ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator

K161065 · Surgitools Pty , Ltd. · HEW · Oct 27, 2016 · Obstetrics/Gynecology

Device Facts

Record IDK161065
Device NameColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator
ApplicantSurgitools Pty , Ltd.
Product CodeHEW · Obstetrics/Gynecology
Decision DateOct 27, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is indicated for use by a surgeon in laparoscopic procedures where uterine manipulation and visualization of the position of the vaginal fornices for colpotomy incisions is required, and for maintaining pneumoperitoneum during vaginal vault closure.

Device Story

Device consists of CerviGrip Uterine Manipulator, ColpoWave Colpotomizer, and optional ColpoWave Balloon; used by surgeons in laparoscopic procedures. Manipulator features integrated cervical screw, sliding uterine tip, and tail screw for locking; colpotomizer is double-ended (30mm/40mm) with raised lips for distance gauging and green color for tissue contrast. Balloon accessory provides vaginal occlusion. Device is single-use, sterile, and non-latex. Surgeon operates device to manipulate uterus and visualize vaginal fornices for colpotomy incisions; balloon maintains pneumoperitoneum during vaginal vault closure. Benefits include improved visualization and maintenance of surgical field integrity during laparoscopic surgery.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including mechanical performance (tensile, manipulation, balloon stability), biocompatibility (ISO 10993), and sterilization validation (ISO 11135-1).

Technological Characteristics

Materials: ABS polymer, 304 stainless steel. Components: Uterine manipulator with cervical screw, sliding tip, colpotomizer (30mm/40mm), pneumatic balloon. Single-use. Sterilization: Ethylene Oxide (EtO) per ISO 11135-1. Biocompatibility: ISO 10993-1, -10, -11. Shelf life: 3 years.

Indications for Use

Indicated for use by surgeons in laparoscopic procedures requiring uterine manipulation, visualization of vaginal fornices for colpotomy incisions, and maintenance of pneumoperitoneum during vaginal vault closure.

