K971305 · Gynecare Innovation Center · HIH · Aug 11, 1997 · Obstetrics/Gynecology
Device Facts
Record ID
K971305
Device Name
FEMRX FOCUSED MONOPOLAR (FMP) OPERA STAR SYSTEM
Applicant
Gynecare Innovation Center
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Aug 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FemRx™ Focused Monopolar (FMP) OPERA STAR™ System is intended for gynecologic hysteroscopic / electro-surgical use by trained professional gynecologists in hospital or office environments. It is used to resect and / or ablate endometrial tissue.
Device Story
Electrosurgical system for hysteroscopic resection/ablation of endometrial tissue; used by gynecologists in hospital/office settings. Device utilizes single active electrode (wire or Star loop) with PEARL (Physiologic endometrial ablation/resection loop) design; loop partially insulated to focus current. Return path provided by either metallic outer morcellator housing or standard dispersive skin electrode. System designed for use with isotonic solutions (e.g., NaCl, Ringers Lactate) to mitigate fluid overload risks. Physician operates device under hysteroscopic visualization to treat uterine pathologies; output effect is tissue coagulation, resection, or ablation.
Clinical Evidence
Clinical testing demonstrated equivalent performance between the FMP OPERATAR System (K964441) and the FMP OPERATAR System with Dispersive Electrode Option. In vitro studies on extirpated uteri compared resection and ablation performance against predicates, confirming substantial equivalence.
Technological Characteristics
Monopolar electrosurgical system; active electrode (wire or Star loop) with partial insulation for current focusing. Return electrode: metallic morcellator housing or dispersive skin pad. Energy source: electrosurgical generator (implied). Compatible with isotonic irrigation fluids. Form factor: hysteroscopic instrument.
Indications for Use
Indicated for gynecologic hysteroscopic resection and/or ablation of endometrial tissue in patients with abnormal uterine bleeding, submucous fibroids, or endometrial polyps. Contraindicated in pregnancy, pelvic infection, cervical malignancy, or previously diagnosed endometrial cancer.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
K111751 — VERSAPOINT II BIPOLAR ELECTROSURGERY GENERATOR, VERSAPOINT II 44MM ANGLED LOOP ELECTRODE, VERSAPOINT II FOOTSWITCH, VERS · ETHICON, Inc. · Dec 2, 2011
K964441 — FEMRX FOCUSED MONOPOLAR (FAP) OPERASTAR SYSTEM [OUTPATIENT ENDOSCOPIC RESECTION AND ABLATION (OPERA)] [SPECIALIZED TISSU · Gynecare Innovation Center · Mar 28, 1997
K962271 — ELECTROSURGICAL ELECRTRODES · Northgate Technologies, Inc. · Aug 8, 1997
Submission Summary (Full Text)
{0}
K971305
510(k) SUMMARY
AUG 11 1997
The FemRx Focused Monopolar (FMP) $^{\text{OPERATAR™}}$ System [with Dispersive Electrode Option] is identical to FemRx’s currently cleared FMP $^{\text{OPERATAR™}}$ System with the exception that it includes the option of using a standard dispersive skin electrode. When used in this way it is basically identical to our original $^{\text{OPERATAR™}}$ System except that it incorporates a PEARL electrode (Physiologic endometrial ablation / resection loop). This loop is partially insulated to “focus” the current toward the tissue and avoid the reduction in effect associated with conventional devices. The use of isotonic solutions (e.g., NaCl or Ringers Lactate) is strongly recommended in order to reduce the risks associated with fluid overload.
The FMP $^{\text{OPERATAR}}$ System differs from currently marketed bi-polar devices in that it continues to have a single active electrode (wire loop or Star loop) as opposed to two active electrodes. The return electrode (the metallic outer Morcellator housing of the device OR a standard dispersive pad located on the patients skin) has such a large surface area relative to the active electrode that there can be no tissue coagulating (much less cutting or ablating) effect at its surface.
*In vitro* extirpated uteri studies comparing the resection and ablation performance of the Focused Monopolar device with the predicates confirm that tissue effects are substantially equivalent to currently marketed devices. In addition, clinical testing has demonstrated equivalent performance between the FMP $^{\text{OPERATAR™}}$ System (recently cleared via K964441) and the FMP $^{\text{OPERATAR™}}$ System [with Dispersive Electrode Option].
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
George M. Savage, M.D.
Senior Vice President
FemRx
1221 Innsbruck Drive
Sunnyvale, California 94089
Re: K971305
Focused Monopolar (FMP) OPERASTAR™ System
with Dispersive Electrode Option
Dated: July 3, 1997
Received: July 11, 1997
Regulatory class: II
21 CFR §884.4160 and §884.1690
Product code: 85 HIH & KNF
Dear Dr. Savage:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,


Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{2}
510(k) Number (if known): K971305
Device Name: FemRx™ FMP OPERA STAR™ (with Dispersive Electrode Option)
Indications For Use:
The FemRx™ Focused Monopolar (FMP) OPERA STAR™ System is intended for gynecologic hysteroscopic / electro-surgical use by trained professional gynecologists in hospital or office environments. It is used to resect and / or ablate endometrial tissue.
INDICATIONS:
Abnormal Uterine Bleeding
Submucous Fibroids
Endometrial Polyps
CONTRAINDICATIONS:
Pregnancy
Pelvic Infection
Cervical Malignancy
Previously diagnosed Endometrial cancer
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Colin M. Pollard
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K971305
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
000009
000009
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.