K192534 · Crossbay Medical · HHK · Mar 25, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K192534
Device Name
CrossBay Endometrial Tissue Sampler (ETS)
Applicant
Crossbay Medical
Product Code
HHK · Obstetrics/Gynecology
Decision Date
Mar 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1175
Device Class
Class 2
Indications for Use
The CrossGlide™ ETS, Endometrial Tissue Sampler, is indicated for use to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.
Device Story
Sterile, single-use endometrial tissue sampler; consists of delivery catheter, everting membrane, inner aspiration catheter, and 3cc syringe. Operates by creating negative pressure via syringe; everting membrane facilitates placement of inner catheter into uterine cavity; device movement scrapes and aspirates mucosal tissue. Used in clinical settings by healthcare providers to obtain histological biopsy samples. Benefits patient by enabling minimally invasive tissue collection for diagnostic evaluation.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993-5, -10), sterilization validation (ISO 11135, ISO 10993-7), packaging/shipping validation (ASTM F88, F2096, D4169), shelf-life (ASTM F1980), and mechanical/functional testing (tensile, flexural, leak, negative pressure, and uterine model comparison).
Technological Characteristics
Manual aspiration device; everting membrane (0.04mm wall thickness); inner aspiration catheter (1.3mm lumen); acorn tip (16mm length, 12.4mm OD). Materials validated per ISO 10993. Sterilization via Ethylene Oxide (ANSI/AAMI/ISO 11135).
Indications for Use
Indicated for patients requiring endometrial tissue biopsy; removes material from the uterus and mucosal lining via scraping and vacuum suction.
Regulatory Classification
Identification
An endometrial suction curette is a device used to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction. This device is used to obtain tissue for biopsy or for menstrual extraction. This generic type of device may include catheters, syringes, and tissue filters or traps.
K102847 — DISPOSABLE ENDOMETRIAL SUCTION CURETTE · Jiangsu Suyun Medical Materials Co., Ltd. · Apr 14, 2011
K992926 — GINRAM ENDOCIC SINGLE OVAL PORT (MODEL NO. 720195) · Ri Mos. S.R.L. · Oct 26, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
CrossBay Medical Inc. % Cindy Domecus, RAC (US & EU) Principal, Domecus Consulting Services Domecus Consulting Services, LLC 1171 Barroihet Drive Hillsborough. CA 94010
Re: K192534
Trade/Device Name: CrossGlide™ ETS, Endometrial Tissue Sampler Regulation Number: 21 CFR 884.1175 Regulation Name: Endometrial Suction Curette and Accessories Regulatory Class: II Product Code: HHK Dated: February 20, 2020 Received: February 24, 2020
Dear Cindy Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Monica D. Garcia. Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192534
Device Name CrossGlideTM ETS, Endometrial Tissue Sampler
Indications for Use (Describe)
The CrossGlideTM ETS, Endometrial Tissue Sampler, is indicated for use to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary- K192534
#### I. SUBMITTER INFORMATION
| Submitter: | CrossBay Medical Inc.<br>13240 Evening Creek Drive, Suite 304<br>San Diego, CA 92128 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Correspondent: | Cindy Domecus, R.A.C. (US & EU)<br>Regulatory Consultant to CrossBay Medical Inc.<br>Phone: 650.343.4813<br>Fax: 650.343.7822<br>Email: domecusconsulting@comcast.net |
| Date Summary Prepared: | March 24, 2020 |
# II. SUBJECT DEVICE INFORMATION
| Device Trade Name: | CrossGlide™ ETS, Endometrial Tissue Sampler |
|--------------------|-------------------------------------------------------|
| Common Name: | Endometrial Sampling Device |
| Regulation Number: | 21 CFR §884.1175 |
| Regulation Name: | Endometrial Suction Curette and Accessories |
| Regulatory Class: | II |
| Product Code: | HHK (Curette, Suction, Endometrial [and accessories]) |
#### III. PREDICATE DEVICE INFORMATION
The predicate device is the Marina Ampler (MAS), K021876.
The predicate device has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The CrossGlide™ ETS, Endometrial Tissue Sampler, is a sterile, disposable, single-use device which enables the removal of mucosal tissue from the uterus for histological biopsy. The CrossGlide™ ETS contains a Delivery Catheter with an everting Membrane and an Inner Aspiration Catheter. The everting Membrane places the Inner Aspiration Catheter into the uterine cavity. Movement of the ETS device, after creating negative pressure with the 3cc syringe, removes material and tissue from the uterine cavity.
#### V. INDICATIONS FOR USE
The CrossGlide™ ETS, Endometrial Tissue Sampler, is indicated for use to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.
