K222492 · Lujena, Inc. · HDX · Jun 27, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K222492
Device Name
Hope&Her Vaginal Dilators
Applicant
Lujena, Inc.
Product Code
HDX · Obstetrics/Gynecology
Decision Date
Jun 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.3900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hope&Her Vaginal Dilators are indicated for women who need vaginal dilation for an examination, in preparation for a surgical procedure, or for the relief of vaginismus.
Device Story
Hope&Her Vaginal Dilators are a series of reusable, non-sterile, single-user rigid dilators; diameters range from 15 mm to 38 mm. Device includes a handle for manual operation. Used by patients to provide passive distention of vaginal tissues in controlled stages. Prescription-only device. Patients use the dilators to gradually increase vaginal diameter. Benefits include facilitating medical examinations, preparing for surgical procedures, and treating vaginismus. Device is cleaned by the user with mild soap and boiling water.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), vaginal irritation (ISO 10993-23), and acute systemic toxicity (ISO 10993-11). Results confirmed device is non-cytotoxic, non-irritating, non-sensitizing, and not systemically toxic. Physical property testing included appearance, diameter, and tensile strength.
Technological Characteristics
Materials: Polyester and polyamide. Design: Conical, rigid dilators with handle. Dimensions: 15-38 mm diameter range. Operating principle: Passive mechanical distention via progressively larger sizes. Maintenance: Manual cleaning with mild soap and boiling water. Non-sterile. Standalone device.
Indications for Use
Indicated for women requiring vaginal dilation for medical examinations, surgical preparation, or relief of vaginismus.
Regulatory Classification
Identification
A vaginal stent is a device used to enlarge the vagina by stretching, or to support the vagina and to hold a skin graft after reconstructive surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 5, 2023
Lujena, Inc. % Alan Donald President Matrix Medical Consulting, Inc. 8880 Rio San Diego Drive, Suite 800 San Diego, CA 92108
Re: K222492 Trade/Device Name: Hope&Her Vaginal Dilators Regulation Number: 21 CFR§ 884.3900 Regulation Name: Vaginal Stent Regulatory Class: II Product Code: HDX
Dear Alan Donald:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 27, 2023. Specifically, FDA is updating this SE Letter to correct the contact information in the 510(k) Summary as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jason Roberts, Ph.D., OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, (240) 402-6400, Jason.Roberts@fda.hhs.gov.
Sincerely,
Reginald K. Avery -S
### for
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG" are written in blue, followed by the word "ADMINISTRATION" in a smaller font size.
June 27, 2023
Lujena, Inc. % Alan Donald President Matrix Medical Consulting, Inc. 8880 Rio San Diego Drive, Suite 800 San Diego, CA 92108
Re: K222492
> Trade/Device Name: Hope&Her Vaginal Dilators Regulation Number: 21 CFR§ 884.3900 Regulation Name: Vaginal Stent Regulatory Class: II Product Code: HDX Dated: Mav 24, 2023 Received: May 24, 2023
Dear Alan Donald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Reginald K. Avery -S
for
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22492
Device Name Hope&Her Vaginal Dilators
Indications for Use (Describe)
Hope&Her Vaginal Dilators are indicated for women who need vaginal dilation for an examination, in preparation for a surgical procedure, or for the relief of vaginismus.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) Summary – K222492 Hope&Her Vaginal Dilators
#### General Information on Submitter l.
| Applicant: | Lujena, Inc. |
|-----------------|---------------------------------------------|
| Address: | 772 Jamacha Road. #80<br>El Cajon, CA 92019 |
| Telephone: | 1 619 449 1200 |
| Contact Person: | Troy Gemmer |
| Contact Title: | President |
| Email: | lujena@icloud.com |
| Date Prepared: | June 15, 2023 |
#### II. General Information on Device
| Name of Device: | Hope&Her Vaginal Dilators |
|--------------------------------|---------------------------|
| Dilators Common of Usual Name: | Vaginal Dilators |
| Regulation Number: | 21 CFR 884.3900 |
| Classification Name: | Vaginal Stent |
| Regulatory Class: | II |
| Product Code: | HDX (Dilator, Vaginal) |
| Classification Panel: | Obstetrics/Gynecology |
#### lll. Predicate Device
| Predicate Device | 510(k) Number |
|------------------|---------------|
| Amielle | K983045 |
This predicate device has not been subject to a design-related recall.
