Central Monitoring System Software is intended to conduct centralized antepartum and intrapartum monitoring of pregnant women's vital sign information from compatible bedside monitors. The software collects, stores, displays and alarms the information provided on the bedside monitor. The monitoring parameters include Electrocardiogram(ECG), Heart Rate(HR), Respiration(RESP), Pulse Oxygen Saturation(SpO2), Pulse Rate(PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Impedance Cardiograph(ICG), TEMP(Temperature), Carbon dioxide (CO2), Anesthetic Gas (AG), Fetal Heart Rate (FHR), Uterine contraction (TOCO) and Fetal Movement (FM). It is intended to be used in the hospital or medical institutions, and it is not intended for home use.
Device Story
Software-only device; collects, stores, displays, and alarms vital sign data from compatible bedside monitors via wired or wireless networks. Inputs include ECG, HR, RESP, SpO2, PR, NIBP, IBP, ICG, TEMP, CO2, AG, FHR, TOCO, and FM. Used in hospitals/medical institutions by clinicians at central nursing stations. Provides real-time visualization of multiple patients; triggers visual alerts for out-of-limit values or poor signal quality. Facilitates clinical decision-making by enabling remote review of patient status and archiving of monitoring data. Benefits include centralized oversight of maternal and fetal health, allowing timely intervention.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing verified compatibility with bedside monitors and acceptable data latency. Software verification and validation conducted per FDA guidance.
Technological Characteristics
Software-only device; operates on networked hardware. Connectivity via wired or wireless network. Complies with IEC 62304 (Software Life Cycle) and IEC 62366 (Usability Engineering).
Indications for Use
Indicated for centralized antepartum and intrapartum monitoring of pregnant women in hospital or medical institutional settings. Not for home use.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three faces in profile, representing health and well-being. The figure is positioned within a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling the top portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2016
Guangdong Biolight Meditech Co., Ltd. % Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai 200237 CHINA
Re: K153580
> Trade/Device Name: Central Monitoring System Software Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal Monitoring System and Accessories Regulatory Class: Class II Product Code: HGM Dated: August 5, 2016 Received: August 8, 2016
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Douglas Silverstein -S For Division 2016.09.07 15:26:22 -04'00' Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153580
Device Name Central Monitoring System Software
#### Indications for Use (Describe)
Central Monitoring System Software is intended to conduct centralized antepartum and intrapartum monitoring of pregnant women's vital sign information from compatible bedside monitors. The software collects, stores, displays and alarms the information provided on the bedside monitor. The monitoring parameters include Electrocardiogram(ECCG), Heart Rate(HR), Respiration(RESP), Pulse Oxygen Saturation(SpO2), Pulse Rate(PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Impedance Cardiograph(ICG), TEMP(Temperature), Carbon dioxide (CO2), Anesthetic Gas (AG), Fetal Heart Rate (FHR), Uterine contraction (TOCO) and Fetal Movement (FM). It is intended to be used in the hospital or medical institutions, and it is not intended for home use.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> | |
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K153580
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
- 1. Date of Preparation: 09/06/2016
- 2. Sponsor Identification
#### Guangdong Biolight Meditech Co., Ltd.
No.2 Innovation First Road, Technology Innovation Coast, Hi-Tech Zone, Zhuhai, Guangdong, 519085, P.R. China Establishment Registration Number: 3007305624 Contact Person: Jin Liang Position: R&D Director Tel: +86-756-3399967 Fax: +86-756-3399989 Email: liang j@blt.com.cn
- Designated Submission Correspondent 3. Ms. Diana Hong (Primary Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
- 4. Identification of Proposed Device Trade Name: Central Monitoring System Software Common Name: Central Monitoring System Software Model(s): Truscope CNS/ M6000C
Regulatory Information
Classification Name: Perinatal monitoring system and accessories Classification: Class II Product Code: HGM, system, monitoring, perinatal Regulation Number: 21 CFR 884.2740 Review Panel: Obstetrics/Gynecology
- 5. Identification of Predicate Device(s)
510(k) Number: K103172 Product Name: Obstetrical Data Management and Monitoring Software Application Model Name: CIV-OB (PLUS)
- Device Description 6.
The Central Monitoring System Software-only device that is intended to conduct centralized antepartum and intrapartum monitoring of pregnant women's vital sign information from compatible bedside monitors connected by a wired or wireless network. The Central Monitoring System consists of
{4}------------------------------------------------
two models: M6000C and Truscope CNS. Both models provide functions including collecting, storing, displaying, and alarming (e.g. when the monitored data exceeds a set value, the device will alarm to alert medical personnel) the information which is received from the bedside monitor(s). Multiple patients can be monitored simultaneously. Parameters monitored include ECG/HR, RESP, SpO2, PULSE, NIBP, IBP, ICG, TEMP, CO2, AG, FHR, TOCO and FM.
