K153475 · Edan Instruments, Inc. · KNG · Jan 29, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K153475
Device Name
SD5 & SD6 Ultrasonic Tabletop Doppler
Applicant
Edan Instruments, Inc.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Jan 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
The SD5 Ultrasonic TableTop Doppler (hereinafter called "SD5") and SD6 Ultrasonic TableTop Doppler (hereinafter called "SD6") are intended to be used by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices. The 2 MHz and/or 3 MHz obstetrical probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. They can also be used to verify fetal heart viability.
Device Story
Hand-held tabletop Doppler device for non-invasive fetal heart rate (FHR) monitoring. Uses 2 MHz or 3 MHz ultrasound probes to transmit acoustic energy to maternal abdomen; reflected ultrasound from fetal heart and blood flow is received by probe and converted to electrical signals. Main unit performs digital processing of waveform data for display on LED screen; audio signals routed to speaker. SD5 uses wired connection; SD6 uses Bluetooth for signal transfer. Powered by AC mains or internal Ni-MH battery; probes powered by internal Lithium battery. Used in hospitals, clinics, and private offices by clinicians to assess fetal well-being and viability. Output provides real-time FHR data to assist clinical decision-making during pregnancy and labor.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, biocompatibility evaluation (ISO 10993-1), electrical safety, EMC, and software verification/validation.
Technological Characteristics
Tabletop Doppler system. Materials: shell of probe and wire jacket identical to K150901. Sensing: Doppler shift of ultrasound (2 MHz/3 MHz). Power: AC (100-240V) or Ni-MH battery (main unit); Lithium battery (probes). Connectivity: Wired (SD5), Bluetooth (SD6). Safety: Class I, Type BF applied parts. Standards: IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2, ISO 10993-1, IEC 62304.
Indications for Use
Indicated for detection of fetal heart rate from early gestation through delivery, general indication of fetal well-being, and verification of fetal heart viability in pregnant patients. Used by healthcare professionals (nurses, midwives, ultrasound technicians, physician assistants) via prescription.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 8, 2016
Edan Instruments, Inc. % Ms. Queena Chen Regulatory Engineer 3/F - B, Nanshan Medical Equipments Park Nanhai Road 1019# Shenzhen, Guangdong 518067 CHINA
Re: K153475
Trade/Device Name: SD5 & SD6 Ultrasonic Tabletop Doppler Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: November 30, 2015 Received: December 2, 2015
Dear Ms. Chen:
This letter corrects our substantially equivalent letter of January 29, 2016. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
SD5&SD6 Ultrasonic Tabletop Doppler
Indications for Use (Describe)
The SD5 Ultrasonic TableTop Doppler (hereinafter called "SDS") and SD6 Ultrasonic TableTop Doppler (hereinafter called "SD6") are intended to be used by health care professionals including registered nurses, midwives, ultrasound technicians, and physician assistants, by prescription in hospitals, clinics and private offices.
The 2 MHz and/or 3 MHz obstetrical probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. They can also be used to verify fetal heart viability.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Ultrasonic Tabletop Doppler SD5
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|----------------------|--------------------------------|-------------------|---|-----|-----|----------------|--------------------|------------------|
| General | Specific | B | M | PWD | CWD | Colour Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalm | Ophthalmic | | | | | | | |
| | Fetal | | | | P | | | |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal | Small Organ (Specify) | | | | | | | |
| Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(Non-Card.) | | | | | | | |
| | Musculo-skel<br>(conventional) | | | | | | | |
| | Musculo-skel<br>(superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-Esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously licensed Additional Comments:
{4}------------------------------------------------
System: Ultrasonic Tabletop Doppler SD5
Transducer: 2 MHz CW fetal probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | Other*<br>(Specify) |
|----------------------|--------------------------------|-------------------|---|-----|-----|-------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Colour<br>Doppler | Combined<br>(Specify) | |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| | Fetal | | | | P | | | |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal | Small Organ (Specify) | | | | | | | |
| Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(Non-Card.) | | | | | | | |
| | Musculo-skel<br>(conventional) | | | | | | | |
| | Musculo-skel<br>(superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-Esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously licensed
{5}------------------------------------------------
System: Ultrasonic Tabletop Doppler SD5