Regulatory Classification

Identification

A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 27, 2016 Surgitools Pty Ltd. % Elizabeth Pugh Regulatory Affairs Consultant Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, Texas 78701 Re: K161065 > Trade/Device Name: ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator Regulation Number: 21 CFR 884.1640 Regulation Name: Culdoscope and accessories Regulatory Class: Class II Product Code: HEW Dated: September 28, 2016 Received: September 28, 2016 Dear Elizabeth Pugh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {1}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, For Division Douglas Silverstein -S 2016.10.27 15:26:25 -04'00' Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161065 #### Device Name ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator Indications for Use (Describe) The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is indicated for use by a surgeon in laparoscopic procedures where uterine manipulation of the position of the vaginal fornices for colpotomy incisions is required, and for maintaining pneumoperitoneum during vaginal vault closure. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------|---------------------------------------------| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # K161065 # 1. Submission Sponsor Surgitools Pty Ltd. 231 Timberlane Dr. Woodvale Western Australia 6026, Australia Phone number: 61-08-93093222 Contact: Jai Singh Title: Director: Global Sales & Marketing # 2. Submission Correspondent Emergo Global Consulting, LLC 2500 Bee Cave Rd, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Contact: Elizabeth Pugh, Consultant, RA Email: project.management@emergogroup.com #### 3. Date Prepared October 14, 2016 # 4. Device Identification | Trade/Proprietary Name: | ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator | |-------------------------|--------------------------------------------------------------------------------| | Common/Usual Name: | Uterine Manipulator and Colpotomizer | | Classification Name: | Culdoscope and accessories | | Regulation Number: | 884.1640 | | Product Code: | HEW, Culdoscope and accessories<br>LKF, Cannula, manipulator/injector, uterine | | Device Class: | Class II | | Classification Panel: | Obstetrics/Gynecology | # 5. Legally Marketed Predicate Device(s) K110819, Singh Colpotomizer System, Surgitools Pty Ltd. #### 6. Device Description The Surgitools Pty Ltd. ColpoWave Colpotomizer and CerviGrip Uterine Manipulator consists of the ColpoWave™ Colpotomizer, CerviGrip™ Uterine Manipulator, and the ColpoWave™ Balloon {4}------------------------------------------------ accessory. All parts are made without natural latex rubber and supplied sterile to be disposed of after use. The single-use CerviGrip™ Uterine Manipulator is designed with an integrated cervical screw, a sliding uterine tip, tail screw that can lock the uterine tip in place, and a screw to hold the colpotomizer in place. The CerviGrip™ Uterine Manipulator is manufactured with ABS polymer and 304 Stainless Steel. The ColpoWave™ Colpotomizer is non-fenestrated and has incorporated two different sizes due to the double-ended colpotomizer design (30mm or 40mm diameter). Cup shape with one partial circumferential raised lip at each end and a second partial raised circumferential lip 20mm from each end providing the surgeon with a known distance to help gauge pelvic distances during dissection. The ColpoWave™ Colpotomizer is manufactured with green ABS polymer to provide better visualization (greater contrast than white against tissue). The ColpoWave™ Balloon is an optional accessory, which is inflated to fit various vaginas. # 7. Indications for Use Statement The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is indicated for use by a surgeon in laparoscopic procedures where uterine manipulation and visualization of the position of the vaginal fornices for colpotomy incisions is required, and for maintaining pneumoperitoneum during vaginal vault closure. #### 8. Predicate Comparison The following table compares the Surgitools Colpotomizer and CerviGrip Uterine Manipulator to the predicate device with respect to indications for use, technological characteristics, and materials: | Trade Name | Singh Colpotomizer System<br>K110819 (Surgitools Pty Ltd.) | ColpoWave Colpotomizer and CerviGrip<br>Uterine Manipulator<br>K161065 (Surgitools Pty Ltd.) | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for use | The Singh Colpotomizer System is<br>indicated for use by a surgeon in<br>laparoscopic procedures requiring<br>uterine manipulation and the<br>visualization of the position of the<br>vaginal fornices for colpotomy incisions<br>into the vaginal vault. The surgeon<br>makes the colpotomy incisions to access<br>or remove intraperitoneal tissue from<br>the pelvic cavity. | The ColpoWave Colpotomizer and<br>CerviGrip Uterine Manipulator is<br>indicated for use by a surgeon in<br>laparoscopic procedures where uterine<br>manipulation and visualization of the<br>position of the vaginal fornices for<br>colpotomy incisions is required, and for<br>maintaining pneumoperitoneum during<br>vaginal vault closure. | {5}------------------------------------------------ | System<br>Components | Uterine manipulator with cervical<br>screw attachment; Sliding uterine tip (inner rod); Sliding and rotating funnels<br>(available in two sizes) with a lip<br>acting as a colpotomizer; O-ring and screw to hold the<br>funnel in place; and Vaginal occluder (plastic<br>vaginal plug) | Uterine manipulator with cervical<br>screw attachment Sliding uterine tip Sliding and rotating colpotomizer<br>(incorporating two sizes) ColpoLock and screw to hold the<br>colpotomizer in place; Pneumatic plug (PneumoBloc); and Vaginal occluder (ColpoWave Balloon | |-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Uterine Tips | Single length inner rod | Single length inner rod | | Vaginal Plug | OD 43mm<br>Length 50mm | ColpoWave Balloon is inflated to fit<br>various vaginas | | Design<br>Colpotomizer | Two colpotomizers of different<br>sizes. 35mm and 40mm Cup shape with partial<br>circumferential raised lip at its end. White. Non-fenestrated. | One colpotomizer incorporating two<br>different sizes. One side diameter<br>30mm and one of diameter 40mm Cup shape with one partial<br>circumferential raised lip at each end<br>and a second partial raised<br>circumferential lip 20mm from each<br>end. Green. Non-fenestrated. | | Uterine<br>manipulator | Uterine manipulator with cervical<br>screw attachment; Sliding uterine tip (inner rod); Tail screw with hydrotubation port;<br>and O-ring and screw to hold the funnel<br>in place | Uterine manipulator with integrated<br>cervical screw and hydrotubation<br>port; Sliding uterine tip; Tail screw that can lock the uterine<br>tip in place; and ColpoLock and screw to hold<br>colpotomizer in place. | | Materials | Stainless steel and polymer | Stainless steel and polymer | | Single use or reusable | Reusable | Single use | | Biocompatible materials | Yes | Yes | The Surgitools ColpoWave Colpotomizer and CerviGrip Uterine Manipulator has the same intended use but different technological characteristics compared to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness. # 9. Non-Clinical Performance Data As part of demonstrating safety and effectiveness of the ColpoWave Colpotomizer and CerviGrip Uterine Manipulator and in showing substantial equivalence to the predicate device, Surgitools Pty Ltd. completed a number of tests and validations. The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator meets all the requirements for overall design, sterilization, and biocompatibility confirming that the design output meets the design inputs and specifications for the device. {6}------------------------------------------------ The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device: - Performance Testing The device successfully passed bench testing to evaluate mechanical and functional properties to ensure the device is robust to withstand forces that exceed the expected forces during use, including tensile tests, manipulation, and balloon strength and stability. - Biocompatibility testing per ISO 10993-1, ISO 10993-10 and ISO 10993-10 and ISO 10993-11 The subject device is an external communicating device that contacts tissue for less than 24 hours. Cytotoxicity, Intracutaneous Reactivity, Sensitization, and Acute Systemic Toxicity Testing were performed. All tests were conducted on the finished device after manufacturing, packaging, and sterilization had been completed to reflect the actual state of the device when exposed to the patient. All patient-contacting device components passed biocompatibility testing required by ISO 10993-1 and FDA General Program Memorandum G95-1. - . Sterilization Validation The sterility of the Surgitools Uterine Manipulator and Colpotomizer system is assured by using a validated sterilization method qualified in accordance with ISO 11135-1. The validation results met the acceptance criteria defined in the standard and demonstrates the ability of the process to deliver SAL of 10 °. Examination of the EtO residuals has been conducted in accordance with ISO 10993-7, which has validated a minimum release period of 7 days. - Shelf Life Testing After the sterilization cycle was complete, the products were accelerated aged and tested to demonstrate no degradation in strength or functionality over the anticipated shelf life period of three (3) years. #### 10. Clinical Performance Data There was no human clinical testing required to support the substantial equivalence of the subject device. #### 11. Conclusion Based on the comparison and analysis above, the Surgitools ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is substantially equivalent to the referenced predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...