← Product Code [KNF](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/KNF) · K222304

# Sonata Transcervical Fibroid Ablation System 2.2 (K222304)

_Gynesonics, Inc. · KNF · Nov 8, 2022 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/KNF/K222304

## Device Facts

- **Applicant:** Gynesonics, Inc.
- **Product Code:** [KNF](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/KNF.md)
- **Decision Date:** Nov 8, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Sonata® Transcervical Fibroid Ablation System 2.2 is intended for diagnostic intrauterine imaging and transcervical treatment of symptomatic uterine fibroids, including those associated with heavy menstrual bleeding.

## Device Story

The Sonata System 2.2 is a transcervical, uterine-sparing device for treating symptomatic uterine fibroids. It integrates an Intrauterine Ultrasound (IUUS) probe for visualization and an RFA handpiece for radiofrequency ablation. The system includes an RF generator, a SMART tablet with graphical guidance software, and a system cart. The clinician uses the IUUS probe to localize fibroids and guide the placement of needle electrodes from the RFA handpiece into the target tissue. The system provides real-time ultrasound imaging and a graphical overlay (SMART Guide) to plan ablation zones and maintain a thermal safety border relative to the uterine serosa. RF energy is delivered to induce thermal fixation and coagulative necrosis. The device is used in a clinical setting by a physician. The output allows for precise, image-guided treatment of fibroids without incisions, potentially reducing patient recovery time and symptoms like heavy menstrual bleeding.

## Clinical Evidence

No new clinical data was required for this submission. The device relies on clinical evidence from the original IDE study (G140114, NCT02228174), which involved 147 subjects across 22 centers in a single-arm cohort study.

## Technological Characteristics

The system uses radiofrequency (RF) energy for ablation and B-mode ultrasound for imaging. Materials include glass fiber filled polyetherimide, glass reinforced vinyl ester, fluorocarbon rubber, silicone, Nitinol, and surgical grade stainless steel. The system is networked via internal data transfer (FireWire). It complies with ISO 10993 biocompatibility standards and electrical safety standards including ANSI/AAMI ES60601-1 and IEC 60601-2-37. The software includes a graphical user interface for treatment planning and automated RF energy control.

## Regulatory Identification

A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.

## Predicate Devices

- Sonata Transcervical Fibroid Ablation System 2.2 ([K211535](/device/K211535.md))

## Reference Devices

- Terason uSmart3200T Ultrasound System ([K150533](/device/K150533.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

November 8, 2022

Gynesonics, Inc. Diane King VP, Global Regulatory Affairs & Healthcare Compliance 600 Chesapeake Drive Redwood City, CA 94063

Re: K222304

> Trade/Device Name: Sonata® Transcervical Fibroid Ablation System 2.2 Regulation Number: 21 CFR§ 884.4160 Regulation Name: Unipolar Endoscopic Coagulator-Cutter and Accessories Regulatory Class: II Product Code: KNF, ITX, IYO Dated: October 7, 2022 Received: October 11, 2022

Dear Diane King:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Jason Roberts -S

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K222304

Device Name

Sonata® Transcervical Fibroid Ablation System 2.2

Indications for Use (Describe)

The Sonata® Transcervical Fibroid Ablation System 2.2 is intended for diagnostic intrauterine imaging and transcervical treatment of symptomatic uterine fibroids, including those associated with heavy menstrual bleeding.

| Type of Use (Select one or both, as applicable)                                                |
|------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Remediation Use (Part 21 CFR 601 Subpart E)</label> |
| <label><input type="checkbox"/> For-Profit Control Use (21 CFR 601 Subpart E)</label>          |

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# gyn

# 510(k) Summary

| Sponsor:        | Gynesonics, Inc.<br>600 Chesapeake Drive<br>Redwood City, CA 94063                                             |
|-----------------|----------------------------------------------------------------------------------------------------------------|
| Contact Person: | Diane King<br>VP Global Regulatory Affairs and Healthcare Compliance<br>dking@gynesonics.com<br>(650) 216-3883 |
| Date Prepared:  | November 2, 2022                                                                                               |

