K110294 · Olympus Medical Systems Corporation · HIF · Nov 18, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K110294
Device Name
HIGH FLOW INSUFFLATION UNIT UHI-4
Applicant
Olympus Medical Systems Corporation
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Nov 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This instrument has been designed for insufflation of abdominal cavity, and provides automatic suction and smoke evacuation to facilitate laparoscopic observation, diagnosis and treatment. This instrument is also designed for controlled CO2 insufflation to create a cavity along the saphenous vein and/or radial artery to facilitate observation during an endoscopic vessel harvesting procedure.
Device Story
UHI-4 is a high-flow CO2 insufflator for laparoscopic surgery and endoscopic vessel harvesting. Device inputs include gas pressure sensors and user-defined settings for flow rate and pressure. It transforms inputs via valve switching and sensor monitoring to maintain cavity pressure; it also provides automated suction and smoke evacuation. Used in OR settings by surgeons/staff. Output includes real-time display of gas flow, cavity pressure, and CO2 volume. A FlexRay interface enables communication with compatible electrosurgical or ultrasonic generators. Safety features include warning lamps, alarms for over-pressure/obstruction, and automated suction activation. Benefits include maintained visualization and controlled cavity creation for minimally invasive procedures.
Clinical Evidence
Bench testing only. Testing verified pressure control in saphenous vein and radial artery models and confirmed accurate CO2 delivery and over-pressurization protection. Risk analysis performed per ISO 14971:2007.
Technological Characteristics
Laparoscopic insufflator; CO2 gas source; sensor-based pressure control with valve switching; FlexRay interface for generator communication; front panel display for flow/pressure/volume; includes insufflation/suction tubes. Applied standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, ISO 14971, IEC 60601-1-8.
Indications for Use
Indicated for patients undergoing laparoscopic observation, diagnosis, and treatment of the abdominal cavity, or endoscopic vessel harvesting procedures requiring cavity creation along the saphenous vein or radial artery. Not intended for pediatric use.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows handwritten text that appears to be a document identifier or page number. The text includes the string 'K110294' on the first line, which likely serves as a unique identifier for the document. Below this, the text 'pg 1 of 4' indicates that this is page 1 of a 4-page document. The handwriting is somewhat stylized, with the numbers and letters clearly distinguishable.
# 510(k) SUMMARY
# HIGH FLOW INSUFFLATION UNIT UHI-4
### October 3, 2011
#### General Information 1
Applicant:
트
트
OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047
- Official Correspondent: Stacy Abbatiello Kluesner, M.S., RAC Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-5405 FAX: 484-896-7128 Email: stacy.kluesner@olympus.com
- Manufacturer: SHIRAKAWA OLYMPUS CO., LTD. 3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura, Nishishirakawa-gun Fukushima, JAPAN 961-8061 Establishment Registration No: 3002808148
#### Device Identification 2
- 비 Device Trade Name: HIGH FLOW INSUFFLATION UNIT UHI-4 ■ Common Name: INSUFFLATOR 트 21 CFR 884.1730 Regulation Number:
ll
- 트 Laparoscopic insufflator Regulation Name:
- Regulatory Class:
- Obstetrics/Gynecology, Cardiovascular ■ Classification Panel:
- 미 Product Code: HIF, OSV
{1}------------------------------------------------
K110294
pg. 2 of 4
#### Predicate Device Information 3
| 1 | ■ Device Name: | 45L CORE Insufflator F114 |
|---|-----------------|--------------------------------------------------------------------------------|
| | ■ Common Name: | Carbon Dioxide Insufflator for Laparoscopy and<br>Endoscopic Vessel Harvesting |
| | ■ Manufacturer: | W. O. M. WORLD OF MEDICINE AG |
| | ■ 510(k) No. | K063367 |
| 2 | ■ Device Name: | HIGH FLOW INSUFFLATION UNIT UHI-3 |
| | ■ Common Name: | INSUFFLATOR |
| | ■ Manufacturer: | OLYMPUS MEDICAL SYSTEM CORP. |
| | ■ 510(k) No. | K014166 |
#### Device Description 4
The subject device , UHI-4, has new insufflation function, small mode, for insufflation of cavities along the saphenous vein and radial artery. Therefore, the insufflation pressure of small mode is designed to be less than the normal mode which is implemented in UHI-3.
- . The small mode is not intended for pediatric use.
- . This subject device has user-friendly displays for indicating gas flow rate, cavity pressure, and insufflated CO2 gas volume on the front panel of UHI-4.
