← Product Code [HFX](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HFX) · K142163

# MALE CIRCUMCISION KITS:RAPIDECLAMP (K142163)

_Medicfit Technology Sdn Bhd · HFX · Apr 20, 2015 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HFX/K142163

## Device Facts

- **Applicant:** Medicfit Technology Sdn Bhd
- **Product Code:** [HFX](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HFX.md)
- **Decision Date:** Apr 20, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Circumcision of foreskin of males from infant to adult by circumferential excision of the foreskin or prepuce at or near the level of coronal sulcus, with minimal amount of preputial skin remaining.

## Device Story

RAPIDECLAMP is a manual surgical instrument kit for male circumcision; replaces conventional methods to facilitate faster procedures, reduce bleeding, and minimize pain. Kit includes guide ring, spike body, anti-slip ring, locking cover, protective shield, measuring plate, and blade/scissors. Physician selects appropriate size (8mm to 29mm) based on patient age (infant to adult). Device clamps foreskin to restrict blood flow; excess tissue is excised with blade/scissors. Device remains in place until healing occurs or is removed by physician via mechanical unlocking of safety locks and disassembly. Used in clinical settings; provides a suture-free alternative to traditional circumcision.

## Clinical Evidence

Clinical testing conducted at two Malaysian clinics (Ar-Raudhah Clinic, JB Specialist). Studies demonstrated ease of application, short procedure time, and no requirement for sutures. Data confirms safety and effectiveness for circumcision in adult males.

## Technological Characteristics

Materials: Polycarbonate (PC) and Acrylonitrile Butadiene Styrene (ABS). Biocompatibility: ISO 10993-1 compliant. Form factor: Multi-component clamp system (guide ring, spike body, anti-slip ring, locking cover, protective shield). Sterilization: Sterile. Single-use. Mechanical operation.

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- Smart KlampR ([K032091](/device/K032091.md))
- Shang RingTM ([K121789](/device/K121789.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract design featuring three stylized human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 20, 2015

Medicfit Technology Sdn Bhd % Jigar Shah Consultant MDI Consultants. Inc. 55 Northern Blvd Great Neck, NY 11021

Re: K142163

Trade/Device Name: Male Circumcision Kits: RapideClamp™ Regulation Number: 21 CFR § 884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HFX Dated: March 17. 2015 Received: March 19, 2015

Dear Jigar Shah,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K142163

Device Name

Male Circumcision Kits: RAPIDECLAMP

Indications for Use (Describe)

Circumcision of foreskin of males from infant to adult by circumferential excision of the foreskin or prepuce at or near the level of coronal sulcus, with minimal amount of preputial skin remaining.

Type of Use (Select one or both, as applicable)

| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 201 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart C)</label> |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

# MEDICFIT TECHNOLOGY SDN BHD Male Circumcision Kits: RAPIDECLAMP™

| Date Prepared:           | April 17, 2015                                                                               |  |  |  |  |
|--------------------------|----------------------------------------------------------------------------------------------|--|--|--|--|
| Submitted by:            | MEDICFIT TECHNOLOGY SDN BHD                                                                  |  |  |  |  |
|                          | No31, Jalan Emas 31, Bandar Sungai Emas,<br>42700 Banting, Selangor Darul Ehsan,<br>Malaysia |  |  |  |  |
| Establishment<br>Number: | N/A                                                                                          |  |  |  |  |
| Representative:          | mdi Consultants, Inc.                                                                        |  |  |  |  |
| Proprietary Name:        | Male Circumcision Kits: RAPIDECLAMP™                                                         |  |  |  |  |
| Common Name:             | Male Circumcision Clamp                                                                      |  |  |  |  |
| Classification Name:     | Clamp, Circumcision, Obstetrics & Gynecology<br>21 CFR § 884.4530                            |  |  |  |  |
| Product Code:            | HFX                                                                                          |  |  |  |  |
|                          |                                                                                              |  |  |  |  |

# Predicate Devices:

| Trade/Proprietary Name | Manufacturer                                                     | 510(k) Number |
|------------------------|------------------------------------------------------------------|---------------|
| Smart KlampR           | Emergo Group Inc.                                                | K032091       |
| Shang RingTM           | Wuhu Snnda Medical<br>Treatment Appliance<br>Technology Co., Ltd | K121789       |

# Device Description:

The Male Circumcision Kits: RAPIDECLAMP™ is a surgical procedure purposely for circumcision which using a set of clamps that involves partial or complete removal of the foreskin (prepuce) of the penis. This device was designed technologically using molding process to replace the conventional method.

