Veloxion System, Veloxion Controller Kit, Veloxion Resecting Device Kit, Veloxion Fluid Control Set, Veloxion Saline Pole
Applicant
Corinth Medtech, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Jul 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Veloxion System is intended for intrauterine use by trained gynecologists for endoscopically controlled tissue chip resection and coagulation, and removal of intrauterine polyps and myomas via suction channel under continuous flow conditions following resection using a bipolar resecting device. It is also intended to distend the uterus by filling with saline to facilitate viewing with a hysteroscope and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.
Device Story
Veloxion System is a bipolar electrosurgical resectoscope for intrauterine procedures. Inputs: saline irrigation fluid; bipolar RF energy for tissue cutting/coagulation. Operation: Controller manages two peristaltic pumps for balanced irrigation/aspiration and monitors intrauterine pressure via dual independent sensors; software regulates fluid deficit and pressure. Output: resected tissue chips (via suction channel); visual feedback on fluid deficit and cavity pressure. Used in clinical settings by gynecologists. System includes controller, footswitch, sterile single-use resecting device, sheath, optical obturator, fluid control set, and saline pole. Benefits: continuous flow visualization, precise pressure control, and automated fluid deficit monitoring to enhance safety during hysteroscopic resection.
Clinical Evidence
Bench testing only. Testing included software verification/validation (IEC 62304), system integrity, functional testing (cut/coagulation, aspiration, irrigation, pressure control), dimensional inspection, simulated use (tissue resection, pressure regulation), durability testing, fluid deficit testing, comparative testing to predicates, electrical safety/EMC (IEC 60601-1, 60601-1-2, 60601-2-18, 60601-2-2), and usability risk evaluation.
Technological Characteristics
Bipolar RF energy; dual peristaltic pumps; saline irrigation. Materials: Tungsten (99.95% purity) electrode, FEP insulation. Dimensions: 26Fr (8.6mm) sheath OD, 193mm working length. Connectivity: Integrated controller with user interface for pressure/deficit monitoring. Sterilization: Sterile, single-use resecting device kit. Software: Rule-based control for fluid/pressure management.
Indications for Use
Indicated for trained gynecologists performing endoscopically controlled intrauterine tissue chip resection, coagulation, and removal of polyps and myomas. Includes uterine distension with saline and fluid deficit monitoring.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
Richard Wolf Resection Pump and Resectoscope for Chip Aspiration (K042523)
Richard Wolf Hystero Pump II 2222 and Fluid Monitoring 2223 (K022449)
ERBE ESU Model VIO 300D with Accessories (K023886)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 25, 2018
Corinth MedTech, Inc. Sandeep Saboo Vice President, Regulatory Affairs and Quality Assurance 1601 S. De Anza Blvd., Suite 221 Cupertino. CA 95014
Re: K180752
Trade/Device Name: Veloxion System Regulation Number: 21 CFR& 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH, HIG, GEI Dated: June 22, 2018 Received: June 25, 2018
Dear Sandeep Saboo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sharon M. Andrews -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K180752
Device Name Veloxion System
#### Indications for Use (Describe)
The Veloxion System is intended for intrauterine use by trained gynecologists for endoscopically controlled tissue chip resection and coagulation, and removal of intrauterine polyps and myomas via suction channel under continuous flow conditions following resection using a bipolar resecting device. It is also intended to distend the uterus by filling with saline to facilitate viewing with a hysteroscope and to monitor the volume differential between into and out of the uterus.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subnart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K180752: 510(k) Summary
## I. Submitter Information
| Submitter name: | Corinth MedTech, Inc.<br>1601 S. De Anza Blvd, Suite 221<br>Cupertino, CA 95014 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Sandeep Saboo<br>Vice President, Regulatory Affairs and Quality Assurance<br>Phone: (408) 996-2517<br>Fax: (408) 996-0621 |
| Date Prepared: | 23 July 2018 |
#### II. Product Classification
| Device Name: | Veloxion System | |
|---------------------------|------------------------------|----------------|
| Common Name: | Resectoscope | |
| Regulation: | 21 CFR 884.1690 | |
| Regulation Name: | Hysteroscope and accessories | Subject Device |
| Class: | II | |
| Product Code: | HIH | |
| Additional Product Codes: | HIG, GEI | |
#### III. Predicate Devices
The predicate device is the system comprised of the following legally marketed devices as used in combination to which substantial equivalence is claimed:
Primary Predicate: Richard Wolf Medical Instruments Corporation:
- · Resection Pump and Resectoscope for Chip Aspiration (K042523)
- · Hystero Pump II 2222 and Fluid Monitoring 2223 (K022449)
Predicate: ERBE, USA Inc., ERBE ESU Model VIO 300D with Accessories (K023886) Predicate: Veloxion System (cleared per K162979)
None of the predicates have been subject of a design related recall.
