K161554 · KARL STORZ Endoscopy-America, Inc. · HIF · Mar 2, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K161554
Device Name
ENDOFLATOR 40, ENDOFLATOR 50
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Mar 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
CO2 insufflators and their accessories are used to create and/or to distend a cavity during diagnostic or therapeutic endoscopic interventions.
Device Story
ENDOFLATOR 40 and 50 are microprocessor-controlled CO2 insufflators; used to create/distend cavities for diagnostic and therapeutic endoscopic procedures. Devices feature redundant pressure/flow monitoring, automatic pressure relief, and graphical user interface. ENDOFLATOR 50 includes integrated gas heater and user-saveable procedure settings. Operated by clinicians in clinical settings. Devices provide real-time display of pressure, flow rates, volume, and operating mode. Output informs clinician of cavity status; enables safe insufflation during procedures. Benefits include controlled, automated cavity distension for improved visualization and access during surgery.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including software verification/validation, electrical safety (IEC 60601-1), and performance testing (pressure/flow accuracy, overpressure safety, gas supply monitoring).
Technological Characteristics
Microprocessor-controlled CO2 insufflator. Features redundant pressure/flow sensors, automatic pressure relief valves, and graphical user interface. ENDOFLATOR 50 includes integrated gas heater. Connectivity: standalone. Software: major level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-18.
Indications for Use
Indicated for creating and/or distending a cavity in adult and pediatric patients during laparoscopy, endoscopy of the lower gastrointestinal tract (e.g., TEO, colonoscopy), and endoscopic vessel harvesting.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
Predicate Devices
World of Medicine's Insufflator 50L FM 134 (K153513)
Olympus Medical Systems Corp's High Flow Insufflation Unit UHI-4 (K122180)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2017
Karl Storz Endoscopy-America, Inc. Nozomi Yagi, MAS, RAC Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245
Re: K161554
Trade/Device Name: ENDOFLATOR 40 & 50 Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF, OSV, FCX Dated: January 26, 2017 Received: January 27, 2017
Dear Nozomi Yagi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161554
Device Name ENDOFLATOR 40&50
Indications for Use (Describe) The ENDOFLATOR 40 & 50 are used to create and/or to distend a cavity in the following diagnostic and therapeutic interventions:
- · Laparoscopy
- · Pediatric Laparoscopy
- · Endoscopy of the Lower Gastrointestinal Tract (e.g., TEO, colonoscopy)
- · Endoscopic Vessel Harvesting
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED:
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Submitter: | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>EI Segundo, CA 90245 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Nozomi Yagi<br>Regulatory Affairs Specialist<br>Phone: (424) 218-8351<br>Fax: (424) 218-8519 |
| Date of<br>Preparation: | February 28, 2017 |
| Type of 510(k)<br>Submission: | Traditional |
| Device<br>Identification: | Trade Name: ENDOFLATOR 40 and ENDOFLATOR 50<br>Classification Name: Insufflator, Laparoscopic (21 CFR Part<br>884.1730) |
| Regulatory<br>Class: | II |
| Product Code: | HIF (Insufflator, Laparoscopic),<br>OSV(Insufflator, Endoscopic Vessel Harvesting),<br>FCX(Insufflator, Automatic Carbon-Dioxide For Endoscope) |
| Guidance<br>Document: | Hysteroscopic and Laparoscopic Insufflators: Submission Guidance<br>for a 510(k) issued on August 1, 1995 for Product Code HIF |
| Predicate<br>Device(s): | Primary Predicate Device: World of Medicine's Insufflator 50L FM<br>134 (K153513)<br>Second Predicate Device: Olympus Medical Systems Corp's High<br>Flow Insufflation Unit UHI-4 (K122180) |
| Device<br>Description: | This submission includes two microprocessor controlled CO2<br>insufflators that have been developed in parallel: ENDOFLATOR 40<br>(Model Number: UI400) and ENDOFLATOR 50 (Model Number:<br>U1500), both of which come with HIGH FLOW and PEDIATRICS<br>modes. The HIGH FLOW mode has the pressure range of 1-30 mmHg<br>and the flow rate range of 1-40 LPM (ENDOFLATOR 40) and 1-50<br>LPM (ENDOFLATOR 50). The PEDIATRICS mode has the pressure |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
