← Product Code [HET](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET) · K200216

# 4K UHD Camera System (K200216)

_Scivita Medical Technology Co., Ltd. · HET · Aug 21, 2020 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K200216

## Device Facts

- **Applicant:** Scivita Medical Technology Co., Ltd.
- **Product Code:** [HET](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET.md)
- **Decision Date:** Aug 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1720
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

4K UHD Camera System is used to provide imaging of the operative area in endoscopic surgery. 4K Camera Control Unit The 4K camera control unit has been designed to be used with endoscopes, 4K camera head, light source, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K camera head is compatible with the 4K camera control unit only. 4K Camera Head The 4K camera head has been designed to be used with endoscopes, 4K camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

## Device Story

4K UHD Camera System captures endoscopic images using a CMOS sensor-equipped camera head; converts optical input to electronic signals; processes signals via a 4K camera control unit (CCU). System outputs 4K/2K video via HDMI/SDI to external monitors for real-time visualization. Used in hospitals/clinics by physicians/trained professionals during endoscopic surgery. Provides high-resolution visualization of operative area; aids clinical diagnosis, treatment, and observation. Data can be recorded to external USB devices. Reusable system; requires cleaning/disinfection between uses.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified via depth of field, resolution, direction/field of view, color reproduction, signal-to-noise ratio, geometric distortion, frame frequency, system delay, and image intensity uniformity tests. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and reprocessing standards (AAMI TIR 30, AAMI TIR 12) confirmed.

## Technological Characteristics

Reusable endoscopic imaging system. Components: 4K camera head (1/3" 3-chip CMOS), CCU (models ES-CS4K100/200). Connectivity: HDMI/SDI video output. Power: 100-240V AC. Reprocessing: Disinfection via glutaraldehyde. Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ASTM D4169, AAMI TIR 30/12.

## Regulatory Identification

A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

## Predicate Devices

- VISERA 4K UHD SYSTEM ([K151011](/device/K151011.md))

## Reference Devices

- WOLF Laparoscope ([K941541](/device/K941541.md))
- HENKE SASS WOLF OF AMERICA ARTHROSCOPE ([K080560](/device/K080560.md))
- SINUSCOPE AND ACCESSORIES ([K981751](/device/K981751.md))
- SONY Monitor ([K150377](/device/K150377.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 21, 2020

Scivita Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co, Ltd P.O. Box 120-119 Shanghai. 200120 CHINA

Re: K200216

> Trade/Device Name: 4K UHD Camera System Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic Laparoscope and Accessories Regulatory Class: II Product Code: HET, EOB, FGB, GCJ, NWB, FET, HRX Dated: July 9, 2020 Received: July 22, 2020

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason R. Roberts, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200216

Device Name 4K UHD Camera System

#### Indications for Use (Describe)

4K UHD Camera System is used to provide imaging of the operative area in endoscopic surgery.

## 4K Camera Control Unit

The 4K camera control unit has been designed to be used with endoscopes, 4K camera head, light source, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K camera head is compatible with the 4K camera control unit only.

4K Camera Head

The 4K camera head has been designed to be used with endoscopes, 4K camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

|X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements per 21 CFR 807.92.

1. Sponsor Identification

Sponsor Identification: Scivita Medical Technology Co., Ltd. No.8, Zhong Tian Xiang, Suzhou Industrial Park, Suzhou, Jiangsu, 215000, China.

Establishment Registration Number: Not yet registered

Contact Person: Ruqin Wu Position: Quality Manager Tel: +86-512-81877788 Fax: +86-512-85187285 Email: wuruqin@scivitamedical.com

#### Designated Submission Correspondent:

Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

Date of Preparation: August 20, 2020

{4}------------------------------------------------

## 2. Device Information

Trade Name: 4K UHD Camera System Common Name: Endoscopic Video Imaging System Model: ES-CS4K100/100C ES-CS4K200/100C Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic Laparoscope And Accessories Regulatory Class: II Product Code: HET, EOB, FGB, GCJ, NWB, FET, HRX

#### 3. Identification of Predicate Device

510(k) Number: K151011 Product Name: Olympus Medical Systems Corp. The predicate device has not been subject to a design-related recall.