Transducer: 3 MHz CW fetal probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|--------------------------------|-------------------|---|-----|-----|-------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Colour<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| | Fetal | | | | P | | | |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal<br>Imaging<br>& Other | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(Non-Card.) | | | | | | | |
| | Musculo-skel<br>(conventional) | | | | | | | |
| | Musculo-skel<br>(superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-Esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously licensed
{6}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Ultrasonic Tabletop Doppler SD6
| Clinical Application | | Mode of Operation | | | | | | |
|----------------------|--------------------------------|-------------------|---|-----|-----|-------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Colour<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| | Fetal | | | | P | | | |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal | Small Organ (Specify) | | | | | | | |
| Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(Non-Card.) | | | | | | | |
| | Musculo-skel<br>(conventional) | | | | | | | |
| | Musculo-skel<br>(superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-Esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P=previously licensed
{7}------------------------------------------------
System: Ultrasonic Tabletop Doppler SD6
Transducer: 2 MHz CW fetal probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|----------------------|--------------------------------|-------------------|---|-----|-----|-------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Colour<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| | Fetal | | | | P | | | |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal | Small Organ (Specify) | | | | | | | |
| Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(Non-Card.) | | | | | | | |
| | Musculo-skel<br>(conventional) | | | | | | | |
| | Musculo-skel<br>(superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-Esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously licensed
{8}------------------------------------------------
System: Ultrasonic Tabletop Doppler SD6
Transducer: 3 MHz CW fetal probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|--------------------------------|-------------------|---|-----|-----|-------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Colour<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalm<br>ic | Ophthalmic | | | | | | | |
| | Fetal | | | | P | | | |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| Fetal<br>Imaging<br>& Other | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph.<br>(Non-Card.) | | | | | | | |
| | Musculo-skel<br>(conventional) | | | | | | | |
| | Musculo-skel<br>(superficial) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-Esoph. (Cardiac) | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P=previously licensed
{9}------------------------------------------------
# 510(k) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc. |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou, Nanshan Shenzhen, 518067 P.R. China |
| | Tel.: (0755) 26858736 |
| | Fax: (0755) 26882223 |
| Contact Person: | Queena Chen |
| Prepare date: | 2015-11-30 |
| 2. Device name<br>and classification: | Device Name: Ultrasonic Tabletop Doppler |
| | Model: SD5, SD6 |
| | Classification Name:<br>21 CFR 884.2660 Fetal ultrasonic monitor and accessories |
| | Product code: KNG<br>Regulatory Class: Class II |
| 3. Reason for<br>Submission | Technology Change. |
| 4. Predicate<br>Device(s): | SD5&SD6 Ultrasonic Tabletop Doppler / K102138 / Edan Instruments, Inc. |
| 5. Device<br>Description: | The SD5&SD6 Ultrasonic Tabletop Doppler is a hand-held device for non-invasive measurement and display of fetal heart rate utilizing the principle of Doppler shift of an ultrasound. The ultrasound is transmitted from the probe to patient body (maternal abdominal wall), and moves through biophysical objects. The acoustic ultrasound is reflected by blood and moving objects such as the fetal heart. The reflected ultrasound is received by the probe and is converted into electric signals. The waveform data are applied to the CPU for all the digital processing on LED Display, operation keys. The audio signal is taken off for the routing to the speaker to generate the analogue signals before digital processing. The system is compatible with 2 MHz or 3 MHz for fetal heart rate. The signal acquired can be transferred to SD5 by wired method, and by Bluetooth to SD6.<br>The main unit can be powered by main supply (100 V-240 V~) or Ni-MH Battery (7.2 V/2000 mAh). And transducers are powered by 3.7 V/800 mAh Lithium Battery only. |
{10}------------------------------------------------
The SD5 Ultrasonic TableTop Doppler (hereinafter called "SD5") and SD6 6. Indications for Ultrasonic TableTop Doppler (hereinafter called "SD6") are intended to be Use: used by health care professionals including registered nurses, practical nurses, midwives, ultrasound technicians, and physician assistants, by prescription from licensed physicians in hospitals, clinics and private offices.
> The 2 MHz and/or 3 MHz obstetrical probes are indicated for the detection of fetal heart rate from early gestation thru delivery and as a general indication of fetal well being. They can also be used to verify fetal heart viability.
## 7. Predicate Device Comparison
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device.
