← Product Code [HET](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET) · K190744

# ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A) (K190744)

_Olympus Winter & Ibe GmbH · HET · Aug 29, 2019 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K190744

## Device Facts

- **Applicant:** Olympus Winter & Ibe GmbH
- **Product Code:** [HET](/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET.md)
- **Decision Date:** Aug 29, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.1720
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

This instrument has been designed to be used with a video system center, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

## Device Story

ENDOEYE HD II is a rigid video telescope for endoscopic diagnosis, treatment, and surgery. Device inputs include optical images from thoracic/abdominal cavities; outputs are high-definition video signals displayed on a monitor. Used in OR settings by surgeons; requires connection to video system center, light source, and monitor. Features include white light or narrow band imaging (NBI) and distal tip heater (select models) to prevent lens fogging. Provides visualization to assist surgeons in performing minimally invasive procedures. Benefits include high-definition visualization of internal anatomy to guide surgical interventions.

## Clinical Evidence

Bench testing only. Testing included electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), thermal safety, mechanical performance (ISO 8600-3/4/5), transport/shipping (ASTM D4169-16), illumination system testing, software verification (IEC 62304), and usability validation of reprocessing instructions. All samples met predefined acceptance criteria.

## Technological Characteristics

Rigid video telescope; color CCD sensor (2 chips); NBI observation capability; distal tip heater (select models). Connectivity: connects to video system center/light source. Sterilization: autoclavable. Standards: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, ISO 8600-3/4/5, ISO 14971. Software: includes PCB with supervisor circuitry.

## Regulatory Identification

A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

## Predicate Devices

- ENDOEYE HD II ([K111788](/device/K111788.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

August 29, 2019

Olympus Winter & Ibe GmbH Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway Center Valley, PA 18034-061

Re: K190744

Trade/Device Name: ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A) Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic laparoscope and accessories Regulatory Class: II Product Code: HET, GCJ, NWB Dated: July 29, 2019 Received: July 30, 2019

Dear Sheri Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190744

Device Name ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A)

Indications for Use (Describe)

This instrument has been designed to be used with a video system center, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

| Type of Use (Select one or both, as applicable)                                      |                                                                      |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

## August 26, 2019

#### General Information 1

| Manufacturer: | Olympus Winter & Ibe GmbH                  |
|---------------|--------------------------------------------|
|               | Kuehnstr. 61                               |
|               | 22045 Hamburg                              |
|               | Germany                                    |
|               | Establishment Registration Number: 9610773 |

| Official Correspondent: | Sheri L Musgnung                           |
|-------------------------|--------------------------------------------|
|                         | Manager, Regulatory Affairs                |
|                         | Olympus Corporation of the Americas        |
|                         | 3500 Corporate Parkway                     |
|                         | Center Valley PA 18034-061                 |
|                         | Phone: 484 896-3147                        |
|                         | Email: sheri.musgnung@olympus.com          |
|                         | Establishment Registration Number: 2429304 |

#### Device Identification 2

| Proprietary /Trade Name: | ENDOEYE HD II (WA50040A, WA50042A,WA50050A, WA50052A) |
|--------------------------|-------------------------------------------------------|
| Classification name:     | Laparoscope, Gynecologic and accessories              |
| Regulations Number:      | 21 CFR 884.1720                                       |
| Regulatory class:        | II                                                    |
| Product code:            | HET<br>GCJ / NWB                                      |
| Review Panel:            | Obstetrics and Gynecology                             |

#### Predicate Device 3

Olympus K11788, ENDOEYE HD II

The predicate has been subject to a design related recall.

{4}------------------------------------------------

#### 4 Product Description

The ENDOEYE HD II - High Definition Digital Video Laparoscope is a rigid video telescope used for endoscopic diagnosis, treatment, video observation and surgery within the thoracic and abdominal cavities including female reproductive organs. For laparoscopic applications, the video telescope is inserted via a trocar into the patient.

The ENDOEYE HD II is used with a video system center, light source and monitor to achieve its intended function.

In addition, the ENDOEYE HD II can be inserted in compatible instrument trays for reprocessing.

The ENDOEYE HD II can provide an image with either white light or narrow band imaging and

WA50040A and WA50042A include a heater function at the distal tip to reduce fogging of the lens.

#### 5 Indications for Use

This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

The ENDOEYE HD II has the same intended use as the predicate device.