# Device Information

| Proprietary Name:     | Sonata® Transcervical Fibroid Ablation System 2.2                                                                                                   |  |  |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Common Name:          | Sonography-Guided Transcervical Fibroid Ablation System                                                                                             |  |  |
| Class:                | Class II                                                                                                                                            |  |  |
| Regulation:           | 21 CFR 884.4160<br>Unipolar endoscopic coagulator-cutter and accessories                                                                            |  |  |
| Product Code:         | KNF Coagulator-Cutter, Endoscopic, Unipolar (And Accessories<br>ITX Transducer, Ultrasonic, Diagnostic<br>IYO Ultrasonic pulsed echo imaging system |  |  |
| Classification Panel: | 85 – Obstetrical & Gynecological                                                                                                                    |  |  |

# Indications for Use

The Sonata® Transcervical Fibroid Ablation System 2.2 is intended for diagnostic intrauterine imaging and transcervical treatment of symptomatic uterine fibroids, including those associated with heavy menstrual bleeding.

{4}------------------------------------------------

# Predicate Devices

The predicate device is listed in Table 1. The predicate device is the unmodified Sonata Transcervical Fibroid Ablation System 2.2.

The predicate device has not been the subject of any design-related recalls.

Table 1 Table of Predicate and Reference Device

| 510(k)  | Product                                             | 510(k) Holder | Clearance Date |
|---------|-----------------------------------------------------|---------------|----------------|
| K211535 | Sonata Transcervical Fibroid Ablation System<br>2.2 | Gynesonics    | June 17, 2021  |

# Device Description:

The Sonata System 2.2 (Figure 1) provides radiofrequency (RF) ablation of uterine fibroids using a transcervical approach that is uterine sparing without incisions or material uterine distension. The system enables a clinician to deliver radiofrequency energy to fibroid tissue resulting in thermal fixation and coagulative necrosis of the tissue.

Image /page/4/Figure/9 description: The image shows the text "Figure 1 Sonata System 2.2". The text is in bold font. The figure number is 1, and the system version is 2.2.

Image /page/4/Figure/10 description: The image shows a medical device system and its components. On the left is a system cart with a SMART Tablet and RF Generator attached. On the right are the components: IUUS Probe, RFA Handpiece, RFA Handpiece Cable, and two Dispersive Electrodes. The reusability of each component is specified.

{5}------------------------------------------------

The system combines two technologies - ultrasound for visualization, and radiofrequency energy for ablative therapy - in a single integrated handpiece (Figure 2).

Figure 2 Intrauterine Ultrasound (IUUS) Probe being connected to the Radiofrequency Ablation (RFA) Handpiece to form a single Treatment Device

Image /page/5/Picture/4 description: The image shows two medical instruments, labeled as "IUUS Probe" and "RFA Handpiece", being held by gloved hands. The IUUS probe is positioned above the RFA handpiece, both against a blue background. A second image shows the two instruments being held together.

The Sonata System is comprised of medical equipment (Figure 1), software, and various single-use and reusable instruments. Sonata System devices and accessories are summarized in Table 2. The devices that are changed or new with this submission are shown in bold purple.

{6}------------------------------------------------

| Catalog Number      | Product Description                                                      |
|---------------------|--------------------------------------------------------------------------|
| Durable Equipment   |                                                                          |
| SONATA2-110         | Sonata Transcervical Fibroid Ablation System, consisting of:             |
| RFG2-110            | Sonata Radiofrequency Generator                                          |
| USCON-2200          | Sonata SMART Tablet                                                      |
| ACCY-002            | Sonata System Cart                                                       |
| External components | Footswitch, mouse, cables                                                |
| System Software     |                                                                          |
| SW-002              | Sonata System Software v2.1.2                                            |
| Reusable Devices    |                                                                          |
| IUSP-002            | Sonata Intrauterine Ultrasound (IUUS) Probe (Non-Sterile) (modified IFU) |
| IUSP-002S           | Sonata Intrauterine Ultrasound (IUUS) Probe (Sterile)                    |
| ACCY-008            | Sonata RFA Handpiece Cable, Reusable                                     |
| Single-Use Devices  |                                                                          |
| RFA-002             | Sonata Radiofrequency Ablation Handpiece (Sterile)                       |
| DE-001              | Sonata Dispersive Electrode (Non-sterile)                                |
| Accessories         |                                                                          |
| SHPR-001            | Sonata Intrauterine Ultrasound Probe Sterile Shipper Kit                 |
| RTN-001             | Sonata Intrauterine Ultrasound Probe Return Kit                          |
| ACCY-018            | Sonata IUUS Probe Connector Protector                                    |
| OM-1000-GS          | Sonata IUUS Probe Reprocessing Tray (Manufacturer: Summit Medical LLC)   |
| 8EC4A               | Endocavity Ultrasound Transducer (Manufacturer: Terason®)                |