#### 4. Device Description

The proposed system, 4K UHD Camera System, comprises the 4K camera head, objective lens and power cable and video cables. The proposed system is reusable device, and provided non-sterile. The proposed camera head and should be cleaned and disinfected after each use.

The video cables include an SDI video cable and an HDMI video cable. The 4K camera control unit has two models (ES-CS4K100 and ES-CS4K200) with the only difference on signal output terminals, the detail difference refers to Table 3 of this document; the 4K camera head only has one model (ES-CS4K100C). Therefore, the 4K UHD Camera System is available in two models in combination of the 4K camera control unit and 4K camera head. The system model and primary components of the proposed system are provided in Table 1.

| 4K UHD Camera System | System model    | Primary component                                                   |
|----------------------|-----------------|---------------------------------------------------------------------|
|                      | ES-CS4K100/100C | 4K camera control unit (ES-CS4K100)<br>4K camera head (ES-CS4K100C) |
|                      | ES-CS4K200/100C | 4K camera control unit (ES-CS4K200)<br>4K camera head (ES-CS4K100C) |

Table 1 System model and primary component

Working Principle and/or Mechanism of Action:

The 4K UHD Camera System can provide 4K images via different signal output terminals. The 4K camera head incorporates a complementary metal oxide semiconductor (CMOS) image sensor to convert optical images into electronical signals. The objective lens on end of the 4K camera head is used to connect the camera head with the endoscope.

{5}------------------------------------------------

The 4K UHD Camera System is designed to be used with endoscopes, a light source (ES-LS110D), monitors, light guide cables and other ancillary equipment for endoscopic diagnosis, treatment and observation.

# Conditions of Use:

4K UHD Camera System is intended to be used in hospitals, clinics and doctor's office by doctors and trained healthcare professionals. Compatible equipment of the proposed system is provided in the Table 2.

| Equipment                                                                                                              | Manufacturer | Model                                                    | K Number         |
|------------------------------------------------------------------------------------------------------------------------|--------------|----------------------------------------------------------|------------------|
| 4K UHD Camera System                                                                                                   | SCIVITA      | ES-CS4K 100/100C                                         | This application |
| HSV LAPAROSCOPY SET<br>(optional select one from all endoscope<br>when connecting with the proposed<br>device to test) | WOLF         | Laparoscope<br>(0°, 30°,45°)                             | K941541          |
| HENKE SASS WOLF OF AMERICA<br>ARTHROSCOPE                                                                              | WOLF         | Arthroscope<br>(0°,30°,70°)                              | K080560          |
| SINUSCOPE AND ACCESSORIES                                                                                              | WOLF         | Sinuscope<br>(0°,30°,70°)                                | K981751          |
| LED LIGHT SOURCE                                                                                                       | SCIVITA      | ES-LS110D                                                |                  |
| Monitor                                                                                                                | SONY         | LMD-X2705MD/<br>LMD-X550MD/<br>LMD-X310MD/<br>LMD-X310MT | K150377          |

Table 2 Recommended compatible equipment of the proposed system

The key performance specifications of the proposed system are provided in the Table 3.

## Table 3 General specifications

| Weight                               | 5Kg                         |                                                                                                                                                                                       |
|--------------------------------------|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimension(W×H×D)                     | 370 (W) ×95 (H)×425.5(D) mm |                                                                                                                                                                                       |
| Camera Head                          | 1/3" Three Chip CMOS        |                                                                                                                                                                                       |
| Signal output (resolution)           | 4096×2160, 1920×1080        |                                                                                                                                                                                       |
| Signal<br>output<br>(interface type) | ES-CS4K100/100C             | There are HDMI×2, SDI-1 BNC terminal×4, SDI-2 BNC<br>terminal×1 seven interfaces in total. This model can output<br>the following three types: 4K SDI × 1, 2K SDI × 1, 4K<br>HDMI × 2 |
|                                      | ES-CS4K200/100C             | There are HDMI×2, SDI-1 BNC terminal×4, SDI-2 BNC<br>terminal×1 seven interfaces in total. This model can                                                                             |

