MODULAP LOOP

K121343 · Atc Technologies, Inc. · KNF · Jun 14, 2013 · Obstetrics/Gynecology

Device Facts

Record IDK121343
Device NameMODULAP LOOP
ApplicantAtc Technologies, Inc.
Product CodeKNF · Obstetrics/Gynecology
Decision DateJun 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.4160
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Modulap Loop is intended for the amputation of the uterus during laparoscopic supracervical hysterectomy and for resection of pedunculated subserosal (abdominal) myomas. It is not intended for use in hysteroscopy or for contraceptive coagulation of the fallopian tubes.

Device Story

Modulap Loop is 5 mm laparoscopic instrument; consists of insulated monopolar electrosurgical probe with flexible wire loop distal tip and sliding insulating sheath. Device connects to standard electrosurgical generators and cables. Used by surgeons in laparoscopic procedures to amputate uterus or resect myomas. Surgeon operates device via laparoscopic access; sliding sheath exposes central 1-inch cutting area of wire loop to apply electrosurgical energy for coagulation and cutting. Single-use, sterile, disposable design. Benefits include precise tissue resection during minimally invasive gynecological surgery.

Clinical Evidence

Bench testing only. Performance, reliability, material properties, biocompatibility, packaging validation, and electrical safety testing were performed to demonstrate substantial equivalence.

Technological Characteristics

5 mm laparoscopic monopolar electrosurgical instrument; flexible wire loop tip with 1-inch active cutting area; sliding insulating sheath; single-use; sterile; compatible with standard electrosurgical generators.

Indications for Use

Indicated for monopolar electrosurgical coagulation/cutting during laparoscopic surgery, specifically for uterine amputation during supracervical hysterectomy and resection of pedunculated subserosal (abdominal) myomas. Contraindicated for hysteroscopy and contraceptive coagulation of fallopian tubes.

Regulatory Classification

Identification

A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## 5. 510(k) Summary | Date Prepared: | March 15, 2012 JUN 1 4 2013 | |--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name: | Modulap Loop | | Common Name: | Unipolar endoscopic coagulator-cutter | | Submitter: | ATC Technologies Inc.<br>30B Upton Drive,<br>Wilmington, MA 01887 | | Contact Person: | John Gillespie (consultant)<br>Phone: (774) 567-0277<br>Fax Number: (781) 939-0726<br>Email: jgillespie@clovermedical.com | | Classification: | Coagulator-Cutter, Endoscopic, Unipolar (and<br>accessories) 21 CFR 884.4160, Product code KNF,<br>Class II | | Intended Use: | The Modulap Loop is intended for the amputation of<br>the uterus during laparoscopic supracervical<br>hysterectomy and for resection of pedunculated<br>subserosal (abdominal) myomas.<br>It is not intended for use in hysteroscopy or for<br>contraceptive coagulation of the fallopian tubes. | #### Description of Device The Modulap Loop is a 5 mm laparoscopic instrument. It consists of a standard insulated monopolar electrosurgical probe, with a flexible wire loop as its distal tip, and a sliding sheath constructed of insulating material. The wire loop is insulated except for a central approximately 1 inch cutting area. The device is a single-use sterile, disposable device. It is compatible with standard electrosurgical generators, and associated cables. ### Summary of Technological Characteristics vs. Predicate The Modulap Loop and the Predicate device (Lina Gold Loop, K070315) are both Monopolar electrosurgical electrode with loop style tip and sliding sheath; sized to pass through 5 mm laparoscopic trocar; and compatible with standard electrosurgical generators. Their indications for use are equivalent, and other technological characteristics such as design, materials, performance, anatomical sites, energy used, human factors, biocompatibility, sterility, electrical safety, mechanical safety and standards met are the same or equivalent. {2}------------------------------------------------ ### Test Data Testing including performance, reliability, materials property testing, biocompatibility, packaging validation and electrical safety testing was performed which supports that the Modulap Loop is equivalent to the predicate with regard to safety and effectiveness. #### Substantial Equivalence Based on the Indication for Use, technological characteristics, performance testing, and comparison to its predicate device we conclude that the Modulap Loop has been shown to be substantially equivalent to a legally marketed predicate device and is safe and effective for its intended use. {3}------------------------------------------------ #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 14, 2013 ATC Technologies, Inc. % Mr. John Gillespie Consultant Clover Medical 79 Haven Street DOVER MA 02030 Re: K121343 Trade/Device Name: Modulap Loop Regulation Number: 21 CFR§ 884.4160 Regulation Name: Unipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: KNF Dated: May 17, 2013 Received: June 7, 2013 Dear Mr. Gillespie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce-prior-to-May-28,-1.976,-the-enactment-date-of-the-Medical-Device-Amendments-or-to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {4}------------------------------------------------ Page 2 - Mr. John Gillespie You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, for Benjamin-R-Fisher-Ph:D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 4. Indications for Use Statement ## Indications for Use 510(k) Number (if known): K121343 Device Name: Modulap Loop Indications for Use: The Modulap Loop is intended for monopolar electrosurgical coagulation/cutting during laparoscopic surgery. It is indicated for the amputation of the uterus during supracervical hysterectomy and for resection of pedunculated subserosal (abdominal) myomas. It is not intended for use in hysteroscopy or for contraceptive coagulation of the fallopian tubes. Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of ____________________________________________________________________________________________________________________________________________________________________ Herbert P. Lerner -S K121343
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...