K201952 · Crossbay Medical · HHK · Aug 5, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K201952
Device Name
CrossGlide ETS Plus
Applicant
Crossbay Medical
Product Code
HHK · Obstetrics/Gynecology
Decision Date
Aug 5, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1175
Device Class
Class 2
Indications for Use
The CrossGlide™ ETS Plus is indicated for use for sonohysterography and to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.
Device Story
Sterile, single-use device for sonohysterography and endometrial biopsy. Comprises delivery catheter with everting membrane and inner aspiration catheter. Physician inserts device into uterine cavity; everting membrane facilitates cervical crossing. Saline infused via 10cc luer-lock syringe for sonohysterography. Same syringe used to aspirate saline and create negative pressure for tissue collection. Manual scraping and vacuum suction remove mucosal tissue. Used in clinical setting by physicians. Output is tissue sample for histological analysis. Benefits include combined diagnostic imaging and biopsy capability in one procedure.
Clinical Evidence
Bench testing only. Evaluated mechanical integrity (tensile, pressurization), functional performance of pinch clamps, and vacuum maintenance with 10cc syringe. Biocompatibility (cytotoxicity, irritation, sensitization) leveraged from predicate. Sterilization validated per ANSI/AAMI/ISO 11135:2014. Packaging/shelf-life validated per ASTM F88, F2096, D4169, and F1980.
Technological Characteristics
Sterile, single-use, manual suction curette. Materials identical to predicate. Everting membrane technology; hydraulic principle for cervical crossing. Components: delivery catheter, everting membrane, inner aspiration catheter, pinch clamps, 10cc syringe. Sterilization: Ethylene Oxide (ANSI/AAMI/ISO 11135:2014). No software or electronic components.
Indications for Use
Indicated for sonohysterography and endometrial tissue biopsy via scraping and vacuum suction in patients requiring uterine mucosal sampling.
Regulatory Classification
Identification
An endometrial suction curette is a device used to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction. This device is used to obtain tissue for biopsy or for menstrual extraction. This generic type of device may include catheters, syringes, and tissue filters or traps.
CrossBay SonoSure™ Sonohysterography and Endometrial Sampling Device (K133144)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 5, 2020
CrossBay Medical % Cindy Domecus Principal Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, CA 94010
Re: K201952
> Trade/Device Name: CrossGlide™ ETS Plus Regulation Number: 21 CFR 884.1175 Regulation Name: Endometrial Suction Curette and Accessories Regulatory Class: II Product Code: HHK Dated: July 13, 2020 Received: July 14, 2020
Dear Cindy Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201952
Device Name CrossGlide™ ETS Plus
Indications for Use (Describe)
The CrossGlide™ ETS Plus is indicated for use for sonohysterography and to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary (21 CFR § 807.92(c))
#### I. SUBMITTER INFORMATION
| Submitter: | CrossBay Medical Inc.<br>13240 Evening Creek Drive, Suite 304<br>San Diego, CA 92128 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Correspondent: | Cindy Domecus, R.A.C. (US & EU)<br>Principal, Domecus Consulting Services LLC<br>Regulatory Consultant to CrossBay Medical Inc.<br>Phone: 650.343.4813<br>Fax: 650.343.7822<br>Email: domecusconsulting@comcast.net |
| Date Summary Prepared: | August 3, 2020 |
## II. SUBJECT DEVICE INFORMATION
| Device Trade Name: | CrossGlide™ ETS Plus |
|--------------------|-------------------------------------------------|
| Common Name: | Curette, Suction, Endometrial (And Accessories) |
| Regulation Number: | 21 CFR §884.1175 |
| Regulation Name: | Endometrial suction curette and accessories |
| Regulatory Class: | II |
| Product Code: | HHK |
#### III. PREDICATE DEVICE INFORMATION
The predicate device is the CrossBay™ Endometrial Tissue Sampler, K192534. The reference device is the CrossBay SonoSure™ Sonohysterography and Endometrial Sampling Device, K133144.
The predicate device and the reference device have not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The CrossGlide™ ETS Plus is a sterile, disposable, single-use device which enables saline infusion sonohysteroraphy and the removal of mucosal tissue from the uterus for histological biopsy. The CrossGlide™ ETS Plus contains a Delivery Catheter with an everting Membrane, an Inner Aspiration Catheter that contains a lumen for saline infusion and aspiration. The everting Membrane places the Inner Aspiration Catheter into the uterine cavity. A proximal luer
{5}------------------------------------------------
connector allows a 10cc luer lock syringe to infuse saline into the uterine cavity for sonohysterography. The 10cc syringe can be used to remove the saline from the uterine cavity and create negative pressure. Movement of the CrossGlide™ ETS Plus device, after creating negative pressure with the 10cc syringe, removes material and tissue from the uterine cavity.
## V. INDICATIONS FOR USE
The CrossGlide™ ETS Plus is indicated for use for sonohysterography and to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction in order to obtain tissue for histological biopsy.