Table 2 Sonata System 2.2 Devices and Accessories

A single-use Radiofrequency Ablation (RFA) Handpiece attaches to a reusable Intrauterine Ultrasound (IUUS) Probe as shown in Figure 2 to provide sonography-guided RF ablation. Once connected, the combination is referred to as the "Treatment Device". The RFA Handpiece connects to the Sonata RF Generator and contains the Needle Electrodes that deliver radiofrequency energy to the target tissue. The IUUS Probe connects to the SMART Tablet and provides diagnostic ultrasound imaging and guidance. Ultrasound guidance is used to localize the fibroids from within the uterine cavity, guide placement of the RFA Handpiece Needle Electrodes into a target fibroid and ensure safety with respect to the serosa. When the Needle Electrodes are anchored within tissue, the physician is able to pivot the IUUS Probe transducer around the Needle Electrodes in order to confirm safety of the uterine serosa through multiple ultrasound planes.

The Sonata System allows for treatment planning through the use of a graphical interface and automated control of RF energy delivery.

{7}------------------------------------------------

Sonata Graphical Guidance Software (GGS) includes the SMART Guide (Figure 3) and integrates treatment planning, targeting, and ablation of fibroids. The SMART Guide displays a real-time graphic overlay on the live ultrasound image for targeting and deployment of radiofrequency ablation.

Image /page/7/Figure/3 description: The image is a figure from the Sonata System SMART Guide. The figure shows two images, one ultrasound and one anatomical, with labels for the IUUS Probe Tip, Introducer Guide, Ablation Zone, and Thermal Safety Border. The text explains that the SMART Guide is a graphical overlay used for sizing and positioning of the Ablation Zone over the targeted fibroid, and that the Thermal Safety Border must be kept within the uterine serosa at all times.

#### Figure 3 Sonata SMART Guide

Two main elements of the SMART Guide are the Ablation Zone and the Thermal Safety Border.

- Ablation Zone (red inner ellipse) a two-dimensional representation of the outer . boundary of the average region of tissue ablation for the selected ablation size
- . Thermal Safety Border (green outer ellipse) - the distance at which tissue outside of the Ablation Zone is safe from the potential of thermal damage.

# Changes triggering this 510(k)

Table 3 summarizes the changes to the Sonata System 2.2 that are driving the need for a new 510(k).

{8}------------------------------------------------

| Sonata System 2.2<br>Component/Accessory                                                                                                           | Description of Change                                                                                                                                                                                                                                                            | Reason for change                                                                                                          | Impact of Change                                                                                                                                                                                                     |
|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sonata Intrauterine<br>Ultrasound (IUUS) Probe<br>IUSP-002                                                                                         | Labeling change: Addition<br>of validated instructions for<br>sterilization of the IUUS<br>Probe by the following<br>liquid chemical sterilant<br>processing systems:<br>STERIS System 1E®<br>(available in the US only)<br>STERIS System 1®<br>EXPRESS<br>(available only OUS). | Expand options for<br>sterilization of reusable<br>device.                                                                 | Sterilization validation<br>and verification of device<br>compatibility required.                                                                                                                                    |
| Addition of unclassified<br>accessory:<br>ACCY-018 Sonata IUUS<br>Probe Connector Protector<br>Referenced within Sonata<br>IUUS Probe IUSP-002 IFU | New accessory supports<br>reprocessing of the IUUS<br>Probe                                                                                                                                                                                                                      | Provides a water-tight<br>cover for the IUUS<br>Probe electrical<br>connector during liquid<br>reprocessing<br>procedures. | Design controls were<br>followed for the new<br>accessory.  See Table 5<br>for summary of design<br>controls and test results<br>relevant to use of the<br>new accessories for<br>reprocessing of the<br>IUUS Probe. |