#### IV. Device Description
The Veloxion System consists of the following components:
- · Veloxion Controller (with Integrated Fluid Control)
- o Footswitch
- · Veloxion Resecting Device Kit:
- o Veloxion Resecting Device
- o Sheath
- 0 Continuous Flow Optical Obturator
- · Veloxion Fluid Control Set
- · Veloxion Saline Pole
The Veloxion System also includes the following Class I accessories for handling of waste collected from the patient (these items apply no aspiration to the patient, they only handle waste after it is already outside the patient), which includes:
{4}------------------------------------------------
K180752
- · Waste Accessory Bag for collection of waste aspirated form the patient and from under patient's buttocks,
- · Waste Accessory Tubing for collection of waste from under the patient's buttocks
- · Waste Management Pump for moving the waste to the Waste Bag.
The Veloxion System provides bipolar resection and coagulation of intrauterine tissue, it distends the uterus by filling with saline, it provides pressure control of the intrauterine cavity for insufflation to facilitate viewing with a hysteroscope and it monitors the fluid deficit (potential fluid absorbed by the patient's body) to the physician established limit. Therefore:
- · The Veloxion Controller provides bipolar radiofrequency outputs (for cut and coagulation) and fluid/pressure control through the use of two integrated peristaltic pumps, provides the user interface to establish the desired set pressure and the fluid deficit limits. The software then monitors, controls and notifies the user when the limits are reached or when specific conditions are met.
- The Veloxion Resecting Device is a sterile single use hand held bipolar radiofrequency device configured for the resection of tissue and aspiration of resected chips. Fluid inflow and aspiration of the resected chips are controlled by the Controller's peristaltic pumps.
- · The Veloxion Fluid Control Set is a sterile single use device that provides conduits for fluid inflow, aspiration of resected tissue and fluids and a pressure sensor that provides two independent intrauterine pressure measurement (for redundancy) for monitoring cavity pressure during the procedure thereby facilitating the insufflation function.
- · The Veloxion Saline Pole enables monitoring of saline remaining in the saline bag and facilitates the fluid deficit function.
## V. Indications for Use
The Veloxion System is intended for intrauterine use by trained gynecologists for endoscopically controlled tissue chip resection and coagulation, and removal of intrauterine polyps and myomas via suction channel under continuous flow conditions following resection using a bipolar resecting device. It is also intended to distend the uterus by filling with saline to facilitate viewing with a hysteroscope and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus.
There are minor differences in the indications for use for the subject device when compared to the predicate system when the devices that comprise the predicate system are used together per their labeling to perform a hysteroscopy procedure. The predicate system is comprised of the following devices:
- Richard Wolf Resectoscope and Resection Pump for Chip Aspiration ●
- Richard Wolf Hystero Pump II (Model 2222) and Fluid Monitoring Display (Model 2223)
- ERBE ESU Model VIO 300 D with Accessories ●
The indications for use for the subject device when compared to the predicate Veloxion System differ with respect to the tissue type.
