Retrospective patient traces were used as part of the verification and validation testing to confirm the performance of the improved ultrasound algorithm and address issues with fetal heart rate signal processing.
Performance of ultrasound algorithm (addressing halving/doubling of fetal heart rate)
Indications for Use
Avalon Fetal Monitor FM20: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households. Avalon Fetal Monitor FM30: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households. Avalon Fetal Monitor FM40: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas. Avalon Fetal Monitor FM50: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Device Story
Perinatal monitoring system; inputs include ultrasound (fetal heart rate), maternal ECG, NIBP, SpO2, and intrauterine pressure (IUP) or direct ECG (DECG) for specific models. Device processes physiological signals to display real-time maternal and fetal data. Used by trained healthcare professionals in labor/delivery, antepartum clinics, and private homes. Output provides continuous monitoring of fetal/maternal status; assists clinicians in assessing fetal well-being and labor progression. Modification includes improved ultrasound performance algorithm to reduce signal artifacts like halving/doubling of fetal heart rate and maternal-fetal heart rate confusion.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level and regression testing, hazard analysis, and testing using previously recorded patient traces to confirm performance improvements in the ultrasound algorithm.
Technological Characteristics
Perinatal monitoring system (Class II, 21 CFR 884.2740). Features ultrasound transducers, maternal ECG, NIBP, SpO2, and IUP/DECG sensors. Software-based signal processing algorithm for fetal heart rate detection. Connectivity includes transducer/TMIF interface cables. System is designed for continuous physiological monitoring.
Indications for Use
Indicated for pregnant women requiring monitoring of uterine activity, heart rate, NIBP, pulse rate, and fetal heart rates (single, twins, triplets) in labor/delivery, antepartum, or home settings.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 (K071800)
Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 (K092028)
Submission Summary (Full Text)
{0}------------------------------------------------
# K10299
MAR 1 8 2011
8.0 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
The submitter of this premarket notification is: . Michael Asmalsky Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany
Tel: +49 7031 463-1277 Fax: +49 7031 463-2442 Email: michael.asmalsky@philips.com
This summary was prepared on March 18th, 2011
{1}------------------------------------------------
8.1 Indications The indications for use of the subject devices for Use has not changed.
#### Indications for Use:
Avalon Fetal Monitor FM20:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
#### Avalon Fetal Monitor FM30:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
## Avalon Fetal Monitor FM40:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas,
## Avalon Fetal Monitor FM50:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
{2}------------------------------------------------
8.2 Device Names
The names of the devices are the Philips Avalon Fetal Monitors FM20 (M2702A), FM30 (M2703A),FM40 (M2704A) and FM50 (M2705A).
legally marketed devices Philips 8.3 Device The Avalon Description Fetal/Maternal Monitors FM20, FM30, FM40 and FM50 offer monitoring of fetal and maternal heart rates, uterine activity, maternal ECG, maternal non-invasive blood pressure (NIBP) and oxygen saturation (Sp02) with pulse rate, during antepartum testing and labor and delivery.
> The Avalon Fetal/Maternal Monitor FM20 provides the following external measurement parameters:
- Up to three Fetal Heart Rates (FHR) via ultrasound (US)
- -Fetal Movement Profile
- Uterine activity via external Toco -
- -Maternal Pulse Rate
- -Maternal Heart Rate (MHR) via maternal ECG
- -Non-invasive blood pressure (NIBP)
Avalon FM30 shares all the features and The capabilities of the Avalon FM20. In addition, the Avalon FM30 provides the following external and internal measurement parameters:
- Single Fetal Heart Rate via direct ECG -(DECG)
- -Twin Fetal Heart Rates via direct ECG (DECG) and/or ultrasound ''
- -Uterine activity via intrauterine pressure (IUP)
- Maternal ECG (MECG) "
- י Pulse oximetry (maternal Sp02).
*) Note: maximum three FHR's can be monitored.
