HTx disposable hysteroscope system is intended to be used for viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
Device Story
The HTx Disposable Hysteroscope System is a single-use endoscopic device for gynecological procedures. It consists of a disposable cannula module (containing a CMOS camera and LED illumination) and a reusable image system module. The device is used in hospitals and physician offices by trained medical professionals. During a procedure, the cannula is inserted into the cervical canal; the CMOS camera captures video data, which is transmitted via an image cable to the reusable image system. The image system processes the video and outputs it to a medical-grade display. The cannula features buttons for video recording, image capture, and LED brightness control. An external PC with patient management software can connect to the image system via LAN for data management. The device provides real-time visualization of the uterine cavity, enabling clinicians to perform diagnostic and operative tasks, potentially improving procedural accuracy and patient outcomes.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601-1, 60601-2-18, 60601-1-2), and optical performance testing (ISO 8600-1, ISO 8600-3, IEC 62471).
Technological Characteristics
Disposable cannula with CMOS sensor (400x400 pixels) and LED illumination. Working length 240mm, overall 330mm. Cannula diameters 4.5mm and 6.2mm. Reusable image system module with DVI output and USB interface. Sterilization via ethylene oxide (ISO 11135:2014). Software is Moderate Level of Concern. Connectivity via LAN for external PC patient management.
Indications for Use
Indicated for viewing the cervical canal and uterine cavity in adults for diagnostic and operative procedures.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 15, 2021
AcuVu, Inc. Sam Mostafavi Regulatory Affairs 4546 El Camino Real #211 Los Altos, CA 94022
Re: K211227
> Trade/Device Name: HTx Disposable Hysteroscope System Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: August 10, 2021 Received: August 12, 2021
Dear Sam Mostafavi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211227
Device Name
HTx Disposable Hysteroscope System
#### Indications for Use (Describe)
HTx disposable hysteroscope system is intended to be used for viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K211227 Page 1 of 5
# 510(k) Summary
#### 510(k) Number: K211227
#### Applicant Information:
| Date Prepared: | September 10, 2021 |
|-------------------|---------------------------------------------------------------|
| Manufacturer: | AcuVu Inc.<br>4546 El Camino Real #211<br>Los Altos, CA 94022 |
| Contact Person: | Sam Mostafavi<br>ipcs11@ymail.com |
| Mobile Number: | 650-670-6972 |
| Office Number: | 650-578-9653 |
| Facsimile Number: | 650-578-9653 |
#### Device Information:
| Trade/Device Name: | HTx Disposable Hysteroscope System |
|--------------------|------------------------------------|
| Regulation Number: | 21 CFR § 884.1690 |
| Regulation Name: | Hysteroscope and accessories |
| Regulatory Class: | Class II |
| Product Code | HIH (Hysteroscope and Accessories) |
| Panel: | Obstetrics/Gynecology |
#### Predicate Device:
AcuVu GDT-1000 (K180096)
The predicate device has not been subject to a design-related recall.
#### Indications for Use
HTx disposable hysteroscope system is intended to be used for viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
#### Device Description
The HTx Disposable Hysteroscope System is a single use hysteroscope endoscope intended for gynecology procedure applications. It includes a disposable cannula module, a reusable imaging system, a medical grade display and optional software for processing patient management.
The disposable cannula contains a miniature complementary metal-oxide-semiconductor (CMOS) camera, a light-emitting diode (LED) illumination module, and a channel for fluid and device. The cannula connects directly to the image system via an image cable. The cannula also contains three (3) buttons for video recording, picture taking, and light-emitting diode (LED)
{5}------------------------------------------------
brightness control. The image system connects to a medical grade display module via DVI port. The external patient management software runs on an external PC that connects with the image system via LAN cable.
#### Disposable Cannula
The disposable cannula module family includes two models. HTx60 has an outer diameter of 6.2mm and adapts to tools up to 3mm, and HTx40 has an outer diameter of 4.5mm and adapts to tools up to 5Fr. The disposable single-use cannula contains a miniature camera module and an LED illumination module at the tip. The cannula connects to the image system through an image cable for image data transfer, camera control, and power supply.
The disposable cannula has a working length of 240mm, and an overall length is 330mm.
Fluid irrigation is achieved through a tube which is connected to an IV bag and pressured via peristaltic pump. The disposable cannula is sterilized and packaged in a sealed pouch.
#### Image System Module
The HTx1000 image system Module is reusable. It connects to the disposable cannula through an image cable. Its proximal end has a USB connection interface and power button. It contains the image processing controller and user interface software to check the live video during the procedure.
The Image System Module contains several function buttons for use during a procedure – an LED adjustment button to adjust brightness, an information retrieve button to review the stored images, and a set of navigation buttons to instruct the user on setting functions.
