A Colposcope is a device designed to permit viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose and examine abnormalities of the vagina and cervix and select areas for biopsy. This generic type of device may include a light source, cables, and components parts.
Device Story
Seiler Colposcopes (Models 935, 955, 985) are telescopic optical devices used by clinicians to provide magnified visualization of vulvar, vaginal, cervical, and anogenital tissues. The device consists of an optical pod with multiple magnification levels (3x to 17x), a halogen light source, and a delivery system (floor stand/base). The 935 model features a column-mounted system; the 955 and 985 models utilize swing-arm mounts to improve workspace clearance. The device includes a dual-mode illumination system allowing for immediate switching to a spare lamp if the primary lamp fails during a procedure. Clinicians use the magnified view to identify tissue abnormalities and guide biopsies or surgical interventions like LEEP and conization. The device is intended for use in clinical settings.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety, EMC, thermal, and optical quality assessments, all of which passed.
Technological Characteristics
Telescopic optical system with 300mm objective lens; magnification ranges (3x-17x); halogen light source with dual-mode illumination; floor-stand delivery system with column or swing-arm mounts. Electrical safety, EMC, and thermal testing performed per TUV certification standards.
Indications for Use
Indicated for magnified visualization of vulvar, vaginal, cervical, and anogenital tissues to evaluate abnormalities, select biopsy sites, and facilitate procedures like LEEP and conization. For prescription use only.
Regulatory Classification
Identification
A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.
Special Controls
*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Seiler Instrument. The logo features a stylized "S" with a crosshair design incorporated into the upper portion. Below the main logo, the word "INSTRUMENT" is written in a smaller, sans-serif font.
K101141
pg. 1 of 2
ISO
9001:2000
Registered
October 2001
INSTRUMENT
300-489-2282 · 314-968-2282 Fax: 314-968-2637 Corporate 314-968-9217 Survey Fax: 314-968-3601 Microscope
Since 1945
Date: July 2010
Section 6.
AUG - 6 2010
## 510(k) Summary
We believe our product is (SE) Substantially Equivalent to a number of other devices currently approved by the FDA. Our device by common name is "COLPOSCOPE" a Class II device, Product Code HEX.
Common Name: Colposcope Proprietary Name: Seiler Colposcope Current models - Model 935 Series, Model 955 Series, Model 985 Series
Intended Use: A Colposcope is a device designed to permit viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose and examine abnormalities of the vagina and cervix and select areas for biopsy. This generic type of device may include a light source, cables, and components parts.
## Indications for Use:
The Seiler colposcope is intended to provide magnified visualization of the tissues of the vulva, vagina, cervix, and anogenital area. It is used to evaluate these tissues, select areas for biopsy, as necessary, and to facilitate related procedures, e.g., LEEP, conization, etc.
## Summary:
The Model 935 colposcope with a 300mm objective lens has 3-magnifications, 3x, 7x, and 17x. The optics are mounted on a column and roll-around base for easy access to the areas of examination.
The Model 955 colposcope with a 300mm objective lens has 5-magnification, 3x, 4x, 7x, 11x, and 17x. The optics are mounted on swing-arms on a roll-around base. This mounting system allows the user to "swing" the optics out of the way when not in use while storing the colposcope in a more permanent location close to the examination table.
{1}------------------------------------------------
The Model 985 colposcope with a 300mm lens also has 5-magnifications. 3x. 4x. 7x. 11x. and T Model 955, the optics are mounted on a swing-arm on a roll-around base, however the optics and arm can be swung into place from a stored position above and to the side of the user. This also allows the optics to be stored in a permanent location. An additional benefit of this mounting system is there are no mounting arms or a column directly in front of the user to utilize examining/surgical instruments in a clear work space directly in front of the areas of examination.
Most predicate colposcopes have an optical pod with various magnifications, a power supply, usually halogen, and a delivery system i.e. floor stand and base. We offer similar features, however our power unit has 2 modes of illumination which allows for the spare illumination to be accessed with a flip of a switch. This quick change feature means that if one of the illuminator lamps fails during an examination or procedure, the examination or procedure will not have to pause while a new lamp is installed.
### ALL ELECTRICAL TESTING INCLUDING ELECTRICAL SAFETY, EMC TESTING, THERMAL TESTING, OPTICAL QUALITY TESTING AND TUV CERTIFICATION WERE PERFORMED BY CERTIFIED TESTING LABORATORIES, THE PRODUCT PASSED ALL TESTS. COPIES OF THE TEST RESULTS ARE LOCATED IN SECTIONS D. AND E. IN OUR SUBMISSION.
There are a number of predicate device colposcopes registered with the FDA. Please see below.
Classification Number: 21 CFR 884.1630 Regulation Name: Colposcope Class II Product Code: HEX
| Company Name | 510(k) Number |
|-------------------|---------------|
| D.F. Vasconcellos | K021854 |
| Jedmed Inst. Co | K884934 |
| Leisegang | K940094 |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES,". The text is in all caps and is in a sans-serif font. The text is centered horizontally in the image. The text is black and the background is white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Daniel Meyer Microscope Consultant Seiler Precision Microscopes Seiler Instrument & Mfg Co., Inc. 3433 Tree Court Industrial Blvd. ST. LOUIS MO 63122
AUG - 6 2010
Re: K101141
> Trade Name: Seiler Colposcopes -- Models 935, 955, & 985 Regulation Number: 21 CFR §884.1630 Regulation Name: Colposcope Regulatory Class: II Product Code: HEX Dated: June 24, 2010 Received: June 25, 2010
Dear Mr. Meyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{3}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Michael Lemond
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K101141
# Indications for Use Form
AUG - 6 2010
510(k) Number (if known): K101141
Device Name: _ Seiler Colposcopes Models_935, 955, & 985
Indications for Use:
The Seiler Colposcope is intended to provide magnified visualization of the tissues of the vulva, vagina, cervix and anogenital area. It is used to evaluate these tissues select areas for biopsy, as necessary, and to facilitate related procedures, e.g., LEEP conization, etc.
X Prescription Use _____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) |
|-------------------------------------------------------------------|
|-------------------------------------------------------------------|

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K101141
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.