LiNA OperaScope with HDMI cable and on-board LCD-single unit, LiNA OperaScope with HDMI cable and on-board LCD-6 units, LiNA OperaScope Recording Module
K171113 · Lina Medical Aps · HIH · Jan 4, 2018 · Obstetrics/Gynecology
Device Facts
Record ID
K171113
Device Name
LiNA OperaScope with HDMI cable and on-board LCD-single unit, LiNA OperaScope with HDMI cable and on-board LCD-6 units, LiNA OperaScope Recording Module
Applicant
Lina Medical Aps
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Jan 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LiNA OperaScope is intended for use in visualization of the cervical canal and uterine cavity during diagnostic and therapeutic gynecological procedures. The types of procedures where the OperaScope could offer visualization include: Assessment of abnormal bleeding, pelvic pain, amenorrhea and abnormal findings from hysterosalpingogram; Assessment of infertility and pregnancy wastage; Confirmation of the presence of intrauterine foreign body; Assist in locating submucosal fibroids and polyps targeted for removal; Provide visual guidance during directed biopsy, submucosal myomectomy, transection of intrauterine and septa.
Device Story
Handheld, battery-operated portable hysteroscope; provides direct visualization of cervical canal and uterine cavity via integrated CMOS camera and LED illumination at distal tip. Features inflow/outflow channels for irrigation. Captures video/images via optional Recording Module (connected via HDMI). Used in hospitals and physician offices by trained medical professionals. Provides real-time visual feedback on on-board LCD or external monitor; assists clinicians in diagnostic assessments and surgical guidance (biopsy, myomectomy). Benefits include portable, single-use sterile design reducing cross-contamination risk and simplifying workflow.
Clinical Evidence
Bench testing only. Performance validated per FDA guidance for hysteroscopes. Testing included optical performance (field of view, spatial resolution, distortion), fluid delivery (inflow/outflow, leak testing), mechanical integrity (bending, tensile, torque), electrical safety (IEC 60601-1, 60601-2-18), EMC (IEC 60601-1-2), biocompatibility (ISO 10993), sterilization validation (ISO 11135, 11137), and software verification/validation (IEC 62304). Visualization and illumination adequacy confirmed via hysteroscopy simulator.
Technological Characteristics
Handheld hysteroscope; CMOS camera; LED illumination; inflow/outflow channels; 4.3-inch LCD display; HDMI connectivity. Materials compliant with ISO 10993. Powered by 3V (2 AA batteries) or 9V DC external supply. Sterilization: Ethylene Oxide (handpiece), Gamma radiation (batteries). Software: Moderate level of concern.
Indications for Use
Indicated for visualization of cervical canal and uterine cavity in adult females undergoing diagnostic or therapeutic gynecological procedures, including assessment of abnormal bleeding, pelvic pain, amenorrhea, infertility, pregnancy wastage, foreign body, fibroids, polyps, and guidance for biopsy, myomectomy, or septa/adhesion transection.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
U-Scope 8000 System with HSC+EMB Cannula (K132384)
Gyrus ACMI Invisio Digital Hysteroscope System (K092278)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
January 4, 2018
LiNA Medical ApS % Jennifer Tribbett Director of Regulatory Affairs Emergo Global Consulting, LLC 2500 Bee Cave Rd. Bldg. 1. Suite 300 Austin, TX 78746
Re: K171113
Trade/Device Name: LiNA OperaScope Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: November 22, 2017 Received: November 30, 2017
Dear Jennifer Tribbett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171113
Device Name LiNA OperaScope
#### Indications for Use (Describe)
The LiNA OperaScope is intended for use in visualization of the cervical canal and uterine cavity during diagnostic and therapeutic gynecological procedures.
