K182675 · Polygon Medical, Inc. · HIH · Jan 4, 2019 · Obstetrics/Gynecology
Device Facts
Record ID
K182675
Device Name
Polygon Resection Device
Applicant
Polygon Medical, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Jan 4, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Polygon Resection Device is intended for intrauterine use by a trained gynecologist to hysteroscopically resect and remove tissue, including focal lesions such as endometrial polyps and retained products of conception.
Device Story
Handheld, manually operated hysteroscopic tissue removal device; used by gynecologists in clinical settings. Device shaft inserts into 3.0 mm hysteroscope working channel. Physician squeezes trigger to actuate cutting mechanism; rotation knob adjusts cutting bay orientation. Barb fitting connects to external vacuum system to aspirate tissue into cutting window for resection. No electrical or powered components. Facilitates removal of intrauterine lesions; improves procedural efficiency; enables tissue collection for pathology.
Clinical Evidence
Clinical usability study conducted with 5 physicians across 32 cases. Evaluated ease of use, successful tissue removal without clogging, and number of activations required. Results demonstrated the device functions as intended per instructions for use.
Technological Characteristics
Manually operated hysteroscopic tissue removal device. Materials evaluated per ISO 10993-1:2009. Sterilization per ISO 11135-1:2014. Features 36cm working length, 7mm cutting window, and in-line trigger. Connects to external vacuum system. No electrical/powered components.
Indications for Use
Indicated for intrauterine tissue resection and removal, including endometrial polyps and retained products of conception, in patients undergoing hysteroscopic procedures performed by trained gynecologists.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 4, 2019
Polygon Medical, Inc. % Christine Santagate Director. Northeast Regional Operations R&Q Solutions 15 Standish Rd Norfolk, Massachusetts 02056
Re: K182675
> Trade/Device Name: Polygon Resection Device Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: Class II Product Code: HIH Dated: December 4, 2018 Received: December 6, 2018
Dear Christine Santagate:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182675
Device Name Polygon Resection Device
Indications for Use (Describe)
The Polygon Resection Device is intended for intrauterine use by a trained gynecologist to hysteroscopically resect and remove tissue, including focal lesions such as endometrial polyps and retained products of conception.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
In accordance with 21 CFR 807.92(a) the following summary of information is provided:
#### Submitter Information
| Submitter's Name: | Ronald Adams |
|--------------------|-------------------------------------------------------------------|
| Address: | Polygon Medical, Inc.<br>18 Hillside Drive<br>Holliston, MA 01746 |
| Telephone: | 775-800-7300 |
| Fax: | 844-225-4600 |
| Contact Person: | Christine Santagate |
| Telephone : | 339-237-8122 |
| Date Prepared: | January 2, 2019 |
| Device Trade Name: | Polygon Resection Device |
| Date Prepared: | January 2, 2019 |
|-----------------------|-----------------------------------------|
| Device Trade Name: | Polygon Resection Device |
| Common/Usual Name: | Hysteroscopic Tissue Removal Device |
| Class: | II |
| Product Code(s): | HIH, Hysteroscope and Accessories |
| Regulation Number(s): | 884.1690 (Hysteroscope and Accessories) |
# Predicate Device(s):
K173901 Device Name: MyoSure MANUAL Tissue Removal Device Manufacturer: Hologic, Inc.
The predicate device has not been subject to a design related recall.
### Device Description:
The Polygon Resection Device is a handheld manually operated device designed for the hysteroscopic removal of intrauterine tissue. The device is provided sterile. The shaft of the device is design to fit into the working channel of any hysteroscope that includes a 3.0 mm (Fr) straight working channel.
The device is activated by the physician by squeezing the trigger. A rotation knob is provided to enable the physician to rotate the cutting bay into an orientation that aligns with the desired specimen. The barb fitting provides an attachment port for a vacuum system. The vacuum system pulls the specimen into the cutting window where it can be resected and aspirated out of the patient.
{4}------------------------------------------------
Image /page/4/Figure/1 description: The image shows a diagram of a surgical tool with labels pointing to different parts. The parts labeled are the barb fitting, rotation knob, shaft, cutting window, handle, and trigger. The tool has a handle with a trigger that is connected to a shaft with a cutting window at the end. The rotation knob is located near the handle and is used to rotate the shaft.
There are no electrical or powered connections necessary for operation of the device.
Figure 1. Polygon Resection Devic
# Indications for Use:
The Polygon Resection Device is intended for intrauterine use by a trained gynecologist to hysteroscopically resect and remove tissue, including focal lesions such as endometrial polyps and retained products of conception.
The subject device intended use is identical to the predicate device.
# Comparison of Technological Characteristics:
The Polygon Resection Device is similar in design, materials, and construction to the Hologic Myosure Manual (predicate device). The principal of operation for both devices is the same.
The primary differences between the subject and predicate devices are as follows:
- The morcellator working length is 36cm for the subject device vs. 32cm for the predicate device.
- . The cutting window dimensions are 7mm for the subject device vs. 10mm for the predicate device.
- The subject device has an in-line trigger orientation, while the predicate device has a ● perpendicular trigger design.
- The subject and predicate devices contain different patient-contacting materials. ●
These differences in technological characteristics between the subject and predicate devices do not raise different questions of safety or effectiveness.
# Non-Clinical Testing Summary:
The following data were provided to demonstrate substantial equivalence and functional performance of the subject device:
- Sterilization validation was conducted in accordance with ISO 11135-1: 2014. ●
{5}------------------------------------------------
- Shelf life validation testing was conducted including package integrity testing and functional testing.
- Biocompatibility evaluation in accordance with ISO 10993-1:2009 Biological . evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
- o Cytotoxicity testing
- Guinea Pig Maximization Sensitization testing o
- O Irritation testing
- Acute Systemic Toxicity testing o
- Design verification testing: ●
- 0 Joint connection strength
- Visual Inspection O
- Tissue trap removal force O
- Functional performance testing was conducted in a simulated use model: ●
- Hysteroscope compatibility O
- o Vacuum system compatibility
- Device cutting performance (successful removal of tissue without clogging, O cutting rate, and fluid usage)
All test samples met the acceptance criteria for each test.
### Clinical Usability Testing Summary:
Usability and human factors were assessed through a clinical usability study. The subject device was used to remove intrauterine tissue by five physicians with a range of experience in a total of 32 cases. The usability study evaluated ease of use, successful removal of tissue without clogging, and number of device activations required to complete the procedure. The usability testing demonstrated the subject device can be used per the Instructions for Use as intended.
### Conclusion:
Based on the intended use, descriptive information and performance provided in this submission, the Polygon Resection Device has been shown to be substantially equivalent to the predicate MyoSure Manual Tissue Removal Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.