← Product Code [HIH](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIH) · K161763

# Smith & Nephew TRUCLEAR ULTRA Mini Tissue Removal Device (K161763)

_Smith & Nephew, Inc. · HIH · Nov 22, 2016 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIH/K161763

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [HIH](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIH.md)
- **Decision Date:** Nov 22, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1690
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The TRUCLEAR Tissue Removal Devices are intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: Submucous myomas Endometrial polyps Retained products of conception

## Device Story

The TRUCLEAR ULTRA Mini Tissue Removal Device is a single-use, mechanical resection instrument used by gynecologists during hysteroscopic procedures. It consists of a stainless steel inner tube that rotates and reciprocates within a stainless steel outer tube, housed in a plastic assembly that connects to a TRUCLEAR Handpiece. The device features a sharp cutting edge at the distal end of the outer tube; as the inner tube moves, it resects tissue while simultaneously aspirating it out of the uterine cavity via suction. The device is used in a clinical setting to treat intrauterine pathologies. By providing a continuous mechanical cutting and aspiration mechanism, it allows for efficient tissue removal, potentially reducing procedure time and improving patient outcomes compared to traditional methods.

## Clinical Evidence

Bench testing only. Testing included: shed testing for blade wear; cutting performance (20-minute activation durability); resection rate evaluation; biocompatibility (cytotoxicity, sensitization, irritation, and chemical extract analysis); and accelerated aging shelf-life testing. All tests met pre-defined acceptance criteria.

## Technological Characteristics

Single-use, mechanical tissue removal device. Materials: 304 stainless steel (outer shaft), 440 stainless steel (inner shaft), chromium coating, and silicone lubricant. Principle: Rotating and reciprocating mechanical resection with simultaneous suction. Dimensions/form factor: Designed for use with TRUCLEAR 5.0 or 5C hysteroscope. Sterilization: Ethylene Oxide (EtO).

## Regulatory Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

## Predicate Devices

- TRUCLEAR ULTRA PLUS Tissue Removal Device ([K132015](/device/K132015.md))

## Reference Devices

- TRUCLEAR INCISOR Tissue Removal Device ([K132015](/device/K132015.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be emerging from a single form. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 22, 2016

Smith & Nephew, Inc. Bradley Heil Regulatory Affairs Manager 150 Minuteman Rd. Andover, MA 01810

> K161763 Trade/Device Name: Truclear Ultra Mini Tissue Removal Device Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: Class II Product Code: HIH Dated: November 11, 2016 Received: November 17, 2016

Dear Bradley Heil,

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K161763

Device Name

TRUCLEAR ULTRA Mini Tissue Removal Device

Indications for Use (Describe) The TRUCLEAR Tissue Removal Devices are intenderine use by trained gynecologists to hysteroscopically resect and remove tissue such as: Submucous myomas Endometrial polyps Retained products of conception

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Smith & Nephew logo. The logo consists of an orange flower-like symbol on the left, followed by the text "We are smith&nephew". The words "We are" are in gray, while "smith&nephew" is in orange.

## 510(k) Summary K161763

| Submitted by:                                | Smith & Nephew, Inc.<br>150 Minuteman Road<br>Andover, MA 01810                                                                    |  |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--|
| Date of Summary:                             | November 18, 2016                                                                                                                  |  |
| Contact Person and Address:                  | Bradley Heil<br>Regulatory Affairs Manager<br>T 901-216-6884<br>F 901-566-7831<br>Bradley.heil@smith-nephew.com                    |  |
| Name of Device:                              | TRUCLEAR ULTRA MINI Tissue<br>Removal Device                                                                                       |  |
| Common Name:                                 | Hysteroscopic Tissue Removal Device                                                                                                |  |
| Device Classification Name and<br>Reference: | 21 CFR 884.1690 Hysteroscope and Accessories                                                                                       |  |
| Device Class:                                | Class II                                                                                                                           |  |
| Panel Code:                                  | Obstetrics and Gynecology                                                                                                          |  |
| Product Code:                                | HIH (Hysteroscope and accessories)                                                                                                 |  |
| Predicate Device                             | TRUCLEAR ULTRA PLUS Tissue Removal Device<br>(K132015)<br>The predicate device has not been subject to a design<br>related recall. |  |
| Reference Device                             | TRUCLEAR INCISOR Tissue Removal Device<br>(K132015)                                                                                |  |

