← Product Code [HIF](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIF) · K022318

# VOMED ENDOSCOPIC AND UROLOGICAL ACCESSORIES (K022318)

_Vomed Volzer Medizintechnik GmbH & Co. · HIF · Sep 19, 2002 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIF/K022318

## Device Facts

- **Applicant:** Vomed Volzer Medizintechnik GmbH & Co.
- **Product Code:** [HIF](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIF.md)
- **Decision Date:** Sep 19, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1730
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

VOMED insufflation needles are reusable devices for temporary use in conjunction with FDA-cleared insufflators to blow gas or powder into a tube, cavity, or organ. They are generally used as accessories to devices that provide access, illumination and visualization of internal structures and for manipulating soft tissue. The devices are indicated for therapeutic and diagnostic laparoscopic and urological procedures. VOMED suction and irrigation tubes and handles are reusable accessories designed for the suction of fluids from wounds and body cavities and/or the irrigation of surgical sites. They are general accessories of devices that provide access, illumination and visualization of internal structures and for manipulating soft tissue. The suction and irrigation devices are indicated for therapeutic laparoscopic and urologic procedures.

## Device Story

VOMED Endoscopic and Urologic Accessories consist of reusable insufflation needles and suction/irrigation tubes/handles. Devices function as accessories to FDA-cleared insufflators and visualization systems. Insufflation needles deliver gas or powder into body cavities; suction/irrigation components remove fluids or irrigate surgical sites. Used in clinical settings (OR) by physicians during laparoscopic and urologic procedures. Devices facilitate access and tissue manipulation. No automated processing or software involved.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Reusable stainless steel or medical-grade material accessories. Form factor includes needles, tubes, and handles. Manual operation; no energy source, software, or connectivity.

## Regulatory Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the edge. Inside the circle is a stylized image of an eagle. The eagle is facing to the right and has three lines representing its wings.

Food and Drug Administration i 9200 Corporate Boulevard Rockville MD 20850

SEP 1 9 2002

VOMED Volzer Medizintechnik GmbH & Co. % Ms. Dagmar S. Mäser Business Support International Amstel 320-I Amsterdam, NETHERLANDS 1017 AP

Re: K022318

Trade/Device Name: VOMED Endoscopic and Urologic Accessories Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: II Product Code: 85 HIF Dated: August 29, 2002 Received: September 3, 2002

Dear Ms. Mäser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (sec above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

This Ictter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to procced to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Henry C. Grigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K 022318 510(k) Number

Device Name

VOMED Insufflation Needles

## INDICATIONS FOR USE

VOMED insufflation needles are reusable devices for temporary use in conjunction with FDA-cleared insufflators to blow gas or powder into a tube, cavity, or organ.

They are generally used as accessories to devices that provide access, illumination and visualization of internal structures and for manipulating soft tissuc.

The devices are indicated for therapeutic and diagnostic laparoscopic and urological procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use  | ✓ | OR | Over-The-Counter Use |                          |
|-------------------|---|----|----------------------|--------------------------|
| (Per CFR 801 109) |   |    |                      | (Optional Format 1-2-96) |

(Division Sign-Off)

Division of Reproductive, Abdominal,

and Radiological Devices

| 510(k) Number | K022318 |
|---------------|---------|
|---------------|---------|

{3}------------------------------------------------

Device Name

| 510(k) Number | K 022318 |
|---------------|----------|
|---------------|----------|

VOMED Suction & Irrigation Devices

INDICATIONS FOR USE

VOMED suction and irrigation tubes and handles are reusable accessories designed for the suction of fluids from wounds and body cavities and/or the irrigation of surgical sites. They are general accessories of devices that provide access, illumination and visualization of internal structures and for manipulating soft tissue.

The suction and irrigation devices are indicated for therapeutic laparoscopic and urologic procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Pcr CFR 801 109) OR

Over-The-Counter Use -----------

Turnt A. Sarr

510(k) Number

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIF/K022318](https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HIF/K022318)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