### Table 3 Summary of Device Changes Triggering New 510(k)

# Comparison to Predicate Device

Table 4 contains a detailed comparison of the modified Sonata System 2.2 to its predicate, the unmodified Sonata System 2.2. The modified Sonata Transcervical Fibroid Ablation System 2.2 employs the same fundamental scientific technology as the currently marketed predicate Sonata Transcervical Fibroid Ablation System 2.2 (K211535). The modifications in the subject device do not raise any new questions of safety and effectiveness.

{9}------------------------------------------------

|                                                                 | Modified Sonata System 2.2<br>(this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Sonata System 2.2<br>(K211535)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Comparison Discussion             |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Characteristics                                                 | Intended Use/Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                   |
| Intended Use                                                    | Ablation of uterine fibroids with diagnostic<br>ultrasound imaging.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Ablation of uterine fibroids with diagnostic<br>ultrasound imaging.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                              |
| Indications for Use                                             | The Sonata System is intended for diagnostic<br>intrauterine imaging and transcervical<br>treatment of symptomatic uterine fibroids,<br>including those associated with heavy<br>menstrual bleeding.                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The Sonata System is intended for diagnostic<br>intrauterine imaging and transcervical<br>treatment of symptomatic uterine fibroids,<br>including those associated with heavy<br>menstrual bleeding.                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                              |
| Regulation Number                                               | §884.4160 Unipolar endoscopic coagulator-<br>cutter and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | §884.4160 Unipolar endoscopic coagulator-<br>cutter and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                              |
| Product Code                                                    | KNF<br>Coagulator-Cutter, Endoscopic, Unipolar (And<br>Accessories)<br><br>ITX<br>Transducer,<br>Ultrasonic, Diagnostic<br><br>IYO<br>Ultrasonic pulsed echo imaging system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | KNF<br>Coagulator-Cutter, Endoscopic, Unipolar (And<br>Accessories)<br><br>ITX<br>Transducer,<br>Ultrasonic, Diagnostic<br><br>IYO<br>Ultrasonic pulsed echo imaging system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                              |
| System Functional/ Operational Features                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                   |
| Principal Mode of Operation                                     | Radiofrequency ablation of fibroid tissue<br>resulting in thermal fixation and coagulative<br>necrosis.<br>B Mode ultrasound imaging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Radiofrequency ablation of fibroid tissue<br>resulting in thermal fixation and coagulative<br>necrosis.<br>B Mode ultrasound imaging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                              |
| Primary user interface                                          | Graphical user interface                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Graphical user interface                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                              |
| Characteristics                                                 | Modified Sonata System 2.2<br>(this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Sonata System 2.2<br>(K211535)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Comparison Discussion             |
| Treatment<br>Planning                                           | Integrated SMART Guide in software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Integrated SMART Guide in software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                              |
| Treatment<br>Approach                                           | In situ delivery and control of RF energy through deployable array needle electrodes with impedance and temperature feedback under visual control.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | In situ delivery and control of RF energy through deployable array needle electrodes with impedance and temperature feedback under visual control.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                              |
| Treatment<br>Guidance                                           | Must be used under ultrasound guidance that is integrated into the system. The Ultrasound Console (SMART Tablet) with IUUS Probe is indicated for intrauterine imaging and guidance for placement of the Needle Electrodes.                                                                                                                                                                                                                                                                                                                                                                                                                                         | Must be used under ultrasound guidance that is integrated into the system. The Ultrasound Console (SMART Tablet) with IUUS Probe is indicated for intrauterine imaging and guidance for placement of the Needle Electrodes.                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                              |
| Route of<br>Access                                              | Transcervical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Transcervical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                              |
| System Components                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                   |
