UHI-3 HIGH FLOW INSUFFLATION UNIT

K014166 · Olympus Optical Co., Ltd. · HIF · Jun 25, 2002 · Obstetrics/Gynecology

Device Facts

Record IDK014166
Device NameUHI-3 HIGH FLOW INSUFFLATION UNIT
ApplicantOlympus Optical Co., Ltd.
Product CodeHIF · Obstetrics/Gynecology
Decision DateJun 25, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This instrument has been designed for insufflation of the abdominal cavity, and automatic suction and smoke evacuation. This instrument uses CO2 gas for distension, and it may be used for diagnostic and/or operative laparoscopy.

Device Story

UHI-3 High Flow Insufflation Unit provides CO2 gas for abdominal distension during diagnostic/operative laparoscopy; includes automatic suction and smoke evacuation. Operates by connecting to CO2 supply; integrates with Olympus high frequency (UES-20/30) or ultrasonic surgery units (SonoSurg-G/G2) for smoke evacuation. Suction controlled via external unit through pinch valve; also functions as secondary system for over-pressure relief. Used in clinical/OR settings by physicians. Provides clear field of view by evacuating surgical smoke; maintains abdominal space for instrumentation. Benefits patient by facilitating laparoscopic access and maintaining visibility during surgery.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing and comparison to predicate devices.

Technological Characteristics

Laparoscopic insufflator; CO2 gas delivery; max flow rate 35 l/min. Includes suction/smoke evacuation functionality. Integrates with external HF or ultrasonic surgical units. Components: UHI-3 unit, suction tube, insufflation tube. Class II device.

Indications for Use

Indicated for patients undergoing diagnostic or operative laparoscopic procedures requiring abdominal cavity insufflation with CO2 gas.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KO14166 ## JUN 2 5 2002 ## 510(k) SUMMARY ## HIGH FLOW INSUFFLATION UNIT UHI-3 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR, Section 807.92. #### A. Submitter's Name, Address, Phone and Fax Number #### 1. Manufacturer of the subject device | Name & Address of Manufacturer; | Olympus Optical Co,. Ltd.<br>2-3-1 Shinjuku Monolis Nishi-shinjuku<br>Shinjuku-ku, Tokyo, 163-0914<br>Japan | |-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------| | Registration No :<br>Address, Phone and Fax Number<br>of R&D Department<br>Endoscope Division | 8010047<br>2951 Ishikawa-cho<br>Hachioji-shi, Tokyo 192-8507<br>Japan<br>TEL 81-426-42-5177<br>FAX 81-426-46-5613 | #### 2. Name of Contact Person Name : Ms.Laura Storms-Tyler Director, Regulatory Affairs Olympus America Inc. Two Corporate Center Drive Melville, NY 11747-3157 TEL (631)844-5688 FAX (631) 844-5416 # B. Device Name, Common Name Address, Phone and Fax : | 1. Device Name : | UHI-3<br>HIGH FLOW INSUFFLATION UNIT | |--------------------------|-----------------------------------------------------| | 2. Common/Usual Name : | HIGH FLOW INSUFFLATION UNIT | | 3. Classification Name : | Laparoscopic Insufflator<br>21CFR884.1730, Class II | {1}------------------------------------------------ #### C. Predicate Devices: | Model | Device Description &<br>510(k)#/ Date Cleared | Manufacturer | |-----------------------------------------------|-----------------------------------------------|------------------------------| | OLYMPUS<br>High Flow Insufflation Unit<br>UHI | #K953162<br>12/19/1995 | Olympus Optical Co., Ltd. | | NORTECH<br>Omniflator 7640<br>Model 7-640-00 | #K983326<br>02/05/1999 | Northgate Technologies, Inc. | #### D. Description of the Device This instrument has been designed for insufflation of the abdominal cavity, and automatic suction and smoke evacuation. This instrument uses CO2 gas for distension, and it may be used for diagnostic and/or operative laparoscopy. This instrument is constituted of High Flow Insufflation Unit UHI-3, Suction Tube and Insufflation Tube, and they are used with Recommended Accessories (Optional) and Ancillary Equipment. And auto smoke evacuating function of UHI-3 works by combining with the Olympus high frequency unit (UES-20 / 30) or ultrasonic surgery unit (SonoSurg-G / G2) #### 1. High Flow Insufflation Unit UHI-3 The High Flow Insufflation Unit UHI-3 has been designed for insufflating CO2 gas into abdominal cavity during a laparoscopic procedure. This provides a clear view and space for laparoscopic instrumentation to be used. The UHI-3 has suction capability for the purpose of evacuating smoke caused by HF instruments. The suction ability can also be used as a secondary system to evacuate the abdominal cavity in an over-pressure condition. The suction is controlled with an external suction unit through the pinch valve. ## 2. Suction Tube The suction tube is designed to be used with the Olympus UHI-3 High Flow Insufflation Unit, and a trocar to assist in maintaining a clear field of view by evacuating smoke from the operating sight during laparoscopic surgery. - 3. Insufflation Tube The Insufflation tube is designed to be connected to the Olympus UHI-3 High Flow Insufflation Unit, to supply CO2 gas to the abdominal cavity. ## E. Intended Use of the device This instrument has been designed for insufflation of the abdominal cavity, and automatio suction and smoke evacuation. This instrument uses CO2 gas for distension, and it may be used for diagnostic and/or operative laparoscopy. ## F. Reason for not requiring clinical data The subject HIGH FLOW INSUFFLATION UNIT UHI -3 is similar to the OLYMPUS High Flow Insufflation Unit (UHI) cleared in the previous 510(k) (#953162). Though the maximum flow rate of the previous device, High Flow Insufflation Unit UHI is 12 I/min, less than 20 I/min, which is the maximum flow rate which FDA has currently cleared for marketing laparoscopic Insufflators, the maximum flow rate of this subject device is 35 l/min over the current FDA's cleared level. However the maximum flow rate of OMNIFLATOR 7640 MODEL 7-640-00 cleared in the previous 510(k) (#983326) is 40 l/min. {2}------------------------------------------------ And when compared to the predicated devices listed above, the Olympus High Flow Insufflation Unit UHI-3 does not incorporate any significant change in intended use, method of operation, material, or design that could affect safety and effectiveness. Then it is judged that clinical data is not needed. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three abstract shapes resembling birds in flight, arranged in a row. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUN 2 5 2002 Ms. Laura Storms-Tyler Director, Regulatory Affairs & Quality Assurance Olympus America, Inc. Two Corporate Center Drive MILVILLE NY 11747-3157 Re: K014166 Trade/Device Name: Olympus UHI-3 High Flow Insufflation Unit Regulation Number: 21 CFR 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: II Product Code: 85 HIF Dated: April 22, 2002 Received: April 29, 2002 Dear Ms. Storms-Tyler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C. Snogdon Nancy C. Progdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATION FOR USE STATEMENT 510(k) Number (if known): HIGH FLOW INSUFFLATION UNIT UHI-3 Device Name: Indications for Use: ાર This instrument has been designed for insufflation of the abdominal cavity, and This instrument has been skees wacuation. This instrument uses CO2 gas for distension, and it may be used for diagnostic and/or operative laparoscopy. > Concurrence of CDRH, Office of Device Evaluation ODE Prescription Use Prescription Use / (Per 21 CFR 801.109) Over-The-Counter Use (Optional Format 1-2-96) David A. Bergson (Division Sign-Off) Division of Reproductive. A and Radiological Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...