomedic International S.L.
Product Code
PBQ · Obstetrics/Gynecology
Decision Date
Oct 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Anchorsure is indicated for attaching suture to ligaments of the pelvic floor.
Device Story
Anchorsure is a manual surgical suture kit used for pelvic floor ligament attachment. The device consists of a monofilament polypropylene suture, a PEEK anchor, a stainless steel/POM anchoring handle, and a stainless steel surgical needle. It is intended for use by clinicians in a surgical setting. The device functions as a manual instrument to secure sutures to pelvic floor ligaments. It does not involve electronic processing, AI, or software. The clinical benefit is the facilitation of suture placement during pelvic floor procedures.
Clinical Evidence
Bench testing only. Testing included sterilization (EO residuals, SAL), packaging (accelerated aging), and mechanical testing (suture strength per USP 881, suture diameter per USP 861, tensile strength of suture-anchor interface per USP 871, and anchor pullout strength). Biocompatibility was supported by data from the Surelift Prolapse System (K102815).
Technological Characteristics
Manual surgical instrument. Components: Polypropylene monofilament suture (USP 0), PEEK anchor, stainless steel AISI 303/POM handle, stainless steel needle. Sterilization via ethylene oxide. No electronic or software components.
Indications for Use
Indicated for attaching suture to ligaments of the pelvic floor.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
D 117
| DIMA | 510K SUMMARY | ANCHORSURE | K120831 |
|--------------------|---------------------------------------------------------------------------------------------|------------|---------|
| SUBMITTER: | Neomedic International, S.L.<br>C/ Maestrat 41-43 1º<br>08225 Terrassa (Barcelona)<br>Spain | | Page |
| CONTACT PERSON: | Jeffrey Shideman, Ph.D. | | |
| DATE PREPARED: | October 10th 2012 | | |
| DEVICE NAME: | ANCHORSURE | | |
| REGULATION NUMBER: | 21 CFR 884.4530 | | |
| REGULATION NAME: | Obstetric-gynecologic specialized manual instrument | | |
| PRODUCT CODE: | PBQ | | |
| PREDICATE DEVICES: | GYNECARE PROLENE FASTENER SYSTEM (K042603) | | |
## Device Description:
Anchorsure is a suture kit that consists of monofilament polypropylene suture, an anchor, an anchoring handle, and a surgical needle.
## Description of material components and physical properties:
| Component | Material |
|------------------|---------------------------------|
| Suture | Polypropylene monofilament |
| Anchor | PEEK |
| Anchoring Handle | Stainless steel AISI 303<br>POM |
| Surgical Needle | Stainless steel |
| Component | Properties |
|-----------------------------------|--------------------------------------------------------------------------------|
| Monofilament polypropylene suture | USP 0<br>Diameter: 0.4 mm<br>Length: 800 mm |
| Anchor | Diameter: 3 mm<br>Length: 7 mm |
| Anchoring Handle | Anchoring Handle tube diameter = 6 mm<br>Anchoring Handle tube length = 204 mm |
| Surgical Needle | Diameter: 1.1 mm<br>Curved diameter: 20 mm |
2 2012
{1}------------------------------------------------
2/3
## 510K SUMMARY
#### Predicate Devices:
The following device has been previously cleared by the FDA in the following 510(K):
| Device | 510 (K)<br>document<br>number | Date Cleared | Indications |
|-------------------------------------|-------------------------------|-----------------------|-------------|
| GYNECARE PROLENE FASTENER<br>SYSTEM | K042603 | December 22nd<br>2004 | Suture kit |
#### Intended Use:
Anchorsure is indicated for attaching suture to ligaments of the pelvic floor.
#### Technological Characteristics:
Anchorsure and the Gynecare Prolene Fastener do not have the same technological characteristics because Anchorsure is a suture kit and the Gynecare Prolene Fastener is a staple. The new technological characteristics of the Anchorsure could affect safety and effectiveness. However, the new characteristics do not raise new issues of safety and effectiveness because both devices are used to attachments to the pelvic floor. Accepted scientific methods exist to assess the effects of the new characteristics - specifically those described in the "Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA" issued on June 3, 2003.
{2}------------------------------------------------
P ২/২
DIMA
Performance tests:
| Performance test | Test description |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization | Ethylene oxide residuals<br>Ethylene chlorohydrins residuals<br>Sterility assurance level (SAL) determination |
| Packaging | Accelerated Aging Study |
| Biocompatibility | Biocompatibility testing completed on the Surelift Prolapse<br>System (K102815) was used to support the<br>biocompatibility of Anchorsure. |
| Mechanical tests | Suture strength evaluated per USP 881.<br>Anchor strength<br>Suture diameter evaluated per USP 861.<br>Pullout strength of the anchor<br>Tensile strength of the suture-anchor interface evaluated<br>per USP 871 |
Results of verification testing indicate that the product meets the established performance requirements and standards.
## Conclusions:
Anchorsure is substantially equivalent to its proposed predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Neomedic International S.L. % Jeffrey R. Shideman, Ph.D. President International Medical Products Corporation 7307 Glouchester Drive EDINA MN 55435
Re: K120831
Trade/Device Name: ANCHORSURE Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: PBQ Dated: September 6, 2012 Received: September 10, 2012
Dear Dr. Shideman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
OCT 12 2012
{4}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin K. Eubanks
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K120831
Device Name: ANCHORSURE
Indications for Use:
Anchorsure is indicated for attaching suture to ligaments of the pelvic floor
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of
Norma M. Whaley
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K120831
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.