K172846 · Beijing Hangtian Kadi Technology R&D Institute · LKF · Jun 26, 2018 · Obstetrics/Gynecology
Device Facts
Record ID
K172846
Device Name
Uterine Manipulator
Applicant
Beijing Hangtian Kadi Technology R&D Institute
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Jun 26, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HK-WQ-I and HK-WQ-II are indicated for manipulation of the uterus and injection of fluids during laparoscopic gynecologic procedures such as laparoscopic supracervical hysterectomy, minilap tubal ligation, laparoscopic tubal occlusion or diagnostic laparoscopy. The HK-ZQ-I and HK-ZQ-II are intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitonial tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.
Device Story
Sterile, single-patient use uterine manipulators; four versions (HK-WQ-I/II, HK-ZQ-I/II). Input: manual operation by surgeon. Mechanism: manipulator shaft with inflatable intrauterine balloon; HK-ZQ models feature dial-actuated shaft flexion; optional cervical cups (ceramic or polyphenylene sulfite) in four sizes. Output: uterine positioning, fluid injection, and vaginal sealing to maintain pneumoperitoneum during laparoscopic surgery. Used in clinical/OR settings by gynecological surgeons. Benefits: facilitates uterine manipulation, fluid delivery, and vaginal access while maintaining pneumoperitoneum during minimally invasive procedures.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity). Mechanical performance testing included dimensional specs, tensile/burst strength, shaft deflection, movement force, inflation cycling, and sealing characteristics. Sterilization validated per ISO 11135:2017. Shelf life and package integrity testing performed.
Technological Characteristics
Manual uterine manipulator; shaft with inflatable intrauterine balloon; optional vaginal sealing balloon; cervical cups (ceramic or polyphenylene sulfite). Materials tested for biocompatibility (ISO 10993). Sterilization via ISO 11135:2017. Mechanical actuation via handle dial (HK-ZQ models).
Indications for Use
Indicated for patients undergoing laparoscopic gynecologic procedures including hysterectomy, tubal ligation, tubal occlusion, or diagnostic laparoscopy. HK-ZQ models are indicated for procedures requiring vaginal fornix delineation and colpotomy/culdotomy access.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 26, 2018
Beijing HangTian KaDi Technology R&D Institute % Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, 200120 China
Re: K172846
> Trade/Device Name: HK-WQ-I and HK-WQ-II Uterine Manipulators HK-ZQ-I and HK-ZQ-II Uterine Manipulators Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: May 11, 2018 Received: May 24, 2018
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael T. Bailey -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172846
#### K172846
Device Name HK-WQ-I and HK-WQ-II Uterine Manipulators HK-ZQ-I and HK-ZQ-II Uterine Manipulators
Indications for Use (Describe)
The HK-WQ-II are indicated for manipulation of the uterus, and injection of fluids during laparoscopic gynecologic procedures such as lapacervical hysterectomy, minilap tubal ligation, laparoscopic tubal occlusion or diagnostic laparoscopy.
The HK-ZQ-I and HK-ZQ-II are intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitonial tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary - K172846
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
Date of Preparation: June 25, 2018
#### Sponsor Identification:
Beijing HangTian KaDi Technology R&D Institute Room 301-08, 09, 10, 11, Third floor, Building No.13 No.15 Jing Sheng Nan Er Street, Tong Zhou District 101102 Beijing, PEOPLE'S REPUBLIC OF CHINA
Contact Person: Lingyi Zhang Position: General Manager Tel: +86-10-56407751 Fax: +86-10-56407795-803 Email: zly1110@139.com
#### Submission Correspondent:
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 3609253199 Email: info@mid-link.net
#### Subject Device Information:
| Trade Name: | HK-WQ-I and HK-WQ-II Uterine Manipulators HK-ZQ-I and HK-ZQ-II Uterine Manipulators |
|--------------------|--------------------------------------------------------------------------------------|
| Common Name: | Uterine Manipulator |
| Regulation Name: | Obstetric-gynecologic specialized manual instrument |
| Regulation Number: | 21 CFR 884.4530 |
| Class: | II |
| Product Code: | LKF (cannula, manipulator/injector, uterine) |
| Review Panel: | Obstetrics and Gynecology |
#### Indications for Use Statement:
The HK-WQ-I and HK-WQ-II are indicated for manipulation of the uterus and injection of fluids during laparoscopic gynecologic procedures such as laparoscopic supracervical hysterectomy, minilap tubal ligation, laparoscopic tubal occlusion or diagnostic laparoscopy.
The HK-ZQ-I and HK-ZQ-II are intended for use in laparoscopic procedures where it is desirable to delineate the vaginal fornices and the surgeon intends to remove or access intraperitonial tissue through the vagina by use of a colpotomy or culdotomy incision; such as laparoscopically assisted vaginal hysterectomies, total laparoscopic hysterectomies, while maintaining pneumoperitoneum by sealing the vagina while a colpotomy is performed.
{5}------------------------------------------------
### Device Description:
There are four versions of the subject device. HK-WQ-I and HK-WO-II have a curved central guide rod, while the HK-ZQ-I and HK-ZQ-II have a central guide rod which can be rotated (or "flexed") up or down using a dial on the handle.