The types of procedures where the OperaScope could offer visualization include:
· Assessment of abnormal bleeding, pelvic pain, amenorthea and abnormal findings from hysterosalpingogram;
- · Assessment of infertility and pregnancy wastage;
· Confirmation of the presence of intrauterine foreign body;
· Assist in locating submucosal fibroids and polyps targeted for removal;
· Provide visual guidance during directed biopsy, submucosal myomectomy, transection of intrauterine and septa.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 12pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------|
| <span style="font-size: 12pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# LiNA OperaScope
## K171113
#### 1. Submission Sponsor
LiNA Medical ApS
Formervangen 5
DK-2600 Glostrup, Denmark
Contact: Natalia Szychulska
Title: Regulatory Affairs Officer
Office Phone: +48 61 222 21 43
e-Mail: nhe@lina-medical.com
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Contact: Jennifer Tribbett
Title: Director of Regulatory Affairs and Quality Assurance
Office Phone: (512) 327.9997
#### 3. Date Prepared
January 2, 2018
#### 4. Device Identification
| Trade/Proprietary Name: | LiNA OperaScope |
|-------------------------|------------------------------|
| Common/Usual Name: | Hysteroscope |
| Classification Name: | Hysteroscope and Accessories |
{5}------------------------------------------------
| Regulation Number: | 884.1690 |
|-----------------------|-----------------------------------|
| Product Code: | HIH, Hysteroscope and Accessories |
| Device Class: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
#### 5. Legally Marketed Predicate Device(s)
- . K132384 (Primary Predicate): U-Scope 8000 System with HSC+EMB Cannula
- K092278: Gyrus ACMI Invisio Digital Hysteroscope System ●
LiNA Medical performed a search of FDA's Medical Device Recall data base and did not discover any recalls for these legally marketed predicate devices.
#### 6. Indication for Use Statement
The LiNA OperaScope is intended for use in visualization of the cervical canal and uterine cavity during diagnostic and therapeutic gynecological procedures.
The types of procedures where the OperaScope could offer visualization include:
• Assessment of abnormal bleeding, pelvic pain, amenorrhea and abnormal findings from hysterosalpingogram;
- Assessment of Infertility and pregnancy wastage;
- Confirmation of the presence of intrauterine foreign body;
- Assist in locating submucosal fibroids and polyps targeted for removal;
• Provide visual guidance during directed biopsy, submucosal myomectomy, transection of intrauterine adhesions and septa.
Table A compares the OperaScope to the predicate devices with respect to the Indications for Use.
The OperaScope Indications for Use statement is basically a combination of the two predicate device Indications for Use. Although the wording is slightly different between the subject and predicate devices, the overall intent is the same and does not expand the clinical use of the subject device.
{6}------------------------------------------------
| Manufacturer | LiNA Medical ApS | EndoSee Corporation | Gyrus ACMI Inc. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | OperaScope | U-Scope 8000 HSC+EMB<br>Cannula | Gyrus ACMI Invisio<br>Digital Hysteroscope<br>System |
| 510(k) Number | K171113 | K132384 | K092278 |
| Indications for Use | The LiNA OperaScope is<br>intended for use in<br>visualization of the<br>cervical canal and uterine<br>cavity during diagnostic<br>and therapeutic<br>gynecological procedures.<br>The types of procedures<br>where the OperaScope<br>could offer visualization<br>include:<br>• Assessment of abnormal<br>bleeding, pelvic pain,<br>amenorrhea and<br>abnormal findings from<br>hysterosalpingogram;<br>• Assessment of Infertility<br>and pregnancy wastage;<br>• Confirmation of the<br>presence of intrauterine<br>foreign body;<br>• Assist in locating<br>submucosal fibroids and<br>polyps targeted for<br>removal;<br>• Provide visual guidance<br>during directed biopsy,<br>submucosal<br>myomectomy, transection<br>of intrauterine<br>adhesions and septa. | Intended to be used to<br>permit viewing of the<br>cervical canal and uterine<br>cavity for the purpose of<br>performing diagnostic<br>procedures and to obtain<br>an endometrial tissue<br>sample (biopsy).