- 7. Indications for Use
Central Monitoring System Software is intended to conduct centralized antepartum and intrapartum monitoring of pregnant women's vital sign information from compatible bedside monitors. The software collects, stores, displays and alarms the information provided on the bedside monitor. The monitoring parameters include Electrocardiogram(ECG). Heart Rate(HR), Respiration(RESP), Pulse Oxygen Saturation(SpO2), Pulse Rate(PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Impedance Cardiograph(ICG), TEMP(Temperature), Carbon dioxide (CO2), Anesthetic Gas (AG), Fetal Heart Rate (FHR), Uterine contraction (TOCO) and Fetal Movement (FM). It is intended to be used in the hospital or medical institutions, and it is not intended for home use.
| 8. | Predicate Comparison |
|----|----------------------|
|----|----------------------|
| Item | Proposed Device(s) | Predicate Device(s) |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | HGM | HGM |
| Regulation Number | 21 CFR 884.2740 | 21 CFR 884.2740 |
| Intended Use | Central Monitoring System Software is<br>intended to conduct centralized<br>antepartum and intrapartum<br>monitoring of pregnant women's vital<br>sign information from compatible<br>bedside monitors. The software<br>collects, stores, displays and alarms the<br>information provided on the bedside<br>monitor. The monitoring parameters<br>include Electrocardiogram(ECG),<br>Heart Rate(HR), Respiration(RESP),<br>Pulse Oxygen Saturation(SpO2), Pulse<br>Rate(PR), Non-invasive Blood<br>Pressure (NIBP), Invasive Blood<br>Pressure(IBP), Impedance<br>Cardiograph(ICG),<br>TEMP(Temperature), Carbon dioxide<br>(CO2), Anesthetic Gas (AG), Fetal<br>Heart Rate (FHR), Uterine contraction<br>(TOCO) and Fetal Movement (FM). It<br>is intended to be used in the hospital or<br>medical institutions, and it is not<br>intended for home use. | The CIVNET CIV-obTM (plus) is<br>a clinical data managing software<br>application and is indicated for<br>antepartumn and intrapartumn<br>monitoring of pregnant women in a<br>healthcare setting.<br>The CIVNET CIV-obT* (plus) is<br>indented to manage perinatal<br>monitoring data acquired from<br>bedside monitors or manual inputs<br>for viewing at the central nursing<br>station. The system also produces<br>an electronic medical record.<br>The CIYNET CIV-ob1~" (plus)<br>has display fields for the following<br>obstetric data:<br>Monitoring site details Patient demographics, Provider notes, fetal heart rate (FHR), uterine activity(via<br>tocodynamometry or IUP) Head Decent Cervix Dilation |
| Sterile | Not applicable | Not applicable |
{5}------------------------------------------------
| Single Use | Not applicable | Not applicable |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | Not applicable | Not applicable |
| Network<br>connecting to<br>bedside monitor | Wired and Wireless | Wired |
| Display | Fetal heart rate, TOCO, maternal vital<br>signs, patient demographic data, and<br>notes.<br>Providing the means to display<br>multiple beds simultaneously. | Fetal heart rate, TOCO, maternal<br>vital signs, patient demographic<br>data, and notes.<br>Providing the means to display<br>multiple beds simultaneously. |
| Print | Print (locally or remotely) CTG,<br>patient record | Print (locally or remotely) CTG,<br>patient records s, and CIV-ob TM<br>(plus) data base definition (e.g.<br>item names). |
| Archive | CTG and maternal vital signs.<br>Providing the ability to archive files<br>to a secondary or tertiary storage<br>medium (i.e. optical disk).<br>Providing automatic archiving of the<br>data. | CTG and maternal vital signs.<br>Providing the ability to archive<br>files to a secondary or tertiary<br>storage medium (i.e. optical disk).<br>Saving data automatically. |
| Alarm | Visual alerts of fetal/maternal monitor<br>such as out-of-limit heart rate or poor<br>signal quality. | Visual alerts of fetal/maternal<br>monitor such as out-of-limit heart<br>rate or poor signal quality. |
| Electronic patient<br>record | Easy interfacing with any IT patient<br>record system for data acquisition,<br>viewing and storage of electronic<br>patient record. | Easy interfacing with any IT<br>patient record system for data<br>acquisition, viewing and storage of<br>electronic patient record. |
| Notes | Providing the user the ability to enter<br>comments and specific data. | Providing the user the ability to<br>enter comments and specific data. |
| Remote Access | Review fetal/maternal monitor data | Review fetal/maternal monitor data |
Central Monitoring System Software and the predicate device do not have identical indication for use statements; however, they have the same intended use (centralized antepartum and intrapartum monitoring of pregnant women's vital sign information).
In addition, there are differences in the device technological characteristics. The predicate device uses a wired network to connect to the bedside units, while the subject device gives the option of a wired or wireless network. Wireless networks are widely used for various applications. There are also differences in the parameters that the subject device and predicate collect, store, display and alarm. These differences do not raise different questions of safety as compared to the predicate device.
- 9. Non-Clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The Central Monitoring System also complies with voluntary standards for medical devices.
The test results demonstrated that the proposed device complies with the following standards:
- IEC 62366: 2007 Medical Devices Application of Usability Engineering to Medical Devices A
- IEC 62304: 2006 Medical Device Software Software Life Cycle Process >
{6}------------------------------------------------
The following data were provided to support the substantial equivalence determination:
- A Performance testing was conducted to verify that the Central Monitoring System Software is compatible with various bedside monitors.
- A Performance testing was conducted to verify that the data latency rates were acceptable for the clinical environment.
- A Software verification and validation testing as recommended in FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005)
- 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and performance data described above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.