INTRAN 500
Device Facts
| Record ID | K961424 |
|---|---|
| Device Name | INTRAN 500 |
| Applicant | Utah Medical Products, Inc. |
| Product Code | HFN · Obstetrics/Gynecology |
| Decision Date | Jul 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.2700 |
| Device Class | Class 2 |
Indications for Use
The UMP Intran 500 is a sensor-tipped intrauterine pressure catheter used by obstetricians in monitoring the frequency and duration of contractions during labor and delivery, primarily indicated in “high risk” births.
Device Story
Sensor-tipped intrauterine pressure catheter; monitors uterine contraction frequency/duration during labor/delivery. Input: intrauterine pressure signals; amniotic fluid samples. Operation: catheter inserted into amniotic sac; pressure sensor transmits data to monitor; new molded clear plastic cylinder (View Port) and clip (Color Tab) added to amnio port for visual inspection of amniotic fluid color. Used in clinical labor/delivery settings by obstetricians. Output: pressure data for clinical monitoring; visual fluid sample for assessment. Benefit: facilitates contraction monitoring and fluid assessment in high-risk births.
Clinical Evidence
No clinical data provided. Safety inferred from historical usage data (1.4 million uses since 1991) and lack of product liability lawsuits.
Technological Characteristics
Sensor-tipped intrauterine pressure catheter. Features molded clear plastic cylinder (View Port) and plastic clip (Color Tab) for fluid visualization. Materials identical to predicate Intran-Plus. Sterilization and biocompatibility data maintained by manufacturer.
Indications for Use
Indicated for obstetricians monitoring frequency and duration of uterine contractions during labor and delivery, specifically for high-risk births.
Regulatory Classification
Identification
An intrauterine pressure monitor is a device designed to detect and measure intrauterine and amniotic fluid pressure with a catheter placed transcervically into the uterine cavity. The device is used to monitor intensity, duration, and frequency of uterine contractions during labor. This generic type of device may include the following accessories: signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
- Intran-400 (K961424)
Related Devices
- K964279 — SOFTRANS INTRAUTERINE PRESSURE CATHETER SYSTEM (IUP 4000 SOFTRANS IUPC) · Graphic Controls Corp. · Jan 22, 1997
- K974389 — KOALA INTRAUTERINE PRESSURE CATHETER, KOALA EXTERNAL INTRAUTERINE PRESSURE CATHETER · Clinical Innovations, Inc. · Feb 19, 1998
- K954955 — STORC INTRAUTERINE PRESSURE CATHETER · Clinical Innovations, Inc. · Apr 17, 1996
- K250897 — Koala Intrauterine Pressure Catheter (IPC-5000E) · Clinical Innovations, LLC · May 22, 2025
- K991235 — SOFTRANS TEMP INTRAUTERINE PRESSURE CATHETER SYSTEM (IUP 5000, SOFTRANS TEMP IUPC) · Ludlow Technical Products · Jul 8, 1999