<br>The sample is used for<br>cytologic and histologic<br>diagnosis. Generally<br>recognized indications<br>for diagnostic<br>hysteroscopy include:<br>abnormal bleeding,<br>infertility and pregnancy<br>wastage, evaluation of<br>abnormal<br>hysterosalpingogram,<br>intrauterine foreign<br>body, amenorrhea and<br>pelvic pain. | Intended to be used to<br>permit viewing of the<br>cervical canal and uterine<br>cavity for the purpose of<br>performing diagnostic<br>and surgical procedures.<br>Diagnostic Hysteroscopy:<br>• Abnormal uterine<br>bleeding<br>• Infertility &<br>pregnancy wastage<br>• Evaluation of<br>abnormal<br>hysterosalpingogram<br>• Intrauterine foreign<br>body<br>• Amenorrhea<br>• Pelvic pain<br>Operative Hysteroscopy:<br>• Directed biopsy<br>• Removal of fibroids<br>and polyps<br>• Transection of<br>intrauterine<br>adhesions<br>• Transection<br>intrauterine septa |
## Table A – Comparison of Indications for Use
{7}------------------------------------------------
#### 7. Device Description
The LiNA OperaScope is a handheld, battery-operated portable hysteroscope which allows direct viewing of the cervical canal and uterine cavity via the on-board LCD display. If the user wants to capture images and/or record video sequences utilizing a USB stick, a Recording Module is available, which connects to the device via the HDMI cable. The OperaScope can also be connected to an external monitor via the HDMI cable. External monitors are not provided by LiNA Medical ApS and therefore, monitors are not the subject of this 510(k) submission.
The OperaScope is supplied as a sterile, single use device and the Recording Module is supplied as a nonsterile reusable unit. The OperaScope contains a miniature Complimentary Medical Oxide Semi-conductor (CMOS) camera and a Light-Emitting Diode (LED) illumination at the tip, a channel for infusion of irrigating fluid and a channel for fluid outflow, a display unit (LCD display), microelectronics, firmware and single use batteries for powering the device. The Recording Module contains microelectronics, software, controls for recording pictures and video sequences, USB port for storage of pic/video, HDMI port for connection of the OperaScope, HDMI port for connecting an external monitor and an input port for external 9V DC power supply.
| Ref. number | Description |
|-------------|--------------------------------------------------------------------|
| OP-201-1 | LiNA OperaScope™ with HDMI cable and<br>on-board LCD – single unit |
| OP-201-6 | LiNA OperaScope™ with HDMI cable and<br>on-board LCD – 6 units |
| OP-RM-1 | LiNA OperaScope™ Recording Module |
#### Model Numbers
#### 8. Substantial Equivalence Discussion
- 1) U-Scope 8000 HSC + EMB Cannula (K132384)
The OperaScope is substantially equivalent to the EndoSee Corporation's U-Scope 8000 HSC+EMB Cannula (K132384) in terms of technological characteristics and intended use.
- 2) Gyrus ACMI Invisio Digital Hysteroscope System (K092278)
The OperaScope is substantially equivalent to the Gyrus ACMI Invisio Digital Hysteroscope System (K092278) in terms of indication for use.
Table B compares the OperaScope to the predicate devices with respect to the mode of operation, technological characteristics, materials, and performance testing. The OperaScope is substantially equivalent to the U-Scope 8000 HSC+EMB Cannula (K132384) in terms of intended use and technological
{8}------------------------------------------------
characteristics such as the CMOS camera, LED illumination, inflow/outflow channel, image capturing features and LCD display unit and to the Gyrus ACMI Invisio Digital Hysteroscope System (K092278) in terms of indication for use.