#### Device Description

Smith & Nephew's TRUCLEAR ULTRA Mini Tissue Removal Device is a single use device that hysteroscopically removes intra-uterine tissue. The device consists of a stainless steel inner tube that rotates and reciprocates inside a stainless steel outer tube. The proximal ends of the tubes are enclosed within a plastic housing. The housing is engaged with the TRUCLEAR Handpiece which drives the inner tube of the device. The inner tube has a sharp cutting edge that resects tissue in the cutting window at the distal end of the outer tube. The resected tissue is simultaneously aspirated out of the uterine cavity using suction.

{4}------------------------------------------------

### Indications for Use

The TRUCLEAR Tissue Removal Devices are intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: Submucous myomas Endometrial polyps Retained products of conception

The indication for use statement for the subject device is identical to the predicate device.

|                           | TRUCLEAR ULTRA<br>Mini (Subject Device)    | TRUCLEAR ULTRA PLUS Device<br>(Predicate Device) |
|---------------------------|--------------------------------------------|--------------------------------------------------|
| Principle of<br>Operation | Mechanical Resection                       | Mechanical Resection                             |
| Style of Device           | Reciprocating                              | Reciprocating                                    |
| Scope used                | TRUCLEAR 5.0 or 5C<br>Hysteroscope         | TRUCLEAR 8.0 Hysteroscope                        |
| Outer and inner<br>shaft  | 304 Stainless Steel<br>440 Stainless steel | 304 Stainless Steel<br>465 Stainless Steel       |
| Coating                   | Chromium coating                           | none                                             |
| Inner shaft Lubricant     | Silicone                                   | Silicone                                         |
| Sterility                 | EtO                                        | EtO                                              |
| Single Use                | Yes                                        | Yes                                              |

#### Comparison of Technological Characteristics

## Substantial Equivalence Information

The subject device has different technological characteristics compared to the predicate device.

The differences between the subject device and the predicate do not raise different types of safety and effectiveness questions, as these differences have been evaluated in previous 510(k) submissions for hysteroscopic tissue removal devices. The differences in technological characteristics can be evaluated through bench testing. The differences in material can be evaluated through biocompatibility assessment.

#### Summary of Pre-clinical Testing

To further support a determination of substantial equivalence, non-clinical bench testing was conducted to support the subject device. The specific types of non-clinical testing conducted are listed below.

{5}------------------------------------------------

- Shed Test- A shed test was conducted to determine the relative wear of the blade during use. Results from the testing showed an acceptable level of shedding.
- · Cutting Performance Test- A cutting performance test was conducted to demonstrate the durability and robustness of the device to show the ability to function for 20 minutes of activation. Results from the testing demonstrated that the subject devices met the acceptance criteria.
- · Resection Rate Test- A resection rate test was conducted to evaluate the performance of the device in terms of an acceptable resection rate. Results from the testing verified that the subject device met the acceptance criteria showing an acceptable resection rate.
- · Biocompatibility Testing- Cytotoxicity, sensitization, and irritation testing were conducted along with a chemical analysis of the extract. Results from the testing show the device to be acceptable for biological use.
- · Shelf Life Testing- Accelerated age testing along with post-aging performance testing was conducted and justifies the shelf life of the device.

The reference predicate, TRUCLEAR INCISOR, addresses any differences in materials and performance testing between the subject and primary predicate devices. In addition, a risk analysis was conducted to show that any residual risks are outweighed by the medical benefit.

## Conclusion

Based on the performance data provided, the subject device is substantially equivalent to the proposed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIH/K161763](https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIH/K161763)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