## VI. COMPARISON OF INDICATIONS FOR USE WITH PREDICATE DEVICE
Table 1 below presents the Indications for Use for the subject and predicate devices, with the differences noted in bold font. The differences between the subject and predicate Indications for Use do not alter the intended use of the device. As a reference device is used in this submission, the Indications for Use for the reference device are also provided.
| Subject Device<br>CrossGlide™ ETS Plus | Predicate Device,<br>K192534<br>CrossGlide™ ETS | Reference Device<br>K133144<br>CrossBay SonoSure™<br>Sonohysterography and<br>Endometrial Sampling Device | | Subject Device<br>CrossGlide™<br>ETS Plus | Predicate<br>Device<br>K192534<br>CrossGlide™<br>ETS | Reference Device<br>K133144<br>CrossBay<br>SonoSure™<br>Sonohysterography<br>and Endometrial<br>Sampling Device | Comparison to<br>Predicate Device |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The CrossGlide™ ETS Plus is<br>indicated for use for<br>sonohysterography and to<br>remove material from the<br>uterus and from the mucosal<br>lining of the uterus by scraping<br>and vacuum suction in order to<br>obtain tissue for histological<br>biopsy. | The CrossGlide™ ETS,<br>Endometrial Tissue<br>Sampler, is indicated for<br>use to remove material<br>from the uterus and from<br>the mucosal lining of the<br>uterus by scraping and<br>vacuum suction in order<br>to obtain tissue for<br>histological biopsy. | The CrossBay SonoSure™<br>Sonohysterography and<br>Endometrial Sampling Device is<br>indicated for use to access the<br>uterine cavity for saline infusion<br>sonohysterography and to<br>obtain endometrial biopsy, if<br>indicated, utilizing the same<br>device. | Device<br>Components | Inner catheter,<br>everting<br>membrane,<br>aspiration with<br>hole, blue and<br>white pinch<br>clamp,<br>compliant tube,<br>syringe | Inner catheter,<br>everting<br>membrane,<br>aspiration with<br>hole, blue and<br>white stopcock,<br>compliant tube,<br>syringe | Delivery catheter,<br>cytology brush, saline<br>infusion bag, handle,<br>acorn tip | Different: The<br>subject device has<br>the same components<br>as compared to the<br>predicate device<br>except the blue and<br>white stopcocks have<br>been substituted by<br>blue and white pinch<br>clamps, which<br>perform the same<br>function. This<br>difference does not<br>raise different<br>questions of safety<br>and effectiveness<br>(S&E). |
| Inflation<br>Method | Manual, with<br>3cc syringe | Manual, with<br>3cc syringe | N/A | Same | | | |
| Aspiration<br>Method | Manual, with<br>10cc syringe | Manual, with<br>3cc syringe | N/A | Same aspiration<br>method using<br>different size of<br>syringe | | | |
| Overall<br>Catheter<br>Length (cm) | 43-48 cm | 43-48 cm | 28 cm | Same | | | |
| Inner<br>Catheter | Length: 315mm<br>(includes<br>proximal<br>connector and<br>pinch clamp)<br>Outer diameter:<br>2mm<br>Internal | Length: 315mm<br>(includes<br>Stopcock)<br>Outer diameter:<br>2mm<br>Internal<br>diameter: 1.3mm | N/A | Same | | | |
### TABLE 1. COMPARISON OF INDICATIONS FOR USE
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technological characteristics of the subject and predicate device are compared below in Table 2.
{6}------------------------------------------------
{7}------------------------------------------------
| | Subject Device<br>CrossGlide™<br>ETS Plus | Predicate<br>Device<br>K192534<br>CrossGlide™<br>ETS | Reference Device<br>K133144<br>CrossBay<br>SonoSure™<br>Sonohysterography<br>and Endometrial<br>Sampling Device | Comparison to<br>Predicate Device |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| | diameter: 1.3mm | | | |
| Delivery<br>Catheter | Length: 156mm<br>(includes Acorn<br>Tip)<br>Outer diameter:<br>4.1mm<br>Internal diameter<br>3.3.mm | Length: 156mm<br>(includes Acorn<br>Tip)<br>Outer diameter:<br>4.1mm<br>Internal diameter<br>3.3.mm | Length: 100mm<br>(includes Acorn Tip) | Same |
| Everting<br>Membrane | Length: 65mm at<br>full deployment<br>Outer diameter:<br>3.5mm @ 3<br>atmospheres of<br>pressure<br>Wall thickness:<br>0.04mm | Length: 65mm<br>at full<br>deployment<br>Outer diameter:<br>3.5mm @ 3<br>atmospheres of<br>pressure<br>Wall thickness:<br>0.04mm | N/A | Same |
| Aspiration<br>Device Side<br>Hole &<br>Internal<br>Lumen | Side hole<br>elliptical<br>opening<br>Internal lumen<br>1.3mm | Side hole<br>elliptical<br>opening<br>Internal lumen<br>1.3mm | N/A | Same |
| Acorn Tip | Length: 16mm<br>Outer diameter:<br>12.4mm | Length: 16mm<br>Outer diameter:<br>12.4mm | Length: 22.5mm<br>Outside diameter:<br>20.5mm | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same |
| Saline<br>Infusion<br>Method | Manual<br>(syringe) | NA | Same | Different: The<br>predicate device is<br>not indicated for<br>saline infusion. As<br>such the reference |
| Subject Device<br>CrossGlideTM<br>ETS Plus | Predicate<br>Device<br>K192534<br>CrossGlideTM<br>ETS | Reference Device<br>K133144<br>CrossBay<br>SonoSureTM<br>Sonohysterography<br>and Endometrial<br>Sampling Device | Comparison to<br>Predicate Device | |
| | | | device is included to<br>provide scientific and<br>technical information<br>to address the safety<br>and effectiveness of<br>the saline infusion<br>capability of the<br>subject device. | |
{8}------------------------------------------------
At a high level, the subject and predicate devices comprise the following same technological elements:
- . Both the subject and predicate device use a delivery catheter to place the aspiration device into the uterine cavity to remove material.