#### Technological Characteristics 6

The ENDOEYE HD II has the same technological characteristics as the predicate device.

| Device        | Subject device     | Predicate device   |
|---------------|--------------------|--------------------|
| Device Name   | EndoEYE HD II      | EndoEYE HD II      |
| Model Number  | WA50040A, WA50042A | WA50040A, WA50042A |
|               | WA50050A, WA50052A | WA50050A, WA50052A |
| 510(k) number | Not yet known      | K111788            |

{5}------------------------------------------------

| Device                                         | Subject device                                                                                                                  | Predicate device                                                                                                                |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Field of view                                  | WA50040A, WA50042A:<br>90° $\pm$ 9° (at 16:9 aspect<br>ratio)<br>WA50050A, WA50052A:<br>80.4° $\pm$ 8° (at 5:4 aspect<br>ratio) | WA50040A, WA50042A:<br>90° $\pm$ 9° (at 16:9 aspect<br>ratio)<br>WA50050A, WA50052A:<br>80.4° $\pm$ 8° (at 5:4 aspect<br>ratio) |
| Depth of field                                 | 21 to 200 mm                                                                                                                    | 21 to 200 mm                                                                                                                    |
| Direction of<br>viewing (forward)              | WA50040A, WA50050A:<br>0° $\pm$ 10°<br>WA50042A, WA50052A:<br>30° $\pm$ 10°                                                     | WA50040A, WA50050A:<br>0° $\pm$ 10°<br>WA50042A, WA50052A:<br>30° $\pm$ 10°                                                     |
| Type of CCD chip                               | Color CCD                                                                                                                       | Color CCD                                                                                                                       |
| Number of CCD<br>chip                          | 2                                                                                                                               | 2                                                                                                                               |
| Outer diameter of<br>distal end                | Ø 10 mm (WA50040A,<br>WA50042A)<br>Ø 5.4 mm (WA50050A,<br>WA50052A)                                                             | Ø 10 mm (WA50040A,<br>WA50042A)<br>Ø 5.4 mm (WA50050A,<br>WA50052A)                                                             |
| Outer diameter of<br>insertion tube<br>maximum | WA50040A: 10.07 mm<br>WA50042A: 10.2 mm<br>WA50050A: 5.45 mm<br>WA50052A: 5.45 mm                                               | WA50040A: 10.07 mm<br>WA50042A: 10.2 mm<br>WA50050A: 5.45 mm<br>WA50052A: 5.45 mm                                               |
| Working length                                 | WA50040A: 325 $\pm$ 5 mm<br>WA50042A: 330 $\pm$ 1 mm<br>WA50050A: 300 $\pm$ 0.5 mm<br>WA50052A: 302 $\pm$ 1 mm                  | WA50040A: 325 $\pm$ 5 mm<br>WA50042A: 330 $\pm$ 1 mm<br>WA50050A: 300 $\pm$ 0.5 mm<br>WA50052A: 302 $\pm$ 1 mm                  |
| NBI observation                                | available                                                                                                                       | available                                                                                                                       |
| Heater<br>functionality                        | WA50040A,<br>WA50042A: Yes<br>WA50050A,<br>WA50052A: No                                                                         | WA50040A,<br>WA50042A: Yes<br>WA50050A,<br>WA50052A: No                                                                         |
| Autoclavability                                | yes                                                                                                                             | yes                                                                                                                             |

The subject devices have equivalent technology, performance, dimensions and materials. The differences to the predicate device ENDOEYE HD II are:

- . Addition of a printed circuit board (PCB) containing a supervisor circuitry and a second temperature sensor
- Updated Instructions for Use. .

The above described differences in technological characteristics do not raise different questions of safety and effectiveness.

{6}------------------------------------------------

#### 7 Performance Data

The following Design verification and validation testing was performed on the subject device:

- Electrical safety as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition 3.0
- Electromagnetic compatibility (EMC) as per IEC 60601-1-2 Edition 4.0
- Thermal Safety as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition ● 3.0
- Mechanical performance testing as per ISO 8600-3 Edition 1, ISO 8600-4 Edition 2 and ISO 8600-5 Edition 1
- Transport and shipping testing as per ASTM D4169-16 ●
- Tests related to the expected service life as per IEC 60601-1 Edition 3.1 and . IEC 60601-2-18 Edition 3.0
- Testing of the illumination system as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition 3.0
- Software Verification as per IEC 62304 Edition 1.1 ●
- . Usability Validation of the Instructions for Use as per FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling

Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007.

All samples tested met their predefined acceptance criteria.

#### 8 Conclusion

The performance data support the safety of the device and demonstrate that the subject devices comply with the recognized standards as specified and support a substantial equivalence determination.

---

**Source:** [https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K190744](https://fda.innolitics.com/submissions/OB/subpart-b%E2%80%94obstetrical-and-gynecological-diagnostic-devices/HET/K190744)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
