TM2000 FETALTRACE TRANSTELEPHONIC RECEIVING CENTER
Applicant
Card Guard Scientific Survival , Ltd.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Aug 12, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The TM2000 FetalTrace Transtelephonic Receiving Center is a software system designed and intended for supporting a remote (transtelephonic) monitoring of patient Fetal/Maternal parameters. The System enables receiving, storing, displaying, updating, printing and retransmitting of patient Fetal/Maternal parameters and other patient related data. (such as demographics, doctors, medical history and status, diagnoses, etc.).
Device Story
The TM2000 FetalTrace is a software-based receiving center for remote monitoring of fetal/maternal parameters. It receives, stores, displays, and prints patient data transmitted via telephone. The system runs on MS Windows, utilizing a database for patient demographics, medical history, and signal data. Healthcare providers use the GUI to review transmitted signals and generate reports. The device facilitates clinical decision-making by providing a centralized platform for monitoring fetal/maternal status remotely. It is intended for use in a clinical setting by medical professionals to manage patient data and monitor fetal health.
Clinical Evidence
Bench testing only. Verification testing was conducted to ensure system compliance with design specifications and to evaluate safety and risk, confirming the device performs as intended without adverse health effects.
Technological Characteristics
Software system running on MS Windows OS. Uses C/C++ native methods and Windows API for GUI and database connectivity. Database engine: Microsoft Access. Maximum database size: 2 GB. Supports localization and report generation. Moderate level of concern software.
Indications for Use
Indicated for remote transtelephonic monitoring of fetal and maternal physiological parameters in patients requiring perinatal monitoring.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
Image /page/0/Figure/0 description: This document is a Special 510(k) Premarket Notification for the TM2000 FetalTrace Transtelephonic Receiving Center. The document is dated August 12, 2002, and includes a summary of safety and effectiveness for the TM2000 FetalTrace Receiving Center Software. The document also includes a restriction on use, duplication, or disclosure of proprietary information.
| | Name | Title | Signature | Date |
|-------------|---------------|-----------------------|-----------|----------------|
| Written by | A. Gonorovsky | Deputy Chief Engineer | | March 26, 2002 |
| Approved by | Amnon Shur | Project Manager | | March 26, 2002 |
| | | | | |
{1}------------------------------------------------
| | Special 510(k) Premarket Notification<br>TM2000 FetalTrace Transtelephonic Receiving Center | K021574 page 2/4 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Submitter: | Card Guard Scientific Survival Ltd.,<br>2 Pekeris St. P.O.B. 527<br>Rehovot 76100, Israel<br>Tel: 972-8-9484600<br>Fax: 972-8-9484605 | |
| Contact Person: | Alex Gonorovsky,<br>Deputy Chief Engineer, Regulatory Affairs<br>Tel: 972-8-9484624<br>E-mail: alexanderg@cardguard.com | |
Date Prepared: 18 April 2002
## Table of Contents
| 1. | Definition and Intended Use | 3 |
|----|-----------------------------------------------------------|---|
| 2. | Device Class | 3 |
| 3. | Applicable Regulatory Documents and Card Guard Procedures | 3 |
| 4. | Operational Characteristics, Features and Functions | 3 |
| 5. | User Interface | 4 |
| 6. | Substantial Equivalence | 4 |
| 7. | Design Controls and Hazard Analysis | 4 |
| 8. | Conclusions | 4 |
{2}------------------------------------------------
KCZ1574 pi34
## 1. Definition and Intended Use
The TM2000 FetalTrace Transtelephonic Receiving Center is a software system designed and intended for supporting a remote (transtelephonic) monitoring of patient Fetal/Maternal parameters. The System enables receiving, storing, displaying, updating, printing and retransmitting of patient Fetal/Maternal parameters and other patient related data. (such as demographics, doctors, medical history and status, diagnoses, etc.).
## 2. Device Class
The TM2000 FetalTrace Transtelephonic Receiving Center system is class II medical device (21 C.F.R. Par. 870.2920 (1992)).
## 3. Applicable Regulatory Documents and Card Guard Procedures
No performance standards have been developed under Section 514 of the Federal Food, Drug and Cosmetic Act for telephone ECG and Spirometric transmitter devices.
The complete list of the Applicable Documents referenced and/or incorporated in this project is provided in the FDA submission package: Chapter 18, Applicable Normative Documents: Standards, Regulations, Guidances, Procedures; Publications
Chapter 18 includes the following sections:
- Applicable FDA Documents ●
- . American National Standards Institute (ANSI), Association for the Advancement of Medical Instrumentation (AAMI)
- . International Electrotechnical Commission (IEC/CEI), International Organization for Standardization (ISO), European Norm (EN)
- . Code of Federal Regulations
- International Special Committee On Radio Interference (CISPR) .