{6}------------------------------------------------

|                     | output the following four types: 4K SDI × 1, 2K SDI × 1,<br>4K HDMI × 1, 2K HDMI × 1                                                                                      |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Video output format | HDMI output: 2160/59.94p, 2160/50p, 1080/59.94p,<br>1080/59.94i, 1080/50p, 1080/50i<br>SDI output: 2160/59.94p, 2160/50p, 1080/59.94p,<br>1080/59.94i, 1080/50p, 1080/50i |
| Image recording     | The data is stored in real time to the external access USB<br>interface device by photographing or recording functions.                                                   |
| Operation mode      | Continuous operation                                                                                                                                                      |
| White balance       | Automatic white balance                                                                                                                                                   |

The level of concern of the software contained in the subject device is determined to be Moderate Level of Concern.

## 5. Indications for use:

4K UHD Camera System is used to provide imaging of the operative area in endoscopic surgery.

## 4K Camera Control Unit

The 4K camera control unit has been designed to be used with endoscopes, 4K camera head, light source, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K camera head is compatible with the 4K camera control unit only.

#### 4K Camera Head

The 4K camera head has been designed to be used with endoscopes, 4K camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation

The indications for use statement of the subject device is similar to that of the video processor component of the predicate device. The main difference in the indication for use is the subject device doesn't include indications for the light source because a light source is not included in this subject device submissions for use of the camera control unit and camera head of the subject device and predicate device are the differences in the indications do not represent a new intended use.

{7}------------------------------------------------

6. Comparison of Technological Characteristics

| ITEM                                  | Subject Device K200216                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device K151011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                  | 4K UHD Camera System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | VISERA 4K UHD SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                          | HET, EOB, FGB, GCJ, NWB, FET,<br>HRX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | HET, GCJ, EOB, EOQ, FGB, NWB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation Number                     | 21 CFR 884.1720                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 21 CFR 884.1720,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Class                                 | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for Use                   | 4K UHD Camera System is used to<br>provide imaging of the operative area in<br>endoscopic surgery.<br><br>4K Camera Control Unit<br>The 4K camera control unit has been<br>designed to be used with endoscopes, 4K<br>camera head, light source, monitors, and<br>other ancillary equipment for endoscopic<br>diagnosis, treatment, and observation.<br>The 4K camera head is compatible with<br>the 4K camera control unit only.<br><br>4K Camera Head<br>The 4K camera head has been designed<br>to be used with endoscopes, 4K camera<br>control unit, and other ancillary<br>equipment for endoscopic diagnosis,<br>treatment, and observation | VISERA 4K UHD CAMERA CONTROL<br>UNIT OLYMPUS OTV-S400<br>The camera control unit has been designed<br>to be used with Olympus endoscopes,<br>camera heads, light source, monitors, and<br>other ancillary equipment for endoscopic<br>diagnosis, treatment, and observation.<br><br>VISERA 4K UHD XENON LIGHT<br>SOURCE OLYMPUS CLV-S400<br>The light source has been designed to be<br>used with Olympus endoscopes, camera<br>control unit, light guide cables, and other<br>ancillary equipment for endoscopic<br>diagnosis, treatment, and observation.<br><br>4K CAMERA HEAD OLYMPUS<br>CH-S400-XZ-EB<br>The camera head has been designed to be<br>used with Olympus endoscopes, camera<br>control unit, and other ancillary equipment<br>for endoscopic diagnosis, treatment, and<br>observation. |
| Single use / Reusable                 | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterile                               | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Configuration<br>(primary components) | 4K Camera Control Unit<br>4K Camera Head                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | light source<br>4K Camera Control Unit<br>4K Camera Head                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| ITEM                                  | Subject Device K200216                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Model                                 | ES-CS4K100<br>ES-CS4K200                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K151011<br>VISERA 4K UHD CAMERA CONTROL<br>UNIT OLYMPUS OTV-S400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Power supply                          | 100-240V AC, 50/60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 100-240V~ ±10%,<br>50/60Hz±1Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Dimension                             | 370(W) ×95(H)×425.5(D) mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 390(W) × 160(H) × 506(D) mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Weight                                | 5Kg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 13.5Kg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Input power                           | 150VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 350VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Resolution                            | 4096×2160pxl,<br>1920×1080pxl                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 4096×2160,<br>3840×2160,<br>1920×1080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Image<br>signal<br>output             | ES-CS4K100                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Total seven signal output terminals<br>HDMI (4K HMDI) ×2,<br>SDI (4K SDI) terminal×4,<br>SDI (3G/HD-SDI) terminal×1;<br>The seven signal output terminals can<br>output three types of signal output: 4K<br>SDI , 2K SDI , 4K HDMI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                       | ES-CS4K200                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Total seven signal output terminals:<br>HDMI (4K HMDI) ×1,<br>HDMI (HMDI)×1,<br>SDI (4K) terminal×4,<br>SDI (3G/HD-SDI) terminal×1;<br>The seven signal output terminals can<br>output four types of signal output: 4K SDI, 2K SDI , 4K HDMI, 2K HDMI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Standard Color Chart<br>Output        | Color bar image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Color bar image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Observation light imaging             | White light imaging(WLI)<br>Six imaging function                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | White light imaging (WLI)<br>Narrow band imaging (NBI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