{5}------------------------------------------------
Veloxion System
Traditional 510(k) Premarket Notification
| Comparison of Indications for Use | |
|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Indications For Use |
| Veloxion System<br>(Subject Device) | The Veloxion System is intended for intrauterine use by trained<br>gynecologists for endoscopically controlled tissue chip resection and<br>coagulation, and removal of intrauterine polyps and myomas via suction<br>channel under continuous flow conditions following resection using a<br>bipolar resecting device. It is also intended to distend the uterus by filling<br>with saline to facilitate viewing with a hysteroscope and to monitor the<br>volume differential between fluid flowing into and out of the uterus. |
| Richard Wolf Fluid<br>Manager<br>(Predicate K022449)<br>Fluid Monitoring<br>Display (Model 2223)<br>and Hystero Pump II<br>(Model 2222) | The Hysteroscopy Fluid Management System is indicated for liquid<br>distension of the uterus for diagnostic and operative hysteroscopic<br>procedures, and monitoring the volume differential between the irrigation<br>fluid flowing into and out of the uterus. |
| Resection Pump and<br>Resectoscope for Chip<br>Resection<br>(Primary Predicate,<br>K042523) | The Resection Pump 2228 with Resectoscope 8659.xxx is used for<br>endoscopically controlled tissue chip resection and removal of intrauterine<br>polyps, intrauterine myoma or prostate adenomas via suction channel<br>under continuous flow conditions following resection using a high-<br>frequency electrode with a Resectoscope. |
| ERBE ESU Model VIO<br>300 D with Accessories<br>(Predicate, K023886) | The ERBE ESU Model VIO 300 D with Accessories is intended to deliver<br>High Frequency (HF) electrical current for the cutting and/or coagulation<br>of tissue. |
| Veloxion System<br>K162979<br>(Predicate Device) | The Veloxion System is intended for endoscopically controlled tissue chip<br>resection and coagulation and removal of prostate adenomas via suction<br>channel under continuous flow conditions following resection using a<br>bipolar resecting device. |
The differences in indication statements do not represent a new intended use.
#### VI. Comparison of Technological Characteristics with the Predicate Device
The Veloxion System (subject device for hysteroscopy) and the predicate system (Richard Wolf Resectoscope and Resection Pump for Chip Aspiration (K042523), ERBE ESU Model VIO 300 D with Accessories (K023886) and Fluid Monitoring Display Model 2223 and Hystero Pump II Model 2222 (K022449)) and the previously cleared Veloxion System (K162979) have the same or similar technological characteristics in terms of basic operating principle and basic design features with one key difference being that Veloxion System includes dual, balanced pump system while the predicate system includes single pump for irrigation only. Both systems include roller peristaltic pumps and are to be used with specifically designed fluid monitoring unit and tube sets.
{6}------------------------------------------------
## Veloxion System Traditional 510(k) Premarket Notification
| Technological Characteristics | VELOXION SYSTEM<br>(Subject and Predicate) | PREDICATE SYSTEM Devices<br>ERBE ESU Model VIO 300 D with Accessories [K023886]<br>Richard Wolf Resection Pump [K042523]<br>Richard Wolf Fluid Manager [K022449] | |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Monopolar or Bipolar | Bipolar | | Bipolar or Monopolar |
| Energy Type: | Radiofrequency, Bipolar | Radiofrequency, Bipolar or Monopolar Resection | |
| Optics: | Compatible with commercially available 30degree Endoscopes. | Compatible with 30degree Richard Wolf Endoscope | |
| Able to Set & Monitor Fluid Deficit? | YES | YES | |
| Able to Set Cavity Pressure on the Device? | YES | YES | |
| Continuous Flow? : | YES | YES | |
| Irrigation fluid: | Saline | Saline | |
| Pump? | Dual Pump (Irrigation, Aspiration) | Single Pump (Irrigation only) | |
| Able to Monitor saline remaining?: | YES | NO | |
| Irrigation Flow Rate: | Programmed irrigation flow rate constant for each mode<br>Steady State = 100ml/min, Cut=380ml/min, Coag=100ml/min for<br>Subject & Predicate<br>ASPIRATE=400ml/min for subject and 750ml/min for predicate. | Max. Irrigation flow rate is Set by the User from 150 to<br>500ml/min. However, actual flow rate depends on the actual<br>cavity pressure and outflow stopcock opening. | |
| Set Pressure Range: | 35 to 125mmHg (subject)<br>15 to 60mmHg (predicate) | 15 to 150mmHg | |
