← Product Code [HCZ](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HCZ) · K134047

# JARIT GYNECOLOGICAL FORCEPS, MILTEX GYNECOLOGICAL FORCEPS, MEISTERHAND GYNECOLOICAL FORCEPS (K134047)

_Integra LifeSciences Corporation · HCZ · Apr 11, 2014 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HCZ/K134047

## Device Facts

- **Applicant:** Integra LifeSciences Corporation
- **Product Code:** [HCZ](/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HCZ.md)
- **Decision Date:** Apr 11, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

Integra® Gynecological Surgical Forceps are hand-held instruments with dual blades that are indicated for pulling, grasping, holding, or compressing tissue during gynecological procedures.

## Device Story

Integra® Gynecological Surgical Forceps are manual, hand-held, reusable surgical instruments featuring dual blades. Designed for use by surgeons during vaginal and abdominal gynecological procedures, these instruments provide a secure grip on tissue while minimizing laceration or perforation. The device functions through mechanical force applied by the operator to pull, grasp, hold, or compress tissue. The forceps are constructed from stainless steel and are provided non-sterile, requiring sterilization by the healthcare facility before use. The device is intended to assist in surgical manipulation of tissue, thereby facilitating gynecological interventions.

## Clinical Evidence

Bench testing only. No clinical data was required or provided. Testing included manual and automated cleaning validation, pre-vacuum and gravity displacement steam sterilization validation, and corrosion testing. All tests passed. Biocompatibility was established via the use of AISI 420 stainless steel, which has a long history of safe use in surgical instruments.

## Technological Characteristics

Manual, hand-held, reusable surgical forceps. Materials: AISI 420 stainless steel (ASTM F899-11, ISO 7153-1). Design: 1-piece, non-malleable, varied sizes. Energy source: None (manual). Sterilization: Steam sterilization (autoclave) per ANSI/AAMI ST79 and ISO 13402. Compliance: ASTM F1089-02 (corrosion).

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- Miltex® Gynecological Forceps (Preamendment)
- Forceps Obstetrical/Gynaecological ([K092840](/device/K092840.md))
- Miniclamps, Stabiliz, Fallopian Tube (Preamendment)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary 1

### 510(k) Summary

# APR : 1 2014

| Submitted by:          | Integra York PA, Inc.                                                                                                                                                                                                                                       |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                        | 589 Davies Drive                                                                                                                                                                                                                                            |
|                        | York, PA 17402 USA                                                                                                                                                                                                                                          |
| Contact Person:        | Ruthanne Vendy                                                                                                                                                                                                                                              |
|                        | Regulatory Affairs Manager                                                                                                                                                                                                                                  |
|                        | Integra York PA, Inc.                                                                                                                                                                                                                                       |
|                        | 589 Davies Drive, York, PA 17402 USA                                                                                                                                                                                                                        |
|                        | Phone: (717) 757-7974                                                                                                                                                                                                                                       |
| Date Prepared:         | January 31, 2014                                                                                                                                                                                                                                            |
| Device Trade Name:     | Integra® Gynecological Forceps                                                                                                                                                                                                                              |
| Brands:                | Jarit®, Miltex®, and MeisterHand®                                                                                                                                                                                                                           |
| Device Common Name(s): | Gynecological Forceps; Vulsellum Forceps; Uterine Vulsellum<br>Forceps; Tumor Forceps; Uterine Elevating Forceps; Hysterectomy<br>Forceps; Uterine Artery Forceps; Membrane Puncturing Forceps;<br>Placenta Forceps; Uterine Polyps Forceps; Forceps; Clamp |
| Classification Name:   | Forceps, Surgical, Gynecological                                                                                                                                                                                                                            |
| Device Class:          | Class II                                                                                                                                                                                                                                                    |
| Product Code:          | HCZ                                                                                                                                                                                                                                                         |
| CFR Classification:    | 21 CFR 884.4530                                                                                                                                                                                                                                             |

#### Device Description:

Integra® Gynecological Surgical Forceps are designed to provide a secure grip on tissue during vaginal and abdominal hysterectomy procedures. These atraumatic instruments are designed to minimize laceration or perforation of the tissue even when required retraction is exerted. Due to differences in anatomy of the site and types of surgical procedures, a variety of configurations and models are offered to provide the surgeon with options to ensure clinical needs are met.

#### Indications For Use:

Integra® Gynecological Surgical Forceps are hand-held instruments with dual blades that are indicated for pulling, grasping, holding, or compressing tissue during gynecological procedures.

{1}------------------------------------------------

#### Predicate Devices:

|              | Device                               | Manufacturer                  |
|--------------|--------------------------------------|-------------------------------|
| Preamendment | Miltex® Gynecological Forceps        | E. Miltenberg, Inc.           |
| K092840      | Forceps Obstetrical/Gynaecological   | Instrumed International, Inc. |
| Preamendment | Miniclamps, Stabiliz, Fallopian Tube | Sklar Corporation             |