{4}------------------------------------------------

It was designed specifically to promote faster circumcision process, prevent severe bleeding with less pain. It contains a set of clamp with a protective shield, RAPIDECLAMP™ measuring plate, and a blade or scissor. This device known as RAPIDECLAMP™ is made up of Polycarbonate (PC) and Acrylonitrile butadiene styrene (ABS).

Generally, RAPIDECLAMP™ consists of Guide Ring, Spike Body-Front and Rear, Anti-Slip Ring, Locking Cover, and Protective Shield.

### Model Numbers and Description:

RC50308 – Male Circumcision Kits: RAPIDECLAMP™ Size: 8mm Patients: Infant to children

RC50311 – Male Circumcision Kits: RAPIDECLAMP™ Size: 11mm Patients: Infant to children

RC50314 – Male Circumcision Kits: RAPIDECLAMP™ Size: 14mm Patients: Infant to children

RC50317 – Male Circumcision Kits: RAPIDECLAMP™ Size: 17mm Patients: Teenagers to adult

RC50320 – Male Circumcision Kits: RAPIDECLAMP™ Size: 20mm Patients: Teenagers to adult

RC50323 – Male Circumcision Kits: RAPIDECLAMP™ Size: 23mm Patients: Teenagers to adult

RC50326 – Male Circumcision Kits: RAPIDECLAMP™ Size: 26mm Patients: Teenagers to adult

RC50329 - Male Circumcision Kits: RAPIDECLAMP™ Size: 29mm Patients: Teenagers to adult

{5}------------------------------------------------

## Indication For Use:

Circumcision of foreskin of males from infant to adult by circumferential excision of the foreskin or prepuce at or near the level of coronal sulcus, with minimal amount of preputial skin remaining.

