← Product Code [LKF](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF) · K122820

# VIKY UP (K122820)

_Endocontrol · LKF · Jun 18, 2013 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF/K122820

## Device Facts

- **Applicant:** Endocontrol
- **Product Code:** [LKF](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF.md)
- **Decision Date:** Jun 18, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The ViKY UP device is indicated for computer-controlled laparoscopic hysterectomies (such as da Vinci-assisted laparoscopic hysterectomies) for the purpose of holding and controlling the movement of a uterine manipulator.

## Device Story

ViKY UP is a motorized, external robotic arm for holding and positioning uterine manipulators during laparoscopic hysterectomies. The device mounts to the operating room table and grasps the manipulator. Surgeons control the arm's movement in three directions via foot pedal or voice commands. The system is used in the OR to provide remote control of the manipulator, reducing the need for manual assistance. It remains external to the patient at all times. By automating manipulator positioning, the device assists the surgeon during computer-controlled procedures, potentially improving procedural efficiency and ergonomics.

## Clinical Evidence

Clinical study conducted on 30 patients (ages 18-80) undergoing laparoscopic hysterectomy. Primary objective: evaluate safety and effectiveness of device for manipulating the uterine manipulator. Secondary objectives: setup time and procedural impact. Results showed no adverse effects or complications; device demonstrated safety and effectiveness for the intended use.

## Technological Characteristics

Motorized robotic arm; electrical energy source; steam sterilizable; external mounting mechanism for OR table; voice and foot-pedal control interfaces.

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- ViKY Endoscope Holder ([K082233](/device/K082233.md))
- Probe Holder System ([K071405](/device/K071405.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Traditional 510(k): ViKY UP device

Image /page/0/Picture/1 description: The image shows a logo with the word "endo" in a bold, sans-serif font. Below "endo" is the word "control" in a smaller font. Above the word "endo" is a circular design that resembles a spiral or a stylized eye. The logo appears to be for a company or product named "endo control."

Mav 2013

# K122820

# ( k ) Summarv

The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act (SMDA) and 21 CFR 807.92.

## Date prepared

May 2013

# 510(k) submitter

EndoControl 5 avenue du Grand Sablon 38700 La Tronche France Phone: +33 (0)4 76 63 75 83

Fax: +33 (0)4 76 54 95 61

# Contact name

Carine Huguel carine.huguel@endocontrol-medical.com Phone: +33 (0)4 76 63 75 82

## Subject device

Trade name Common name · Requlation number Requlation name Requiatory class Product code

ViKY UP Holder, Manipulator, Positioner, Arm Unclassified Unclassified Unclassified LKF

#### Predicate devices

- · ViKY Endoscope Holder (K082233), EndoControl
- · Probe Holder System (K071405), Intuitive Surgical

#### Purpose

The purpose of this 510(k) submission is to demonstrate that the subject device is substantially equivalent to the above listed predicate devices.

## Device description

ViKY UP is a motorized uterine manipulator positioner for computer-controlled laparoscopic hysterectomies. ViKY UP holds and moves the uterine manipulator in three directions according to surgeon orders by means of either a foot controller or a voice controller.

Use of the ViKY UP allows surgeon to remotely gain control on uterine manipulator positioning from his operating position.

The ViKY UP includes two attachment mechanisms: one at its first end for mounting the system to the operating room table and the other at the second end for securely grasping a uterine manipulator.

The ViKY UP remains external to the patient's body at all times and can be steam sterilized.

//////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////// EndoContral S.A.S. - 5, av. du Gd Sablon - 36702 La Tranche - France -- Tel: +33 (0)4 76 64 95 61 SAS au capital de 157 6706 - nº SIRET: 492 059 217 00025 - code APE: 71128

{1}------------------------------------------------

# Traditional 510(k): ViKY UP device

### Indications for use

The ViKY UP device is indicated for computer-controlled laparoscopic hysterectomies (such as da Vinci-assisted laparoscopic hysterectomies) for the purpose of holding and controlling the . movement of a uterine manipulator.

#### Comparison to predicate devices

The Probe Holder System uses hydraulic energy whereas ViKY EP and ViKY UP use electrical energy. The three systems include a foot pedal to actuate the mechanism, ViKY EP and ViKY UP also include voice actuation. ViKY EP and ViKY UP can be remote controlled, the Probe Holder System cannot.

Based upon available technical information, intended use, performance information and method of use provided in this pre-market notification, the ViKY UP device is substantially equivalent to Probe Holder System (K071405) and has the same technological characteristics as the ViKY EP device (K082233).

Compared to both predicate devices, ViKY UP does not create new risk, does not raise new issues on safety and effectiveness.

#### Performance data

Design analysis and testing has been conducted to confirm that basis functional characteristics of the subject device are substantially equivalent to those of the predicate devices.

Use has been validated through bench testing and clinical testing that demonstrated that the ViKY UP device can be used safely and efficiently for the claimed intended use.

#### Non-clinical tests

Non-clinical testing consisted of bench and anatomical tests.

These tests validated usability and ergonomy of the ViKY UP device and confirmed that basic functional characteristics are in compliance with marketing system requirements, and that design output meets the design input requirements.

#### Clinical tests

A clinical study has been carried out on 30 patients candidate for laparoscopic hysterectorny procedure and aged between 18 and 80. Exclusion criteria were principally comprised of patients for whom anatomy precludes the use of a uterine manipulator.

The primary objective was to evaluate the safety and effectiveness of using the ViKY UP device to manipulate the uterine manipulator during computer-controlled laparoscopic hysterectomies. Secondary objectives included set up time and overall procedural impact of the ViKY UP device for uterine manipulation.

No adverse effects or complication were observed during the study. The results showed safety and effectiveness of the device for computer-controlled laparoscopic hysterectomies and did not raise any issue.

### Conclusion

Data included in this submission demonstrate that the evaluation of ViKY UP does not raise any additional concerns regarding safety and effectiveness and can be considered substantially equivalent to 510{k} cleared predicate devices.

xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx EndoControl S.A.S. - 5, av. du Gd Sablon - 38700 La Tranche - Tel: +33 (0) 476 63 75 83 - Fax: +33 (0)4 76 6.4 95 61 SAS au capílal de 1.57 670€ - nº SIRET: 492 059 217 00025 - code APE: 71.28

Page 212

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings, facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle symbol.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Letter dated: June 18, 2013

EndoControl % Ms. Carine Huguel 5 avenue du Grand Sablon 38700 La Tronche, France

Re: K122820

Trade/Device Name: VIKY UP Regulatory Class: Unclassified Product Code: LKF Dated: April 30, 2013 Received: May 23, 2013

# Dear Ms. Huguel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding andadulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Carine Huguel

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for use statement

# 510(k) number K 122820

Device name ViKY UP

#### Indications for use

The ViKY UP device is indicated for computer-controlled laparoscopic hysterectomies (such as da Vinci-assisted laparoscopic hysterectomies) for the purpose of holding and controlling the movement of a uterine manipulator.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21-CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Joshua C. Nipper -S

(Division Sign-Off) Division of Surgical Devices 510 (k) Number: K122820

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF/K122820](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/LKF/K122820)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