#### Functional and Technological Comparison
Table below includes a functional and technological comparison between AcuVu HTx System and AcuVu GDT-1000 System (K180096).
| Feature | Subject Device | Predicate Device | Comparison |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| | AcuVu Inc. Inc.<br>HTx Disposable Hysteroscope<br>System | AcuVu Inc.<br>GDT-1000 System (K180096) | |
| Classification | Class II | Class II | Same |
| Regulation Name | 21 CFR § 884.1690 | 21 CFR § 884.1690 | Same |
| Product code | HIH | HIH | Same |
| Indications for use | The HTx disposable hysteroscope<br>system is intended to permit viewing<br>of the cervical canal and uterine<br>cavity for the purpose of performing | AcuVu GDT-1000 System is<br>used to permit viewing of the<br>adult cervical canal and uterine<br>cavity for the purpose of | Same |
| | diagnostic and operative procedure. | performing diagnostic and<br>operative procedure. | |
| Site of use | Hospitals and physician's office. | Hospitals and physician's<br>office. | Same |
| Intended users | Trained Medical Professionals. | Trained Medical Professionals. | Same |
| Device Features:<br>Components | (1) image-capturing system and<br>display; (2) attachable and disposable<br>cannula with inflow and outflow<br>ports. | (1) image-capturing hand piece<br>and tower; (2) attachable<br>cannula configuration with<br>inflow and outflow ports. | Different<br>technology;<br>similar<br>performance |
| Cannula outer<br>diameter | 4.5mm-6.2mm | 4.8 - 5.3 mm (operative)<br>4.8 - 5.3 mm (diagnostic) | Similar to<br>predicate device |
| Cannula length | Working length: 240 mm<br>Overall length: 330 mm | Working length: 220 - 240<br>mm<br>Overall length: 300 - 350 mm | Similar; within<br>range of<br>predicate |
| Illumination light<br>source | LEDs | LEDs | Same |
| Image resolution | The CMOS sensor consists of 400 x<br>400 (160,000) pixels. | The CMOS sensor consists of<br>400 x 400 (160,000) pixels. | Same |
| Optical image | Digital CMOS | Digital CMOS | Same |
| Disposable/ Reusable | Cannula is provided sterile for single<br>use and the image system and the<br>monitor is reusable | Cannula is provided sterile for<br>single use and the handle and<br>tower is reusable. | Same |
| Battery operated | No | No | Same |
| Duration of use | < 24 hours | < 24 hours | Same |
| Field of View and<br>Direction of Viewing | FOV>115 degrees,<br>DOV: HTx 60: 12 Degrees<br>HTx40: 8 degrees | FOV>115 degrees,<br>DOV 0 degree | Same FOV<br>Different DOV |
Table-1: Functional and Technological Comparison
{6}------------------------------------------------
Differences in technological characteristics do not raise different questions of safety and effectiveness.
{7}------------------------------------------------
# Performance Data
Non-clinical tests were performed in accordance with the applicable requirements of the following areas. Verification and validation activities are performed to ensure that productperformance meets design requirements.
## Sterilization
The device is provided sterile via ethylene oxide sterilization validation was performed in accordance with ISO 11135:2014.
## Biocompatibility
The biocompatibility evaluation was conducted in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Testing included:
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010) ●
- Irritation (ISO 10993-10:2010)
- Acute Systemic Toxicity (ISO 10993-11:2017) ●
- Material Mediated Pyrogenicity (ISO 10993-11:2017) .
The results of testing demonstrated the subject device is non-cytotoxic, non-sensitizing, nonirritating, not systemically toxic, and non-pyrogenetic.
## Electromagnetic Compatibility/Electrical Safety
The device was tested and found to be compliant with the following standards for electrical safety and EMC: IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2
#### Performance Testing
Optical performance (direction of view, field of view, distortion/resolution, luminous flux, color performance, photobiological safety), optical safety, and thermal safety testing were conducted in accordance with the standards listed below:
- ISO 8600-1:2015 Medical endoscopes and endotherapy devices Part 1: General ● requirements
- ISO 8600-3:2019 Medical endoscopes and endotherapy devices Part 3: Determination of ● Field of View and Direction of View of Endoscopes with Optics
- IEC 62471:2006 (First Edition) Photobiological safety of lamps and lamp systems
- IEC 60601-2-18:2009 Medical electrical equipment, Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
#### Software
Software documentation for a Moderate Level of Concern device is provided in support of the subject device per FDA's 2005 Guidance for the Content of Premarket Submissions for Software Contained in Medical Device.
{8}------------------------------------------------
# Conclusion:
The performance testing summarized above support a substantial equivalence determination. The performance testing demonstrates that the subject device is as safe and as effective as the legally marketed predicate device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.