| Characteristic           | Predicate<br>(K032091)                                                                                                                                                                                                                                                         | Predicate<br>(K121789)                                                                                                                                                                                            | Applicant                                                                                                                                                                                                                                                                                                                      |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use             | The device is<br>indicated for<br>circumcision<br>of newborns and<br>older mates,<br>defined as<br>circumferential<br>excision of the<br>foreskin or<br>prepuce at or<br>near the level of<br>the coronal<br>sulcus, with<br>minimal amount<br>of preputial skin<br>remaining. | The ShangRing<br>device is<br>indicated for<br>circumcision of<br>adult males,<br>defined as<br>circumferential<br>excision of the<br>foreskin or<br>prepuce at or<br>near the level or<br>the coronal<br>sulcus. | Circumcision of<br>foreskin of males<br>from infant to adult<br>by circumferential<br>excision of the<br>foreskin or<br>prepuce at or near<br>the level of coronal<br>sulcus, with<br>minimal amount of<br>preputial skin<br>remaining.                                                                                        |
| Clamping method          | Clamp on the<br>foreskin to<br>prevent severe<br>bleeding                                                                                                                                                                                                                      | Same                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                           |
| Patient population       | Newborn to<br>adult males                                                                                                                                                                                                                                                      | Adult males                                                                                                                                                                                                       | Infant to adult<br>males                                                                                                                                                                                                                                                                                                       |
| Size                     | 14 sizes (2.6,<br>2.8, 2.9, 3.0,<br>3.1, 3.2, 3.3,<br>3.4, 3.5, 3.6,<br>3.7, 3.8, 3.9,<br>and 4.0 cm)                                                                                                                                                                          | 4 sizes (1.0, 1.3,<br>1.6 and 2.1cm)                                                                                                                                                                              | 8 sizes (8, 11, 14,<br>17, 20, 23, 26,<br>29mm)                                                                                                                                                                                                                                                                                |
| Materials                | Cylinder:<br>transparent<br>polycarbonate<br>Outer clamp:<br>nylon                                                                                                                                                                                                             | Inner and outer<br>rings:<br>Polycarbonate<br>Soft lining of<br>surrounding                                                                                                                                       | Guide ring, Spike<br>Body, Anti-Slip<br>Ring, Protective<br>Shield:<br>Polycarbonate                                                                                                                                                                                                                                           |
|                          |                                                                                                                                                                                                                                                                                | inner ring:                                                                                                                                                                                                       | Locking Cover:                                                                                                                                                                                                                                                                                                                 |
|                          |                                                                                                                                                                                                                                                                                | Silicone                                                                                                                                                                                                          | ABS                                                                                                                                                                                                                                                                                                                            |
| Sterilization            | Sterile                                                                                                                                                                                                                                                                        | Sterile                                                                                                                                                                                                           | Sterile                                                                                                                                                                                                                                                                                                                        |
| Single use               | Yes                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                            |
| Need for suture          | No                                                                                                                                                                                                                                                                             | No                                                                                                                                                                                                                | No                                                                                                                                                                                                                                                                                                                             |
| Removal of clamp         | Remove by<br>doctor                                                                                                                                                                                                                                                            | Remove by<br>doctor                                                                                                                                                                                               | 1. Automatically<br>drop off after<br>healed.<br>2. Using doctor.<br>The doctor must<br>break the safety<br>locks on the Guide<br>Ring (A) with<br>pliers, unscrew the<br>Locking Cover (D),<br>remove the Anti-<br>Slip Ring, cut of<br>gangreneous<br>tissue, remove<br>Spike Body (B1)<br>and removed the<br>Guide Ring (A) |
| Biocompatibility         | Compliant with<br>ISO 10993                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                           |
| Mechanism of<br>Action   | The device is<br>clamped on the<br>foreskin to<br>prevent blood<br>flow into<br>prepuce                                                                                                                                                                                        | Same                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                           |
| Dorsal Incision          | The excess<br>foreskin is<br>incised using<br>medical blade or<br>scissor                                                                                                                                                                                                      | The excess<br>foreskin is<br>incised using<br>medical blade or<br>scissor                                                                                                                                         | The excess<br>foreskin is incised<br>using medical<br>blade or scissor                                                                                                                                                                                                                                                         |
| Additional<br>Instrument | No any<br>additional<br>instrument. Only<br>a set of<br>Smartklamp,<br>Size O-meter,<br>medical scissors<br>or blade                                                                                                                                                           | No any<br>additional<br>instrument. Only<br>a set of<br>Shangring,<br>Special<br>measuring<br>tape, medical<br>scissors or<br>blade                                                                               | No any additional<br>instrument. Only a<br>set of<br>RAPIDECLAMP™, measuring plate<br>and medical<br>scissors or blade.                                                                                                                                                                                                        |

# Technological Characteristics Compared To Predicate Device:

{6}------------------------------------------------

{7}------------------------------------------------

# Summary of Non-clinical Testing:

MEDICFIT TECHNOLOGY has conducted mechanical testing: Locking Test to determine a known force required to confirm that the circumcision clamp is secured properly during the use. The force and break tests were also performed on the RAPIDECLAMP.

This test was conducted to supplement the clinical data collected to demonstrate the safety and effectiveness of the RAPIDECLAMP.

All materials used in the RAPIDECLAMP have been evaluated according to tests outlined in ISO 10993-1 and meet the requirements of Bluebook Memo, General Program Memorandum G95-1 biocompatibility testing for cytotoxicity, sensitization, and irritation. The materials used are Acrylonitrile Butadiene Styrene (ABS) and Polycarbonate (PC)

## Summary of Clinical Testing:

Clinical test were conducted in Malaysia at following clinics/ Hospitals:

- a) Ar-Raudhah Clinic
- b) JB Specialist

These studies have demonstrated that the RAPIDECLAMP is easily applied in a short surgical procedure with no suturing required. The clinical data collected confirms that the RAPIDECLAMP is a safe and effective means of performing circumcision on adult males.

## Conclusions Drawn From Non-Clinical And Clinical Tests:

The results of the non-clinical and clinical tests show that RAPIDECLAMP meets all safety requirements, and is substantially equivalent in clamping method, sterility, single use, no use of suture and intended use to the predicate devices.

The data collected confirms that the differences in the design between the RAPIDECLAMP and the predicate devices do not raise any new issues of safety and effectiveness. The information and data collected support a claim of substantial equivalence of the RAPIDECLAMP to the specified predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HFX/K142163](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HFX/K142163)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