#### IV. Description of Device
Hope&Her Vaginal Dilators is a series of reusable, non-sterile, single user vaginal dilators that are used by patients to provide passive distention of the vaginal tissues using various sized dilators. The subject device has various sized rigid dilators that range from 15 mm to 38 mm in diameter (15, 22, 26, 30, 34, and 38 mm) and a handle to hold the dilator. The subject device is prescription-only. The subject device is composed polyester and polyamide. The subject device is distributed in heat-seal plastic bags.
#### V. Indications for Use
{5}------------------------------------------------
Hope&Her Vaqinal Dilators are indicated for women who need vaginal dilation for an examination, in preparation for a surgical procedure, or for the relief of vaginismus.
Differences in Indications for Use:
Subject device: "Hope&Her Vaqinal Dilators are indicated for women who need vaqinal dilation for an examination, in preparation for a surgical procedure, or for the relief of vaginismus."
Predicate device: "The device is intended to treat women suffering from vaginismus and dyspareunia. Vaginismus is the involuntary spasm of the muscles in the vaqinal wall which then inhibits sexual intercourse by making it painful or impossible. Dyspareunia is the pain experienced during sexual intercourse caused by physical and/or emotional problems. The device comes in varying sizes, the most appropriate is then selected by the physician for use by the patient and the patient's partner as an assistant if appropriate. It is used as a tool to dilate the vaqina in controlled stages. The device can be autoclaved or sterilized by normal methods."
The difference between the subject and predicate device indications for use is that the predicate device includes additional indications for dyspareunia and also includes technological description of the device. As the subject device indications are a subset of the predicate device indications, this difference does not raise intended use concerns.
| | SUBJECT DEVICE<br>K222492 | PREDICATE DEVICE<br>K983045 |
|-----------------------------|------------------------------------|----------------------------------------------------|
| Device Name | Hope&Her Vaginal Dilators | Amielle |
| Regulation<br>Number | 21 CFR§ 884.3900 | 21 CFR§ 884.3900 |
| Regulation<br>Name | Vaginal Stent | Vaginal Stent |
| Regulatory Class | II | II |
| Over<br>the<br>Counter | No | No |
| Feature | Vaginal dilator, controlled stages | Vaginal dilator, controlled stages |
| Target<br>Population | Women suffering from vaginismus | Women suffering from vaginismus and<br>dyspareunia |
| Anatomical Site | Vagina | Vagina |
| Single<br>Patient<br>Device | Yes | Yes |
| Reusable | Yes | Yes |
#### VI. Comparison of Technological Characteristics with the Predicate Device
{6}------------------------------------------------
| Sterile | Non-sterile | Non-sterile |
|-----------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------|
| Device Design | Conical | Conical |
| Materials | Polyester and polyamide | Polybutylene terephthalate dilator |
| Diameter | Available from 15 - 38 mm;<br>increases diameter in metrics of 4 - 7 mm | Available from 15 - 35 mm;<br>increases diameter in metrics of 5 mm |
| Packaging | Packaged in heat-sealed bag with<br>instructions for use. | Packaged in a cardboard box with<br>instructions for use. |
| Operating<br>Principle | Dilate the vagina in controlled stages | Dilate the vagina in controlled stages |
| Resistive<br>component | Progressively larger dilators | Progressively larger dilators |
| Maintenance | Clean with mild soap and boiling water.<br>Towel dry. | Clean with mild soap and hot water. Towel dry. |
| Color | White and turquoise | White |
| Biocompatibility<br>Testing | Yes | Yes |
As noted in the table above, the subject device and predicate device are similar in all aspects, except for differences in target population, materials, dimensions, packaging, maintenance, and color. These differences do not raise different questions of safety or effectiveness and can be evaluated through performance testing.
#### VII. Summary of Non-Clinical Performance Testing
### Biocompatibility
Biocompatibility testing on the vaginal dilators was performed in accordance with the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2009 as follows:
- Cytotoxicity (ISO 10993-5:2009/(R)2014) ●
- Guinea Pig Maximization Sensitization Test (ISO 10993-10: 2010) ●
- Vaginal Irritation (ISO 10993-23: 2021) ●
- Acute Systemic Toxicity (ISO 10993-11:2017) ●
The results of this testing demonstrate that the subject device is non-cytotoxic, nonirritating, non-sensitizing, and not systemically toxic.
### Reprocessing
{7}------------------------------------------------
The device labeling included reprocessing information per FDA's 2015 guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling".
# Physical Properties
The following device characteristics were evaluated for the subject device:
- Appearance
- Diameter
- Tensile strength ●
#### VIII. Conclusion
Based on the results of the performance testing described above, the Hope&Her Vaginal Dilators are as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.