K231407 · Premium Medical Technology, LLC · HIF · Aug 29, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K231407
Device Name
StarFin
Applicant
Premium Medical Technology, LLC
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Aug 29, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The StarFin device is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperioneum for laparoscopic procedures prior to the placement of trocars. The StarFin device is also used for suture capture, passing, externalization and internalization for wound closure after laparoscopic surgery.
Device Story
StarFin is a single-use, sterile, dual-function laparoscopic needle; combines Veress needle functionality with suture passer capabilities. Device features stainless steel cannula with beveled cutting point and spring-loaded inner stylet with blunt tip for puncture protection. During insufflation, hollow inner stylet delivers CO2 to create pneumoperitoneum. For wound closure, inner stylet tip captures, passes, externalizes, and internalizes suture threads. Used in professional healthcare settings (OR) by surgeons. Device simplifies workflow by integrating insufflation and closure tools into one instrument; reduces number of devices required for laparoscopic access and closure; benefits patient by potentially minimizing trocar site trauma and procedure time.
Clinical Evidence
Bench testing only. Performance evaluated via appearance, gas flow/leak rates, puncture force, suture retention, and stylet pull strength. Accelerated aging (ASTM F1980) confirmed 1-year shelf-life. Simulated use testing performed on artificial skin/fascial samples by laparoscopic surgeons.
Technological Characteristics
Materials: 304 stainless steel cannula/stylet; polyphenylsulfone handle. Principle: Spring-loaded inner stylet with blunt tip for puncture protection; valve-controlled gas flow. Sterilization: Ethylene Oxide (ISO 11135:2014). Biocompatibility: ISO 10993-5, -10, -11. Shelf-life: 1 year (ASTM F1980).
Indications for Use
Indicated for patients undergoing laparoscopic procedures requiring pneumoperitoneum establishment via CO2 insufflation and subsequent trocar site wound closure.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Endo-Close™ Auto-Suture™ Trocar Site Closure Device (K954853)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 29, 2023
Premium Medical Technology LLC Kuowei Chang Managing Director 1377 Main Street 2nd Floor Waltham, Massachusetts 02451
Re: K231407
Trade/Device Name: StarFin Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF, GCJ Dated: July 25, 2023 Received: August 1, 2023
Dear Kuowei Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231407
Device Name StarFin
Indications for Use (Describe)
The StarFin device is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperioneum for laparoscopic procedures prior to the placement of trocars. The StarFin device is also used for suture capture, externalization and internalization for wound closure after laparoscopic surgery.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary
#### 1. Submitter Information
#### . 510(K) Submitter:
Premium Medical Technology LLC 1377 Main Street, 2nd Floor Waltham, MA 02451-1624, USA
- Contact: . Kuowei Chang, Ph.D. Chief Scientific Officer Phone: 1-781-891-4201 Email: kchang@PremiumMedTech.com
- Date Prepared: . August 28, 2023
- 2. Device Information
Name: StarFin Model: StarFin-120 Common Name: Veress needle and Endoscopic Suture Passer Regulation Number: 21 CFR 884 1730 Regulation Name: Laparoscopic Insufflator Product Code: HIF and GCJ Regulatory Class: Class II
- 3. Predicate Devices
510(K): K193339 Device Name: GTK Veress Needle
510(K): K954853 Device Name: Endo-Close™ Auto-Suture™ Trocar Site Closure Device
The predicates have not been subject to a design-related recall
- 4. Device Description
{4}------------------------------------------------
The subject device is a single-use, sterile, laparoscopic needle that is used for percutaneous introduction of a hollow stylet into the abdomen for gas insufflation and establishment of pneumoperitoneum prior to abdominal endoscopy and for closure of trocar site wounds with suture threads.
The subject device is equipped with a stainless steel cannular consisting of a beveled needle point for cutting through tissues. Inside the cannula, there is a spring-loaded inner stylet with a rounded blunt tip used for puncture protection. The hollow stainless steel inner stylet provides passing carbon dioxide gas to create pneumoperitoneum. The inner stylet tip is designed for suture capture, passing, externalization, and internalization for wound closure after laparoscopic surgery.