| Positive Action to Increase Above 100mmHg: | YES (subject)<br>40mmHg (predicate) | YES | |
| Pressure Sensor: | Dual, independent, direct sensing of cavity pressure | Dual, independent, pressure sensing of Irrigation tube (indirect<br>measurement) | |
| Over-pressure condition detection: | YES | YES | |
| Maximum Allowable Actual Cavity Pressure | 135mmHg for 5 seconds (subject)<br>75mmHg (predicate) | 200mmHg for 5 seconds | |
| Ultimate mitigation for risk of over-pressurization of<br>uterus if all designed mitigations are unsuccessful | • Non-defeatable, continuous notification tone<br>• Notification displayed:<br>"Remove device from cavity. Check for Clog." | • Continuous notification tone<br>• Over-pressure display icon is illuminated and Pressure<br>Display flashes | |
| Set Fluid Deficit Range: | From 250 to 2500ml. Device stops when limit is reached (subject<br>only) | From 0 to 2000ml but user can exceed 2500ml, if needed after<br>reaching 2000ml limit. | |
| Fluid Deficit Rate ≥ 300ml/min notification to user? | YES (subject only) | YES | |
| Notification to user with pause when Actual fluid<br>deficit is 1000ml? | YES (subject only) | NO | |
| Notification to user when Actual fluid deficit is 250ml<br>less than Set fluid deficit? | YES (subject only) | NO | |
| Shaft OD (Sheath): | With Sheath Assembled: 26Fr (8.6mm) | With Sheath Assembled: 27Fr (9.0mm) | |
| Technological Characteristics | VELOXION SYSTEM<br>(Subject and Predicate) | PREDICATE SYSTEM Devices<br>ERBE ESU Model VIO 300 D with Accessories [K023886]<br>Richard Wolf Resection Pump [K042523]<br>Richard Wolf Fluid Manager [K022449] | |
| Working Length: | Sheath Working Length: 193mm | Sheath Working Length: 186mm | |
| Materials (Electrode/ Insulation): | Electrode: Tungsten (99.95% purity)<br>Insulation: FEP | Electrode: Tungsten<br>Electrode Insulation: PTFE | |
| How Supplied (Sterility): | Sterile, Single Use<br>(Veloxion Resecting Device Kit) | Sterile component:<br>Non-Sterile, Reusable Components: | Electrode<br>Working Element<br>Outer Sheath<br>Obturator |
{7}------------------------------------------------
## Veloxion System Traditional 510(k) Premarket Notification
{8}------------------------------------------------
K180752
The main technological differences between the subject Veloxion System and the predicate system are the following:
- . The Veloxion System includes a dual, balanced pump system for a more controlled cavity pressure.
- The Veloxion System includes programmed flow rates that are equivalent to flow rate delivered using the predicate system which also varies based on user adjustment of outflow stopcock.
- . The patient contacting components of the Veloxion System are sterile, single use while the predicate system components are sterile, single use for some components and reusable for other components.
The differences outlined were evaluated through performance testing to demonstrate safety and effectiveness of the subject Veloxion System.
## VII. Performance Data
The following performance data have been provided in support of the substantial equivalence determination.
- Software Verification and Validation Testing performed per IEC 62304 and ● documentation provided per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
- Other Tests were performed per approved test protocols which included: ●
- Integrity: System withstands operating pressures o
- Functional Testing: Cut and coagulation, aspiration, irrigation, pressure control O
- Dimensional Inspection and Testing O
- Simulated Use: Tissue resection and spot coagulation, regulation of cavity pressure O
- Durability Testing: Electrode durability testing for tissue resection and coagulation. o
- Fluid deficit testing O
- o Comparative testing to predicate for electrode durability, pressure control and fluid control.
- Electrical Safety & EMC: In accordance with IEC 60601-1:2005, IEC 60601-1-O 2:2014, IEC 60601-2-18:2009 and IEC 60601-2-2:2009.
- Usability Testing: Use related risk evaluation o
## VIII. Conclusions
Based on the results of performance tests, the subject Veloxion System is considered to be substantially equivalent and as safe and effective as the predicate system:
- · Richard Wolf Resectoscope and Resection Pump for Chip Aspiration,
- · ERBE ESU Model VIO 300 D with Accessories, and
- · Richard Wolf Fluid Manager (Fluid Monitoring Display (Model 2223) and Hystero Pump II (Model 2222)).
as well as the previously cleared Veloxion System (K162979).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.