{4}------------------------------------------------
### VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| THE PREDICATE DEVICE | | | |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device &<br>Predicate<br>Device | K192534 | K021876 | Comparison |
| Indications for<br>Use | The CrossGlide™ ETS,<br>Endometrial Tissue<br>Sampler, is indicated<br>for use to remove<br>material from the<br>uterus and from the<br>mucosal lining of the<br>uterus by scraping and<br>vacuum suction in<br>order to obtain tissue<br>for histological biopsy. | The device is<br>indicated for use to<br>remove material<br>from the uterus and<br>from the mucosal<br>lining of the uterus<br>by scraping and<br>vacuum suction in<br>order to obtain<br>tissue for<br>histological biopsy<br>or for menstrual<br>extraction. | Different: The predicate<br>device can also be used for<br>menstrual extraction.<br>However, both devices have<br>the same intended use to<br>remove endometrial tissue for<br>biopsy samples |
| Device<br>Components | Inner catheter,<br>everting membrane,<br>aspiration with hole,<br>blue and white<br>stopcock compliant<br>tube, syringe | Catheter with<br>syringe<br>Catheter without<br>syringe | Different: The subject device<br>includes additional<br>components as compared to<br>the predicate device. These<br>differences do not raise<br>different questions of safety<br>and effectiveness (S&E). |
| Aspiration<br>Method | Manual | Manual | Same |
| Overall<br>Catheter<br>Length (cm) | 43-48 cm | Unknown | Different: The length of the<br>predicate device is not known.<br>Differences in catheter length<br>do not raise different<br>questions of S&E. |
| Inner Catheter<br>Delivery<br>Catheter | Length: 315mm<br>(includes Stopcock)<br>Outer diameter: 2mm<br>Internal diameter:<br>1.3mm<br>Length: 156mm<br>(includes Acorn Tip) | Not Applicable<br>Length Unknown<br>3-3.5 mm (OD) | Different: The predicate<br>device does not include a<br>separate inner catheter.<br>Presence of an inner catheter<br>and its length do not raise<br>different questions of S&E.<br>Different: The length of the<br>predicate device is not known; |
| | Outer diameter:<br>4.1mm<br>Internal diameter<br>3.3.mm | | however, its outer diameter is<br>smaller than the subject<br>device. The length and outer<br>diameter of the subject device<br>do not raise different<br>questions of S&E. |
| Everting<br>Membrane | Length: 65mm at full<br>deployment<br>Outer diameter:<br>3.5mm @ 3<br>atmospheres of<br>pressure<br>Wall thickness:<br>0.04mm | Not Applicable | Different: The predicate<br>device does not include an<br>everting membrane.<br>Presence of an everting<br>membrane in the subject<br>device does not raise different<br>questions of S&E. |
| Aspiration<br>Device Side<br>Hole &<br>Internal<br>Lumen | Side hole elliptical<br>opening<br>Internal lumen 1.3mm | Unknown | Different: The design of the<br>tip opening and inner lumen of<br>the predicate device is not<br>known. Differences in tip<br>design and internal lumen<br>diameter do not raise different<br>questions of S&E. |
| Acorn Tip | Length: 16mm<br>Outer diameter:<br>12.4mm | Not Applicable | Different: The predicate<br>device does not include an<br>acorn tip. The presence of an<br>acorn tip does not raise<br>different questions of S&E. |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same |
{5}------------------------------------------------
As noted in the table above, there are differences in the indications for use statements and technological features of the subject and predicate devices. However, as stated in the table, the differences in indications for use do not represent a new intended use, and the differences in technological features do not raise different questions of safety and effectiveness.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
{6}------------------------------------------------
## Biocompatibility testing
Biocompatibility testing was conducted according to ISO 10993-1 "Biological Evaluation of Medical Devices" and FDA's guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." Testing included the following:
- Cytotoxicity (ISO 10993-5:2009) .
- . Vaginal Irritation (ISO 10993-10:2010)
- . Sensitization (ISO 10993-10:2010)
## Sterilization Validation
The CrossGlide™ ETS was evaluated for adoption into the current ethylene oxide (EO) sterilization cycle of the contract sterilizer. The evaluation process was performed utilizing the AAMI Guidance contained in TIR28:2016 Product adoption and process equivalence for ethylene oxide sterilization. The EO Sterilization Cycle was validated in accordance with ANSI/AAMI/ISO 11135:2014. Residuals from the sterilization process were assessed and shown to be in accordance with ISO 10993-7:2008.
## Packaging, Shipping Validation, and Shelf-Life
Packaging and shipping validation studies were conducted pursuant to the applicable ASTM guidelines (ASTM F88/F88M - 15 "Standard Test Method for Seal Strength of Flexible Barrier Materials"; and, ASTM F 2096-11 "Standard Test Methods for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble leaks)"; ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems). The shelf-life is supported by packaging and performance tests conducted on samples exposed to accelerated aging conditions pursuant to ASTM F1980 – 16 "Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices". Additional assessments to support device function over its shelf-life included the following:
- . Dimensional assessments
- Compliant tube ability to be distended and maintain targeted pressure ●
- Assessment of device eversion feature ●
- Evaluation of device marking to confirm correct deployment distance of the inner ● catheter beyond the everting membrane
- . Negative pressure measurement to confirm ability to aspirate tissue samples
- Mechanical testing, including tensile joint strength, flexural testing and butt testing
- Leak testing
- . Functional testing of the subject and predicate device to compare uterine deployment and specimen collection using a uterine model
## VIII: CONCLUSIONS
The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the
{7}------------------------------------------------
subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.