← Product Code [OSP](/submissions/OB/subpart-c%E2%80%94obstetrical-and-gynecological-monitoring-devices/OSP) · K153262

# PUREtrace (K153262)

_Nemo Healthcare BV · OSP · Feb 7, 2017 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-c%E2%80%94obstetrical-and-gynecological-monitoring-devices/OSP/K153262

## Device Facts

- **Applicant:** Nemo Healthcare BV
- **Product Code:** [OSP](/submissions/OB/subpart-c%E2%80%94obstetrical-and-gynecological-monitoring-devices/OSP.md)
- **Decision Date:** Feb 7, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.2720
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

PUREtrace™ is an electromyography (EMG) device that non-invasively measures intrapartum uterine activity. PUREtrace™ is indicated for use on women who are at term (≥37 completed weeks), in labor, with singleton pregnancies. The signal is acquired from surface electrodes that are placed on the maternal abdomen. PUREtrace™ is intended for use by healthcare professionals in a clinical setting.

## Device Story

Device registers electrical activity of uterine muscle (myometrium) in laboring patients; utilizes multi-surface electrode patch (Graphium patch) applied to maternal abdomen; acquires electromyography (EMG) signals (electrohysterography, EHG) generated by uterine muscle contractions; signals transmitted to PUREtrace hardware module; module converts EMG signals into uterine activity measurements; output sent to compatible external fetal monitor for display as tocogram; used by healthcare professionals in clinical settings; provides continuous uterine activity monitoring; assists clinicians in labor management; benefits patients by providing non-invasive alternative to invasive monitoring.

## Clinical Evidence

Observational study of 42 women at term in active labor. 40 usable UA registrations compared against tocodynamometry (TOCO) and intrauterine pressure catheter (IUPC). Three blinded clinical investigators assessed tracing interpretability and annotated uterine contractions. Results demonstrated PUREtrace performance is substantially equivalent to the predicate device for displaying uterine activity.

## Technological Characteristics

System includes PUREtrace hardware module, patch cable, and Graphium patch (disposable surface electrodes). Operates via transabdominal electromyography. Powered by AC mains (Class II medical grade). Connectivity via output connectors to external fetal monitors. Biocompatibility per ISO 10993-1; electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.

## Regulatory Identification

An external uterine contraction monitor (i.e., the tokodynamometer) is a device used to monitor the progress of labor. It measures the duration, frequency, and relative pressure of uterine contractions with a transducer strapped to the maternal abdomen. This generic type of device may include an external pressure transducer, support straps, and other patient and equipment supports.

## Predicate Devices

- Monica AN24 ([K112390](/device/K112390.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 7, 2017

Nemo Healthcare BV Aron Wierts Project Leader De Run 4630 5504 DB, Veldhoven The Netherlands

Re: K153262

Trade/Device Name: PUREtrace TM Regulation Number: 21 CFR§ 884.2720 Regulation Name: External Uterine Contraction Monitor and Accessories Regulatory Class: II Product Code: OSP Dated: January 26, 2017 Received: January 30, 2017

Dear Aron Wierts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Joyce M. Whang -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K153262

Device Name PUREtrace™

Indications for Use (Describe)

PUREtrace™ is an electromyography (EMG) device that non-invasively measures intrapartum uterine activity. PUREtrace™ is indicated for use on women who are at term (≥7 completed weeks), in labor, with singleton pregnancies. The signal is acquired from surface electrodes that are placed on the maternal abdomen. PUREtrace™ is intended for use by healthcare professionals in a clinical setting.

Type of Use (Select one or both, as applicable)

| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K153262

| Submitter                                       | Nemo Healthcare B.V.<br>De Run 4630<br>5504 DB, Veldhoven<br>The Netherlands<br>Phone: +31 40 751 6150<br>Fax: +31 40 751 6159                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:                                 | Aron Wierts                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Date Prepared:                                  | 03-Feb-17                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Proprietary or Trade Name:                      | PUREtrace™                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common/Usual Name:                              | External uterine contraction monitor                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification:                                 | 21 CFR 884.2720                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name:                            | External uterine contraction monitor and accessories                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulatory Class:                               | II                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product code:                                   | OSP, uterine electromyographic monitor                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                 | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                 | PUREtrace                                                                                                                                                                                                                                                                                                                                                                                                                                      | Monica AN24                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                 | K153262                                                                                                                                                                                                                                                                                                                                                                                                                                        | K112390                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Indications for<br>Use                          | PUREtrace™ is an<br>electromyography (EMG) device<br>that non- invasively measures<br>intrapartum uterine activity.<br>PUREtrace™ is indicated for use on<br>women who are at term (≥37<br>completed weeks), in labor, with<br>singleton pregnancies. The signal is<br>acquired from surface electrodes<br>that are placed on the maternal<br>abdomen. PUREtrace™ is intended<br>for use by healthcare professionals<br>in a clinical setting. | The Monica AN24 is an intrapartum<br>maternal-fetal monitor that non-<br>invasively measures and displays fetal<br>heart rate (FHR), uterine activity (UA),<br>and maternal heart rate (MHR). The<br>AN24 acquires and displays the FHR<br>tracing from abdominal surface<br>electrodes that pick up the fetal ECG<br>(fECG) signal. Using the same surface<br>electrodes, the AN24 also acquires and<br>displays the UA tracing from the uterine<br>electromyography (EMG) signal and the<br>MHR tracing from the maternal ECG<br>signal (mECG). The AN24 maternal-fetal<br>monitor is intended for use by<br>healthcare professionals in a clinical<br>setting. |
| Operating<br>principle (for UA<br>measurement)  | Transabdominal electromyography                                                                                                                                                                                                                                                                                                                                                                                                                | Transabdominal electromyography                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| System<br>components /<br>parts                 | • PUREtrace module (hardware<br>module)<br>• Patch cable<br>• Graphium patch (sensor)                                                                                                                                                                                                                                                                                                                                                          | • Monica AN24 recorder/ monitor<br>(hardware module)<br>• Electrode cable connector<br>• Electrodes (sensor)<br>• Battery charger AN24                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Patient interface<br>(materials)                | Surface electrodes incorporated in<br>single electrode patch                                                                                                                                                                                                                                                                                                                                                                                   | Surface electrodes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Patient interface<br>(use<br>characteristics)   | Single patient use, disposable                                                                                                                                                                                                                                                                                                                                                                                                                 | Single patient use, disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Power source                                    | AC Mains power (50/60Hz)<br>Class II medical grade power supply<br>unit                                                                                                                                                                                                                                                                                                                                                                        | Battery operated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Display of<br>(uterine activity)<br>information | Information to be displayed<br>on an<br>external fetal monitor                                                                                                                                                                                                                                                                                                                                                                                 | Information to be displayed on an<br>external device, such as a fetal monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

Predicate Device:

Monica Healthcare AN24 (K112390)

## Device Description:

PUREtrace is a device for registering the electrical activity of the uterine muscle (myometrium) of laboring patients. PUREtrace consists of a multi surface-electrode patch (Graphium patch) and a hardware module (PUREtrace module). The Graphium patch is applied to the abdomen of the patient and acquires the electromyography (EMG) signal of the uterine muscle (electrohysterography, EHG). During a contraction, a large number of uterine muscle cells contract in a controlled manner and the changes in potential can be measured as changes in electrical on the skin of the pregnant woman's abdomen. This signal is sent to the PUREtrace module, converted into a measure for uterine activity and subsequently sent to a compatible fetal monitor where it is displayed as a tocogram. PUREtrace is available in multiple variants with different output connectors to facilitate interfacing with various fetal monitors.

#### Indications for use:

PUREtrace™ is an electromyography (EMG) device that non-invasively measures intrapartum uterine activity. PUREtrace™ is indicated for use on women who are at term (≥37 completed weeks), in labor, with singleton pregnancies. The signal is acquired from surface electrodes that are placed on the maternal abdomen. PUREtrace™ is intended for use by healthcare professionals in a clinical setting.

{4}------------------------------------------------

# Predicate Comparison

The following table compares PUREtrace to the predicate device with respect to indications for use and technological characteristics:

{5}------------------------------------------------

PUREtrace has the same intended use but different technological characteristics compared to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness.

# Summary of non-clinical performance testing

Non-clinical tests were conducted to verify that PUREtrace met all design specifications and is substantially equivalent to the predicate. The test results demonstrated that the proposed device complies with the following standards:

- . Biocompatibility testing per ISO 10993-1:2009 including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2010) and irritation (ISO 10993-10:2010)
- . Electrical Safety testing per IEC 60601-1:2005 + CORR. 1 (2006) + CORR.2 (2007) with US deviations per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012
- Electromagnetic Compatibility testing per IEC 60601-1-2:2007

Software verification and validation testing was also conducted as recommended in FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005).

# Summary of clinical performance testing

An observational clinical study has been conducted with an enrollment of 42 women at term and in active labor, who were simultaneously monitored on uterine activity (UA) with PUREtrace, tocodynamometry (TOCO) and intrauterine pressure catheter (IUPC). 40 UA registrations that consisted of sufficient useable data were assessed to evaluate the performance of PUREtrace. The tracings were reviewed by three individual clinical investigators, who were blinded to the technology used and the study subjects corresponding with the tracings. The clinical investigators assessed the interpretability of the tracings. In addition, they annotated the tracings by marking all deflections they assessed to be uterine contractions. The study demonstrated that PUREtrace performed substantially equivalent to the predicate device for displaying uterine activity of term pregnant women in active labor.

## Substantial Equivalence Conclusion

Based on the comparison and analysis above, PUREtrace is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-c%E2%80%94obstetrical-and-gynecological-monitoring-devices/OSP/K153262](https://fda.innolitics.com/submissions/OB/subpart-c%E2%80%94obstetrical-and-gynecological-monitoring-devices/OSP/K153262)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