| Manufacturer | LiNA Medical ApS | EndoSee Corporation | Gyrus ACMI Inc. |
|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | OperaScope | U-Scope 8000 HSC+EMB<br>Cannula | Gyrus ACMI Invisio<br>Digital Hysteroscope<br>System |
| 510(k) Number | K171113 | K132384 | K092278 |
| Product Code | HIH | SAME | SAME |
| Regulation Number | 884.1690 | SAME | SAME |
| Regulation Name | Hysteroscope and<br>Accessories | SAME | SAME |
| Intended Users | Trained Medical<br>Professionals | SAME | SAME |
| Site of Use | Hospitals and Physician<br>Offices | SAME | SAME |
| Device Features and<br>Components | ● Sterile/Single-Use<br>Handpiece<br>(Hysteroscope) with<br>CMOS camera, LED<br>illumination, LCD<br>display unit, channel<br>for infusion of<br>irrigating fluid,<br>channel for fluid<br>outflow,<br>microelectronics,<br>firmware and single<br>use batteries for<br>powering the device.<br>● Non-Sterile/Reusable<br>Recording Module<br>with software,<br>microelectronics and<br>controls for recording<br>pictures/video<br>sequences. USB port<br>for storage of<br>pictures/video, HDMI<br>ports (2) for | ● Sterile/Single-Use<br>Biopsy Cannula with<br>CMOS camera, LED<br>illumination, channel<br>for irrigating fluid,<br>channel for<br>aspiration of tissue.<br>● Non-Sterile/Reusable<br>HandTower with<br>connector and<br>locking mechanism<br>to attach/detach<br>cannula, LCD display<br>unit, power button,<br>illumination<br>brightness adjuster,<br>video processor,<br>electronics,<br>microcontrollers,<br>firmware for capture<br>of single image/video<br>procedures,<br>rechargeable battery. | ● Non-Sterile<br>(sterilization by the<br>user)/Reusable<br>Hysteroscope with<br>CMOS sensor, LED<br>light source,<br>electrical wiring,<br>printed circuit board<br>(PCB), electrical cord<br>to connect to the<br>Camera Control Unit<br>(CCU)<br>● Non-Sterile/Reusable<br>Camera Control Unit<br>(CCU) with PCBs,<br>software, power<br>supply, power cables<br>to process and<br>display images<br>transmitted by the<br>camera. |
| Manufacturer | LiNA Medical ApS | EndoSee Corporation | Gyrus ACMI Inc. |
| Trade Name | OperaScope | U-Scope 8000 HSC+EMB<br>Cannula | Gyrus ACMI Invisio<br>Digital Hysteroscope<br>System |
| | connection to<br>handpiece<br>(hysteroscope) and<br>external monitor and<br>input port for external<br>9V DC power supply. | | |
| Optical Image | Digital CMOS Technology | SAME | SAME |
| Image Resolution | CMOS camera consists of<br>approximately 160,000<br>pixels | SAME | Approximately 50,000<br>pixels |
| Illumination Light<br>Source | LED at tip of cannula | SAME | LED integrated in handle |
| Light at 20mm from<br>target. Range high<br>to low level [LUX] | 322 - 66 LUX | 325 - 74 LUX | 675 - 670 LUX |
| Inflow/outflow<br>channel design | Inflow: Inflow channel and<br>working channel<br>combined in one channel<br>above the camera.<br>Outflow: One outflow<br>channel with two<br>entrances placed on top<br>and bottom of distal tip. | Inflow: Inflow occurs<br>through two smaller<br>inflow channels on each<br>side of the camera.<br>Outflow: No outflow<br>channel. | Inflow: Inflow channel<br>and working channel<br>combined in one channel<br>below the camera.<br>Outflow: No outflow<br>channel. |
| Cannula tip design | Rounded polymer tip.<br>Angled shaft proximal to<br>tip (~20°). Steerable by<br>rotation knob. | Rounded polymer tip. 25°<br>angled shaft proximal to<br>tip. Steerable by rotation<br>of device. | Rounded stainless steep<br>tip. Flexible. Sideward<br>steerable 0 - 30° by<br>control means on handle. |
| Image Transmission | Transmits images from<br>CMOS video camera to<br>local LCD display unit on<br>the handpiece and remote<br>monitor | Transmits images from<br>CMOS video camera to<br>local LCD display unit on<br>the HandTower | Transmits images from<br>CMOS video camera to<br>remote monitor |
