The Philips M2720A Cordless Fetal Transducer System is indicated for pregnant women from about 22 weeks of gestation on through delivery. It is intended for continuous cordless fetal monitoring in connection with a fetal monitor for resting or ambulating patients, also usable during hydrotherapy, for antepartum testing and labor and delivery (intrapartum). The system should only be used by, or under the direct supervision of, a licensed physician or other health care practitioner who is trained in the use of fetal heart rate monitors and in the interpretation of fetal heart rate traces. It is not intended for home use.
Device Story
System measures, processes, and transmits fetal parameters to Philips Series 50 fetal monitors via radio telemetry. Comprises base station (M2720A) and three wireless transducers: Toco (M2725A) for external uterine activity; Ultrasound (M2726A) for fetal heart rate/movement; optional ECG/auxiliary (M2727A) for direct fetal scalp ECG or maternal ECG. Designed for ambulatory patients, including use in water/hydrotherapy. Replaces wired connections of predicate systems with integrated front-end circuitry in transducers. Operated by clinicians in clinical settings; output displayed on fetal monitor for trace interpretation to support labor/delivery management. Benefits include patient mobility during monitoring.
Clinical Evidence
Bench testing only. Verification and validation activities included environmental, electrical safety, EMC, radio telemetry, biocompatibility, and user evaluations for consumer accuracy. Hardware and software verification and cable interface testing were performed. No clinical trial data presented.
Technological Characteristics
Wireless perinatal monitoring system using radio telemetry. Includes base station and three transducers (Toco, Ultrasound, ECG/Aux). Features integrated front-end circuitry within transducers for direct signal processing. Designed for use in clinical environments, including hydrotherapy. Software-based verification conducted.
Indications for Use
Indicated for pregnant women from approximately 22 weeks gestation through delivery. Used for continuous fetal monitoring during antepartum testing and labor/delivery (intrapartum) for resting or ambulatory patients, including during hydrotherapy. Contraindicated for home use; requires supervision by trained healthcare practitioners.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{1}------------------------------------------------
## KD23931 p 1/2
## 510(k) Summary (As required by 21 C.F.R. §807.92)
| Submitted by: | Markus Stacha<br>Philips Medizin Systeme Boeblingen GmbH<br>Cardiac and Monitoring Systems<br>Hewlett-Packard Str.2<br>71034 Boeblingen<br>Germany | | FEB 2 4 2003 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|--------------|
| Date of Summary: | November 22, 2002 | | |
| Device Name | Philips Avalon CTS Cordless Fetal Transducer System | | |
| Common Name | Perinatal Monitoring System | | |
| Classification Name | Regulation Number | Classification Name | |
| | 21 C.F.R §884.2740 | Perinatal Monitoring System and<br>Accessories | |
| Predicate Devices | Philips M1310A Fetal Telemetry System-Series 50T (K942887),<br>Philips M1350A Intrapartum Fetal Monitor (K900480), Philips<br>M1350B Perinatal Monitoring System (K954351), Philips<br>M1351A Perinatal Monitoring System (K921957), Philips<br>M1353A Fetal Monitor (K921956) and the Philips M2600A<br>Telemetry Systems (K993516, K980429, and K961165). | | |
| Modifications | The modification incorporates the front-end circuitry for each<br>parameter into the transducer, thereby replacing the cable used in<br>the predicate versions by a direct hardware connection. There are<br>no modifications to the device design that affect safety &<br>effectiveness of the Perinatal Monitoring System. | | |
| Device Description | The Perinatal Monitoring System operates by measuring,<br>processing, and transmitting standard fetal parameters to any<br>Philips Series 50 fetal monitor using radio transmission<br>technology. A group of three wireless transducers is available to<br>obtain specific fetal data in the antepartum period and during labor<br>and delivery. The transducers are designed to operate while the<br>patient is ambulatory, including activity taking place in a<br>submerged water environment, like bathing and hydrotherapy.<br>The devices are identified by the following model numbers:<br>System and Base Station M2720A; Toco transducer M2725A<br>(measures uterine activity externally); US transducer M2726A<br>(measures fetal heart rate and fetal movement profile); and the<br>optional ECG/auxiliary transducer M2727A (measures DECG or<br>direct fetal scalp ECG and MECG or maternal ECG). | | |
{2}------------------------------------------------
K023931 p 2/2
| Intended Use | The intended use of the modified device, as described in its<br>labeling, has not changed from that of the predicate device as a<br>result of the modifications. The Philips M2720A Cordless Fetal<br>Transducer System is indicated for pregnant women from about 22<br>weeks of gestation on through delivery. It is intended for<br>continuous cordless fetal monitoring in connection with a fetal<br>monitor for resting or ambulating patients, also usable during<br>hydrotherapy, for antepartum testing and labor and delivery<br>(intrapartum).<br><br>The system should only be used by, or under the direct supervision<br>of, a licensed physician or other health care practitioner who is<br>trained in the use of fetal heart rate monitors and in the<br>interpretation of fetal heart rate traces. It is not intended for home<br>use. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics | The Philips Avalon CTS Cordless Fetal Transducer System and<br>accessories have the same technological characteristics as the<br>legally marketed predicate devices. |
| Testing | Verification and validation testing activities were conducted to<br>establish the performance and reliability characteristics of the<br>modified device.<br><br>Testing involved environmental, electrical safety, EMC, radio<br>telemetry, biocompatibility and user evaluations for consumer<br>accuracy. Hardware and software verification testing and cable<br>interface verification testing were also conducted. Test results<br>showed substantial equivalence. |
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with its wings spread, rendered in a simple, bold line drawing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 4 2003
Mr. Markus Stacha Regulatory Affairs Engineer Philips Medizin Systeme Böblingen GmbH Hewlett-Packard-Str.2 71034 Böblingen GERMANY
Re: K023931
Trade/Device Name: M2720A Avalon CTS Cordless Fetal Transducer System Regulation Number: 21 CFR §884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: 85 HGM Dated: November 22, 2002 Received: November 26, 2002
Dear Mr. Stacha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known)
六023931
Device Name: The Philips Avalon CTS Cordless Fetal Transducer System (M2720A) and Accessories
Indications for Use: The Philips Avalon CTS Cordless Fetal Transducer System (M2720A) and Accessories is indicated for pregnant women from about 22 weeks of gestation on through delivery. It is intended for continuous cordless fetal monitoring in connection with a fetal monitor for resting or ambulating patients, also usable during hydrotherapy, for antepartum testing and labor and delivery (intrapartum).
The system should only be used by, or under the direct supervision of, a licensed physician or other health care practitioner who is trained in the use of fetal heart rate monitors and in the interpretation of fetal heart rate traces. It is not intended for home use.
## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use.
OR
Over-The-Counter Use
David h. Ingram
---
(Division Sign-Off)
Division of Reproductive, A and Radiological Devices 510(k) Number
Page __ of ___________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.