| RF Generator                                                    | An RF Generator provides RF energy to the RFA Handpiece through the handpiece cable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | An RF Generator provides RF energy to the RFA Handpiece through the handpiece cable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                              |
| Treatment<br>Device                                             | Single-use RFA handpiece with trocar-pointed shaft and 7 deployable needle electrodes, with reusable Cable. Combines with the reusable Intrauterine Ultrasound Probe to form the "Treatment Device".                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Single-use RFA handpiece with trocar-pointed shaft and 7 deployable needle electrodes, with single-use Cable. Combines with the reusable Intrauterine Ultrasound Probe to form the "Treatment Device".                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                              |
| Dispersive<br>Electrodes                                        | Dispersive Electrodes, quantity 2, with cables; provides return path for the RF energy delivered by the Handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Dispersive Electrodes, quantity 2, with cables; provides return path for the RF energy delivered by the Handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                              |
| Characteristics                                                 | Modified Sonata System 2.2 (this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Sonata System 2.2 (K211535)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Comparison Discussion             |
| Ultrasound Console                                              | Incorporates the Terason uSmart3200T Ultrasound System (K150533) with addition of Sonata Graphical Guidance software. The uSmart3200T is a tablet with 11.6" LED backlit display, lithium-polymer battery. Uses a medical-grade power supply. Data transferred internally from the ultrasound engine to the laptop computer over a FireWire (aka IEEE 1394)                                                                                                                                                                                                                                                                                                         | Incorporates the Terason uSmart3200T Ultrasound System (K150533) with addition of Sonata Graphical Guidance software. The uSmart3200T is a tablet with 11.6" LED backlit display, lithium-polymer battery. Uses a medical-grade power supply. Data transferred internally from the ultrasound engine to the laptop computer over a FireWire (aka IEEE 1394)                                                                                                                                                                                                                                                                                                         | Same                              |
| Ultrasound Transducer                                           | Gynesonics Sonata Intrauterine Ultrasound (IUUS) Probe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Gynesonics Sonata Intrauterine Ultrasound (IUUS) Probe                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                              |
| Ultrasound Transducer                                           | Claimed compatibility with commercially available Terason 8EC4A transducer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Claimed compatibility with commercially available Terason 8EC4A transducer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                              |
| Power cord                                                      | Power cord - A medical grade power cord that provides AC power to the power strip on the System Cart. The power strip in turn powers the RF Generator and the Ultrasound Console.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Power cord - A medical grade power cord that provides AC power to the power strip on the System Cart. The power strip in turn powers the RF Generator and the Ultrasound Console.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                              |
| Footswitch                                                      | Pneumatic footswitch with PVC tubing used to activate and terminate delivery of RF energy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Pneumatic footswitch with PVC tubing used to activate and terminate delivery of RF energy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                              |
| Optical Mouse                                                   | Optical Mouse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Optical Mouse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                              |
| System Cart                                                     | Cart, accommodates tablet                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Cart, accommodates tablet                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                              |
| Materials                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                   |
| Materials - Patient Contact - IUUS Probe                        | Glass fiber filled polyetherimide, glass reinforced vinyl ester, fluorocarbon rubber, UV adhesive, Silicone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Glass fiber filled polyetherimide, glass reinforced vinyl ester, fluorocarbon rubber, UV adhesive, Silicone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                              |
| Patient Contact Materials - active electrode                    | Medical grade metal alloys and plastic polymers (i.e. Nitinol®, surgical grade stainless steel)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Medical grade metal alloys and plastic polymers (i.e. Nitinol®, surgical grade stainless steel)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                              |
| Characteristics                                                 | Modified Sonata System 2.2<br>(this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Sonata System 2.2<br>(K211535)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Comparison Discussion             |
