K123335 · Sunray Medical Apparatus Co., Ltd. · HGM · Mar 21, 2013 · Obstetrics/Gynecology
Device Facts
Record ID
K123335
Device Name
SUNRAY FETAL MONITOR
Applicant
Sunray Medical Apparatus Co., Ltd.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Mar 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
SRF618B5 Fetal Monitor detects and displays single or twin (optional) Fetal Heart Rate (FHR), Fetal Movement (FM) and Uterine Activity (UA) in a real time on the color LCD viewer, and also provides the fetal heart beat sound with internal speaker. Ten hours of tracing may be stored and later retrieved for printing. It is intended for antepartum use by trained healthcare personnel. It is not intended for home use.
Device Story
Fetal monitor SRF618B5 provides continuous antepartum monitoring of fetal heart rate (FHR), fetal movement (FM), and uterine activity (UA). Device uses ultrasound Doppler transducer on maternal abdomen to detect fetal heart signals; TOCO sensor measures uterine activity. System processes signals to display waveforms and numerical data on color LCD; provides audible fetal heart sounds via internal speaker. Data stored in internal flash memory for later review and printing. Operated by trained healthcare personnel in clinical environments. Output assists clinicians in assessing fetal well-being during antepartum period.
Clinical Evidence
Bench testing only. Evaluated for safety and performance per IEC 60601-1 (general safety), IEC 60601-2-37 (ultrasonic diagnostic/monitoring equipment), IEC 60601-1-2 (EMC), and ISO 10993-1, -5, -10 (biocompatibility). No clinical data presented.
Technological Characteristics
Fetal monitor using 2MHz PW ultrasound transducer for Doppler FHR detection and TOCO sensor for UA. Features color LCD display, internal speaker, and internal flash memory for 10-hour data storage. Complies with IEC 60601-1, IEC 60601-2-37, and ISO 10993 standards. Connectivity includes printing capability.
Indications for Use
Indicated for antepartum monitoring of fetal heart rate (FHR), fetal movement (FM), and uterine activity (UA) in pregnant women, including single or twin pregnancies. Intended for use by trained healthcare personnel in clinical settings; not for home use.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
Sunray Medical Apparatus Co., Ltd. Sponsor: Fetal Monitor, model: SRF618B5 Subject Device: 510(k) submission report (V1.0), Chapter 4 File No.:
# MAR 2 1 2013
### Chapter 4. 510(k) Summary
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
### 1. Submitter Information
Sponsor Name: Sunray Medical Apparatus Co., Ltd. Address: 4/F No. 242 Tianhe Dong Road, Guangzhou, PR China Contact Person: Rong Jingbo (R&D Director) Tel: +86-20-87570362 / 87502927 Fax: +86-20-87583004 / 87514127 Email: rongjb@sunray.cn
## Application Correspondent Information:
MEDLAB (Shenzhen) Information Service Co., Ltd. Address: Rm. 2706-Building A, Zhongfang Jingyuan, Futian District, Shenzhen, PR China, 518034 Contact Person: Ms. Sabrina Wei (Manager) Tel: +86-755-83089699 Fax: +86-755-83089760 Email: sabrinawei@hotmail.com
#### 2. Subject Device Information
| Type of 510(k) submission: | Traditional |
|----------------------------|---------------------------------------------|
| Common Name: | Fetal Monitor |
| Trade Name: | Fetal Monitor, model: SRF618B5 |
| Classification Name: | Perinatal monitoring system and accessories |
| Review Panel: | Obstetrical and Gynecological |
| Product Code: | HGM, HGL |
| Regulation Number: | 884.2740 |
| Regulation Class: | 2 |
## 3. Predicate Device Information
Report by Sunray Medical Apparatus Co., Ltd.
{1}------------------------------------------------
| Sponsor: | Sunray Medical Apparatus Co., Ltd. | |
|--------------------|--------------------------------------------|------------------|
| Subject Device: | Fetal Monitor, model: SRF618B5 | |
| File No.: | 510(k) submission report (V1.0), Chapter 4 | |
| Sponsor: | Bionet Co., Ltd. | Bionet Co., Ltd. |
| Device Name: | FC-700 | FC-1400 |
| 510(k) Number: | K043597 | K043598 |
| Product Code: | HGM | HGM |
| Regulation Number: | 884.2740 | 884.2740 |
| Regulation Class: | 2 | 2 |
## 4. Device Description
SRF618B5 is a fetal monitor, providing continuous monitoring, displaying, printing and recording of single or twin (optional) Fetal Heart Rate (FHR), Fetal Movement (FM) and Uterine Activity (UA) for antepartum testing and monitoring. SRF618B5 irradiates the ultrasound wave to maternal abdomen, and detects the Doppler effect signal reflected from the heart of the fetus. SRF618B5 extracts FHR and FM from this signal and provides the fetal heart beat sound with internal speaker.
SRF618B5 measures the UA of a pregnant woman using TOCO sensor. SRF618B5 displays FHR, UA and FM with waveforms and numbers on the color LCD screen, saves them in internal flash memory and prints parts of them to review in details.