Table 4 General Comparison

Table 5 Comparison of ES-CS4K100/200 Camera Control Unit

{8}------------------------------------------------

| Table 6 Comparison of Camera Head |  |
|-----------------------------------|--|
|                                   |  |

| ITEM                  | Subject Device K200216    | Predicate Device K151011                |
|-----------------------|---------------------------|-----------------------------------------|
| Model                 | ES-CS4K100C               | 4K CAMERA HEAD OLYMPUS<br>CH-S400-XZ-EB |
| Camera Head dimension | W48 mm × H64 mm × L168 mm | W43.6 mm × H49.5 mm × L122.5 mm         |
| Cable dimension       | φ5mm× 2.9m                | 5.1 mm x 3 m                            |

{9}------------------------------------------------

| Wight(excluding cable) | 231.4g (excluding cable)                         | 280 g (excluding cable) |
|------------------------|--------------------------------------------------|-------------------------|
| Reprocessing           | End user reprocess                               | End user sterilized     |
|                        | Disinfection by using glutaraldehyde<br>solution | EOG/ STERRAD            |

The differences in technological characteristics do not raise different questions of safety and effectiveness.

- 7. Summary of Non-Clinical Performance Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > IEC 60601-1:2005+AMD1:2012 Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance.
- > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests.
- IEC 60601-2-18:2009 Medical electrical equipment Part 2-18: Particular requirements for the basic safety and > essential performance of endoscopic equipment
- > ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems.
- > AAMI TIR 30:2016 A compendium of processes, materials, and acceptance criteria for cleaning reusable medical devices
- > AAMI TIR 12:2010 Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers

The software of the proposed device was validated as Moderate level of concern (LoC) in accordance with the following guidance documents: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices -Guidance for Industry and FDA Staff

The following performance testing were also conducted on the subject device and the subject device met all predefined acceptance criteria:

- A Depth of Field Test
- A Resolution Test
- A Direction of View and Field of View Test
- Color Reproduction and Signal Noise Ratio Test >
- > Geometric Distortion Test
- > Image Frame Frequency and System Delay
- > Image Intensity Uniformity

{10}------------------------------------------------

# 8. Conclusion

The performance testing summarized above support a substantial equivalence determination. The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K200216](https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K200216)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