| Patient Contact<br>Materials<br>- dispersive<br>electrode       | Acrylate-polymer based hydrogel, polyester<br>fabric with poly film and medical grade acrylic<br>adhesive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Acrylate-polymer based hydrogel, polyester<br>fabric with poly film and medical grade acrylic<br>adhesive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                              |
| Biocompatibility                                                | Is biocompatible with intended use in<br>compliance with<br>• ISO 10993-1 5th Ed. 2018-08<br>• ISO 10993-5 3rd ed. 2009-06-01<br>• ISO 10993-10 3rd ed. 2010-08-01<br>• ISO 10993-11 3rd ed. 2017-09                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Is biocompatible with intended use in<br>compliance with<br>• ISO 10993-1 5th Ed. 2018-08<br>• ISO 10993-5 3rd ed. 2009-06-01<br>• ISO 10993-10 3rd ed. 2010-08-01<br>• ISO 10993-11 3rd ed. 2017-09                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                              |
| Safety and Performance                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                   |
| Electrical Safety<br>& EMC                                      | ANSI/AAMI ES60601-1:2005/(R)2012 And A1:2012,<br>C1:2009 / (R)2012 And A2:2010/(R)2012<br>IEC 60601-1-2 Ed 4: 2014-02<br>IEC 60601-1-6 Ed 3.1 2013-10<br>ANSI AAMI IEC 62366-1:2015<br>IEC 60601-1-8: Edition 2.1 2012-11<br>ANSI/AAMI IEC 60601-2-2:2017<br>IEC 60601-2-37 Ed 2.1 2015<br>IEC 62304 Ed. 1.1 2015-06                                                                                                                                                                                                                                                                                                                                                | ANSI/AAMI ES60601-1:2005/(R)2012 And A1:2012,<br>C1:2009 / (R)2012 And A2:2010/(R)2012<br>IEC 60601-1-2 Ed 4: 2014-02<br>IEC 60601-1-6 Ed 3.1 2013-10<br>ANSI AAMI IEC 62366-1:2015<br>IEC 60601-1-8: Edition 2.1 2012-11<br>ANSI/AAMI IEC 60601-2-2:2017<br>IEC 60601-2-37 Ed 2.1 2015<br>IEC 62304 Ed. 1.1 2015-06                                                                                                                                                                                                                                                                                                                                                | Same                              |
| Performance<br>testing –<br>System level<br>(bench)             | • Shelf-life / Service life<br>• Confirmatory verification to specifications,<br>standards, and guidance documents.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | • Shelf-life / Service life<br>• Confirmatory verification to specifications,<br>standards and guidance documents.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same including design<br>controls |
| Characteristics                                                 | Modified Sonata System 2.2<br>(this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Sonata System 2.2<br>(K211535)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Comparison Discussion             |
| Performance<br>testing –<br>Ablation                            | Ablation output RF Generator safety features Handpiece retention forces Ultrasound visibility of the handpiece Dispersive Electrode adhesion RF Generator software and hardware verification and validation, including GUI, alerts, communication between components, real-time feedback to user via device sensors, power control, and software/hardware interface              Successfully demonstrated through early clinical and bench ablation testing that the system performs as intended and per specifications. Ablation capability was confirmed and the radiofrequency ablation provides a reproducible, discretely demarcated zone of tissue necrosis. | Ablation output RF Generator safety features Handpiece retention forces Ultrasound visibility of the handpiece Dispersive Electrode adhesion RF Generator software and hardware verification and validation, including GUI, alerts, communication between components, real-time feedback to user via device sensors, power control, and software/hardware interface              Successfully demonstrated through early clinical and bench ablation testing that the system performs as intended and per specifications. Ablation capability was confirmed and the radiofrequency ablation provides a reproducible, discretely demarcated zone of tissue necrosis. | Same                              |
| Acoustic Output<br>Measurement<br>Standard                      | NEMA UD 2-2004 (R2009)<br>IEC 60601-2-37 Ed 2.1 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | NEMA UD 2-2004 (R2009)<br>IEC 60601-2-37 Ed 2.1 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                              |
| Acoustic Output<br>Global<br>Maximum<br>B Mode:                 | ISPTA ≤ limit of 720 mW/cm²<br>Value: 162 mW/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | ISPTA ≤ limit of 720 mW/cm²<br>Value: 162 mW/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                              |
|                                                                 | MI ≤ limit of 1.9<br>Value: 1.7                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | MI ≤ limit of 1.9<br>Value: 1.7                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                   |
| Characteristics                                                 | Modified Sonata System 2.2<br>(this submission)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Sonata System 2.2<br>(K211535)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Comparison Discussion             |
| Usability and                                                   | IEC 60601-1-6 Ed 3.1 2013-10                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | IEC 60601-1-6 Ed 3.1 2013-10…

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/KNF/K222304](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/KNF/K222304)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