The Avalon Fetal/Maternal Monitor FM40 provides the following external measurement parameters:
- Up to three Fetal Heart Rates (FHR) via " ultrasound (US)
- Fetal Movement Profile .
- Uterine activity via external Toco י
- -Maternal Pulse Rate
- Maternal Heart Rate (MHR) via maternal ECG י
- Non-invasive blood pressure (NIBP). י
- -Pulse oximetry (maternal Sp02).
CONFIDENTIAL
Page 137
{3}------------------------------------------------
The Avalon FM50 shares all the features and
capabilities of the Avalon FM40. In addition, the Avalon FM50 provides the following external and internal measurement parameters:
- 'Single Fetal Heart Rate via direct ECG *) (DECG)
- Twin Fetal Heart Rates , via direct ECG (DECG) and/or ultrasound "
- Uterine activity via intrauterine pressure , (IUP)
- Maternal ECG (MECG) ,
" Note: maximum three FHR's can be monitored.
{4}------------------------------------------------
8.4 Device The Philips Avalon Fetal Monitors FM20 (M2702A), Classification FM30(M2703A), FM40(M2704A) and FM50(M2705A) and where applicable including their accessories are classified under ProCcode HGM, classification 884.2740, ·Perinatal monitoring system and their accessories.
| Device Panel | Classification | ProCode | Description |
|--------------------------------------------------------|----------------|---------|------------------------------------------------|
| Obstetrical and<br>Gynecological<br>Monitoring Devices | \$884.2740, II | HGM | Perinatal monitoring<br>system and accessories |
The subject devices Philips Avalon Fetal Monitors 8.5 Subject FM30, FM40 and FM50 are substantially devices FM20, equivalent to previously cleared Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 marketed pursuant to K071800 and K092028.
modification of the Philips Avalon Fetal 8.6 The Modifications Monitors FM20, FM30, FM40 and FM50 only provides an improved ultrasound performance algorithm of the released ultrasound transducers and TMIF interface cables to be used with the Philips Avalon Fetal monitors FM20, FM30, FM40 and FM50. The software changes were made to address problems with halving and doubling of the fetal heart. rate and mistaken display of maternal heart rate for fetal heart rate.
8.7 The subject devices Philips Avalon Fetal Monitors Technological FM20, FM30, FM40 and FM50 have the same Characteristics characteristics as the legally marketed predicate devices Avalon Fetal Monitors FM20, FM30, FM40 and FM50 .
{5}------------------------------------------------
K102958 page 6 of 6
8.8 Verification and Validation
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified devices with respect to the predicate. Testing involved system level and regression tests as well as testing from the hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence. Some tests included the use of previously recorded patient traces.
The results demonstrate that the Philips Avalon Fetal Monitors meet all reliability requirements and performance claims.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Michael Asmalsky Regulatory Affairs Engineer Philips Medizin Systeme Boblingen GmbH Hewlett-Packard Str. 2 71034 BOBLINGEN GERMANY
MAR 1 8 2011
Re: K102958
> Trade Name: Avalon Fetal/Maternal Monitors (FM20, FM30, FM40, & FM50) Regulation Number: 21 CFR §884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: February 21, 2011 Received: March 3, 2011
Dear Mr. Asmalsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{7}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for . the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Lenoir MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
## Enclosure
{8}------------------------------------------------
## Indications for Use
## 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: Philips Avalon Fetal Monitors FM20 (M2702A) , FM30 (M2703A) , FM40 (M2704A) and FM50 (M2705A) .
## Indications for Use:
## Avalon Fetal Monitor FM20:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parmeters uterine activity, heart rate, non-invasive blood pressure, palae rate of pregnant women and the fetal heart rates of single feuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
## Avalon Fetal Monitor FM30:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
### Avalon Fetal Monitor FM40:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
#### Avalon Fetal Monitor FM50:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Prescription Use ___Yes Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use __ No_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign Off)
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K102958
Page 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.