The environment of use is professional healthcare.
#### 5. Indication for Use
The StarFin device is intended for percutaneous insertion into the peritoneal cavity for the purpose of insufflation with carbon dioxide to establish pneumoperitoneum for laparoscopic procedures prior to the placement of trocars. The StarFin device is also used for suture capture, passing, externalization and internalization for wound closure after laparoscopic surgery.
## 6. Comparison of the Technological Characteristics
## Table below provides comparison of technological characteristics between Subject Device and Predicate Devices:
| Items | Subject Device<br>Dual-Function | Predicate K193339<br>Single-Function | Predicate K954853<br>Single-Function | Comments |
|------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------|--------------------------------------|----------|
| Product<br>Description<br>&<br>Intended<br>Use | Combination Veress<br>Needle<br>&<br>Trocar Site Wound<br>Closure | Veress Needle | Trocar Site Wound<br>Closure | Same |
| Model<br>Number | StarFin-120 | VN-120 | 173022 | / |
| Manufacturer | Premium Medical<br>Technology LLC | Guangzhou T.K<br>Medical Instrument<br>Co. Ltd. | Covidien LLC | / |
| Classification | Class II | Class II | Class II | Same |
| Product Code | HIF & GCJ | HIF & FHO | GCJ | / |
| Cutting Point | C=Bevel | C=Bevel | C=Bevel | Same |
{5}------------------------------------------------
| Spring loaded<br>(Inner Stylet) | Yes | Yes | Yes | Same |
|---------------------------------|-------------------------------------|---------------------|-------------------------------------|-----------|
| Valve<br>Controlled | Yes | Yes | No | Same |
| Needle<br>Material | 304 Stainless Steel | 304 Stainless Steel | 304 Stainless Steel | Same |
| Handle<br>Material | polyphenylsulfone | polycarbonate | ABS plastic and<br>polyethylene | Different |
| Shape of the<br>suture passer | A needle with a dull,<br>blunt hook | N/A | A needle with a dull,<br>blunt hook | Same |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same |
The subject and predicate devices have the same intended use. The subject device's intended use is the combination of the intended use of the predicate devices. The subject and predicate device have different technological features, including the needle diameter, length, and handle material. These technological differences do not raise different questions of safety or effectiveness.
#### 7. Non-Clinical Performance Data
The following data were provided in support of the substantial equivalence determination.
#### a. Biocompatibility Test
The stainless-steel outer cannula and inner stylet of the subject device are the components in contact with human tissue for less than 24-hours. The following tests were completed and met the requirements, in accordance with:
ISO 10993-5, 2009, Cytotoxicity Test
ISO 10993-10, 2021, Maximization/Sensitization Test
ISO 10993-10, 2021, Irritation/Intracutaneous Reactivity Test
ISO 10993-11, 2017, Acute Systemic Toxicity Test
ISO 10993-11, 2017/USP<151>, Material Mediated Pyrogenicity Test
#### b. Sterilization Validation
The sterilization validation was performed and verified according to ISO 11135:2014. The EO and ECH residual of the subject device were evaluated and verified in accordance with ISO10993-7:2008.
#### c. Bench Performance Tests
The subject device was tested to demonstrate it meets stated performance expectations and compared with the predicate devices. The tests included appearance, size, gas flow rate, gas leak rate, puncture force, suture retention force, stylet pull strength. The tests were also conducted with aged sample (accelerated
{6}------------------------------------------------
aging equivalent to 1 year per ASTM F1980) and compared with non-aged samples to demonstrate 1 year shelf-life.
The simulated use of subject device was evaluated and compared with the predicate devices on an artificial skin/fascial sample by qualified and trained laparoscopic surgeons.
#### 8. Conclusions
The non-clinical performance data demonstrate that the subject device is as safe and effective as the predicate devices and support the subject device is substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.