Each device is a sterile, disposable, single-patient use device which consists of a manipulator shaft with an inflatable intrauterine balloon at the proximal end. The intrauterine balloon is inflated by liquid via a standard syringe injected through the check valve. The internal injection tip of the manipulator shaft is open to allow direct intrauterine introduction of medicine liquid via the injector port. The device also incorporates a vaginal sealing balloon which is inflated by liquid via a standard syringe injected through the check valve to seal the vagina and maintain pneumoperitoneum. The location of the uterus can be adjusted by the handle.
Each device version can be used with a cervical cup. During use, the open-end of the cup rests on the cervix. There are two versions of the cup- a ceramic version and a polyphenylene sulfite version. For both materials, the cup is available in four different sizes - 30mm, 35mm, 40mm and 46mm.
## Predicate Device Information:
Predicate Device 1 for the HK-WQ Curved version: 510(k) Number: K093556 Trade Name: Vcare Dx Sponsor: Conmed Corporation
Predicate Device 2 for the HK-ZQ Flexed version: 510(k) Number: K131781 Trade Name: Clearview Total Sponsor: Clinical Innovations
The predicate devices have not been subject to a design related recall.
### Substantially Equivalent (SE) Comparison:
Table 1 Comparison of Technology Characteristics for HK-WQ
| Item | Proposed Device | Predicate Device 1<br>K093556 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | LKF | LKF |
| Regulation Number | 21 CFR 884.4530 | 21 CFR 884.4530 |
| Indications for Use | The HK-WQ-I and HK-WQ-II are<br>indicated for manipulation of the uterus<br>and injection of fluids during<br>laparoscopic gynecologic procedures<br>such as laparoscopic supracervical<br>hysterectomy, minilap tubal ligation,<br>laparoscopic tubal occlusion or<br>diagnostic laparoscopy. | VCARE Dx uterine<br>manipulator/injector cannula is<br>indicated for manipulation of the<br>uterus, and injection of fluids during<br>laparoscopic gynecologic procedures<br>such as laparoscopic supracervical<br>hysterectomy, minilap tubal ligation,<br>laparoscopic tubal occlusion or<br>diagnostic laparoscopy. |
| Curved Shaft | Yes | Yes |
| Liquid introduction | Yes | Yes |
{6}------------------------------------------------
| Balloon inflation | Yes | Yes |
|--------------------------------------|-----|-----|
| Cervical cup | Yes | Yes |
| Prevents loss of<br>pneumoperitoneum | Yes | Yes |
Table 2 Comparison of Technology Characteristics for HK-ZQ
| Item | Proposed Device | Predicate Device 2<br>K131718 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | LKF | LKF |
| Regulation Number | 21 CFR 884.4530 | 21 CFR 884.4530 |
| Indications for Use | The HK-ZQ-I and HK-ZQ-II are<br>intended for use in laparoscopic<br>procedures where it is desirable to<br>delineate the vaginal fornices and the<br>surgeon intends to remove or access<br>intraperitonial tissue through the<br>vagina by use of a colpotomy or<br>culdotomy incision; such as<br>laparoscopically assisted vaginal<br>hysterectomies, total laparoscopic<br>hysterectomies, while maintaining<br>pneumoperitoneum by sealing the<br>vagina while a colpotomy is performed. | The ClearView Uterine Manipulator<br>Device with ColpCup and Occluder<br>is intended for use in laparoscopic<br>procedures where it is desirable to<br>delineate the vaginal fornices and the<br>surgeon intends to remove or access<br>intraperitonial tissue through the<br>vagina by use of a colpotomy or<br>culdotomy incision; such as<br>laparoscopically assisted vaginal<br>hysterectomies, total laparoscopic<br>hysterectomies, while maintaining<br>pneumoperitoneum by sealing the<br>vagina while a colpotomy is<br>performed. |
| Bending Angle | 180 degree total range of motion | 210 degree total range of motion |
| Liquid introduction | Yes | Yes |
| Balloon inflation | Yes | Yes |
| Cervical cup | Yes | Yes |
| Prevents loss of<br>pneumoperitoneum | Yes | Yes |
Each version of the subject device has the same intended use as its respective predicate. Each subject device also has different technological characteristics (design, material composition) compared to its respective predicate device; however, the differences in technological characteristics do not raise different questions of safety or effectiveness.
## Summary of Non-Clinical Performance Tests:
The following non-clinical tests were conducted on the subject devices:
- Biocompatibility include cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010), irritation ● (ISO 10993-10:2010), acute systemic toxicity (ISO 10993-11:2006), and pyrogenicity (USP <151> and ISO 10993-11:2006)
- Mechanical performance testing
{7}------------------------------------------------
- For the HK-WQ-I, HK-WQ-II, HK-ZQ-I, and HK-ZQ-II, the following tests were performed: o
- dimensional specifications ■
- ı intrauterine balloon tensile and burst strength tests
- " uterus shaft deflection force test
- movement force test
- . intrauterine balloon inflation cycling
- For the versions which include the vaginal balloon (i.e., HK-WQ-II and HK-ZQ-II) the O following additional tests were performed:
- vaginal sealing balloon maximum inflation volume test l
- vaginal sealing balloon tensile strength under maximum inflation volume l
- vaginal sealing balloon sealing characteristics
- . vaginal sealing balloon air leakage test
- For the cervical cup, the following tests were performed o
- dimensional specifications 트
- l structural strength
- .. tension resistance
- 트 locking force
- I movement force
- Sterilization validation per ISO 11135:2017 ●
- . Shelf life testing including package integrity and mechanical performance testing
#### Conclusion:
The results of performance testing demonstrate the subject devices are substantially equivalent to their respective predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.