#### 5. Intended Use
SRF618B5 Fetal Monitor detects and displays single or twin (optional) Fetal Heart Rate (FHR), Fetal Movement (FM) and Uterine Activity (UA) in a real time on the color LCD viewer, and also provides the fetal heart beat sound with internal speaker. Ten hours of tracing may be stored and later retrieved for printing. It is intended for antepartum use by trained healthcare personnel. It is not intend for home use.
#### 6. Test Summary
SRF618B5 Fetal Monitor has been evaluated the safety and performance by lab bench testing according to the following standards:
- IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, + 1988+A1:1991+A2:1995
- IEC 60601-2-37, Medical electrical equipment Part 2-37: Particular requirements for the � basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2001+A1:2004+A2:2005
- IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic safety �
Report by Sunray Medical Apparatus Co., Ltd.
Page 4-2
{2}------------------------------------------------
Sponsor: Sunray Medical Apparatus Co., Ltd. Fetal Monitor, model: SRF618B5 Subject Device: 510(k) submission report (V1.0), Chapter 4 File No.:
and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, 2001+A1:2004
- ISO 10993-1, Biological evaluation of medical devices Part 1: Evaluation and testing, 2003 �
- ISO 10993-5, Biological evaluation of medical devices Part 5: Tests for In Vitro cytotoxicity, . 2009
- ISO 10993-10, Biological evaluation of medical devices Part 10: Tests for irritation and . delaved-type hypersensitivity, 2002+A1:2006
#### 7. Comparison to Predicate Device
Compare with predicate devices, the subject device is very similar in design principle, intended use, indication for use, functions, material and the applicable standards. The following differences between subject device and predicate devices do not raise and new questions of safety or effectiveness.
(1) Although the Power Supply, Working and Storage Environment, Dimensions and Weight, and some Safety Degree of subject device are a little different from predicate devices, they are both compliance with IEC 60601-1 requirements.
(2) Although some specifications of FHR (Fetal Heart Rate) and UC (Uterine Contraction Pressure) measurement for subject device are a little different from predicate devices, they can conduct their function normally.
(3) Although some specifications of Printer & Recorder, Display & Sound for subject device are a little different from predicate devices, these are only assistant functions.
(4) Although some Ultrasound Transducer Specification of subject device is a little different from predicate devices, they are both compliance with Track 1 requirement of "Guidance for Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers".
Conclusions: The subject device has all features of the predicate devices. The differences do not affect the safety and effectiveness of the subject device.
#### 8. Conclusion
The subject device has all features of the predicate devices. The few differences do not affect the safety and effectiveness of the subject device.
Thus, the subject device is substantially equivalent to the predicate devices.
#### 9. Summary Prepared Date: 17 January 2013
Report by Sunray Medical Apparatus Co., Ltd.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2013
Sunray Medical Apparatus Co., Ltd. % Ms. Sabrina Wei Project Manager MEDLAB (Shenzhen) Information Service Co., Ltd. B102, Nanshan Medical Devices Industrial Park No.1019 of Nanhai Ave. SHENZHEN GUANGDONG 518067 P.R. CHINA
Re: K123335
Trade/Device Name: Fetal Monitor, model: SRF618B5 Regulation Number: 21 CFR§ 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM, HGL Dated: January 17, 2013 Received: March 4, 2013
Dear Ms. Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Chapter 3. Indications for Use
# Indications for Use
510(k) Number (if known): K123335
Device Name: Fetal Monitor, model: SRF618B5
Indications for Use:
SRF618B5 Fetal Monitor detects and displays single or twin (optional) Fetal Heart Rate (FHR), Fetal Movement (FM) and Uterine Activity (UA) in a real time on the color LCD viewer, and also provides the fetal heart beat sound with internal speaker. Ten hours of tracing may be stored and later retrieved for printing. It is intended for antepartum use by trained healthcare personnel. It is not intended for home use.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert - S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K123335 510(k) Number
510(k) Number
{6}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
## System: Fetal Monitor, model: SRF618B5
Transducer: 2MHz PW Probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body for antepartum use as follows:
| | Mode of Operation | | | | | | | |
|---------------------------------|-------------------|---|---|-----|-----|---------------|-----------------------|---------------------|
| Clinical Application | | B | M | PWD | CWD | Color Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | | | | | | | | |
| Fetal / Obstetrics | | | | N | | | | |
| Abdominal | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | |
| Intra-operative (Neurological) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ (Specify) | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac | | | | | | | | |
| Intravascular | | | | | | | | |
| Peripheral vascular | | | | | | | | |
| Other (Urology) | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional Comment: The above probe is a 2 MHz PW transducer for the fetal heart rate (FHR) detection. There are two FHR transducers in this device. These two transducers are the same. Use one for single fetus, use two for twins.
Image /page/6/Picture/7 description: The image shows the text "Herbert P Lerner-S 2013.03.21 15:41:47". The text is in a simple, sans-serif font and is printed in black ink. The background of the image is white. The text appears to be a timestamp or a record of some kind.
Page 2 of 2
Report by Sunray Medical Apparatus Co.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.