| | range of 1-15 mmHg and the flow range of 0.1-15 LPM. | | | | |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--|
| | The ENDOFLATOR 40 &50 have redundant pressure and flow<br>monitoring, automatic pressure relief, user control via graphical user<br>interface and display of pressures/flow rates/ volume/operating mode. | | | | |
| | The ENDOFLATOR 50 has three extra features not included in the<br>the integrated gas heater, it can only be used with non-heated<br>insufflation tubing set (Model Number: 031200-01). On the other<br>hand, the ENDOFLATOR 50 (Model Number: UI500) can be used<br>with either the non-heated insufflation tubing set (Model Number:<br>031200-01) or the heatable insufflation tubing set (Model Number:<br>031210-01). | | ENDOFLATOR 40: (1) the higher maximum flow rate (50LPM), (2)<br>the integrated gas heater, and (3) the ability for the user to save up to<br>30 procedure settings. Because the ENDOFLATOR 40 does not have | | |
| Intended Use: | CO2 insufflators and their accessories are used to create and/or to<br>distend a cavity during diagnostic or therapeutic endoscopic<br>interventions. | | | | |
| Indications For<br>Use, | Subject Device | Primary predicate<br>device, K153513 | Secondary predicate<br>device, K122180 | | |
| Comparison | ENDOFLATOR<br>40&50 | W.O.M. Insufflator<br>50L FM134 | Olympus High Flow<br>Insufflation Unit UHI-<br>4 | | |
| | Indication for Use | | | | |
| | The ENDOFLATOR<br>40&50 are used to create<br>and/or to distend a cavity<br>in the following diagnostic<br>and therapeutic<br>interventions: | The Standard, Pediatric<br>and Bariatric operating<br>modes of the device are<br>indicated to fill and<br>distend a peritoneal cavity<br>with gas during a<br>laparoscopic procedure. | This instrument is<br>intended to insufflate the<br>abdominal cavity and<br>provides automatic suction<br>and smoke evacuation to<br>facilitate laparoscopic<br>observation, diagnosis and | | |
| | Technological<br>Characteristics: | W.O.M. Insufflator 50L FM134 was chosen for having the Indication<br>for Use and Technological Characteristics most similar to the subject<br>device. The subject device differs from the primary predicate in that<br>an operation mode is not specific to indicated intervention; the subject<br>device has HIGH FLOW mode to cover laparoscopy, endoscopy of<br>lower GI tract, and endoscopic vessel harvesting indications and<br>PEDIATRICS mode to cover pediatric laparoscopy indication.<br>Olympus' High Flow Insufflation Unit UHI-4 was chosen as a<br>secondary predicate device to support the substantial equivalence of<br>the additional anatomical location of use (i.e., colon) that the primary<br>predicate does not include.<br>The subject device has the same Intended Use as both primary and<br>secondary predicate devices.<br>The subject and both primary and secondary predicate devices are CO2<br>insufflators that are used to create and maintain the cavity. There are<br>similarities and differences in the technological characteristics<br>including but not limited to the following: | | | |
| | | Subject Device | Primary predicate<br>device, K153513 | Secondary<br>predicate device,<br>K122180 | |
| | Distension<br>Medium | CO2 | Same as the subject<br>device | Same as the subject<br>device | |
| | Indicated<br>Population | Adult and Pediatric | Same as the subject<br>device | Adult | |
| | Modes | High Flow,<br>Pediatrics | Bariatric,<br>Standard,<br>Vessel Harvest,<br>Pediatric | Normal Cavity,<br>Small Cavity | |
| | Maximum<br>Flow Rate | 50 LPM | Same as the subject<br>device | 45 LPM | |
| | Maximum<br>Pressure | 30 mmHg | Same as the subject<br>device | 25 mmHg | |
| | Overpressure<br>Action | Visual and audible<br>alarms followed by<br>pressure relief after 5<br>seconds | Visual and audible<br>alarms followed by<br>pressure relief after 2-<br>5 seconds | Visual and audible<br>alarms followed by<br>pressure relief after 0<br>or 4.5 seconds | |
| | Gas Supply<br>Shortage<br>Action | When cylinder<br>pressure is below 1<br>bar, visual and<br>audible alarms will be<br>activated and<br>insufflation stops | At <15 bar, visual and<br>audible alarms will be<br>activated and<br>insufflation stops | When cylinder<br>pressure is below 0.1<br>MPa (1 bar), visual<br>and audible alarms<br>will be activated and<br>insufflation stops | |
| | Over<br>Temperature<br>Action | At >41°C,<br>Information signal<br>indicates that heating<br>is switched off | At >42°C, visual and<br>audible alarm | N/A, gas is not heated | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.