#### Technological Comparison of Subject Devices to Predicates:

| Integra York PA, Inc.<br>Gynecological Forceps | E. Miltenberg, Inc.<br>(now Integra York<br>PA, Inc.)<br>Gynecological<br>Forceps | Instrumed<br>International, Inc.<br>OB/GYN Forceps                         | Sklar Corporation<br>Gynecological<br>Forceps                              |                                                                            |
|------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| 510(k) #                                       | Subject of Submission                                                             | Preamendment                                                               | K092840                                                                    | Preamendment                                                               |
| Class                                          | II                                                                                | II                                                                         | II                                                                         | II                                                                         |
| Pro Code                                       | HCZ                                                                               | HCZ                                                                        | HCZ                                                                        | HCZ                                                                        |
| Regulation #                                   | 884.4530                                                                          | 884.4530                                                                   | 884.4530                                                                   | 884.4530                                                                   |
| Classification                                 | Obstetric-gynecologic<br>manual instrument                                        | Obstetric-gynecologic<br>manual instrument                                 | Obstetric-gynecologic<br>manual instrument                                 | Obstetric-gynecologic<br>manual instrument                                 |
| Intended Use                                   | used to pull, grasp, or<br>compress during<br>gynecological<br>examination        | used to pull, grasp, or<br>compress during<br>gynecological<br>examination | used to pull, grasp, or<br>compress during<br>gynecological<br>examination | used to pull, grasp, or<br>compress during<br>gynecological<br>examination |
| Patient<br>Contact<br>Materials                | Stainless steel                                                                   | Stainless steel                                                            | Stainless steel                                                            | Stainless steel                                                            |
| Malleability                                   | Non-malleable                                                                     | Non-malleable                                                              | Non-malleable                                                              | Non-malleable                                                              |
| Sterility                                      | Non-sterile                                                                       | Non-sterile                                                                | Non-sterile                                                                | Non-sterile                                                                |
| Sterilization                                  | Sterilizable                                                                      | Sterilizable                                                               | Sterilizable                                                               | Sterilizable                                                               |
| Utility                                        | Reusable                                                                          | Reusable                                                                   | Reusable                                                                   | Reusable                                                                   |
| Energy<br>Source                               | N/A (manual)                                                                      | N/A (manual)                                                               | N/A (manual)                                                               | N/A (manual)                                                               |
| Device Sizes                                   | Varied                                                                            | Varied                                                                     | Varied                                                                     | Varied                                                                     |
| Components                                     | 1-piece                                                                           | 1-piece                                                                    | 1-piece                                                                    | 1-piece                                                                    |
| Substantially<br>Equivalent                    | N/A                                                                               | Yes1                                                                       | Yes1                                                                       | Yes1                                                                       |

1 Evaluative Statement: These devices are completely comparable to the predicate instruments. There are no significant differences in lechnology, intended use, or design between the predicates selected. The minor differences between patterns have no impact on equivalence.

#### Performance Standards:

No performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, the following standards were used to assess the Integra® Gynecological Forceps:

- . AAMI TIR12:2010 - Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers
- . AAMI TIR30:2011 - A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices

{2}------------------------------------------------

- ANSI/AAMI ST79:2010/A1:2010/A2:2011/A3:2012 Comprehensive guide to steam . sterilization and sterility assurance in health care facilities
- . ASTM F 899-11 - Standard Specification for Wrought Stainless Steels for Surgical Instruments
- . ASTM F 1089-02 - Standard Test Method for Corrosion of Surgical Instruments
- ISO 13402:1995 Surgical and dental hand instruments -- Determination of resistance . against autoclaving, corrosion and thermal exposure
- . ISO 7153-1:2001 - Surgical instruments -- Metallic materials -- Part 1: Stainless steel

| Non-Clinical Testing Performed                                                      | Results |
|-------------------------------------------------------------------------------------|---------|
| Manual Cleaning Validation                                                          | Pass    |
| Automated Cleaning Validation                                                       | Pass    |
| Pre-Vacuum Steam Sterilization Validation at 270°F (132°C) for 4 minutes            | Pass    |
| Gravity Displacement Steam Sterilization Validation at 270°F (132°C) for 15 minutes | Pass    |
| Corrosion Testing                                                                   | Pass    |

No biocompatibility testing was performed on the proposed devices, as stainless steel AISI 420 is recognized as a suitable stainless steel for surgical instruments per ASTM F899-11 and ISO 7153-1. AlSI 420 has a long history of safe and effective use, and has been used in other medical devices that have been cleared by the FDA (e.g., K971977).

#### Conclusions drawn from Non-Clinical Data:

All necessary testing has been performed on the Integra® Gynecological Forceps and the results support the conclusion that the subject devices are substantially equivalent to the legally marketed predicate devices based on intended use, materials, technology, and design and as such, do not raise any concerns of safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular fashion around a stylized eagle emblem. The eagle is depicted with three curved lines representing its body and wings. The text is in all caps.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2014

Integra LifeSciences Corporation Integra York PA, Inc. Ruthanne Vendy, RAC Regulatory Affairs Manager 589 Davies Drive York, PA 17402

Re: K134047

> Trade/Device Name: Integra® Gynecological Forceps Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HCZ Dated: January 28, 2014 Received: February 3, 2014

Dear Ruthanne Vendy,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Ruthanne Vendy, RAC

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/4/Picture/7 description: The image shows the name "Herbert P. Lerner - S" in a bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and the text is easy to read. The text appears to be a signature or a nameplate.

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

K134047 510(k) Number (if known):

Integra® Gynecological Forceps Device Name:

Indications for Use:

Integra® Gynecological Surgical Forceps are hand-held instruments with dual blades that are indicated for pulling, grasping, holding, or compressing tissue during gynecological procedures.

Prescription Use_ V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter-Use (Part 21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner 2014.04.11 15:35:24 -04'00'

Image /page/5/Picture/11 description: The image shows the word "INTEGRA" in bold, black letters. Below the word, there is some smaller text that is difficult to read. To the right of the word, there is a small graphic that appears to be a collection of squares and circles.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HCZ/K134047](https://fda.innolitics.com/submissions/OB/subpart-e%E2%80%94obstetrical-and-gynecological-surgical-devices/HCZ/K134047)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
