GALILEO HYSTEROSCOPES AND LAPAROSCOPES

K962256 · Galileo Electro-Optics Corp. · HIH · Dec 6, 1996 · Obstetrics/Gynecology

Device Facts

Record IDK962256
Device NameGALILEO HYSTEROSCOPES AND LAPAROSCOPES
ApplicantGalileo Electro-Optics Corp.
Product CodeHIH · Obstetrics/Gynecology
Decision DateDec 6, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1690
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Galileo Hysteroscopes are rigid endoscopes intended for direct visualization of cervical canal and uterine cavity for diagnostic and surgical procedures during gynecological procedures. The Galileo Laparoscopes are intended for direct visualization of the organs within the peritoneum for diagnostic and surgical procedures on the female genital organs. The Galileo Laparoscopes are designed to be introduced through natural body cavities or through introducers, needles, trocars, catheters, sheaths or other devices with thru-lumens having inside diameters larger than the outside diameter of the endoscope.

Device Story

Reusable rigid endoscopes; rod-lens or fiber-optic imaging bundles; external high-intensity light source transmits light via optical fibers to distal tip; objective lens captures target image; image transmitted to proximal eyepiece for direct viewing or video monitor connection. Used in clinical settings by physicians for gynecological diagnostic and surgical procedures. Enables visualization of body cavities, tissues, and organs; optional working channels allow passage of surgical instruments. Benefits include minimally invasive access to internal anatomy for diagnosis and treatment.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Rigid endoscope; stainless steel construction; rod-lens or fiber-optic imaging bundles; external high-intensity light source; optical fiber illumination; optional working channels for instruments; various lengths and diameters; reusable.

Indications for Use

Indicated for direct visualization of the cervical canal, uterine cavity, and organs within the peritoneum for diagnostic and surgical procedures on female genital organs.

Regulatory Classification

Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

Predicate Devices

Submission Summary (Full Text)

{1} DEC - 6 1996 # 510(k) Summary ## Galileo Electro-Optics Corporation ## Galileo Hysteroscopes and Laparoscopes 1. Sponsor/Applicant Name and Address K962256 p193 Galileo Electro-Optics Corporation Galileo Park P.O. Box 550 Sturbridge, MA 01566 Telephone (508) 347-9191 Contact Person Kin M. Wong, Director of Quality Assurance Date of Summary Preparation June 11, 1996 2. Device Name Proprietary Name: Galileo Hysteroscopes and Laparoscopes Common/Usual Name: Hysteroscopes and Laparoscopes Classification Name: Hysteroscopes and accessories Laparoscopes and accessories 3. Identification of Predicate or Legally Marketed Device(s) The Galileo Hysteroscopes and Laparoscopes are substantially equivalent to several legally marketed endoscopes including the Karl Storz Hysteroscopes and Laparoscopes, (K88270 and K935277) and the Richard Wolf Hysteroscopes and Laparoscopes (K880314 and K770378). Galileo Hysteroscopes and Laparoscopes 510(k) 6/11/96 Page D-1 {2} K962256 p293 ## 4. Device Description The Galileo Hysteroscopes and Laparoscopes are a line of reusable rigid endoscopes based on existing endoscope technology. The Galileo Hysteroscopes and Laparoscopes will be available in various lengths, diameters and configurations for the convenience of the user. Two designs of endoscopes are available: ### Galileo Rod and Lens Design Endoscope: The rod and lens design Galileo Endoscope functions by light being transmitted from a standard external high intensity light source through optical fibers to the distal tip of the endoscope. The image of the target is then transmitted from the distal end via an objective lens and an alternating set of rod and lenses to a proximal eyepiece. The image can be viewed directly or it may be transmitted through a video camera to a video monitor. ### Galileo Fiber Optic Design Endoscope: The fiber optic design Galileo Endoscope functions by light being transmitted from a standard external high intensity light source through optical fibers to the distal tip of the endoscope. The image of the target is then transmitted from the distal end via an objective lens and a fiberoptic imaging bundle (instead of a rod and lens) to a proximal eyepiece. The image can be viewed directly or it may be transmitted through a video camera to a video monitor. ## 5. Intended Use The Galileo Hysteroscopes are rigid endoscopes intended for direct visualization of cervical canal and uterine cavity for diagnostic and surgical procedures during gynecological procedures. The Galileo Laparoscopes are intended for direct visualization of the organs within the peritoneum for diagnostic and surgical procedures on the female genital organs. The Galileo Laparoscopes are designed to be introduced through natural body cavities or through introducers, needles, trocars, catheters, sheaths or other devices with thru-lumens having inside diameters larger than the outside diameter of the endoscope. Galileo Hysteroscopes and Laparoscopes 510(k) 6/11/96 Page D-2 {3} K962257 p 3 of 3 # 6. Comparison of Technological Characteristics The Galileo Hysteroscopes and Laparoscopes and the substantially equivalent devices are identical in intended use in that they are all rigid endoscopes intended to be passed through a lumen of an introducer or into natural body cavities for visualization of body cavities, tissues, organs or canals. The Galileo Hysteroscopes and Laparoscopes and the substantially equivalent devices are similar in designs in that they all offer various configurations including rod/lens or fiberoptic design, optional working channels, several OD’s and lengths, and use external light sources, and similar stainless steel materials. The Galileo Hysteroscopes and Laparoscopes and the substantially equivalent devices are similar in technological characteristics in that they offer a channel for either viewing body cavities, tissues, organs or canals and an optional channel for passing instruments into the desired anatomical sites. Galileo Hysteroscopes and Laparoscopes 510(k) 6/11/96 Page D-3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...