- Institute of Electrical and Electronics Engineers (IEEE) .
- Card Guard Ltd: Applicable Procedures .
- Card Guard Ltd: Product Definitive Reference .
- . Articles, Publications
- 4. Operational Characteristics, Features and Functions
- 1. Runs on any MS Windows operating system
- 2. Maximum database file size of 2 GB. Storage of up to 10 - 40 thousand transmissions (depending on transmitter type)
- 3. Accessing and updating the receiving center DB. Manual entry of patient and physician detail
- 4. Receiving and processing patients' transmitted signals and medical data
- 5. Graphic representation of FM signals
- 6. Intuitive, user friendly HMI
- 7. Capability for signal/data transmission over Web (Planned for implementation in future)
- 8. Localization
- 9. Reports generation
{3}------------------------------------------------
Image /page/3/Figure/0 description: The image contains the logo for Card Guard Scientific Survival LTD on the left. On the right, the text reads "Special 510(k) Premarket Notification TM2000 FetalTrace Transtelephonic Receiving Center". The text appears to be the title or heading of a document or presentation. The logo includes a stylized graphic resembling a heartbeat.
- 10. Reports previewing and printing
- KOZ1574 page 4/4 11. Manageable security, fail-safe protection against unauthorized access. Discretionary access control, restrictions based on privileges (Not yet implemented)
#### 5. User Interface
The TM2000 Graphic User Interface (GUI) features pull down menus and dialog boxes for representation and updating of data. Generation of GUI and implementation of the related DB connectivity mechanism utilizes the C/C++ native methods in conjunction with the Windows API functions.
#### 6. Substantial Equivalence
The TM2000 FetalTrace is substantially equivalent to its predicate TM2000 K013879 Receiving Center, cleared on Nov 23, 2001.
The TM2000 FetalTrace has the Fetal/Maternal service, with the user interface forms totally identical to those of TM2000 K013879.
The TM2000 FetalTrace differs from the TM2000 K013879 in two aspects:
- 1. Unlike the TM2000 K013879, the TM2000 FetalTrace has no ECG and Spiro services.
- 2. The TM2000 FetalTrace uses the Access DB engine while the TM2000 K013879 uses the Oracle DB engine. This aspect will be discussed in further detail Substantial Equivalence 2 TM2000 FetalTrace vs. TM2000 EasyTrace since the latter also uses Access.
The proof of substantial equivalence in all that concerns the intended use, principles of operation, features and technological characteristics is provided in Chapter 7. Substantial Equivalence to Cleared Devices.
### 7. Design Controls and Hazard Analysis
The Card Guard's product design procedure, and QA and QC policy, formalize the design and production process and assure that all the respective requirements are met. In the framework of the Design Controls the testing was conducted to verify the system compliance with all its design specifications.
The device Level of Concern criteria were evaluated and the system was determined to be a moderate level of concern system.
The rigorous design evaluation and the System Safety and Risk analysis expose potential failures or possible system flaws which could directly or indirectly effect the patient.
#### 8. Conclusions
The system constitutes a safe and reliable means for receiving, storing, displaying, analyzing, updating, printing and re-transmitting of patient Fetal/Maternal parameters and other patient related data.
Its operation present no adverse health effect or safety risks to patients when used as intended.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal for the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus features a staff with two snakes coiled around it, topped with a pair of wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 1 2 2002
Mr. Alex Gonorovsky Deputy Chief Engineer, Regulatory Affairs Card Guard Scientific Survival, Ltd. 2 Pekeris St. P.O.B. 527 Rehovot 76100 ISRAEL
Re: K021574
Trade/Device Name: TM2000 Fetal Trace Transtelephonic Receiving Center Software Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories
Regulatory Class: II Product Code: 85 HGM Dated: May 8, 2002 Received: May 14, 2002
Dear Mr. Gonorovsky :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (sec above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Card Guard Scientific Survival LTD. The logo is in black and white and features the words "CARD GUARD" in bold letters, with the words "Scientific Survival LTD." underneath in a smaller font. To the right of the text is a graphic that resembles an EKG readout.
Indications For Use TM2000 Fetal Trace Receiving Center Software
The TM2000 FetalTrace Telemedicine Receiving Center is intended for supporting transtelephonic monitoring of the Fetal/Maternal parameters of patients.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
V Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use
(Optional Format 1-2-96)
David A. Ingram
(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.