- Both the subject and predicate device employ a syringe to apply vacuum on the aspiration . device.
- . Both the subject and predicate device use similar materials in the construction of the catheter.
- . Both the subject device and predicate device utilize everting Membrane technology using the same materials, configuration, dimensions, and hydraulic principle for crossing the cervical canal.
- . Both the subject device and the predicate device have the same sterilization method, packaging construction and materials.
The following technological differences exist between the subject and predicate device:
- The subject device uses a 10cc syringe to apply vacuum suction to the Aspiration Device. The predicate device uses a 3cc syringe to apply vacuum suction.
- . The subject device utilizes a Blue Pinch Clamp and a White Pinch Clamp as one-way valves (on/off). The predicate device utilizes a Blue Stopcock and White Stopcock as one-way valves in the same locations on the catheter. The stopcock and pinch clamp share the same function and performance characteristics.
- . The subject device utilizes a 10cc syringe to infuse saline trough the catheter and into the uterine cavity.
{9}------------------------------------------------
The differences in technology do not raise different questions of safety and effectiveness, as both the subject and predicate device operate to transfer fluid/tissue through the catheter via manual pressure.
# VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Bench Testing
Physical bench testing confirmed that the CrossGlide™ ETS Plus performs according to the product specifications. Device evaluation consisted of physical and functional testing performed pursuant to test protocols that were used to support clearance of the predicate or reference device. Mechanical testing, including tensile and pressurization testing, was conducted to determine if the pinch clamps could operate during expected functional conditions. Additional pressurization testing was conducted to determine if the use of a 10cc syringe can maintain adequate negative pressure required for sample collection. The subject device successfully passed all functional bench testing (i.e., met predefined acceptance criteria).
### Biocompatibility testing
The subject device utilizes identical materials and manufacturing as the predicate device. Therefore, biocompatibility testing on the predicate device, which was conducted according to ISO 10993-1 "Biological Evaluation of Medical Devices" and FDA's guidance "Use of International Standard ISO 10993-1", final document issued on June 16, 2016 can be leveraged to support the biocompatibility of the subject device. Testing included the following with passing results: 1) Cytotoxicity, 2) Vaginal Irritation, and 3) Sensitization.
## Sterilization Validation
The subject device utilizes the same product design and packaging as the predicate device. The predicate device was evaluated for adoption into the current ethylene oxide (EO) sterilization cycle of the contract sterilizer. The evaluation process was performed utilizing the AAMI Guidance contained in TIR28:2016 Product adoption and process equivalence for ethylene oxide sterilization. The EO Sterilization Cycle was originally validated in 2016, using the overkill method, in accordance with ANSI/AAMI/ISO 11135:2014. The results and conclusion of this evaluation confirmed that the subject device could be adopted into the previously validated cycle.
#### Packaging, Shipping Validation, and Shelf-Life
The subject device utilizes the same product design and packaging as the predicate device. Packaging and shipping validation studies were conducted pursuant to the applicable ASTM guidelines (ASTM F88/F88M - 15 "Standard Test Method for Seal Strength of Flexible Barrier
{10}------------------------------------------------
Materials"; and, ASTM F 2096-11 "Standard Test Methods for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble leaks)"; ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems, using Distribution Cycle 13. The shelf-life is supported by packaging and performance tests conducted on samples exposed to accelerated aging conditions pursuant to ASTM F1980 - 16 "Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices".
## IX: CONCLUSIONS
The CrossGlide™ ETS Plus Sonohysterography and Endometrial Tissue Sampler has the same intended use as the predicate device. There are differences in the indications for use statements for the subject and predicate devices. The differences between the subject and predicate Indications for Use do not alter the intended use of the device. In addition, the subject device has the similar technological characteristics as the predicate device and the differences do not raise different questions of safety and effectiveness. Finally, the submitted testing demonstrates that the subject device is as safe, as effective, and performs as well as the predicate device. Therefore, the CrossGlide™ ETS Plus is substantially equivalent to the cleared predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.