TENACORE TRANSDUCERS FOR FETAL ULTRASONIC AND TOKODYNAMOMETER MONITORING
K043075 · Tenacore Holdings, Inc. · HGM · Aug 18, 2005 · Obstetrics/Gynecology
Device Facts
Record ID
K043075
Device Name
TENACORE TRANSDUCERS FOR FETAL ULTRASONIC AND TOKODYNAMOMETER MONITORING
Applicant
Tenacore Holdings, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Aug 18, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
These transducers are intended to be used as replacement accessories for Hewlett Packard and Corometrics fetal monitors. These transducers measure fetal heart rate or uterine contractions in the gravid patient.
Device Story
Tenacore transducers serve as direct replacement accessories for existing perinatal monitoring systems. The device includes two types of transducers: an ultrasound transducer utilizing Doppler shift technology to detect fetal heart rate, and a tokodynamometer transducer utilizing a strain gauge to measure the duration, frequency, and relative pressure of uterine contractions. These transducers connect to compatible fetal monitors (e.g., Hewlett Packard 1350 series, Corometrics 115-150 series) which display the graphical relationship between labor and fetal heart rate. Used in clinical settings by healthcare providers, the output provides real-time physiological data to assist in labor management and fetal monitoring, potentially benefiting the patient by enabling continuous assessment of fetal well-being and labor progression.
Clinical Evidence
Bench testing only. The manufacturer declared conformance to consensus performance standards for electrical, electromagnetic compatibility, mechanical, efficacy, safety, and biocompatibility.
Technological Characteristics
Transducers for fetal ultrasound and tokodynamometer monitoring. Ultrasound transducer uses Doppler shift technology; tokodynamometer uses strain gauge technology. Designed as direct replacements for specific Hewlett Packard and Corometrics fetal monitor series. Conforms to consensus performance standards for electrical, electromagnetic, mechanical, and biocompatibility.
Indications for Use
Indicated for use in gravid patients during labor to monitor fetal heart rate and uterine contractions via compatible fetal monitoring systems. Prescription use only.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
Corometrics 116 (5700 & 2260) (K891S95)
Epic Medical Equipment Services transducer (K992811)
Submission Summary (Full Text)
{0}------------------------------------------------
043075
## enacore Holding
647 East Young Street - Santa Ana, CA 92705
Ph: 714-444-4643 Fx: 714-549-73
## Section 9: 510(k) Summary
AUG 1 8 2005
| Contact Person: | Brand Caso |
|-----------------|------------------------------|
| | Director Product Development |
| | Tenacore Holdings, Inc |
| | 647 East Young Street |
| | Santa Ana, CA 92705 |
March 1, 2004 Date Prepared:
Product Classification: Product Classification. Class 11-Transducers for use with perinatal monitoring systems as accessories as described in Title 21. Part 884, Subpart C, Classification 85 HGL, Regulation number Obstetrical Ultrasonic Transducer and Accessories and in Title 21, Part 884, Subpart C, Classification #85 HFM, Regulation number 884.2720 External Uterine Contraction Monitor and Accessories
Tenacore Transducers for Fetal Ultrasonic and Tokodynamometer Monitoring. Trade Name:
Common Name: Transducers for Fetal Ultrasonic and Tokodynamometer Monitoring
Predicate devices:
Corometrics 116 (5700 & 2260), 510 (k) # K891S95 by GE Marquette a. Medical Systems Epic's transducer, 510(k) #992811 by Epic Medical Equipment Services. b.
Tenacore Holdings, Inc. Transducers for Fetal Ultrasound and Tokodynamometer Description: Monitoring are direct replacements for similar transducers manufactured by GE Marquette Medical Systems and Epic Medical Equipment Services. A Transducer employing ultrasound Doppler Shift Technology is used for the detection of Fetal Heart Rate during labor and a Tokodynamometer Transducer, which actually is a strain gauge, is used to evaluate and measure the duration, frequency, and relative pressure of uterine contractions during labor.
Intended Use: Tenacore Transducers for Fetal Ultrasound and Tokodynamometer Monitoring arc intended to be used only as direct replacement accessories for appropriate Perinatal Monitoring Systems that display the graphical relationship between labor and fetal heart rate.
Performance
Tenacore Holdings, Inc has declared to conform to consensus performance standards Standards: concerning Electrical Electromagnetic Compatibility / Mechanical / Efficacy / Safety and Biocompatibility aspects of the product.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the HHS logo in the center. The logo is a stylized depiction of a human figure with three lines extending from the head, representing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 8 2005
Mr. Brand Caso Director of Product Development Tenacore Holdings, Inc. 647 East Young St. SANTA ANA CA 92705
Re: K043075
Trade/Device Name: Transducers for use with Perinatal Monitoring Systems and Accessories Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories
Regulatory Class: II Product Code: HGM Dated: July 1, 2005 Received: July 18, 2005
Dear Mr. Caso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device; subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
4
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotton will and in yourse of substantial equivalence of your device to a legally premated notined device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the 00 : 77). I Samall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known): K043075/S2
Device Name: Transducers for use with perinatal monitoring systems as accessories
Indications For Use:
These transducers are intended to be used as replacement accessories for Hewlett Packard and Corometrics fetal monitors. These transducers measure fetal heart rate or uterine contractions in the gravid patient.
## Tenacore
Compatible Monitors Transducer
| Model TFH102 | For use with Hewlett Packard 1350 Series Fetal Monitors |
|--------------|----------------------------------------------------------------------------------|
| Model TFC102 | For use with Corometrics Model 115, 116, 118, 120, and 150 Series Fetal Monitors |
| Model TFH101 | For use with Hewlett Packard 1350 Series Fetal Monitors |
| Model TFC101 | For use with Corometrics Model 115, 116, 118, 120, and 150 Series Fetal Monitors |
Prescription required.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
| AND/OR | |
| Over-The-Counter Use<br>(21 CFR 807 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _1_
David h. Legnon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K043075
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.