SOUNMED COLOPOSCOPE COLPO-99/99PLUS

K070845 · Sounmed, Inc. · HEX · Oct 12, 2007 · Obstetrics/Gynecology

Device Facts

Record IDK070845
Device NameSOUNMED COLOPOSCOPE COLPO-99/99PLUS
ApplicantSounmed, Inc.
Product CodeHEX · Obstetrics/Gynecology
Decision DateOct 12, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1630
Device ClassClass 2

Indications for Use

The Sounmed Colposcope is a device designed to permit viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. A Colposcope is used to diagnose abnormalities and select areas for biopsy.

Device Story

Sounmed COLPO-99/99 PLUS is a stereoscopic optical colposcope for gynecologic examinations. Device uses telescopic optics to provide magnified views of vaginal and cervical tissues; features include 12.5X eyepiece magnification with diopter adjustment, 300mm focal length, and green light filter for tissue contrast. System includes halogen illumination with adjustable brightness and a ball joint mount for 360-degree positioning. Operated by clinicians in a clinical setting; provides direct visual output to the physician. Enables identification of tissue abnormalities and biopsy site selection. Benefits include enhanced visualization of anatomical structures to support diagnostic accuracy.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Stereoscopic optical colposcope; 12.5X magnification; 300mm focal length; halogen illumination; green light filter; ball joint mount for 360-degree motion. Complies with EN 60601-1-1, EN 60601-1-2, and ISO 14971:2000.

Indications for Use

Indicated for the direct visualization of vaginal and cervical tissues to diagnose abnormalities and select areas for biopsy in patients undergoing gynecologic examination.

Regulatory Classification

Identification

A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.

Special Controls

*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ pg. 1 of 2 # 510(K) Summary 0CT 1 2 2007 This summary of 510(k) safety and effectives information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. K170845 The assigned 510(k) number is: Submitter / Distributor: Sounmed, Inc. 6800 N.W. 77th Court Miami, FI 33166 Telephone: 305-477-0986 305-477-5351 Fax: Contact Person: . > Jhoana Lores Email: jhoana@sounmed.com Manufacturer: POLOS S.A 10, Amissou str., GR 551 Kalamaria Thessaloniki Greece Date Prepared: . March 23, 2007 Name of the device: | Trade/Proprietary Name: | Sounmed Colposcopes<br>COLPO-99<br>COLPO 99 PLUS | |-------------------------|--------------------------------------------------| | Common Name: | Colposcope | | Classification: | | |-----------------|--------| | Device Class: | II | | Product Code: | 85 HEX | | Regulation #: | 884. | ## Legally Marketed Predicate Device: DVF CPG Colposcope, K021854, DF Vasconcellos, SA 1630 {1}------------------------------------------------ K070845 pg. 20f2 ## Device Description: The ColposcopeS COLPO-99 / 99 PLUS is a precise optical instrument designed specially for the gynecologic examination. The Sounmed colposcope can be used to view vaginal en cervical tissues using stereoscopic optics. The COLPO-99/ 99 PLUS Colposcope has detailed features that include a wide field of view, long focal length, uniform illumination, adjustable brightness, ease operation, and exceptional optics. It is an essential instrument for any gynecologic examination. The general features of the device are as follows: - Eyepiece Magnification: 12.5X with diopter adjustment - Focal Length: . - 300mm - . Green filter light. - Easy replacement of halogen light bulb . - Ball joint gives you 360° easy motion for gross focus. . #### Statement of Intended Use: The Sounmed Colposcope is a device designed to permit viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. A Colposcope is used to diagnose abnormalities and select areas for biopsy. #### Comparison to predicate devices: The Sounmed (subject device), and Vasconcello Colposcope are intended to permit direct viewing and imaging of the tissues of the vagina, cervix and external genital to diagnose abnormalities and select areas for biopsy. The Vansconcello colposcope rely on similar style of zoom optical systems and objectives that provide the necessary working distances required for patient observation. Both devices are offered with similar floor stand. ## Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence are as follows: The following standards were met: - EN 60601-1-1 - - EN 60601-1-2 - - -ISO 14971:2000 # Discussion of Clinical Tests Performed: Not Applicable. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines extending from its head, all enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 1 2 2007 Mr. Ruben Olivares CEO Sounmed, Inc. Regulatory Affairs Department 6800 N.W. 770 Court MIAMI FL 33166 Re: K070845 Trade Name: Colposcope COLPO-99/99 Plus Regulation Number: 21 CFR §884.1630 Regulation Name: Colposcope Regulatory Class: II Product Code: HEX Dated: October 8, 2007 Received: October 12, 2007 Dear Mr. Olivares: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K070845 Device Name: Sounmed Colposcope COLPO-99 / 99 PLUS Indications For Use: The Sounmed Colposcope COLPO-99 is intended to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy C Brogdon (Division Sign-Off) (Division Sign-Off)) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K070845 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...