- lt has several safety functions, such as follows: .
- warning lamps
- · alarms for excessive pressure and tube obstruction
- suction activated for excessive pressure.
- Gas pressure is controlled by reading sensors and switching valves. .
- lt has a function for evacuating smoke (clearing the field of vision) when either an Electro . Surgical Instrument or Ultrasonic Surgical Instrument is used during the operation. This
{2}------------------------------------------------
K110294
Pg. 3 of 4
function may be controlled with an optional foot switch
- A FlexRay interface enables the UHI-4 to communicate its setting data, measured values, . and other parameters between UHI-4 and the SonoSurg Generator SonoSurg-G2 or the ELECTROSURGICAL UNIT UES-40.
The standard set of HIGH FLOW INSUFFLATION UNIT UHI-4 consists of an INSUFFLATION TUBE MAJ-590, a SUCTION TUBE MAJ-591 and a CYLINDER HOSE (PIN) FOR UHI-3 MAJ-1080.
#### ട Indications for Use
This instrument has been designed for insufflation of abdominal cavity, and provides automatic suction and smoke evacuation to facilitate laparoscopic observation, diagnosis and treatment. This instrument is also designed for controlled CO2 insufflation to create a cavity along the saphenous vein and/or radial artery to facilitate observation during an endoscopic vessel harvesting procedure.
#### Comparison of Technological Characteristics 6
The HIGH FLOW INSUFFLATION UNIT UHI-4 is basically identical to the predicate device UHI-3 (K014166) in intended use except for the ability to create small cavities along the saphenous vein/or and radial artery. It is also similar in specifications except for the addition of small mode and expanding the upper flow rate limit in normal mode.
The HIGH FLOW INSUFFLATION UNIT UHI-4 is basically identical to the predicate device 45L CORE Insufflator F114 (K063367) for the Indication for Use to create a cavity along the saphenous vein and/or radial artery during endoscopic vessel harvesting procedures.
#### Summary of nonclinical testing 7
Bench testing demonstrates that UHI-4 performs safely and functionally by proving that UHI-4 can control each proper setting pressure as demonstrated in clinical articles. Bench tests were conducted to verify the functionality of the UHI-4 setting used for each anatomical site indicated. Brief summaries of the tests are listed below.
- 미 Insufflation in endoscopic harvest of saphenous vein via CABG Applicator.
It was confirmed that UHI-4 can control the pressure in an environment saphenous vein model
{3}------------------------------------------------
K110294
pg. 1 of 4
and the acceptance criteria was met.
Insufflation in endoscopic harvest of radial artery via CABG Applicator. ■ It was confirmed that UHI-4 can control the pressure in an environment radial artery model and the acceptance criteria was met.
- = Simulated use testing to demonstrate that the UHI-4 can accurately deliver CO2 at the specified flow rates and pressure settings and that the UHI-4 provides protection from over pressurization.
The results from the tests above demonstrate that UHI-4 can properly control each setting pressure and function as intended.
Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The device software is considered a "Major Level of Concern."
The following standards have been applied to the subject UHI-4:
- IEC 60601-1
- IEC 60601-1-1
- IEC 60601-1-2
- י ISO 14971
- IEC 60601-1-8 .
#### 8 Conclusion
When compared to the predicate devices, the HIGH FLOW INSUFFLATION UNIT UHI-4 does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Olympus Medical Systems Corporation % Ms. Stacy Abbatiello Kluesner, M.S., RAC Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway, P.O. Box 610 CENTER VALLEY PA 18034
NOV 1 8 2011
Re: K110294
Trade/Device Name: HIGH FLOW INSUFFLATION UNIT UHI-4 Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: II Product Code: HIF, OSV Dated: November 16, 2011 Received: November 17, 2011
· Dear Ms. Kluesner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls: Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{5}------------------------------------------------
## Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert Lewin MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
## K110294 510(k) Number (if known): Device Name: HIGH FLOW INSUFFLATION UNIT UHI-4
## Indications For Use:
This instrument has been designed for insufflation of abdominal cavity, and provides automatic suction and smoke evacuation to facilitate laparoscopic observation, diagnosis and treatment.
This instrument is also designed for controlled CO2 insufflation to create a cavity along the saphenous vein and/or radial artery to facilitate observation during an endoscopic vessel harvesting procedure.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
_1
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-------------------------------------------------------------------|---------|
| Division of Reproductive, Gastro-Renal, and<br>Urological Devices | |
| 510(k) Number | K110294 |
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.