Please refer to following table to find differences between the subject device and predicate device. All the differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
| Item | SD5/SD6 | SD5/SD6 | Comparison<br>Result |
|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacturer/K# | EDAN Instruments/<br>K102138 | Current Submission | --- |
| Intended Use/Indications for Use | | | |
| Intended<br>Use/Indications for<br>Use | The SD5 Ultrasonic<br>TableTop Doppler<br>(hereinafter called “SD5”)<br>and SD6 Ultrasonic<br>TableTop Doppler<br>(hereinafter called “SD6”)<br>are intended to be used by<br>health care professionals<br>including registered nurses,<br>practical nurses, midwives,<br>ultrasound technicians, and<br>physician assistants, by<br>prescription from licensed<br>physicians in hospitals,<br>clinics and private offices.<br>The 2 MHz and/or 3 MHz<br>obstetrical probes are<br>indicated for the detection of<br>fetal heart rate from early<br>gestation thru delivery and as<br>a general indication of fetal<br>well being. They can also be<br>used to verify fetal heart<br>viability following patient<br>trauma.<br>The 4 MHz, 5 MHz and/or 8<br>MHz vascular probes are<br>indicated for the detection of<br>blood flow in veins and<br>arteries for assisting in the<br>detection of peripheral<br>vascular disease. | The SD5 Ultrasonic TableTop<br>Doppler (hereinafter called<br>“SD5”) and SD6 Ultrasonic<br>TableTop Doppler (hereinafter<br>called “SD6”) are intended to be<br>used by health care professionals<br>including registered nurses,<br>practical nurses, midwives,<br>ultrasound technicians, and<br>physician assistants, by<br>prescription from licensed<br>physicians in hospitals, clinics<br>and private offices.<br>The 2 MHz and/or 3 MHz<br>obstetrical probes are indicated<br>for the detection of fetal heart rate<br>from early gestation thru delivery<br>and as a general indication of fetal<br>well being. They can also be used<br>to verify fetal heart viability. | Different |
| Operation characteristic | | | |
| Mode of operation | Continuous operation | Continuous operation | Same |
| Physical Specifications | | | |
| Dimensions | Main Unit: 265 mm x 155<br>mm x 180 mm | Main Unit: 265 mm x 155 mm x<br>180 mm | Same |
| Weight | Main Unit: 1.7 kg (with<br>battery and probe) | Main Unit: 1.7 kg (with battery<br>and probe) | Same |
| Display: | Effective display area: 62.0<br>mm (W) x 44.0mm(H)<br>Number of characters:<br>128*64 Dots<br>Dot size: 0.39 mm (W) x<br>0.55 mm (H)<br>Dot pitch: 0.44 mm (W) x<br>0.60 mm (H) | Effective display area: 62.0 mm<br>(W) x 44.0mm(H)<br>Number of characters: 128*64<br>Dots<br>Dot size: 0.39 mm (W) x 0.55<br>mm (H)<br>Dot pitch: 0.44 mm (W) x 0.60<br>mm (H) | Same |
| Environmental Specifications | | | |
| Working | Temperature: + 5 °C ~ + 40°C<br>( +41°F ~ +104°F)<br>Relative Humidity: 25% RH<br>~ 80% RH (non-condensing)<br>Atmospheric Pressure: 86<br>kPa~106 kPa | Temperature:0°C ~ + 40°C<br>(+32°F ~ +104°F)<br>Relative Humidity: 15% RH ~<br>95% RH (non-condensing)<br>Atmospheric Pressure: 86 kPa ~<br>106 kPa | Different |
| Transport and<br>Storage | Temperature: -20°C ~ +55°C<br>(-4°F ~ +131°F)<br>Relative Humidity: 25% RH<br>~ 93% RH (non-condensing)<br>Atmospheric Pressure: 70<br>kPa~106 kPa | Temperature: -20°C ~ +55°C (-4°F<br>~ +131°F)<br>Relative Humidity: 15% RH ~<br>95% RH (non-condensing)<br>Atmospheric Pressure: 70 kPa ~<br>106 kPa | Same |
| Safety Classifications | | | |
| Anti-electric Shock<br>Type | Class I equipment with<br>internal power supply | Class I equipment with internal<br>power supply | Same |
| Anti-electric Shock<br>Degree | Type B applied parts | Type BF applied parts | Different |
| Degree of Protection<br>against Harmful<br>Ingress of Water | Main Unit: Ordinary<br>equipment (Sealed equipment<br>without liquid proof)<br>Probe: IPX4 | Main Unit: Ordinary equipment<br>(Sealed equipment without liquid<br>proof)<br>Probe: IPX4…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.