The Stryker Urology and Gynecology Hardware System is intended to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples of use of the product include the visualization and manipulation of anatomy, ablation, biopsy, incision, and resection of tissue, and/or as the surgeon deems appropriate. The system is intended for use in general urological and gynecological surgery through the minimally invasive approach, by utilizing natural orifices to access the surgical site. The system's use is intended for, but not limited to the following types of procedures: Dilation of the urethra, and cold-slitting of urethral strictures Trans-urethral incision and resection of the prostate Trans-urethral removal of bladder tumors Trans-cervical resection and ablation of the endometrium Trans-cervical resection of fibroids
Device Story
System comprises endoscopes, sheaths, electrodes, and surgical accessories for minimally invasive urological and gynecological procedures. Device provides visualization and manipulation of anatomy; enables ablation, biopsy, incision, and resection of tissue. Operated by surgeons in clinical settings via natural orifices. System facilitates procedures including urethral dilation, prostate resection, bladder tumor removal, and endometrial/fibroid resection. Output is direct visualization and mechanical/electrosurgical tissue interaction. Benefits include minimally invasive access to surgical sites.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing and adherence to voluntary safety and performance standards including IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18, ANSI/AAMI HF-18, and ISO 10993.
Technological Characteristics
System includes endoscopes, sheaths, electrodes, and mechanical accessories. Materials evaluated per ISO 10993. Electrosurgical components comply with ANSI/AAMI HF-18. Electrical safety per IEC 60601-1, 60601-2-2, and 60601-2-18. Connectivity: None (mechanical/optical/electrosurgical).
Indications for Use
Indicated for patients undergoing endoscopic diagnostic and therapeutic urological and gynecological procedures via natural orifices. Contraindicated in patients with acute pelvic inflammatory disorder (PID). Hysteroscopy may be contraindicated by inability to distend the uterus, cervical stenosis, cervical/vaginal infection, uterine bleeding/menses, known pregnancy, invasive carcinoma of the cervix, or recent uterine perforation.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Submission Summary (Full Text)
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Document Control Center - WO66-G609
Silver Spring, MD 20993-0002
Stryker Endoscopy
Mr. Christopher L. Cook
Quality Engineer
5900 Optical Court
San Jose, CA 95138
JUL 27 2015
Re: K040390
Trade/Device Name: Stryker Urology and Gynecology Hardware System
Regulation Number: 21 CFR 884.1690
Regulation Name: Hysteroscope and accessories
Regulatory Class: II
Product Code: HIH, FAJ, FAS and KQT
Dated (Date on orig SE ltr): February 10, 2004
Received (Date on orig SE ltr): February 17, 2004
Dear Mr. Cook,
This letter corrects our substantially equivalent letter of May 17, 2004.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the *Federal Register*.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D.
Director
Division of Reproductive, Gastro-Renal, and Urological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
Device Name: Stryker Urology and Gynecology Hardware System
Indications for Use:
The Stryker Urology and Gynecology Hardware System is intended to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples of use of the product include the visualization and manipulation of anatomy, ablation, biopsy, incision, and resection of tissue, and/or as the surgeon deems appropriate. The system is intended for use in general urological and gynecological surgery through the minimally invasive approach, by utilizing natural orifices to access the surgical site. The system's use is intended for, but not limited to the following types of procedures:
- Dilation of the urethra, and cold-slitting of urethral strictures
- Trans-urethral incision and resection of the prostate
- Trans-urethral removal of bladder tumors
- Trans-cervical resection and ablation of the endometrium
- Trans-cervical resection of fibroids
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
Over-The-Counter-Use ☐ (Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number K040390
Stryker Urology and Gynecology Hardware System 510(k) submission
3 of 15
{3}
MAY 17 2004
K040390
Page 1 of 2
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## Device Name
Proprietary Name: Stryker Urology and Gynecology Hardware System
Common and Usual Name: Scope, Obturator, Working Element, Sheath, Bridge, Electrode, Albarran Deflector, Timberlake Obturator, Cutting Loop, Roller Ball, Cold Knives, Dilator/Sound/Bougie, Bladder Syringe, Ellik Evacuator, Forceps
Classification Name: Endoscope and accessories, Cystoscopes, Hysteroscope, Resectoscope, Sheaths, Electrode, Urethrotome, Dilators, Evacuator, Non-Electric Biopsy Forceps, G-U, Surgical Instruments
The Stryker Urology and Gynecology Hardware System is substantially equivalent in terms of safety and effectiveness to many currently marketed devices and surgery systems currently marketed by Henke Sass Wolf, Omnitech Systems, GIMMI, and ACMI.
The Stryker Urology and Gynecology Hardware System is an extension of the Hystero-Resectoscope and Accessories as currently marketed by Stryker and OEM suppliers Henke Sass Wolf and Omnitech Systems. The Stryker Urology and Gynecology Hardware System is composed of endoscopes, sheaths, accessories, and applied parts which provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures.
The Stryker Urology and Gynecology Hardware System is intended to provide the user with the means for endoscopic diagnostic and therapeutic surgical procedures. Examples of use of the product include the visualization and manipulation of anatomy, ablation, biopsy, incision, and resection of tissue, and/or as the surgeon deems appropriate. The system is intended for use in general urological and gynecological surgery through the minimally invasive approach, by utilizing natural orifices to access the surgical site. The system's use is intended for, but not limited to the following types of procedures:
- Dilation of the urethra, and cold-slitting of urethral strictures
- Trans-urethral incision and resection of the prostate
- Trans-urethral removal of bladder tumors
- Trans-cervical resection and ablation of the endometrium
- Trans-cervical resection of fibroids
## Contraindications:
- Acute Pelvic Inflammatory Disorder (PID)
Hysteroscopy may be contraindicated by the following conditions, depending on their severity or extent:
- Inability to distend the uterus
- Cervical stenosis
- Cervical/vaginal infection
- Uterine bleeding or menses
- Known pregnancy
- Invasive carcinoma of the cervix
- Recent uterine perforation
The Stryker Urology and Gynecology Hardware System conforms to the following voluntary safety and performance standards: IEC 60601-1
Medical Electrical Equipment – General Requirements for Safety, IEC 60601-2-2 Medical Electrical Equipment – General Requirements for Safety, Collateral Standard, Electromagnetic Compatibility, IEC 60601-2-18 Particular Requirements for the Safety of Endoscopic Equipment, ANSI/AAMI HF-18 Electrosurgical Devices, ISO 10993 Biological Evaluation of Medical Devices.
{4}
K'04039c
Page 242
There are no significant technological or performance differences between the Stryker Urology and Gynecology Hardware System and the identified predicate devices and surgery systems, nor are there any new questions raised regarding safety or effectiveness, therefore, the Stryker Urology and Gynecology Hardware System is substantially equivalent to the identified predicate devices and surgery systems.
Contact:
Christopher L. Cook
Date:
Stryker Endoscopy
5900 Optical Court
San Jose, CA 95138
Phone: 408-754-2288
Fax: 408-754-2521
Email: chris.cook@stryker.com
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.