| Maximum Insertion<br>Diameter (Tip) | 4.3 mm | 4.8 mm | 5.5 mm with continuous<br>flow sheath |
| Shaft/Cannula<br>Length | 240mm (24 cm) | 270mm (27 cm) | Information not available |
| Manufacturer | LiNA Medical ApS | EndoSee Corporation | Gyrus ACMI Inc. |
| Trade Name | OperaScope | U-Scope 8000 HSC+EMB<br>Cannula | Gyrus ACMI Invisio<br>Digital Hysteroscope<br>System |
| LCD Display Size | 4.3 inches (diagonal) | 3.5 inches (diagonal) | Not applicable |
| Energy Source | Handpiece: 3V (2 AA batteries) Recording Module: 9V DC external power adapter | 3.7V Battery | CCU: 110-240V |
| Sterilization | Handpiece (Hysteroscope): Ethylene Oxide (EO) Handpiece Batteries: Radiation (gamma) Recording Module: Non-Sterile | Cannula: Ethylene Oxide (EO) HandTower: Non-Sterile | Hysteroscope: Non-Sterile CCU: Non-Sterile |
| Tissue Contact<br>Materials | Compliant with ISO 10993 | SAME | SAME |
#### Table B – Comparison of Characteristics
{9}------------------------------------------------
{10}------------------------------------------------
The OperaScope shares the same or similar mode of operall technical and functional capabilities as the predicate devices. In addition, the OperaScope is similar in design and function to the U-Scope 8000 HSC + EMB Cannula (K132384) in terms of mode of operation and use. The minor differences in the image transmission, cannula length and battery voltage do not raise any new issues of safety or effectiveness.
#### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the OperaScope and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, LiNA completed a number of non-clinical performance tests. The OperaScope meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The OperaScope was subjected to safety and optical performance testing in accordance with the FDA's "Hysteroscopes and Gynecological Laparoscopes Submission Guidance for a 510(k)" (March 7, 1996).
The OperaScope passed all applicable testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
{11}------------------------------------------------
- . Field of view and direction of view per ISO 8600-3
- . Maximum width of insertion portion per ISO 8600-4
- Biocompatibility testing per ISO 10993-1, ISO 10993-10, ISO 10993-10, ISO 10993-12
- . EO and ECH residual testing per ISO 10993-7
- . Electromagnetic Compatibility per EN/IEC 60601-1-2 (EMC and Immunity)
- Electrical Safety per AAMI/ANSI ES60601-1 and EN 60601-2-18
- . Sterilization Validation per ISO 11135 (Ethylene Oxide – OperaScope), ISO 11137-1 and ISO 11137-2 (Radiation – OperaScope Batteries) and maintenance of sterilization per ISO 11737-2.
- Mechanical testing for bending (displacement), tensile strength and torque.
- . Spatial Resolution utilizing ISO 3334 (test chart No. 2)
- Visualization and illumination study to determine the adequacy of imaging while in use in the uterine cavity utilizing a hysteroscopy simulator.
- . In-flow and out-flow fluid delivery, including leak testing
- Fuctionality testing to verify compatibility of instruments and accessories with respect to dimensions
- . Shelf-Life and Simulated shipping conditions per ASTM D4169
- Battery lifetime
- Software Validation and Verification per IEC 62304
- . FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices [Moderate Level of Concern Applied to the OperaScope]
- . FDA's guidance document Hysteroscopes and Gynecologic Laparoscopes Submission Guidance for 510(K), including section V(E)(6b)/distortion characteristics.
#### 10. Statement of Substantial Equivalence
The OperaScope, as designed and manufactured, is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.