| | The differences do not raise different questions of safety and<br>effectiveness. | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Non-Clinical<br>Performance<br>Data: | The ENDOFLATOR 40 & 50 is tested according to the following<br>standards:<br>• IEC 60601-1:2005 + CORR.1:2006 + CORR.2:2007 + AM1<br>(2012)<br>• IEC 60601-1-2:2014<br>• IEC 60601-1-18:2006/AMD1:2012<br>Software verification and validation testing were conducted and<br>documentation was provided as recommended by FDA's Guidance for<br>Industry and FDA Staff titled Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical Devices. The software<br>for this device was considered as a "major" level of concern, since a<br>failure or latent flaw in the software could directly result in serious<br>injury or death to the patient or operator.<br>In addition, the following bench testing has been performed for<br>following safety and performance requirements as well as for<br>comparative purposes: | | |
| | Bench Testing Summary<br>Verification Test<br>(Test#) | Acceptance Criteria | Conclusion |
| | Safety - Overpressure | The device informs user, if an unintended<br>high cavity pressure is measured and<br>activates appropriate valves if the<br>activation threshold is exceeded. | Pass |
| | Performance -- Pressure | The difference between the measured and<br>the set point pressures is within the<br>specified limits. | Pass |
| | Safety -Maximum pressure | The measured pressure shall not exceed the<br>mode related specified limits. | Pass |
| | Safety - Overpressure | The gas flow is stopped if a cavity pressure<br>is measured is higher than mode related<br>limits. | Pass |
| | Performance - Gas supply | The device informs the user if the pressure<br>gas supply falls under specified limits. | Pass |
| | Performance -- Flow | The difference between the measured and<br>the set point flow is within the specified<br>limits. | Pass |
| | Safety - Pressure Controller | The pressure controller temperature shall<br>stain within the specified limits. | Pass |
| | Safety - Overtemperature | (ENDOFLATOR 50 only)<br>The heater is switched off if the fault<br>arises. | Pass |
| | | | |
| | Performance - Heater<br>connection symbol | specified limits.<br>(ENDOFLATOR 50 only)<br>The heater connection symbol is correctly<br>displayed. | Pass |
| | Comparative -- Adult and<br>Pediatric Laparoscopy | To support laparoscopy indication of the<br>subject device, the pressure control of High<br>Flow mode must be substantially<br>equivalent to the primary predicate<br>device's Standard/Bariatric modes and the<br>secondary predicate device's Normal<br>Cavity mode.<br>Similarly, to support Pediatric Laparoscopy<br>indication of the subject device, the<br>pressure control of Pediatrics mode must be<br>substantially equivalent to the primary<br>predicate device's Pediatric mode. | SE to<br>predicate<br>devices |
| | Comparative - Lower<br>Gastrointestinal Tract | In order to support lower gastrointestinal<br>indication of the subject device, the<br>pressure control of High Flow mode must<br>be substantially equivalent to the secondary<br>predicate device's Small Cavity mode. | SE to<br>predicate<br>devices |
| | Comparative - Endoscopic<br>Vessel Harvesting | In order to support Endoscopic Vessel<br>Harvesting indication of the subject device,<br>the pressure control of High Flow mode<br>must be substantially equivalent to the<br>primary predicate device's Vessel<br>Harvesting mode and the secondary<br>predicate device's Small Cavity mode. | SE to<br>predicate<br>devices |
| Clinical<br>Performance<br>Data: | Clinical testing was not required to demonstrate the substantial<br>equivalence to the predicate devices. Non-clinical bench testing was<br>sufficient to assess safety and effectiveness and to establish the<br>substantial equivalence of the modifications. | | |
| Conclusion: | The conclusions drawn from the nonclinical tests such as the software<br>data, the electrical safety and performance data, as well as the bench<br>top performance data, demonstrated the subject device is as safe as<br>and as effective as the primary and secondary predicate devices. As<br>such, we concluded that the substantial equivalence of the subject and<br>both predicate devices has been met. | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in smaller letters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.