K112134 · Smith & Nephew, Inc. · HIH · Nov 22, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K112134
Device Name
TRUCLEAR OPERATIVE HYSTEROSCOPE
Applicant
Smith & Nephew, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Nov 22, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Smith & Nephew TRUCLEAR Operative Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
Device Story
Rigid hysteroscope for office, hospital, and ambulatory surgical center use; operated by physicians. Device utilizes optical fiber bundle design to reduce needle diameter while maintaining working channel space. Internal design features 'D' shaped working channel serving as inflow channel. Imaging bundle transmits visual data from objective lens to eyepiece. Enables visualization of cervical canal and uterine cavity; facilitates diagnostic and surgical procedures. Benefits include reduced diameter for improved access.
Clinical Evidence
Bench testing only. Optical performance testing (brightness, vignetting, resolution, focus, image distortion) and mechanical testing (sheath force, instrument compatibility) demonstrated performance comparable to predicate. Biocompatibility verified per ISO 10993-1. Electrical safety verified per IEC 60601 and IEC 60601-2-18. Cleaning and sterilization validated per AAMI TIR 30 and AAMI TIR 12.
Technological Characteristics
Rigid hysteroscope; optical fiber bundle imaging; 'D' shaped working/inflow channel. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601 and IEC 60601-2-18. Sterilization validated per AAMI TIR 30 and AAMI TIR 12.
Indications for Use
Indicated for patients requiring diagnostic or surgical visualization of the cervical canal and uterine cavity.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
K112134
We are smith&nephew
Endoscopy Smith & Nenhew Inc 150 Minuteman Road Andover, MA 01810 USA
T 978 749 1000 F 978 740 1500 www.smith-nephew.com
# NOV 2 2 2011
## SECTION IV
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based. Smith & Nephew TRUCLEAR Operative Hysteroscope Date Prepared: July 15, 2011
## A. Submitter's Name:
Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road. Andover MA. 01810
### B. Company Contact
Janice Haselton Sr. Regulatory Affairs Specialist T 978-749-1494 F 978-749-1443
### C. Device Name
| Trade Name: | Smith & Nephew TRUCLEAR Operative Hysteroscope |
|----------------------|------------------------------------------------|
| Common Name: | Hysteroscope |
| Classification Name: | Hysteroscope and Accessories per CFR §884.1690 |
#### D. Predicate Device
The predicate device is the Smith & Nephew Operative Hysteroscope cleared in K013870.
#### E. Description of Device
Smith & Nephew's TRUCLEAR Operative Hyteroscope is a rigid hysteroscope that is intended for use in office based, hospital and ambulatory surgical centers. The operative hysteroscope incorporates an optical fiber bundle design in order to reduce the overall diameter of the needle portion of the hysteroscope and still provide adequate space in the working channel for instrumentation. The Smith & Nephew TRUCLEAR Operative Hysteroscope internal design consists of a "D" shaped working channel into the needle portion of the hysteroscope. The working channel also acts as an inflow-channel.
#### F. Intended Use
The Smith & Nephew TRUCLEAR Operative Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
{1}------------------------------------------------
# >< We are smith&nephew
#### Comparison of Technological Characteristics ে
The Smith & Nephew TRUCLEAR Operative Hysteroscope has the same fundamental technological characteristics as the unmodified predicate device and is substantially equivalent in design, materials and intended use as the unmodified predicate device.
The proposed Smith & Nephew TRUCLEAR Operative Hysteroscope has the following similarities as the predicate device cleared in K013870:
- the same indications for use .
- utilizes the same operating principle .
- . incorporates the same basic mechanical design.
- . manufactured under the same Quality System.
The Smith & Nephew TRUCLEAR Operative Hysteroscope differs from the predicate Smith & Nephew Operative Hysteroscope cleared in K013870 in that:
- . The overall diameter of the operative hysteroscope with sheath has been reduced from 9 mm to a minimum of 5.5 mm.
- The rod/lens design has been replaced by an imaging bundle. The imaging bundle . transmit the image similar to the rod/lens design as both designs relay an image through a fiber bundle or rod/lenses from the objective lens to the eyepiece.
- . The working length of the operative hysteroscope and sheath has been increased.
There are no significant differences between the proposed and predicate devices that raise new questions of safety or efficacy.
#### H. Summary Performance Data
The Smith & Nephew TRUCLEAR Operative Hysteroscope optical performance testing has demonstrated that the operative hysteroscope is substantially equivalent to the predicate Smith & Nephew Operative Hysteroscope (K013870). Optical performance tests included brightness, vignetting, resolution, near and far focus, image run out, depth of field and image distortion testing.
Performance tests included sheath force testing and working channel instrument compatibility testing.
Optical and performance testing demonstrated that the Smith & Nephew TRUCLEAR Operative performed comparably under similar conditions.
Additionally, the Smith & Nephew TRUCLEAR Operative Hysteroscope complies with applicable portions of IEC 60601-2-18 and IEC 60601 electrical standards and has successfully passed cleaning and sterilization validations per AAMI TIR 30 and AAMI TIR 12.
All materials have been demonstrated to meet biocompatibility requirements according to FDA General Program Memorandum #G95-1, Use of International Standards ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing".
In conclusion, the optical and performance testing demonstrates that the proposed device is as safe, effective and performs as well as or better than the predicate.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the top half of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Janice Haselton Sr. Regulatory Affairs Specialist Smith & Nephew. Inc. 150 Minuteman Road ANDOVER MA 01810
NOV 2 2 2011
Re: K112134
Trade/Device Name: Smith & Nephew TRUCLEAR Operative Hysteroscope Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: October 25, 2011 Received: October 26, 2011
Dear Ms. Haselton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{3}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Hubert Lewin, MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Smith & Nephew TRUCLEAR Operative Hysteroscope Device Name:
Indications For Use:
The Smith & Nephew TRUCLEAR Operative Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hebert Kern
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K112134
Page 1 of 1